Lifescan blood glucose monitor manuale d’uso
- Visualizza on-line o scarica il manuale
- 74 pagine
- 1.75 mb
Vai alla pagina of
Manuali d’uso simili
-
Blood Glucose Meter
Lifescan OneTouch UltraMini Blood Glucose Modelling System
2 pagine 0.73 mb -
Blood Glucose Meter
Lifescan OneTouch Basic Basic Blood Glucose Monitoring System
98 pagine 2.07 mb -
Blood Glucose Meter
Lifescan Blood Glucose Modelling System
2 pagine 0.7 mb -
Blood Glucose Meter
Lifescan SureStep Blood Glucose Monitor
89 pagine 2.63 mb -
Blood Glucose Meter
Lifescan Blood Glucose Monitor System
56 pagine 3.86 mb -
Blood Glucose Meter
Lifescan Blood Glucose Monitor
89 pagine 2.51 mb -
Blood Glucose Meter
Lifescan SureStep
90 pagine 2.72 mb -
Blood Glucose Meter
Lifescan Blood Glucose Monitoring System
103 pagine 1.51 mb
Un buon manuale d’uso
Le regole impongono al rivenditore l'obbligo di fornire all'acquirente, insieme alle merci, il manuale d’uso Lifescan blood glucose monitor. La mancanza del manuale d’uso o le informazioni errate fornite al consumatore sono la base di una denuncia in caso di inosservanza del dispositivo con il contratto. Secondo la legge, l’inclusione del manuale d’uso in una forma diversa da quella cartacea è permessa, che viene spesso utilizzato recentemente, includendo una forma grafica o elettronica Lifescan blood glucose monitor o video didattici per gli utenti. La condizione è il suo carattere leggibile e comprensibile.
Che cosa è il manuale d’uso?
La parola deriva dal latino "instructio", cioè organizzare. Così, il manuale d’uso Lifescan blood glucose monitor descrive le fasi del procedimento. Lo scopo del manuale d’uso è istruire, facilitare lo avviamento, l'uso di attrezzature o l’esecuzione di determinate azioni. Il manuale è una raccolta di informazioni sull'oggetto/servizio, un suggerimento.
Purtroppo, pochi utenti prendono il tempo di leggere il manuale d’uso, e un buono manuale non solo permette di conoscere una serie di funzionalità aggiuntive del dispositivo acquistato, ma anche evitare la maggioranza dei guasti.
Quindi cosa dovrebbe contenere il manuale perfetto?
Innanzitutto, il manuale d’uso Lifescan blood glucose monitor dovrebbe contenere:
- informazioni sui dati tecnici del dispositivo Lifescan blood glucose monitor
- nome del fabbricante e anno di fabbricazione Lifescan blood glucose monitor
- istruzioni per l'uso, la regolazione e la manutenzione delle attrezzature Lifescan blood glucose monitor
- segnaletica di sicurezza e certificati che confermano la conformità con le norme pertinenti
Perché non leggiamo i manuali d’uso?
Generalmente questo è dovuto alla mancanza di tempo e certezza per quanto riguarda la funzionalità specifica delle attrezzature acquistate. Purtroppo, la connessione e l’avvio Lifescan blood glucose monitor non sono sufficienti. Questo manuale contiene una serie di linee guida per funzionalità specifiche, la sicurezza, metodi di manutenzione (anche i mezzi che dovrebbero essere usati), eventuali difetti Lifescan blood glucose monitor e modi per risolvere i problemi più comuni durante l'uso. Infine, il manuale contiene le coordinate del servizio Lifescan in assenza dell'efficacia delle soluzioni proposte. Attualmente, i manuali d’uso sotto forma di animazioni interessanti e video didattici che sono migliori che la brochure suscitano un interesse considerevole. Questo tipo di manuale permette all'utente di visualizzare tutto il video didattico senza saltare le specifiche e complicate descrizioni tecniche Lifescan blood glucose monitor, come nel caso della versione cartacea.
Perché leggere il manuale d’uso?
Prima di tutto, contiene la risposta sulla struttura, le possibilità del dispositivo Lifescan blood glucose monitor, l'uso di vari accessori ed una serie di informazioni per sfruttare totalmente tutte le caratteristiche e servizi.
