HP (Hewlett-Packard) 5890 II Bedienungsanleitung

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244

Zur Seite of

Richtige Gebrauchsanleitung

Die Vorschriften verpflichten den Verkäufer zur Übertragung der Gebrauchsanleitung HP (Hewlett-Packard) 5890 II an den Erwerber, zusammen mit der Ware. Eine fehlende Anleitung oder falsche Informationen, die dem Verbraucher übertragen werden, bilden eine Grundlage für eine Reklamation aufgrund Unstimmigkeit des Geräts mit dem Vertrag. Rechtsmäßig lässt man das Anfügen einer Gebrauchsanleitung in anderer Form als Papierform zu, was letztens sehr oft genutzt wird, indem man eine grafische oder elektronische Anleitung von HP (Hewlett-Packard) 5890 II, sowie Anleitungsvideos für Nutzer beifügt. Die Bedingung ist, dass ihre Form leserlich und verständlich ist.

Was ist eine Gebrauchsanleitung?

Das Wort kommt vom lateinischen „instructio”, d.h. ordnen. Demnach kann man in der Anleitung HP (Hewlett-Packard) 5890 II die Beschreibung der Etappen der Vorgehensweisen finden. Das Ziel der Anleitung ist die Belehrung, Vereinfachung des Starts, der Nutzung des Geräts oder auch der Ausführung bestimmter Tätigkeiten. Die Anleitung ist eine Sammlung von Informationen über ein Gegenstand/eine Dienstleistung, ein Hinweis.

Leider widmen nicht viele Nutzer ihre Zeit der Gebrauchsanleitung HP (Hewlett-Packard) 5890 II. Eine gute Gebrauchsanleitung erlaubt nicht nur eine Reihe zusätzlicher Funktionen des gekauften Geräts kennenzulernen, sondern hilft dabei viele Fehler zu vermeiden.

Was sollte also eine ideale Gebrauchsanleitung beinhalten?

Die Gebrauchsanleitung HP (Hewlett-Packard) 5890 II sollte vor allem folgendes enthalten:
- Informationen über technische Daten des Geräts HP (Hewlett-Packard) 5890 II
- Den Namen des Produzenten und das Produktionsjahr des Geräts HP (Hewlett-Packard) 5890 II
- Grundsätze der Bedienung, Regulierung und Wartung des Geräts HP (Hewlett-Packard) 5890 II
- Sicherheitszeichen und Zertifikate, die die Übereinstimmung mit entsprechenden Normen bestätigen

Warum lesen wir keine Gebrauchsanleitungen?

Der Grund dafür ist die fehlende Zeit und die Sicherheit, was die bestimmten Funktionen der gekauften Geräte angeht. Leider ist das Anschließen und Starten von HP (Hewlett-Packard) 5890 II zu wenig. Eine Anleitung beinhaltet eine Reihe von Hinweisen bezüglich bestimmter Funktionen, Sicherheitsgrundsätze, Wartungsarten (sogar das, welche Mittel man benutzen sollte), eventueller Fehler von HP (Hewlett-Packard) 5890 II und Lösungsarten für Probleme, die während der Nutzung auftreten könnten. Immerhin kann man in der Gebrauchsanleitung die Kontaktnummer zum Service HP (Hewlett-Packard) finden, wenn die vorgeschlagenen Lösungen nicht wirksam sind. Aktuell erfreuen sich Anleitungen in Form von interessanten Animationen oder Videoanleitungen an Popularität, die den Nutzer besser ansprechen als eine Broschüre. Diese Art von Anleitung gibt garantiert, dass der Nutzer sich das ganze Video anschaut, ohne die spezifizierten und komplizierten technischen Beschreibungen von HP (Hewlett-Packard) 5890 II zu überspringen, wie es bei der Papierform passiert.

Warum sollte man Gebrauchsanleitungen lesen?

In der Gebrauchsanleitung finden wir vor allem die Antwort über den Bau sowie die Möglichkeiten des Geräts HP (Hewlett-Packard) 5890 II, über die Nutzung bestimmter Accessoires und eine Reihe von Informationen, die erlauben, jegliche Funktionen und Bequemlichkeiten zu nutzen.

Nach dem gelungenen Kauf des Geräts, sollte man einige Zeit für das Kennenlernen jedes Teils der Anleitung von HP (Hewlett-Packard) 5890 II widmen. Aktuell sind sie genau vorbereitet oder übersetzt, damit sie nicht nur verständlich für die Nutzer sind, aber auch ihre grundliegende Hilfs-Informations-Funktion erfüllen.

Inhaltsverzeichnis der Gebrauchsanleitungen

  • Seite 1

    Reference Manual HP 5890 Series II and HP 5890 Series II Plus[...]

  • Seite 2

    Little Falls Site Hewlett-Packard Company 2850 Centerville Road Wilmington, DE 19808-1610 Hewlett-Packard Company 1989, 1990, 1991, 1993, 1994 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws. HP part number 05890-90271 First edition—Jun 1989 Printed In U.S.[...]

  • Seite 3

    Contents Chapter 1 — Columns and Fittings 9 ................. Column oven 11 .......................................................... Column placement 12 ................................................. Packed column 12 .................................................... Hewlett•Packardcapillary columns 13 ..................................[...]

  • Seite 4

    Contents Chapter 4 — Electronic Flow Sensing 57 ............... Displaying gas flow rate 58 ............................................... Designating gas type 59 .................................................. Electronic flow sensor (EFS) calibration 60 ................................. Preparation 61 ........................................[...]

  • Seite 5

    Contents Chapter 6 — Inlet Systems 99 ......................... Packed column inlet 100 ................................................... Electronic flow sensor 102 .............................................. Septum•purgedpacked column inlet 103 ................................. Problems at high inlet temperatures 104 .....................[...]

  • Seite 6

    Contents Chapter 8 — Preventive Maintenance 155 .............. Conditioning columns 156 .................................................. (Re)Packing columns 158 .................................................. Packed column inlet 159 ................................................... Changing septa 159 .......................................[...]

  • Seite 7

    Contents Chapter 9 — Chromatographic Troubleshooting 201 ...... Introduction 202 .......................................................... Baseline symptoms 202 .................................................... Position 202 ........................................................... W ander and drift 203 ......................................[...]

  • Seite 8

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 9

    1 Columns and Fittings[...]

  • Seite 10

    10 Columns and Fittings The HP 5890 SERIES II (hereafter referred to as HP 5890) provides flexibility in choices among inlets, columns, and detectors through use of liners and adapters, allowing any standard column to be used without sacrificing performance. Additional flexibility is gained through positions of inlets and detectors relative to each[...]

  • Seite 11

    Columns and Fittings Column oven 11 Column oven Inlet Ftg Det Ftg Nut Plate Figure 1-1 The Column Oven The oven door latch, located beneath the lower right corner of the door , is pressed upward to open the door . Motor •driven flaps at the rear of the oven admit room air for cool down or near •ambientoperation, so the door is kept closed excep[...]

  • Seite 12

    Columns and Fittings Column oven 12 Column placement Generally , a column may be installed between any inlet and detector . A rigid 1/4•inchpacked glass column, however , if installed in the B (rear •most)inlet, can only be installed in the B (rear•most) detector . Distance relationships among inlets and detectors are shown in Figure 1•2. T[...]

  • Seite 13

    Columns and Fittings Column oven 13 Hewlett-Packard capillary columns Hewlett•Packardcapillary columns are wound on wire frames which mount on a pair of brackets which slip into slots at the top of the oven interior . Figure 1-3 Typical Hewlett-Packard Capillary Columns[...]

  • Seite 14

    Columns and Fittings Fittings 14 Column Hanger Part No. 1460-1914 Column Installed Installed Bracket for Hewlett-Packard Capillary Columns Figure 1-4. The bracket has two positions from which to hang the column wire frame. Depending upon frame diameter , use the position which best centers the column in the oven. Column ends should come off the bot[...]

  • Seite 15

    Columns and Fittings Fittings 15 Graphite O•ringsor ferrules have excellent sealing quality and long service life, can be used continuously to 400 C, and are generally recommended for most applications, particularly capillary and glass columns. They are also recommended for inlet and detector liners, and for split/splitless capillary inlet insert[...]

  • Seite 16

    Columns and Fittings Fittings 16 Type Description Typical Use Part No. 1/4-inch swage, nut 1/4-inch packed metal columns 5080-8753 stainless steel, front ferrule pkg, 20 of each back ferrule 1/8-inch swage, nut 1/8-inch packed metal columns 5080-8751 stainless steel, front ferrule pkg, 20 of each back ferrule 1/4-inch swage, nut 1/4-inch packed met[...]

  • Seite 17

    Columns and Fittings Liners/adapters and inserts, general 17 Liners/adapters and inserts, general A liner/adapter is installed from below , inside the oven; it serves both as an adapter to mate the particular column to the inlet or detector and to provide correct internal volume for proper operation. Inserts are used with inlets only , and, when re[...]

  • Seite 18

    Columns and Fittings Liners/adapters and inserts, general 18 1/8-inch Metal 1/4-inch Metal 1/4-inch Glass Recommended 1/8-inch 1/4-inch 1/4-inch swage- Column Fittings swage-type nut swage-type nut type nut and and ferrules 3 and ferrules 3 graphic ferrule or silicone O-ring(s) Packed Column 19243-80510 1 19243-80520 1 None Inlet Liners or or 19243[...]

  • Seite 19

    Columns and Fittings Liners/adapters and inserts, general 19 HP Series 530 320 m ID 200 m ID Metal/ Glass Recommended Capillary column Capillary column Same as 320 m Same as Column Fittings nut and 1.0-mm nut and 0.5-mm HP Series graphite ferrule, or graphite or 530 silicone O-ring(s) silicone O-ring(s) Packed Column 19244-80540 1 Not Not Not Inlet[...]

  • Seite 20

    Columns and Fittings Inlet/detector liners/adapters 20 Inlet/detector liners/adapters Interchangeable stainless steel liners/adapters, installed from inside the oven, are used with the packed column inlet, and with all detectors, depending upon the column to be installed. Packed column inlet liners Liner Installed Liner, Packed Column Inlet Figure [...]

  • Seite 21

    Columns and Fittings Inlet/detector liners/adapters 21 In addition, liners for the packed column inlet are available to accept glass inserts (discussed later) for reduced reactivity , to trap nonvolatile residues, or for use with an HP Series 530 capillary column. No liner is used with 1/4•inchpacked glass columns. The long leg of the column fits[...]

  • Seite 22

    Columns and Fittings Inlet/detector liners/adapters 22 Detector liners/adapters Liner/Adapter Typical Installed Detector Liner/Adapter Figure 1-6 Detectors require a liner/adapter to be installed when used with packed metal columns (either 1/8• or 1/4•inch),and with any type of capillary column. Normally , no liner is required with 1/4•inchpa[...]

  • Seite 23

    Columns and Fittings Inlet/detector liners/adapters 23 ECD and TCD adapters A makeup gas adapter must be installed in the ECD or TCD base to install a capillary column, and to augment carrier flow through the column with additional gas flow needed for optimal detector operation. The adapter must be removed for packed column applications. In additio[...]

  • Seite 24

    Columns and Fittings Inlet/detector liners/adapters 24 Liner/adapter installation 1/4-inch Ferrule Liner Liner Retainer Nut 1-mm Graphite Ferrule Capillary Column Nut Packed Column Inlet Liner for HP Series 530 Capillary Column Use Nut and Ferrule Installed on a Liner/Adapter Figure 1-7 W ith one exception, liners/adapters are installed in the same[...]

  • Seite 25

    Columns and Fittings Inlet inserts 25 1. Assemble a brass nut and graphite ferrule onto the liner/adapter . 2. Insert the liner/adapter straight into the detector base as far as possible. 3. Holding the liner/adapter in this position, tighten the nut finger •tight. 4. Use a wrench to tighten the nut an additional 1/4 turn. 5. Install the column; [...]

  • Seite 26

    Columns and Fittings Inlet inserts 26 Exercise care! the oven, and/or inlet, or detector fittings may be hot enough to cause burns. WARNING Flared End Insert Installing a Glass Insert in a Packed Column Inlet Figure 1-9 1. In handling the insert, avoid contaminating its surface (particularly its interior). 2. Remove the septum retainer nut and sept[...]

  • Seite 27

    Columns and Fittings Inlet inserts 27 Note: For the liner and insert for an HP Series 530 capillary column, if the column is already installed, a new insert may not seat properly in the liner; the column may prevent it from dropping completely into the liner . If the insert does not drop completely into the liner , do not force it (either the liner[...]

  • Seite 28

    Columns and Fittings Inlet inserts 28 The split insert contains packing material (10% OV •1on 80/100 High Performance Chromosorb•W),held in place by silanized glass wool plugs, located immediately above a mixing chamber . This ensures proper volatilization and homogeneous mixing of the sample prior to its entry into the column. Exercise care! T[...]

  • Seite 29

    Columns and Fittings Inlet inserts 29 Installation, Split/Splitless Capillary Inlet Insert Figure 1-11 3. Using tweezers, forceps, or similar tool, remove any insert already in place. 4. Inspect the new insert to be installed: For a split mode insert, the end with the mixing chamber and packing is inserted first into the inlet. 5. Place a graphite [...]

  • Seite 30

    Columns and Fittings Jet replacement, FIDs or NPDs 30 Jet replacement, FIDs or NPDs Depending upon the column type (packed versus capillary) to be used, and/or analyses to be performed, exchanging the jet in an FID or NPD may be necessary . This must be done prior to column installation, and is particularly important in optimizing FID performance. [...]

  • Seite 31

    2 Keyboard and Displays[...]

  • Seite 32

    32 Keyboard and Displays Alphanumeric Display Instrument Status Run Control Temperature Control Miscellaneous Functions Numeric and Modifier Keys Signal Definition and Control Oven Status SIG 2 ZERO TCD SENS DET SIG 1 B A OFF ON ENTER 7 8 9 6 5 1 2 3 - 0 . 4 CLEAR VALVE FLOW TIME FLOW CRYO PARAM PARAM PURGE FINAL VALUE INJ A TEMP INJ B TEMP DET A T[...]

  • Seite 33

    Keyboard and Displays Displaying setpoints 33 HP 5890 SERIES II (hereafter referred to as HP 5890) operation is monitored and controlled through its front panel keyboard, and alphanumeric and LED displays. Some instrument functions are monitored continuously: signal levels, temperatures, carrier gas flow rates (if electronic flow sensing is install[...]

  • Seite 34

    Keyboard and Displays Entering setpoints 34 Examples of possible displays are provided where appropriate throughout the manual. If a particular function is not installed in your instrument, an appropriate message is displayed when the key corresponding to the function is pressed. For example, if no heated zone controlled by DET B TEMP is installed,[...]

  • Seite 35

    Keyboard and Displays Entering setpoints 35 T o display the function and its setpoint: necessary for a few instrument functions (Instrument Function Key) ( or ) A B then, EITHER ( through , , ) 0 9 - . to enter a new setpoint value OR, for a few functions, (o r) to switch the function on or off ON OFF ENTER Steps in Entering a Setpoint Value Figure[...]

  • Seite 36

    Keyboard and Displays Entering setpoints 36 CLEAR can be used anytime during an entry , prior to pressing ENTER ,t o erase the entry in progress. The * disappears, and the original setpoint display is restored. Rules regarding keyboard usage are summarized below: An instrument function key , when pressed, is shown in the display along with its curr[...]

  • Seite 37

    Keyboard and Displays Keyboard operation, INET control 37 CLEAR is used anytime during setpoint entry , prior to pressing ENTER , to erase the entry in progress. CLEAR , if pressed when no setpoint entry is in progress, displays HP 5890 readiness . Run Control Key START , if pressed while a setpoint entry is in progress, causes the entry to be abor[...]

  • Seite 38

    Keyboard and Displays Protecting setpoints 38 Additional information regarding INET control is available in Chapter 5, Signal Output . Servicing may be required for one or more devices on the INET loop if communication cannot be established. Protecting setpoints The HP 5890 provides a keyboard lock feature to minimize possibility of stored setpoint[...]

  • Seite 39

    Keyboard and Displays Loading default setpoints 39 W ith the keyboard locked, Figure 2•6 shows the display occurring if a setpoint entry is attempted: ACTUAL SETPOINT KEYBOARD LOCKED KEYBOARD LOCKED Message Display Figure 2-6 If the HP 5890 keyboard is locked while the instrument is under INET control, a setpoint file may be loaded into HP 5890 m[...]

  • Seite 40

    Keyboard and Displays Loading default setpoints 40 Upon pressing ENTER , default setpoints are loaded into memory , erasing setpoints already present. T able 2•1 lists resulting HP 5890 default setpoints. Function Default Setpoint Inj Temp (A & B): 50 C, OFF Det Temp (A & B): 50 C, OFF Oven Temp: 50 C, OFF Oven Max Temp: 400 C Cryo Coolin[...]

  • Seite 41

    Keyboard and Displays Loading default setpoints 41 Note that if the battery protecting memory should fail when main power is turned off, the default setpoints are loaded into memory when the battery is replaced. In addition, calibration constants for oven temperature control and gas flow rate monitoring are also reset to default values.[...]

  • Seite 42

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 43

    3 Temperature Control[...]

  • Seite 44

    44 Temperature Control Oven temperature, and temperatures of up to five separate heated zones (detectors, inlets, and/or heated valves), are controlled through keys shown in Figure 3•1. Oven Control Heated Zone Control INIT VALUE FINAL VALUE FINAL TIME INJ A TEMP INJ B TEMP DET A TEMP OVEN MAX EQUIB TIME OVEN TEMP INIT TIME RATE DET B TEMP AUX TE[...]

  • Seite 45

    Temperature Control 45 Note that the ACTUAL value is a measured quantity , while the SETPOINT value is user •defined: in this example, the setpoint value for oven temperature might recently have been changed from 250 to 350 C, and the oven is now heating to the new setpoint. Given sufficient time for equilibration, ACTUAL and SETPOINT values beco[...]