Dopo l'acquisto di successo di attrezzature/dispositivo, prendere un momento per familiarizzare con tutte le parti del manuale d'uso Lifescan blood glucose monitor. Attualmente, sono preparati con cura e tradotti per essere comprensibili non solo per gli utenti, ma per svolgere la loro funzione di base di informazioni e di aiuto.
Sommario del manuale d’uso
-
Pagina 1
™ ™ Owner’ s Booklet Complete Diabetes T racking System For diabetes and life.[...]
-
Pagina 2
CA UTION Before using any product to test your blood glucose (blood sugar), read all instructions and practise the test. Do all quality control checks as directed and consult with a diabetes health- care professional. The One T ouch ™ Profile ™ S ystem is intended for use outside the body (in vitro diagnostic use). I t should be used only for t[...]
-
Pagina 3
Introduction 4 Precautions and limitations 5 One Touch ™ Profile ™ Blood Glucose Meter 6 One Touch ™ Test Strips 8 SETTING THE SYSTEM Coding the meter 10 Check strip test 12 Control solution test 16 TESTING YOUR BLOOD Getting a drop of blood 22 Step- by- step test procedure 29 Event labelling 35 USING THE METER MEMORY Entering the memory 37 R[...]
-
Pagina 4
Time and date 43 Event averaging 44 Insulin prompt 45 Insulin programming 46 Display language 48 Time format 49 Date format 49 Decimal separator 50 Unit of measure 51 Exiting the set-up mode 51 SOLVING PROBLEMS Test procedure messages 52 Error messages 56 TAKING CARE OF YOUR METER Cleaning the meter 63 Replacing the batteries 66 SPECIFICATIONS Perf[...]
-
Pagina 5
INTR ODUCTION The One T ouch ™ Profile ™ G lucose M onitoring System is for people with diabetes who need to check the level of glucose (sugar) in their blood. T o per- form a blood glucose test, a One T ouch ™ T est Strip is inserted into the meter and a small sample of blood is applied to the test strip. The meter then measures the blood gl[...]
-
Pagina 6
PRECA UTIONS AND LIMIT A TIONS • If you experience symptoms that are not consistent with your blood glucose results, and you have carefully followed the procedure described in your owner’s booklet, contact your healthcare professional immediately. • DO NOT make significant changes to your medication program or ignore physical symptoms without[...]
-
Pagina 7
ONE T OUCH ™ PR OFILE ™ BL OOD GLUC OSE METER S tudy this diagram to become familiar with all of the parts of your One T ouch ™ P rofile ™ M eter . 6 Battery Compartment (on back) Data Port C M C Button Test Strip Holder Display Test Area with protective coating M Button On/Off Button[...]
-
Pagina 8
On/off button. This button turns the meter on and off. Display. This is wher e you read the simple messages that guide you through test- ing, setting options, and using the memory . Y our test results will be dis- played here. M Button. The One T ouch ™ P rofile ™ M eter automatically stores your test results. U se the M button to recall the in[...]
-
Pagina 9
ONE T OUCH ™ TEST STRIPS When blood is applied to the white test spot on the test strip, compounds on the test strip react with the blood to form a blue color . The meter reads the color to determine the blood glucose level. 8 TEST SPOT OR THIS NO TE: Y our test strip may look like THIS Consult your test strip package insert for specific test str[...]
-
Pagina 10
IMPOR T ANT TEST STRIP INFORMA TION • One Touch ™ Test Strips are sensitive to heat and light. Keep them sealed in the origi- nal vial and store in a cool, dry place below 30 º C (86 º F). Do not refrigerate or place in direct heat or sunlight. • DO NOT use One Touch Test Strips beyond the expiration date on the vial. • DO NOT use One Tou[...]
-
Pagina 11
INSERT STRIP CODE 9 CODE 9 7:26AM 1/06/00 CODING THE METER Code numbers are used to calibrate the meter with the test strips for reliable results. Make sure y ou set the code: • Before using the meter for the first time. • E very time you change to another vial of test strips. Check that the code number on the display matches the code number on[...]
-
Pagina 12
Step 2: M atch the Code N umbers. If the code number on the display matches the code number on your test strip vial, you may begin testing. If the two code numbers do not match, follow S tep 3 to code the meter . Step 3: C ode the Meter . After the word INSER T is dis- played on the meter , press and release the C button; the code number will appea[...]