  • Seite 46

    Temperature Control Valid setpoint ranges 46 Valid setpoint ranges T able 3•1 lists valid setpoint ranges for the 13 keys controlling oven and heated zone temperatures. NOTE: TOTAL run time will not exceed 650.00 minutes regardless of values enter e , , and . INIT TIME RATE FINAL TIME -80 to 450 1 C Oven Control -80 to 450 1 C Oven Control 0 to 6[...]

  • Seite 47

    Temperature Control Cryogenic (sub-ambient) oven control 47 Cryogenic (sub-ambient) oven control Liquid N or liquid CO cryogenic options are for operation at temperatures less than about 7 C above ambient. This is done through operation of a valve which opens when coolant is demanded and closes when the setpoint temperature is reached. When you pre[...]

  • Seite 48

    Temperature Control Cryogenic (sub-ambient) oven control 48 Oven profile using CRYO, for operation during runs at subambient temperatures 25 50 75 (CRYO ON) CRYO OFF at ambient +15 CRYO ON at ambient + 25 Figure 3-4 Oven profile using CRYO BLAST, for very fast cool down between runs 40 80 120 CRYO BLAST OFF (30 sec. modulation) CRYO BLAST ON ambien[...]

  • Seite 49

    Temperature Control Programming oven temperature 49 Programming oven temperature HP 5890 oven temperature programming allows up to three ramps, in any combination of heating or cooling. Keys defining an oven temperature program include: INIT TEMP A setpoint temperature value at which the oven is maintained at the beginning of a temperature•progra[...]

  • Seite 50

    Temperature Control Oven status 50 In isothermal operation ( RATE = 0 ), if INIT TIME is set equal to 0 (zero), the HP 5890 internally sets run time to the maximum, 650 minutes. A is included in key sequences defining parameters for a second ramp; B is included in key sequences defining parameters for a third ramp. In isothermal operation, and in o[...]

  • Seite 51

    Temperature Control Oven safety 51 In complex two•or three•rampoven temperature programs, information as to the part of the program in progress is monitored by pressing OVEN TEMP . Note that, during a ramp, the SETPOINT value displayed is that calculated to be the correct temperature, based upon specified heating/cooling rate, and initial and f[...]

  • Seite 52

    Temperature Control Fault: messages 52 The message displayed when this occurs is shown in Figure 3•6. ACTUAL SETPOINT WARN: OVEN SHUT OFF Message, Oven SHUT DOWN Figure 3-6 The oven remains off until switched on again via the keyboard ( OVEN TEMP ON ), unless a FAULT: message is displayed (see below , Fault: messages ). Power to the instrument mu[...]

  • Seite 53

    Temperature Control After a power failure . . . 53 ACTUAL SETPOINT FAULT: OVEN > MAX+20 ACTUAL SETPOINT FAULT: OVEN TEMP RDG ACTUAL SETPOINT FAULT: DETA TEMP RDG ACTUAL SETPOINT FAULT: INJA TEMP RDG ACTUAL SETPOINT FAULT: LINE SENSE ACTUAL SETPOINT FAULT: ADC OFFSET Thermal Control FAULT: Messages Figure 3-7 In addition to the message, the red N[...]

  • Seite 54

    Temperature Control Oven temperature calibration 54 INITIAL TIME RATE OVEN FINAL TIME STATUS RUN NOT READY ACTUAL SETPOINT PASSED SELF TEST Message Display, Power Failure and Recovery Figure 3-8 Heated zones return to their respective setpoint values, after which the oven returns to its setpoint value. If OVEN TEMP is displayed after recovery from [...]

  • Seite 55

    Temperature Control Oven temperature calibration 55 The HP 5890 provides the means to (if necessary) reset oven temperature monitoring so the displayed ACTUAL value accurately represents the correct temperature. Oven temperature calibration requires entering the difference ( delta ) value (in C) between an independently measured temperature value v[...]

  • Seite 56

    Temperature Control Oven temperature calibration 56 3. CALIB DELTA is displayed until ENTER is pressed; then oven temperature recalibration occurs. Note that, after calibration, the displayed oven temperature value should match closely the measured value. Any delta value within the range -10.00 through +10.00 C may be entered. If a value outside th[...]

  • Seite 57

    4 Electronic Flow Sensing[...]

  • Seite 58

    58 Electronic Flow Sensing T wo channels of electronic flow rate sensing continuously monitor gas flow rates (usually carrier) in the HP 5890 SERIES II. Proper scaling of displayed values for different commonly used gases is defined through keyboard entries. The two flow channels are distinguished through A and B . If carrier gas flows are monitore[...]

  • Seite 59

    Electronic Flow Sensing Designating gas type 59 Designating gas type T o scale the displayed flow rate value properly , one of four commonly used gases must be designated. The appropriate gas type is selected according to T able 4•1: 1 He (Helium) TCD 2N 2 (Nitrogen) General 3H 2 (Hydrogen) Capillary 4 Ar/CH 4 (Methane in Argon) ECD Number Gas Ty[...]

  • Seite 60

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 60 Electronic flow sensor (EFS) calibration Electronic flow sensor (EFS) calibration may be performed any time to ensure displayed flow rate accurately represents real gas flow rate through the sensor . The EFS is factory•calibrated for four standard gases, H , He, N , and Ar/CH , w[...]

  • Seite 61

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 61 Preparation 1. Access the EFS by removing the left side panel; remove two screws along its lower edge, slide the panel toward the rear of the instrument, and then lift. 2. Through the keyboard, select CALIB AND TEST mode, function 2: CLEAR . 2 ENTER GAIN A is displayed, followed by[...]

  • Seite 62

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 62 3. Locate the EFS module and note its labelling: CHANNEL A/ CHANNEL B, IN/OUT . For the channel being calibrated, locate and disconnect its OUT fitting; use two wrenches in opposition to prevent twisting the tubes. Outlet Line, Channel B EFS Module Outlet Line, Channel A Detail, El[...]

  • Seite 63

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 63 EFS Flow-Measuring Adapter (Part No. 05890-80620) Figure 4-3 5. Assuming there is no gas flow through the channel being calibrated, press ENTER at the keyboard. This updates the zero calibration value. Setting the GAIN calibration value After the zero calibration value is set at ze[...]

  • Seite 64

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 64 Note: The HP 5890 has a timer function that may be used as an aid in measuring flow rate (see the Operating Manual , Chapter 4). Press TIME to access the timer function. After obtaining the desired flow rate, press: CLEAR . 2 ENTER to return to setting the gain value. EFS channel A[...]

  • Seite 65

    Electronic Flow Sensing Electronic flow sensor (EFS) calibration 65 Entering specific ZERO and GAIN values Calibration values for zero and gain should be recorded when a particular channel is calibrated. They can then be reentered through the keyboard if necessary , without repeating the entire calibration procedure. T o enter specific zero and gai[...]

  • Seite 66

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 67

    5 Signal Output[...]

  • Seite 68

    68 Signal Output A standard signal channel, controlled via SIG 1 , always is provided. A second signal channel, controlled via SIG 2 , is provided if Option 550/ Accessory 19242A ( Communications Interface Board ), or Option 560/ Accessory 19254A ( RS•232 ), is installed. Output sources include detector signal(s), heated zone or oven temperatures[...]

  • Seite 69

    Signal Output Zeroing signal output 69 The function of ZERO is to subtract a constant background signal from the detector signal. Background signal sources include the detector itself (background level depending upon detector type), column bleed, or contaminants in supply gas(es). Displaying current ZERO setpoint Current ZERO setpoint value is disp[...]

  • Seite 70

    Signal Output Zeroing signal output 70 Self- ZERO setpoint Referencing Figure 5•2for the +1 V output, note that using ZERO can increase dynamic range available for signal output by shifting an existing constant offset signal to a lower level (usually electrical zero). There are limits to this, however , so it is good practice to have background r[...]

  • Seite 71

    Signal Output Zeroing signal output 71 Figure 5-3 1.0 mV maximum output level + 1.000 mV + 0.100 mV + 0.006 mV 0m V Constant 0.1 mV detector background HP 5890 SERIES II electrical zero signal 0.9 mV usable dynamic range dynamic range 1.0 mV usable O U T P U V O L T A G E T 1 mV Output: Canceling Baseline Offset (the self- function) pressed ENTER Z[...]

  • Seite 72

    Signal Output Signal attenuation 72 Note: If a self• ZERO determination is performed on an active signal exceeding the maximum permitted setpoint value for ZERO (see User•defined ZERO setpoint ), the maximum setpoint value is assigned and the message SIG 1 (or 2 ) ZERO TOO HIGH is displayed. User-defined ZERO setpoint If the self• ZERO setpoi[...]

  • Seite 73

    Signal Output Signal attenuation 73 Thus, signal output level at the +1 mV analog output may be set separately from that at the +1 V output. T able 5•2 gives values permitted for either function, and the output affected. Permitted Affected Output 0 to 13 Both +1 mV & +1 V 0 TO 10 Only +1 mV , Setpoints Key RANGE 2 () ATTN 2 () ON OFF Table 5-[...]

  • Seite 74

    Signal Output Signal attenuation 74 For analytical information from a detector , proper settings for RANGE 2 () and ATTN 2 () are determined such that peaks of interest are on scale at the integrator or chart recorder: peaks of interest must neither flat top by exceeding the allowed maximum output level, nor be too small to be measured. T able 5•[...]

  • Seite 75

    Signal Output Signal attenuation 75 From T able 5•3, note that for a TCD, RANGE 2 () = 0 is suitable for virtually all applications since the entire linear output range of the detector is included. Likewise, RANGE 2 () settings from 0 through 5 cover the entire useful output range for an ECD. Only an FID or NPD may require use of the higher RANGE[...]

  • Seite 76

    Signal Output Signal attenuation 76 Note that if RANGE 2 () or ATTN 2 () is pressed without first pressing SIG 1 or SIG 2 , SIGNAL 1 channel is assumed (and displayed). If desired, SIG 2 can then be pressed to display the same function for the SIGNAL 2 channel. Entering RANGE 2 () / ATTN 2 () setpoints A new setpoint value is entered for either RAN[...]

  • Seite 77

    Signal Output Test signal output 77 Test signal output A test chromatogram, consisting of three peaks, is permanently stored in the HP 5890. Each peak is approximately 1/10 the height of the previous peak, with the first (tallest) peak having a height value of about 125 mV at RANGE 2 () =0( +1 V analog output); half•heightwidth of this peak is ab[...]

  • Seite 78

    Signal Output Test signal output 78 T o access this function, the following key sequence is entered: SIG 1 ( or SIG 2 ) 9 ENTER T est plot mode is confirmed by the display SIGNAL 1 (or 2 ) TEST PLOT . Pressing SIG 1 (or SIG 2 ) a second time displays the current signal level value (which is 0.0 initially). This permits monitoring the output signal.[...]

  • Seite 79

    Signal Output Instrument network (INET) 79 Instrument network (INET) The Instrument Network (INET) is a path for various devices to communicate with each other (data and/or commands). INET permits a group of devices (consisting of a controller , and some number of data Producers and data Consumers ) to function as a single, unified system. INET per[...]

  • Seite 80

    Signal Output Instrument network (INET) 80 Figure 5-6 OUT IN IN OUT OUT IN Sampler S/ECM Controller & Integrator 5890 HP 5890 SERIES II Gas Chromatograph Typical INET Loop Each INET must have one (and only one) device defined as the controller . The controller is responsible for network configuration when the network is first connected and powe[...]

  • Seite 81

    Signal Output Instrument network (INET) 81 configuration, consult appropriate manual(s) for the controller device (the HP 5890 is never a controller ). An instrument Addresses An instrument is a device, housing together a collection of functions, and having a single model number . It has a single pair ( IN and OUT )o f INET cable connections. The I[...]

  • Seite 82

    Signal Output Instrument network (INET) 82 Except for the controller , each instrument handles setpoints for instrument(s) other than itself only as blocks of data to be transmitted, but not altered. Active workspace Each device in an INET loop provides storage area for its own setpoints and parameters. These individual storage areas (each containi[...]

  • Seite 83

    Signal Output Instrument network (INET) 83 INET operation In using the INET function, chromatographic parameters are entered normally through the HP 5890 keyboard. Integration parameters are entered at the controller . Parameters for other devices on the INET loop may be entered at the controller , or at their own keyboards. Collectively , the sepa[...]

  • Seite 84

    Signal Output Instrument network (INET) 84 If a setpoint entry at the HP 5890 keyboard is in progress when a workfile or method is stored or listed at the controller , the entry is aborted. After the operation finishes, the HP 5890 returns to the same setpoint display . When a stored workfile or method is recalled to active workspace at the control[...]

  • Seite 85

    Signal Output INET configuration 85 Automatic INET reconfiguration In the following circumstances, INET automatically reconfigures under direction of the controller: Recovery from a power failure. Recovery from any particular device on the loop being switched off, then on again. Recovery from a disconnected (then reconnected) loop cable. Consult ap[...]

  • Seite 86

    Signal Output INET configuration 86 Figure 5•8shows displays resulting from the key sequence: CLEAR . 3 ENTER Switching between Global and Local W ith regard to the INET function at the HP 5890, there are two operating modes: global or local .I n global mode (default mode), HP 5890 START and STOP keys, when pressed, affect other devices on the IN[...]

  • Seite 87

    Signal Output INET configuration 87 Note that global mode has two states: if GLOBAL flashes (default mode) when displayed, the HP 5890 is in global mode, but not configured into the INET system. When the HP 5890 is properly configured into the INET system, GLOBAL is displayed continuously . This feature provides a convenient diagnostic to determine[...]

  • Seite 88

    Signal Output INET configuration 88 The specific number shown depends upon how INET cables are connected among devices included in the loop. The value shown in the example ( 8 ) implies the HP 5890 is the first instrument on the loop, starting from the OUT receptacle on the controller device (the controller is always defined as 0 ). A 9 indicates t[...]

  • Seite 89

    Signal Output INET configuration 89 Figure 5•1 1 shows resulting displays. INET-HP 5890 Signal Definition ACTUAL SETPOINT GLOBAL ADDR: 8,31 ACTUAL SETPOINT SIG 2 OFF ACTUAL SETPOINT SIG 1 ON FULL RANGE ACTUAL SETPOINT SIG 1 ON RANGED INET Signal Definition Displays Figure 5-11 From the displays, the following may be noted: HP 5890 signal channels[...]

  • Seite 90

    Signal Output HP-IL loopback test 90 RANGED versus FULL RANGE indicates the dynamic range for the data to be transmitted to other devices on the loop; dynamic range for RANGED data is set at the HP 5890 according to the setpoint for RANGE 2 () . Dynamic range for FULL RANGE data is limited only by the detector itself. The choice of the type of data[...]

  • Seite 91

    Signal Output HP-IL loopback test 91 ACTUAL SETPOINT PASSED SELF TEST ACTUAL SETPOINT FAILED SELF TEST ACTUAL SETPOINT HPIL LOOPBACK TEST HPIL LOOPBACK TEST Displays Figure 5-12 The message PASSED SELF TEST indicates INET , at least with respect to the HP 5890, is performing satisfactorily . If FAILED SELF-TEST is displayed, a bad cable may be indi[...]

  • Seite 92

    Signal Output Warn: and fault: messages 92 Warn: and fault: messages ACTUAL SETPOINT FAULT: INET CPU RAM ACTUAL SETPOINT FAULT: INET CPU ACTUAL SETPOINT FAULT: INET RAM TEST ACTUAL SETPOINT FAULT: INET ROM TEST ACTUAL SETPOINT FAULT: ATTN1 TEST ACTUAL SETPOINT FAULT: DAO1 TEST ACTUAL SETPOINT WARN: SIGNAL CHANGED ACTUAL SETPOINT WARN: NO DETECTORS [...]

  • Seite 93

    Signal Output Warn: and fault: messages 93 WARN: SIGNAL CHANGED and/or WARN: NO DETECTORS is displayed if a detector previously assigned to a particular signal channel is found to be absent; for example, if the signal board for a given detector should fail or be removed for service. FAULT: INET CPU is displayed if the processor (and/or its associat[...]

  • Seite 94

    Signal Output File compatibility with data handling devices 94 File compatibility with data handling devices Y ou must have the HP 5890 SERIES II in the proper mode for file compatibility with your data handling device. What are the modes? There are 2 file transfer modes: HP 5890A and HP 5890 SERIES II. HP 5890A File Transfer Mode HP 5890 SERIES II[...]

  • Seite 95

    Signal Output File compatibility with data handling devices 95 ACTUAL SETPOINT HP 5890A mode ACTUAL SETPOINT HP 5890 SERIES II mode EMULATION MODE OK PASSED SELF TEST GC Displays for File Transfer Modes Figure 5-14 How do I change modes? 1. T urn power off . 2. Remove the GC side panel, and locate the main PC board. Main PC Board Top Hinge for Grou[...]

  • Seite 96

    Signal Output File compatibility with data handling devices 96 P15 Main PC Board P13 P12 P5 P6 P2 P3 Finding component P15 on the Main PC Board. Figure 5-16 4. Set the jumper (Part No. 1258•0141) for the proper mode. T o avoid electrostatic damage to the main board, ground yourself to the GC chassis with an ESD strap, or touch an unpainted area o[...]

  • Seite 97

    Signal Output File compatibility with data handling devices 97 How to convert HP 339X Integrator workfiles from 5890A to SERIES II mode: 1. T urn the GC off . 2. Follow the previous instructions to set the GC for 5890A mode (use proper grounding). 3. Download the workfile from the integrator . 4. T urn GC power off . 5. Remove the P15 jumper . (Now[...]

  • Seite 98

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 99

    6 Inlet Systems[...]

  • Seite 100

    100 Inlet Systems This chapter provides information for the following HP 5890 SERIES II (hereafter referred to as HP 5890) inlet systems: Packed column inlet Septum•purged packed column inlet Split/splitless capillary inlet For cool on•columninformation, see the manual Programmable Cool On•ColumnInlet. Maintenance information is provided in C[...]

  • Seite 101

    Inlet Systems Packed column inlet 101 Septum Liner Glass Insert Carrier Gas Column Septum Retainer Nut Graphite Ferrule Swage-type Nut and Ferrules Packed Column Inlet Figure 6-1[...]