-
Pagina 13
I t is very impor tant that you do the fol- lowing two simple checks routinely to make sure you get accurate r esults. CHECK STRIP TEST A purple and white check strip is included in your meter case. Before doing a check strip test, make sure the test strip holder , test area, and check strip are clean, dry , and lint-free. Do the check strip test a[...]
-
Pagina 14
Step 1: T urn on the meter . The code number will appear on the display . Next, INSER T CODE 10 (for example) STRIP will appear . Step 2: I nsert the check strip Slide the notched end of the check strip into the test strip holder with the purple side facing up. After the check strip has been insert- ed, W AIT PLEASE appears for a mo- ment. Step 3: [...]
-
Pagina 15
The display will show ✓ OK if your result falls within the correct check strip range printed on the back of your meter . If the reading is not within the recommended range, the meter display will show ✓ NO T OK REDO ✓ . Repeat the check strip test until your result falls within the correct range. Step 4: T urn the check strip over and insert [...]
-
Pagina 16
IMPOR T ANT CHECK STRIP INFORMA TION • If your check strip result is r epeatedly out of range, your meter may not be working properly . Do not use the meter to test your blood until you get a check strip result that is within the correct range. • Make sure both sides of the check strip ar e clean. If necessary , wipe the check strip with a soft[...]
-
Pagina 17
CONTR OL SOLUTION TEST One T ouch ™ N ormal Control Solution is used instead of blood to tell you that the meter and test strips are working togeth- er properly . The control solution con- tains a known amount of glucose that reacts with a One T ouch ™ T est Strip . Do a control solution test: • Before using your meter for the first time. •[...]
-
Pagina 18
IMPOR T ANT CONTROL SOL UTION TEST INFORMA TION • U se only One T ouch ™ Normal Control S olution. • Check the expiration date of the control solution. D o not use if expired. Recor d the discard date on the vial. Discar d the vial 90 days after first opening. • The control solution range printed on the test strip vial is for One T ouch N o[...]
-
Pagina 19
The control solution test follows the normal test procedure, using control solution instead of blood. Step 1: T urn on the meter . The code number , time, and date will appear . The code number should match the code number on the test strip vial. Step 2: I nsert a test strip. When INSER T CODE 10 (for exam- ple) STRIP appears, slide a test strip in[...]
-
Pagina 20
Step 3: A pply control solution. After the test strip is in the test strip holder , W AIT PLEASE appears for a moment, then APPL Y SAMPLE appears. Shake the control solution vial well before r emoving the cap and adding a drop of control solution to the middle of the test spot, covering it completely . 19 C M N o r m a l G lu c o s e C o n tro l S [...]
-
Pagina 21
20 Step 4: The r esult appears within 45 seconds. Compare the control solution test r esult with the control solution range printed on the One T ouch ™ T est Strip vial. When the control solution test result lies within the control range, you are r eady to test your blood. C M 100 mg/dL CONTRL. 7:52AM 1/06/00 Recor d control solution tests in the[...]
-
Pagina 22
21 TESTING YOUR BL OOD DET AILED TEST PROCEDURE Choose a clean, dry work surface. Make sure you have all the items needed to test. Release Button Ejection /Cocking Control Depth Adjustment Knob Cap Depth Settings Depth Indicator Penlet ™ Plus Adjustable Blood Sampler C M Meter One Touch ™ Test Strip Sterile Lancet[...]
-
Pagina 23
22 CA UTION: T o reduce the chance of infection: • N ev er share a lancet with another person. • Always use a new , sterile LifeScan Lancet. Lancets are for single use only . • The P enlet ™ Plus Sampler should be used by only one person. H owever , in the event the sampler is shared with another person, always use a new lancet and a new or[...]
-
Pagina 24
23 Step 2: Adjust P unctur e Depth. T wist the depth adjustment knob located at the end of the P enlet ™ Plus Sam pler to desired puncture setting. S maller bumps are for shallower p un c- tures, while larger bumps are use d f or deeper punctures. Gently twist off the lancet’ s protective disk. Make two full turns to ensure that the lancet sepa[...]
-
Pagina 25
24 Step 3: C ock the P enlet ™ Plus Sampler . Slide the ejection/cocking control back until you hear a click. If it does not click, the P enlet Plus Sampler may have been cocked when the lancet was inserted. Step 4: W ash Y our Hands. W ash your hands with soap and warm water . W arm water stimulates blood flow to the fingers. Rinse and dry hands[...]
-
Pagina 26
25 • Choose a spot on the side of a different finger each time you test. Repeated punctures in the same spot may cause soreness and calluses. Step 5: G et a Drop of Blood. Helpful tips: • T o increase blood flo w to your finger- tips, massage the hand from the wrist to the fingertip two or three times without touching the puncture site. • Han[...]
-
Pagina 27
26 Gently massage your finger to obtain the required blood v olume. P roceed with your blood glucose test (see “S tep- by-S tep T est Procedur e ”, page 29). Hold the P enlet ™ Plus S ampler firmly against the side of y ou r fi nger with the cap resting on your finger . (G reater pressure of the device against the finger will cause a deeper p[...]
-
Pagina 28
27 Optional: T o replace the protective disk on the lancet, place the disk on a hard surface and push the exposed tip into the protective disk. Caution: Always use caution when remo ving the lancet. Step 6: R emove the Lancet. T o remo ve the P enlet ™ Plus Cap, twist the cap counterclockwise.[...]
-
Pagina 29
28 Return the ejection/cocking control to the middle position. Replace the cap . P oint the lancet down and away from you. P ush forward on the ejection/cocking control and eject the lancet directly into an appropriate container for sharp objects. CLEANING THE PENLET ™ PL US SAMPLER • Clean the P enlet Plus Sampler with soap and water as needed[...]
-
Pagina 30
STEP- BY - STEP TEST PR OCEDURE Step 1: T urn on the meter . P ress the pow er button. A code number and the date and time will appear . The code number must match the code number on the test strip vial. If it does not, change the meter code number fol- lowing the instructions on page 11. 29 MC CODE 10 8:00AM 1/06/00 CODE 10 8:00AM 1/06/00[...]
-
Pagina 31
WAIT PLEASE APPL Y SAMPLE Step 2: I nsert a test strip. N ext, INSER T CODE 10 (for exam- ple) STRIP appears. Slide a test strip into the strip holder , notched end first, with the test spot facing up. M ake sure you push the test strip in all the way until it stops. After W AIT PLEASE, the meter shows APPL Y SAMPLE. Y ou then have 5 minutes to app[...]
-
Pagina 32
Many times, the meter can detect if there is not enough blood and will show but you should also look at the test strip to make sure. Step 3: A pply blood to the test spot Apply blood to the test spot so as to co ver it completely with a round drop . It is v ery important that enough blood is applied. Do not: • Smear blood onto the test spot. • [...]
-
Pagina 33
32 Checking the amount of blood on the test strip. • Remo ve the test strip from the meter and look at the test spot. I t should still have a wet shiny drop that complete- ly covers the cir cle. If the blood sample has a dull, dry appearance, you may not have applied enough blood or you may have smeared it. • Look at the back of the strip. Y ou[...]
-
Pagina 34
The meter displays results abo ve 600 mg/dL or 33.3 mmol/L as HI DANGER . A result that reads HI DANGER means very high blood glucose (severe hyperglycemia). 33 Accur ate results in 45 seconds. The meter then counts down from 45 to 0 seconds, beeps, and displays your result. Remo ve the test strip only after the result is displayed. C M 106 mg/dL 8[...]
-
Pagina 35
T est results below 60 mg/dL or 3.3 mmol/L mean your blood glucose is low (hypogly cemia). IMPOR T ANT : If you get a HI DAN- GER result or a result belo w 60 mg/dL or 3.3 mmol/L and your test result does not reflect how y ou feel, follow the guidelines on pages 16-20 to ensure testing accuracy . Then repeat the test using a new test strip. Recor d[...]
-
Pagina 36
EVENT LABELLING Immediately after testing your blood, you may want to mark your glucose readings with event labels. The One T ouch ™ P rofile ™ meter has 15 sepa- rate event labels. 35 • FASTING • PRE BREAKFAST • AFTER BREAKFAST • PRE NOON MEAL • AFTER NOON MEAL • PRE DINNER • AFTER DINNER • DIFFERENT FOOD • BEDTIME • DURING[...]
-
Pagina 37
I t will appear on the meter display with your test result, alternating with the time and date. 36 T o select an event label, press the C but- ton while the blood glucose test result is still showing on the meter dis- play . Each time you press the C button, a new event label will appear on the display . Hold the C button down to mo ve more rapidly[...]