  • Seite 102

    Inlet Systems Packed column inlet 102 Trap(s) External Plumbing Internal Plumbing Pressure Gauge Packed Column Inlet Electronic Flow Sensor (optional) Mass Flow Controller Column To Detector Carrier Gas Flow Diagram, Packed Column Inlet (with electronic flow sensor) Figure 6-2 Liquid sample is rapidly volatilized inside the inlet. T o ensure comple[...]

  • Seite 103

    Inlet Systems Packed column inlet 103 Assuming the system to be leak•free(and if total flow is < 200 ml/min), after setting the desired column flow rate, total flow through the system should be noted via the EFS. The original column flow rate is reestablished simply by adjusting the mass flow controller so the original total system flow rate v[...]

  • Seite 104

    Inlet Systems Packed column inlet 104 Problems at high inlet temperatures A common problem with conventional packed column inlets operated at high temperatures is septum bleed and the associated ghost peaks. T o minimize this effect, some inlet systems are designed with steep temperature gradients throughout the entire upper length of the inlet to [...]

  • Seite 105

    Inlet Systems Packed column inlet 105 10 20 30 40 60 70 80 90 50 50 100 150 200 250 300 350 400 Bottom of Sep- tum Syringe Tip Base of Injection Port Temperature in Gas Stream — C 35 C Ove n 150 C Oven 300 C Oven Injection Port Setpoint Temperature 350 C Thermal Profiles Figure 6-4 This optimized thermal profile allows very reproducible results a[...]

  • Seite 106

    Inlet Systems Packed column inlet 106 When operating the inlet with septum purge, low bleed septa are unnecessary and the selection of septa should be made primarily for good sealing and extended septa life reasons. On a periodic basis (every 1 to 2 months), the T eflon•coatedO•ring sealing the purge cavity should be replaced. Electronic flow s[...]

  • Seite 107

    Inlet Systems Split/splitless capillary inlet 107 Split/splitless capillary inlet 1 23 Insert 1, 2, or 3 Sealing O-Ring Split Direct Injection Splitless 1/4” Packed Glass Column Packed Metal Column Sealing O-Ring 4 A. CAPILLARY COLUMN B. PACKED COLUMN Item Part No. Description Number 1 Split Insert (packed) 18740-60840 1 Split Insert (unpacked) 1[...]

  • Seite 108

    Inlet Systems Split/splitless capillary inlet 108 The multiple•mode split/splitless capillary inlet system may be used with any of the common types of capillary columns (fused silica, quartz, glass, metal). Specific sampling modes include: Split, for major•componentanalyses. Purged splitless, for trace•component analyses. Each mode requires i[...]

  • Seite 109

    Inlet Systems Split/splitless capillary inlet 109 In general, the carrier gas is chosen to maximize component resolution and detector performance while minimizing overall analysis time. Figure 6•7, a family of van Deemter curves for common carrier gases, illustrates the effect of gas choice and linear velocity ( ) on column efficiency (HETP , Hei[...]

  • Seite 110

    Inlet Systems Split/splitless capillary inlet 110 V an Deemter curves demonstrate advantages of using either He or H as carrier gas. From the curves, several observations may be made: Minima for He and H 2 occur at much higher average linear velocities than N 2 . Thus, He, or even better , H 2 , can be used at far higher velocities than N 2 with on[...]

  • Seite 111

    Inlet Systems Split/splitless capillary inlet 111 Nominal ID (mm) 12 25 50 0.20 135 223 347 0.32 45 82 137 0.53 11 23 42 Nominal Length (m) It must be emphasized that values in this table are recommended as starting points only ! Values listed are independent of carrier gas used. Table 6-1. Suggested Initial Column Pressures (kPa) for Various Capil[...]

  • Seite 112

    Inlet Systems Split/splitless capillary inlet 112 IN OUT IN OUT GA External Plumbing Internal Plumbing Trap(s) Carrier Gas Mass Flow Controller Electronic Flow Sensor (optional) To Detector Solenoid Valve Back- pressure Regulator Split Vent Septum Purge Vent Pressure Gauge Septum Purge Control Capillary Inlet Column (C) (P) (S) COM N.O N.C Flow Dia[...]

  • Seite 113

    Inlet Systems Split/splitless capillary inlet 113 The split ratio is an indicator of the fraction of total sample entering the column: the higher the value, the less sample enters the column. For setting flow for split sampling, see Chapter 4 of the HP 5890 Operating Manual. Verifying inlet purge status V erify that inlet purge flow is currently on[...]

  • Seite 114

    Inlet Systems Split/splitless capillary inlet 114 Splitless sampling For splitless operation, the dilute sample is vaporized inside the inlet insert. Most of the sample is then swept onto the column. For full column efficiency , vaporized sample components must reconcentrate at the head of the column prior to separation; without reconcentration, pe[...]

  • Seite 115

    Inlet Systems Split/splitless capillary inlet 115 Low Volatility Solute High Volatility Solute Column Carrier Gas Needle (a) (b) (c) (d) The Solvent Effect Figure 6-9 The solvent effect is described in great detail elsewhere: see Grob, K. and Grob, K., Jr ., Journal of Chromatography ,9 4 , page 53 (1974); Grob, K. and Grob, G., Chromatographia, 5 [...]

  • Seite 116

    Inlet Systems Split/splitless capillary inlet 116 Solvent Boiling Point ( C) Suggested Initial Oven Temperature Range ( C) Diethyl Ether 36 10 to 25 n-Pentane 36 10 to 25 Methylene Chloride 40 10 to 30 Carbon Disulfide 46 10 to 35 Chloroform* 61 25 to 50 Methanol* 65 35 to 55 n-Hexane 69 40 to 60 Ethyl Acetate* 77 45 to 65 Acetonitrile 82 50 to 70 [...]

  • Seite 117

    Inlet Systems Split/splitless capillary inlet 117 A general guideline is that components boiling at least 150 C above the column temperature will be reconcentrated by cold trapping at the head of the column. Components with lower boiling points are reconcentrated via the solvent effect. Temperature programming Multiple•ramp oven temperature progr[...]

  • Seite 118

    Inlet Systems Split/splitless capillary inlet 118 A recommended procedure is to perform a series of analyses at increasingly higher inlet temperatures using components representative of those of interest, and analyzed using the conditions for later sample analyses. The optimum temperature is where maximum area counts are obtained, and there is no e[...]

  • Seite 119

    Inlet Systems Split/splitless capillary inlet 119 10 20 30 40 50 60 Area Counts + 1.2% Deviation + 1.2% Deviation ~ 20 ppm n-C 14 (Cold Trapped) ~ 10 ppm n-C 11 (Solvent Effect) Purge Activation Time, Sec Solvent: Isooctane Column: 16.5 m x 0.25 mm SE-54 80 C (0.5 min) 170 C@1 5 /min Sample Size: 1.3 ml (Manual Injection) Injection Rate: 1 ml/sec P[...]

  • Seite 120

    Inlet Systems Split/splitless capillary inlet 120 IN OUT IN OUT GA External Plumbing Internal Plumbing Trap(s) Carrier Gas Mass Flow Controller Electronic Flow Sensor (optional) To Detector Solenoid Valve Back- pressure Regulator Split Vent Septum Purge Vent Pressure Gauge Septum Purge Control Capil- lary Inlet Column (C) (P) (S) N.O. N.C. COM Flow[...]

  • Seite 121

    Inlet Systems Split/splitless capillary inlet 121 Noting Figures 6•1 1 and 6•12, the splitless sampling process is as follows: Before Injection: Carrier gas flow enters through the mass flow controller , into the top of the inlet. A small fraction is split off to purge the septum and insert seal, then flows on to the purge vent. The remainder f[...]

  • Seite 122

    Inlet Systems Split/splitless capillary inlet 122 2. W ipe excess solvent from the syringe needle. 3. W ithout introducing air , draw in excess sample. 4. Position the syringe plunger for the required injection volume. W ipe excess sample from the needle. 5. Draw in air until the sample/solvent is entirely within the syringe barrel. The sample is r[...]

  • Seite 123

    7 Detector Systems[...]

  • Seite 124

    124 Detector Systems This chapter provides information for the five HP 5890 SERIES II (hereafter referred to as HP 5890) detector systems: Flame Ionization Detector (FID) Nitrogen•Phosphorus Detector (NPD) Electron Capture Detector (ECD) Thermal Conductivity Detector (TCD) Flame Photometric Detector (FPD) Capillary makeup gas flow rate Detectors [...]

  • Seite 125

    Detector Systems FID and NPD jets 125 Supply pressure for capillary makeup gas should be set to about 276 kPa (40 psi). FID and NPD jets Depending upon the column type to be used, and/or analyses to be performed, exchanging the jet in an FID or NPD may be necessary . T able 7•1lists available jets. Note: If switching from packed column operation [...]

  • Seite 126

    Detector Systems Flame ionization detector (FID) 126 Flame ionization detector (FID) Inlet H 2 Inlet FID Collector Assembly Jet Flame Ionization Detector (FID) Figure 7-1 The flame ionization detector (FID) responds to compounds that produce ions when burned in a H •airflame. These include all organic compounds, although a few (e.g., formic acid,[...]