-
Pagina 38
ENTERING THE MEMOR Y T urn on the meter . P ress the M button to enter the meter memory . The word MEMOR Y will appear briefly . MEMORY MC The One T ouch ™ Profile ™ M eter stores up to 250 test results with date and time. E vent labels and insulin recor ds are also stored. When the memory is full, the oldest recor d is dropped as a new one is [...]
-
Pagina 39
The very first time you turn on your meter and press the M button, the fol- lowing message will appear , showing that there are no r esults stored in the memory: RECALLING RESUL TS T urn on the meter . P ress the M button to enter the memory . P ress the M button again and again to see each of your blood glucose, check strip, control solution and i[...]
-
Pagina 40
Y ou can change or add event labels or insulin information stored in memory while they are displayed on the meter . T o delete an insulin record, simply select ‘00’ for the units. See pages 35 on event labelling and pages 45-47 on the insulin prompt and insulin programming. 45 UNITS INTERM 8:05AM 1/06/00 39 Any event labels added to blood gluco[...]
-
Pagina 41
SET -UP OPTIONS MODE Y ou can change, or in some cases turn off, 11 features of your One T ouch ™ P rofile ™ M eter to suit your individual needs. • Beep prompts •T i m e • Date • Event av eraging • Insulin prompt • Insulin programming • Display language • T ime format • Date format • Unit of measur e • Decimal separator E[...]
-
Pagina 42
• U se the M button to scroll through the options or move within an option. • If you want to change an option set- ting, use the C button. • Once you have finished setting an op- tion, turn the meter off, or pr ess the M button to move to the next option. N ote: Y ou can exit the options mode at any time by turning off the meter . If after tw[...]
-
Pagina 43
BEEP PR OMPTS The One T ouch ™ Profile ™ M eter uses beeping tones to signal various steps in the meter procedures. The beep option is seen immediately after entering the set-up mode. The current set- ting will automatically appear on the display . With the “ON” setting, the meter will continue to provide beep pr ompts dur- ing the test pro[...]
-
Pagina 44
TIME AND DA TE In the set-up mode, use the M button to choose the Time option. The current time will show . Pr ess the C button to change the time; the hour and AM or PM will flash. P ress the C button again to advance the hour to the correct set- ting. N ow press the M button and the minutes will flash. U se the C button to change the minutes. P r[...]
-
Pagina 45
EVENT A VERAGING In the set-up mode, use the M button to choose the event averaging option. P ress the C button to turn on the event averag- ing option. ON AVG M-MOVE C-CHANGE OFF AVG M-MOVE C-CHANGE 44 A verages for each event are displayed by pressing the M button to mo ve from one event to the next. A verages for up to 15 separate events, and th[...]
-
Pagina 46
Step 1: Change the insulin type. When a type of insulin, SHOR T for example, is flashing on the meter , press the C button until your first type of insulin appears on the display . 00 UNITS SHORT M-MOVE C-CHANGE 00 UNITS INTERM M-MOVE C-CHANGE INSULIN PR OMPT Y ou can add information about your insulin type and dosage after each blood glucose test [...]
-
Pagina 47
Step 2: Change the insulin units. When the units area is flashing ‘00’, press the C button until your insulin dose is displayed. P ress the M button again and follow the same steps to enter your second type of insulin and the dose. 00 UNITS SHORT M-MOVE C-CHANGE INSULN SECOND TYPE 00 UNITS INTERM M-MOVE C-CHANGE 45 UNITS INTERM M-MOVE C-CHANGE [...]
-
Pagina 48
After you have mo ved through each area, the complete recor d will be dis- played for two seconds, follow ed by another insulin programming screen. Y ou may then press the C button to enter another insulin recor d. 47 CARBOHYDRA TE COUNTING Y ou can use the insulin programming feature to recor d the number of grams of carbohydrate in your diet. F o[...]
-
Pagina 49
The One T ouch Profile M eter is pre-set to prompt you in English. T o change the display language, enter the set-up mode. U se the M button to choose the display language option. P ress the C button until the language you want appears on the display . 48 DISPLA Y L ANGU AGE The One T ouch ™ Profile ™ M eter can display messages in 19 languages[...]
-
Pagina 50
DA TE FORMA T The meter is pre-set to display the date as month-day-year . F or example: J une 01, 2000 reads as 6-01-00. Y ou can set the meter to display day-month-year . T o change the date format, enter the options mode. U se the M button to choose the date format option. P ress the C button to change the date format. TIME FORMA T The meter is [...]