  • Seite 127

    Detector Systems Flame ionization detector (FID) 127 Compounds producing little or no response include: Rare gases N N Nitrogen Oxides CO *CCl Silicon Halides CO H OC S NH O * Measured at the jet tip. This selectivity can be advantageous: for example, H Oo rC S , used as solvent, do not produce large solvent peaks. The system is linear for most org[...]

  • Seite 128

    Detector Systems Flame ionization detector (FID) 128 FID flameout problems When using pressure programming with large id columns (i.e. 530 columns) it is possible to blow the FID flame out if pressure (flow) becomes too high. If this occurs, either lower the pressure ramp or switch to a more restrictive column (longer and/or smaller id). Compressed[...]

  • Seite 129

    Detector Systems Nitrogen-phosphorus detector (NPD) 129 Nitrogen-phosphorus detector (NPD) NPD Collector Assembly Air Inlet H 2 Inlet NPD Collector Active Element Jet Nitrogen-Phosphorus Detector (NPD) Figure 7-3 The nitrogen•phosphorus detector uses a jet and collector similar to the FID; however , the collector contains a small alumina cylinder[...]

  • Seite 130

    Detector Systems Nitrogen-phosphorus detector (NPD) 130 H and air are required, but at flows significantly less than those for an FID. Normal FID•type ionizations are therefore minimal, so response to compounds not containing nitrogen or phosphorus is reduced. Thus, the detector is both sensitive to and selective toward only compounds containing [...]

  • Seite 131

    Detector Systems Nitrogen-phosphorus detector (NPD) 131 Other gas flow effects of too high flow rates of the hydrogen may allow a true flame to exist around the active element. This would overheat the active element severely and destroy the specific response. T oo low flow rates of air tend to quench the background response of the active element, a[...]

  • Seite 132

    Detector Systems Nitrogen-phosphorus detector (NPD) 132 Performance considerations Contamination V ery little contamination can create serious NPD problems. Common sources include: Columns and/or glass wool treated with H PO (phosphoric acid) Phosphate•containingdetergents Cyano•substitutedsilicone columns (XE•60,OV •225, etc.) Other nitrog[...]

  • Seite 133

    Detector Systems Nitrogen-phosphorus detector (NPD) 133 Residual silanizing reagents from derivatization, and/or bleed from silicone columns, may coat the active element with silicon dioxide. This decreases ionization efficiency , reducing sensitivity . If silanizing is necessary , remove excess reagent before injection. Silicone columns should be [...]

  • Seite 134

    Detector Systems Nitrogen-phosphorus detector (NPD) 134 Both detector baseline and sensitivity change with carrier flow rate due to change in temperature of the active element. This is the reason for the baseline drift in pressure•controlled inlet systems (capillary inlets) when temperature•programming the column. The amount of change in the de[...]

  • Seite 135

    Detector Systems Electron capture detector (ECD) 135 Electron capture detector (ECD) The effluent gas stream from the detector must be vented to a fume hood to prevent possible contamination of the laboratory with radioactive material. Specific cleaning procedures are provided in Chapter 8, Preventive Maintenance . WARNING Requirements for USA owne[...]

  • Seite 136

    Detector Systems Electron capture detector (ECD) 136 In the extremely unlikely event that both the oven and the ECD heated zone should go into thermal runaway (maximum, uncontrolled heating in excess of 400 C) at the same time, and that the ECD remains exposed to this condition for more than 12 hours, the following must be done: After turning off m[...]

  • Seite 137

    Detector Systems Electron capture detector (ECD) 137 Anode Purge Vent Makeup Gas Adapter Electron Capture Detector (ECD) Figure 7-5 Nickel Plating Fused Silica Liner Makeup Gas Column Plated 63 Ni Surface Anode The electron capture detector (ECD) cell contains Ni, a radioactive isotope emitting high•energyelectrons ( •particles).These undergo r[...]

  • Seite 138

    Detector Systems Electron capture detector (ECD) 138 Uncaptured electrons are collected periodically by applying short•term voltage pulses to cell electrodes. This cell current is measured and compared to a reference current, and the pulse interval is then adjusted to maintain constant cell current. Therefore, pulse rate (frequency) rises when an[...]

  • Seite 139

    Detector Systems Electron capture detector (ECD) 139 Chemical Type Hydrocarbons 1 Ethers, esters 10 Aliphatic alcohols, ketones, amines; 100 mono-Cl, mono-F compounds Mono-Br, di-Cl and di-F compounds 1000 Anhydrides and tri-Cl compounds 10 4 Mono-I, di-Br and nitro compounds 10 5 Di-I, tri-Br, poly-Cl and poly-F compounds 10 6 Relative Sensitivity[...]

  • Seite 140

    Detector Systems Electron capture detector (ECD) 140 Considerations for packed column operation Either N or Ar containing 5 or 10% CH , may be used as carrier gas. N yields somewhat higher sensitivity , but it is accompanied by higher noise; minimum detectable limit is about the same. N sometimes produces a negative solvent peak. Ar/CH gives greate[...]

  • Seite 141

    Detector Systems Electron capture detector (ECD) 141 Background level If the ECD system becomes contaminated, whether from impurities in the carrier (or makeup) gas, or from column or septum bleed, a significant fraction of detector dynamic range may be lost. In addition, the output signal becomes noisy . T o check background level, allow ample tim[...]

  • Seite 142

    Detector Systems Electron capture detector (ECD) 142 A very clean system may produce a value below the low end of 10 (100 Hz). T o correct this condition, an adjustment is made to the present potentiometer located on the ECD electronics board. ECD Potentiometer Switch and Adjustment ECD Potentiometer Switch Up: Adj Down: Fixed ECD Potentiometer Adj[...]

  • Seite 143

    Detector Systems Thermal conductivity detector (TCD) 143 Thermal conductivity detector (TCD) 0 ml/min Switching Flow 1 (off) 30 ml/min Column Flow 30 ml/min Switching Flow 2 (on) 30 ml/min Switching Flow 1 (on) 30 ml/min Column Flow 0 ml/min Switching Flow 2 (off) VENT (60 ml/ min) VENT (60 ml/ min) 24 6 6 36 24 36 24 6 36 HP 5890 SERIES II TCD Cel[...]

  • Seite 144

    Detector Systems Thermal conductivity detector (TCD) 144 The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components) and a reference flow of carrier gas alone; it produces voltage proportional to this difference. The voltage then becomes the output signal to t[...]

  • Seite 145

    Detector Systems Thermal conductivity detector (TCD) 145 Because of its exceptionally high thermal conductivity and chemical inertness, He is the recommended carrier gas: it gives large thermal conductivity differences with all compounds except H (considerations necessary in H analyses are discussed later). W ith He as carrier , the TCD exhibits un[...]

  • Seite 146

    Detector Systems Thermal conductivity detector (TCD) 146 Optimizing performance The following sections aid in choosing operating parameters to obtain optimal TCD performance. Temperature TCD sensitivity increases as the temperature difference between the detector filament (automatically set) and the surrounding detector body (chosen detector zone t[...]

  • Seite 147

    Detector Systems Thermal conductivity detector (TCD) 147 As Figure 7•9shows, however , the lower the detector zone temperature, the greater is the temperature difference between the filament versus the surrounding detector body temperature. Thus, for maximum sensitivity , the detector zone should be operated at the lowest temperature possible (li[...]

  • Seite 148

    Detector Systems Thermal conductivity detector (TCD) 148 Note that TCD response becomes relatively flat (insensitive) to reference gas flow rates equal to, or somewhat greater than, flow rate through the column. Analyzing for hydrogen, special considerations Only H has thermal conductivity greater than He. However , binary mixtures of small amounts[...]

  • Seite 149

    Detector Systems Thermal conductivity detector (TCD) 149 TCD-to-FID series connection The following describes, for a TCD whose exhaust vent returns to the inside of the oven, connecting the TCD to an FID. If necessary (see NOTE below), exchange the standard FID jet for the 0.030•inchjet (Part No. 18789•80070). Information about jet exchange is [...]

  • Seite 150

    Detector Systems Thermal conductivity detector (TCD) 150 filament. The immediate symptom is a permanent change in detector sensitivity due to change in filament resistance. If possible, such offending materials should be avoided. If this is not possible, the filament may have to be replaced frequently . Capillary column considerations The TCD cell [...]

  • Seite 151

    Detector Systems Flame photometric detector (FPD) 151 Flame photometric detector (FPD) Optimizing FPD sensitivity and selectivity FPD sensitivity and selectivity are affected by several important parameters. These are listed below , with suggested ways to optimize for each application. A. FPD Flow Rates. FPD flow rates are the most critical for opt[...]

  • Seite 152

    Detector Systems Flame photometric detector (FPD) 152 140 120 100 80 60 40 20 0 0 50 100 150 200 250 300 350 400 450 500 10 20 30 40 50 60 70 = Hydrogen = Nitrogen = Oxygen + = Air Pressure-psig Pressure-kPa Flow ml/min + + + + + FPD Flows versus Supply Pressures Figure 7-11 B. Detector T emperature. Detector heated zone temperature can have a sign[...]

  • Seite 153

    Detector Systems Flame photometric detector (FPD) 153 Flame ignition problems T wo common flame ignition problems are: A loud pop results on ignition and the flame will not light or stay lit. If a loud pop occurs on ignition, it is usually caused by an incorrect ignition sequence. The correct ignition sequence is: 1. Open the auxiliary Nitrogen V a[...]

  • Seite 154

    Detector Systems Flame photometric detector (FPD) 154 3. Under some operating conditions, it is important to continue to hold the ignitor switch in for several seconds after opening the hydrogen valve fully counterclockwise. 4. Under some operating conditions, the flame may be more easily lit with the rubber drip tube removed. After lighting the fl[...]

  • Seite 155

    8 Preventive Maintenance[...]

  • Seite 156

    156 Preventive Maintenance This chapter includes maintenance, cleaning, and leak•testingHP 5890 SERIES II (hereafter referred to as HP 5890) inlet and detector systems. Conditioning columns Columns may contain contaminants; conditioning drives off unwanted volatiles, making the column fit for analytical use. New packed columns should be condition[...]

  • Seite 157

    Preventive Maintenance Conditioning columns 157 back of nut). Adjust the septum purge flow rate to no more than 6 ml/min. c. Cap inlet fittings into detector(s) to prevent entry of air and/or contaminants. 3. Establish a stable flow of carrier gas through the column. He is preferred; however , N is adequate for conditioning packed columns. Do not u[...]

  • Seite 158

    Preventive Maintenance (Re)Packing columns 158 (Re)Packing columns In packing columns (particularly 1/4•inchglass columns), one must consider the type of packing, column bore, and type (metal or glass), the method of sample introduction (flash vaporization or on•column),inlet or detector base requirements. The method of sample introduction and/[...]

  • Seite 159

    Preventive Maintenance Packed column inlet 159 Packed column inlet Changing septa Septum lifetime is dependent upon frequency of use and upon needle quality; burrs, sharp edges, rough surfaces, or a blunt end on the needle decreases septum lifetime. A leaking septum is evidenced by longer retention times, loss of response, and/or loss of column hea[...]

  • Seite 160

    Preventive Maintenance Packed column inlet 160 Caution Column flow is interrupted while changing septa; since some columns may be damaged at elevated temperature without carrier flow , cool the oven to ambient before proceeding. Exercise care! The oven and/or inlet or detector fittings may be hot enough to cause burns. WARNING T urn off carrier flo[...]

  • Seite 161

    Preventive Maintenance Packed column inlet 161 4. Fully open the mass flow controller counterclockwise and wait 1 to 2 minutes to ensure equilibrium. 5. T urn off gas to the inlet at its source. 6. W ait 10 minutes while observing carrier source pressure. If it drops less than 7 to 14 kPa (1 to 2 psi), the system (through the inlet column fitting) [...]

  • Seite 162

    Preventive Maintenance Packed column inlet 162 Packed Column Inlet, Leak-Checking the Septum Figure 8-3 Cleaning T urn off the heated zone for the inlet and allow it to cool. Remove the septum retainer nut and septum; remove also the column and inlet liner . Using a suitable light source, illuminate the inside of the inlet from inside the oven whil[...]

  • Seite 163

    Preventive Maintenance Split/splitless capillary inlets 163 Split/splitless capillary inlets Changing septa For a conventional disk•typeseptum, lifetime is dependent upon needle quality; needles should be sharply pointed and free of burrs or rough surfaces. Choice of septum material is less critical than with a packed column inlet since the septu[...]

  • Seite 164

    Preventive Maintenance Split/splitless capillary inlets 164 1. Loosen and remove the septum retainer nut. Remove and discard the old septum, found either in the top of the inlet or inside the septum retainer nut. Capillary Inlet Septum Replacement, Split/Splitless and Split-Only Capillary Inlet Figure 8-4 2. The new septum is placed in the top of t[...]

  • Seite 165

    Preventive Maintenance Split/splitless capillary inlets 165 Leaks For proper inlet operation, it is essential the entire system be leak•tight. The following procedure should be performed in initial checkout, or any time a leak is suspected. 1. Switch off detector! 2. Install an inlet plug (a paper clip or similar •gaugewire) in the same manner [...]

  • Seite 166

    Preventive Maintenance Split/splitless capillary inlets 166 6. T urn off flow to the inlet by turning off carrier gas at the flow controller ( fully clockwise, turning it only until it bottoms , and then no further). 7. Adjust the back pressure regulator clockwise , an additional 1/4•turn or set the electronic pressure control to 145 kPa (21 psi)[...]

  • Seite 167

    Preventive Maintenance Split/splitless capillary inlets 167 Use leak detection fluid to check for leakage at the column nut. If leakage is observed, try tightening the nut first. If leakage continues, replace the ferrule. Note that if the inlet is hot, leak detection fluid may boil, giving false indication of a leak. If the septum and column nut pr[...]

  • Seite 168

    Preventive Maintenance Split/splitless capillary inlets 168 Solenoid Valve Assembly Solenoid Valve, Split/Splitless Capillary Inlet Figure 8-7 Cleaning T urn off the heated zone for the inlet and allow it to cool. Remove septum retainer nut, septum, insert retainer nut, and inlet insert; also remove the column. Using a suitable light source, illumi[...]

  • Seite 169

    Preventive Maintenance Liner and/or insert care 169 Liner and/or insert care Regardless of the inlet system, inlet inserts and/or liners must be kept clean for optimum performance, particularly their interiors from which contamination may enter the column and/or interact with sample components. Note: Excessive contamination anywhere on an insert or[...]

  • Seite 170

    Preventive Maintenance Liner and/or insert care 170 Repacking a split insert Since, for a split insert, its packing material is discarded in cleaning, the insert must be repacked. Note: Repacking with small•diameterglass beads is not recommended: they are usually contaminated with metal filings due to sieving procedures used. If they must be used[...]

  • Seite 171

    Preventive Maintenance Flame ionization detector (FID) 171 Metal inserts and/or liners Do not use concentrated acid(s) on metal inserts or liners! The insert is washed with noncorrosive solvents (H O, CH OH (methanol), (CH ) CO (acetone), CH Cl (methylene chloride), etc), and then dried thoroughly in an oven at 105 C. After washing and drying, a su[...]

  • Seite 172

    Preventive Maintenance Flame ionization detector (FID) 172 Jet exchange/replacement Depending upon the column type to be used, and/or analyses to be performed, exchanging the jet in an FID may be necessary . Flame Ionization Detector Figure 8-9 Note: The proper jet must be installed prior to column installation. If switching from packed column oper[...]

  • Seite 173

    Preventive Maintenance Flame ionization detector (FID) 173 18789-80070 0.030 Packed Column Only (FID only: Simulated Distillation, TCD-to-FID series operation) 18710-20119 0.018 Packed Column (Standard, FID and NPD) 19244-80560 0.011 Capillary Column (FID and NPD) (FID: high sensitivity, packed column) Part No. Use Jet Tip ID (inch)* NOTE: The 0.01[...]

  • Seite 174

    Preventive Maintenance Flame ionization detector (FID) 174 Collector Assembly Cover Removed, Flame Ionization Detector (FID) Figure 8-10 T urn off the detector and its heated zone; also turn off gases to the detector (particularly H !). Allow time for the detector zone to cool. Open the top cover at its front edge to access the detector . 1. Using [...]

  • Seite 175

    Preventive Maintenance Flame ionization detector (FID) 175 W ash the collector in distilled water , hexane, and/or CH OH (methanol). Dry in an oven at 70 C for at least 1/2•hour . FID Collector Assembly Figure 8-11 4. Using a 1/4•inchhex nut driver , unscrew (counterclockwise) and remove the jet from the detector base.[...]

  • Seite 176

    Preventive Maintenance Flame ionization detector (FID) 176 Je t FID Jet Figure 8-12 5. The jet exists in three sizes: 0.030•, 0.018•, or 0.011•inch. Use a cleaning wire (0.016•inchod, 12•inchlength, Part No. 18765•20070) to loosen/remove internal deposits. Be careful in using the wire with the 0.01 1•inchjet. W ash both the internal b[...]

  • Seite 177

    Preventive Maintenance Flame ionization detector (FID) 177 Interconnect Sprin g FID Signal Board Interconnect Figure 8-13 9. Reassemble the detector cover . Ignition problems Before proceeding, make sure that gases are plumbed correctly , the system is leak•free,flow rates are set correctly , and external lines have been well purged. Note: If Hel[...]

  • Seite 178

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 178 is best to have a new jet on hand to exchange if a damaged jet is suspected. Nitrogen-phosphorus detector (NPD) In addition to the detector itself, other systems associated with the detector may also require routine maintenance. Nitrogen•phosphorus detectors use H 2 gas as fuel. If H 2[...]

  • Seite 179

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 179 T urn off the detector and its heated zone; also turn off gases to the detector (particularly H ! ). Allow time for the detector zone to cool. Open the top cover at its front edge to access the detector . 1. Using a Pozidriv•typescrewdriver , remove three screws around the detector cov[...]