-
Pagina 51
DECIMAL SEP ARA T OR The meter is pre-set to display mil- limoles per litre (mmol/L) and basal insulin rate values using a decimal point. This option allows you to select a comma instead of a decimal point as a decimal separator . 0.0 mmol/L DEC PT M-MOVE C-CHANGE 0,0 mmol/L COMMA M-MOVE C-CHANGE 50 T o change the decimal separator , enter the set-[...]
-
Pagina 52
EXITING THE SET -UP MODE After you have completed viewing or changing your option settings, simply turn the meter off. The meter will auto- matically exit the set-up mode and turn itself off after two minutes. 51 UNIT OF MEASURE The meter is pre-set to display test results in milligrams per decilitre (mg/dL). Y ou have the option to select millimol[...]
-
Pagina 53
The display messages that appear on your One T ouch ™ Pr ofile ™ Meter will guide y ou through testing procedures, alert you to any pr oblems as they occur , and give you informa- tion about test results stored in the meter ’ s memor y . This section provides you with a com- plete listing of the display messages, what they mean, and what to d[...]
-
Pagina 54
45. . . . 0 106 MG/DL or 5.9 MMOL/L A blood test has been completed. This indicates a blood glucose (for example) resultwhen the meter is set for mg/dL or for mmol/L. 8:01 AM 6-01-00 100 MG/DL or 5.6 MMOL/L (for example) CONTRL 7:52 AM 6-01-00 INSERT SIDE 2 80 MG/DL or 4.4 MMOL/L (for example) ✓ OK 7:40 AM 6-01-00 53 A control solution test has b[...]
-
Pagina 55
60 MG/DL or 3.3 MMOL/L (for example) 8:01 AM 6-01-00 alternating with 60 MG/DL or 3.3 MMOL/L (for example) DO YOU NEED A SNACK 240 MG/DL or 13.3 MMOL/L (for example) 8:01 AM 6-01-00 alternating with 240 MG/DL or 13.3 MMOL/L CHECK KETONES 54 MESSAGE APPEARS WHEN Your blood glucose test result is 60 mg/dL (3.3 mmol/L).The message DO YOU NEED A SNACK [...]
-
Pagina 56
55 MESSAGE APPEARS WHEN HI MG/DL or HI MMOL/L DANGER 8:01 AM 6-01-00 alternating with HI MG/DL or HI MMOL/L CALL. DR. TEST KETONES Your blood glucose test result is 600 mg/dL (33.3 mmol/L). This level is severe hyperglycemia. Contact your doctor immediately. Appears briefly when the meter is turned on. You can see that all of the display segments a[...]
-
Pagina 57
When any of the following messages appear , there is a problem with y our One T ouch ™ P rofile ™ M eter or the way in which you are performing a test. In most cases, pr oblems are easy to fix. If you still hav e trouble, call your authorized LifeScan r epresentative. ERR OR MESSA GES PROBLEM When this symbol appears and stays on the screen, it[...]
-
Pagina 58
57 PROBLEM 1. There is dirt, blood, or lint on the test area. 2. Your hand or an object cov- ered the test area during testing. 3. The meter was used in very bright light. The test strip was inserted before the word INSERT appeared on the display. If the word CONTRL appears after a blood test, your blood sample was too small, smeared, or another dr[...]
-
Pagina 59
58 PROBLEM The sample was applied before the words APPLY SAMPLE appeared. 1. The test strip moved during the test. 2. The test strip was not insert- ed correctly. 3. The test strip was removed before the test was completed. 4. There was not enough blood on the test strip. 5. The meter was used in very bright light. WHAT TO DO Repeat the test with a[...]
-
Pagina 60
59 PROBLEM 6. The check strip procedure was incorrect. 7. The meter may not be operating correctly. 1. The blood or control solu- tion sample was too small or smeared. 2. Your hand or an object cov- ered the test area while the meter was turned on. 3. The test strip was not insert- ed far enough into the test strip holder. WHAT TO DO 6. Repeat the [...]
-
Pagina 61
60 PROBLEM 1. If appearing when you turn your meter on, your meter may have an electronic problem. 2. If appearing at the end of a test, your meter may have been moved while applying blood or during the test process. 3. If appearing during the test, the strip has been removed at the WAIT PLEASE screen. The check strip test result is outside the acc[...]
-
Pagina 62
61 PROBLEM The last check strip test was outside the acceptable range and an acceptable repeat test was not performed. The last test failed after the countdown started. The meter has lost some important information, including the code number. Some parts of the display are not working. WHAT TO DO Repeat the checkstrip test. Repeat the test. Use the [...]
-
Pagina 63
T o keep your One T ouch ™ Pr ofile ™ Blood Glucose M eter in good operating condition, follow these simple rules: • Keep the test strip holder and test area clean. • Keep your meter dry . If you acciden- tally get water or other liquids inside the meter , check the meter by doing check strip tests (see page 12) and control solution tests ([...]
-
Pagina 64
CLEANING THE METER DO A DAIL Y CHECK: Look through the small hole in the test strip holder to make sure there is no lint, dirt, or blood blocking it. If there is anything blocking the hole, or if the test area has dirt, lint, or blood on it, remo ve the test strip holder and clean the test area by follo wing the instructions given below . Then do a[...]
-
Pagina 65
Step 2: Clean the test strip holder . W ash the test strip holder with soap and water . Clean the underside of the test strip hold- er . U sing a cotton swab, clean the small hole to remo ve any dirt, blood, or lint. Rinse well. Dry completely with a soft cloth or tissue. Step 3: Clean the test ar ea. Check the clear , protective coating over the t[...]
-
Pagina 66
Rub the test area with a cotton swab or soft cloth dampened with water to remo ve all blood, dirt, or lint from the test area. If necessary , a mild liquid dish washing detergent mixed with water may also be used. Do not apply full- strength detergent to the test area. Be careful not to scratch the test area. Dry the test area with a soft, dry tiss[...]
-
Pagina 67
P ress forward on the base of the test strip holder to be sure it is properly in place. Y ou are now r eady to do a check strip test. If the check strip will not slide into the test strip holder , repeat S tep 4. REPLACING THE BA TTERIES The One T ouch ™ Profile ™ M eter comes with batteries already installed. When the batteries need to be repl[...]
-
Pagina 68
When BA TTR Y` REPLACE NO W appears flashing on the display , the meter will no longer give results and you must r eplace the batter- ies before you can perform another test. Dead batteries or battery removal will not affect the information stored in the meter memory . The time and date will be saved in the meter for approximately an hour without t[...]
-
Pagina 69
Step 3: R emove the old batteries. Remo ve the old batteries by gently slapping the meter against the palm of your hand. N ever tap the meter against a hard surface. Step 4: I nsert the new AAA batteries into the batter y compartment. Make sure that the plus (+) and minus (–) signs on the batteries are cor- rectly aligned. T o replace the battery[...]
-
Pagina 70
PERFORMANCE CHARA CTERISTICS Power Supply: 2 x 1.5V. alkaline batteries (AAA) Battery Life: Approximately one year when testing up to four times a day. Result Range: 0–600 mg/dL (0–33.3 mmol/L). Values of 600 mg/dL (33.3 mmol/L) and above displayed as HI . Display Type: Liquid crystal. Blood Source: Whole blood (capillary or venous). Do not use[...]
-
Pagina 71
W ARRANTY Five-Y ear W arranty If, at any time during the first fiv e years after purchase, the meter does not work for any reason (except for ob vious abuse), LifeScan will replace it with a new meter or equivalent product free of charge. The warranty policy applies only to the original purchaser of this meter and does not include the batteries su[...]
-
Pagina 72
SHORT Short insulin INTERM Intermediate insulin LONG Long insulin XLONG Extra long insulin 30/70 Mixture 30% S hort and 70% Intermediate and 50/50 Mixture 50% S hort and 50% Intermediate MIXTUR Mixture other than mixtur es identified OTHER Other insulin types not specified BOLUS Insulin pump setting, insulin taken over baseline amount CARBOH Carboh[...]
-
Pagina 73
1 995 The enclosed LifeScan Blood Glucose Meter, as an electro-medical device, has been thoroughly tested and found to comply with the essential requirements of the ElectroMagnetic Compatibility Directive (EMC 89/336/EEC), 1995.[...]
-
Pagina 74
For further information on LifeScan products, please call your authorized LifeScan representative. © LifeScan, Inc. 2000 Milpitas, California 95035, U .S.A. U.S . and foreign patents; other patents pending. IFD/OB/OTP Rev .: 02/2000[...]