  • Seite 180

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 180 2. a. Using compressed air or N , blow out loose material from inside the collector . Do this carefully so as not to disturb the active element. Caution Do not attempt to clean the inside of the collector by inserting objects such as wires or brushes; to do so may damage the active eleme[...]

  • Seite 181

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 181 Caution Do not overtighten the jet! Overtightening may permanently deform and damage the jet, the detector base, or both. 8. Replace the NPD collector , and transformer and cover assembly . Be certain the spring contact to the signal board is in good contact with the groove on the collec[...]

  • Seite 182

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 182 Type A Type B NPD Collector and Collector Assembly Figure 8-17 Detector Cover Trans- former Brass Collar Steel Spring Spacer Teflon Spacer Transformer Strap Collector Collector Body O-ring Whenever the collector must be removed from the detector cover of a T ype A NPD, the following proc[...]

  • Seite 183

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 183 3. Remove the T eflon spacer and stainless steel spring spacer from the top of the collector body . 4. Loosen the setscrew in the T eflon portion of the collector body . 5. Grasping the collector at its top end (to avoid contaminating its detecting end), withdraw it from the collector bo[...]

  • Seite 184

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 184 lead on the collector body . Tighten the setscrew to secure the wire and collar . Type B NPD transformer/collector assembly Type B NPD Detector Assembly Figure 8-18. Remove the transformer/collector assembly from the T ype B NPD cover as follows: Caution During disassembly do not touch t[...]

  • Seite 185

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 185 4. Remove the collector from the collector assembly as follows: Loosen the 1.5•mmscrew holding the transformer secondary wire to the top of the collector and disconnect the wire. The hex key wrench required is a 1.5•mmsize and was provided with the instrument. Loosen the 1.5 hex key [...]

  • Seite 186

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 186 Reinstallation 1. Reinstall the jet in the detector base (using a 1/4•inchnut driver). Make sure that the threads are clean and free of burrs that could cause damage. If there is any binding, the cause should be determined and corrected before proceeding. If a torque wrench is availabl[...]

  • Seite 187

    Preventive Maintenance Nitrogen-phosphorus detector (NPD) 187 All collectors should be washed off with GE grade hexane or a similar solvent before reinstalling in the instrument to remove any grease, fingerprints, or other contaminants. Soak the entire collector in a vial of hexane for several minutes (2-10). Remove the collector , touching only th[...]

  • Seite 188

    Preventive Maintenance Electron capture detector (ECD) 188 Electron capture detector (ECD) Frequency test Note: For high sensitivity operation, and starting from a cold system, 24 hours may be necessary before baseline is completely stabilized. Use low•bleedsepta and condition a new septum prior to use in an unused inlet for several hours with 1 [...]

  • Seite 189

    Preventive Maintenance Electron capture detector (ECD) 189 remove the column to the ECD. If a capillary column was installed, remove also the makeup gas adapter in the detector base. 2. Disconnect the carrier gas source line at its fitting on the HP 5890. 3. Using a V espel ferrule, and adapters as necessary , connect the carrier source line to the[...]

  • Seite 190

    Preventive Maintenance Electron capture detector (ECD) 190 flow through the system is available. Allow time for the system to become fully pressurized. 4. Close carrier gas flow at its source and monitor system pressure. 5. The system may be assumed to be leak•freeif no pressure drop is observed over a 10•minuteperiod. 6. If leakage is observed[...]

  • Seite 191

    Preventive Maintenance Electron capture detector (ECD) 191 Packed column: 1. Close the anode purge on/off valve. 2. Remove the column from the detector; install in its place an empty glass column. 3. Establish normal carrier gas flow rate (20 to 30 ml/min); set oven temperature to 250 C. 4. Open the anode purge on/off valve. 5. Heat the ECD to 350 [...]

  • Seite 192

    Preventive Maintenance Thermal conductivity detector (TCD) 192 Radioactivity leak test (wipe test) ECDs must be tested for radioactive leakage at least every six months. Records of tests and results must be maintained for possible inspection by the Nuclear Regulatory Commission and/or responsible state agency . More frequent tests may be conducted [...]

  • Seite 193

    Preventive Maintenance Flame photometric detector 193 Caution Failure to turn off the TCD and to cap the detector column fitting may cause irreparable damage to the filament due to O 2 entering the detector . 3. Establish normal reference gas flow rate (20 to 30 ml/min) through the detector (set oven temperature to 250 C). 4. Heat the detector to 4[...]

  • Seite 194

    Preventive Maintenance Flame photometric detector 194 Likewise, damage to the PMT window cannot be tolerated; if necessary , replace the PMT or call Hewlett•Packardsupport. 1. Remove four screws to remove the PMT adapter flange. Remove the adapter carefully; a quartz window is exposed and may fall out. The window is cleaned in a manner similar to[...]

  • Seite 195

    Preventive Maintenance Flame photometric detector 195 AB C DE 8 5 2 1 11 4 6 7 9 10 16 6 7 15 14 12 13 19 20 21 22 3 27 4 Places O-ring (8) 3t o6m m Subassembly Parts Identification Figure 8-20.[...]

  • Seite 196

    Preventive Maintenance Flame photometric detector 196 F G H 11 29 24 25 23 17 26 28 30 (4 Places) Item Description Part No. Qty 1 Weldment, Base 19256-80540 1 2 Gigabore Liner/Ferrule Assembly (see note) 19256-60590 1 3 Lockwasher 2190-0108 4 4 Lower Heater Block 19256-20500 1 5 Weldment, Transfer Tube 19256-80550 1 6 Nut, Brass, 1/4-inch id 0100-0[...]

  • Seite 197

    Preventive Maintenance Flame photometric detector 197 NOTE: Once installed, the ferrule cannot be removed from the liner for reuse unless both parts are still warm. Cleaning/replacing the FPD jet If a response problem is encountered (sensitivity , noise, selectivity), the FPD jet should be inspected for deposits and, if necessary , cleaned or repla[...]

  • Seite 198

    Preventive Maintenance Flame photometric detector 198 6. Use compressed gas, air , or N 2 to blow out loose particles from the jet and/or detector module body . 7. Inspect and clean deposits from the jet bore and from the threads using a suitable wire. If the jet is damaged in any way , it should be replaced. It is good practice to replace the jet [...]

  • Seite 199

    Preventive Maintenance Flame photometric detector 199 this indicates a leak in the system. Begin checking possible leak sources and monitor the EFS to determine when the leak has been eliminated. Possible leak sources, in order of probability are: 1. septum 2. column fittings 3. supply line swage•type plumbing connections 4. detector block O•ri[...]

  • Seite 200

    Preventive Maintenance Conditioning chemical traps 200 Conditioning chemical traps Remove the trap from its installed location and attach it to a clean, dry gas source (helium or nitrogen). Attach the 1/8•inchend (male) of the chemical trap assembly to the reconditioning gas source using a graphite or a graphitized V espel ferrule (Part No. 0100-[...]

  • Seite 201

    9 Chromatographic Troubleshooting[...]

  • Seite 202

    202 Chromatographic Troubleshooting Introduction This chapter is concerned with diagnosis: the process of going from unexpected behavior of the HP 5890 SERIES II (hereafter referred to as HP 5890) (symptoms) to the probable location of the difficulty (causes). Problems arise from many causes. Some of these are: Electronic or mechanical failure Cont[...]

  • Seite 203

    Chromatographic Troubleshooting Baseline symptoms 203 It can also result from valve operations: If valves are being switched during a run, examine the valve time program to see if the change coincides with a valve operation. This symptom also can occur if the septum suddenly begins to leak; A void the problem by changing septa regularly . Offset•[...]

  • Seite 204

    Chromatographic Troubleshooting Baseline symptoms 204 2. Baseline is erratic, moves up and down (wander): Suspect a leak in the system: Check septum condition and replace if necessary . Check column connections. If the leak is at the detector end of the column, retention times are stable from run to run, but sensitivity is reduced. If it is at the [...]

  • Seite 205

    Chromatographic Troubleshooting Baseline symptoms 205 Contaminated detector gases (hydrogen and air). Air currents from a fan or air conditioner blowing across the top of the instrument may interfere with gas exiting from the detector . This is a possible, though not very likely , cause of noise since detectors are well protected. Switching off the[...]

  • Seite 206

    Chromatographic Troubleshooting Baseline symptoms 206 Spiking Spikes are isolated baseline disturbances, usually as sudden (and large) upscale movements. If accompanied by noise, the noise problem should be solved first, since spiking may disappear at the same time. 1. Spikes appear whenever the chart is running: The cause is almost always electron[...]

  • Seite 207

    Chromatographic Troubleshooting Retention time symptoms 207 Retention time symptoms Retention time drift Retention time drift is a steady increase or decrease of retention times in successive runs. Erratic times (both directions) are discussed below as retention time wander . 1. In a series of runs, retention times suddenly increase: This may be du[...]

  • Seite 208

    Chromatographic Troubleshooting Retention time symptoms 208 2. Reproducibility is good early in the run but not toward the end: This may occur in temperature•programminga very densely packed column; as column contents expand with heating, resistance to flow may be so great that a mass flow controller cannot maintain constant flow . T ry increasin[...]

  • Seite 209

    Chromatographic Troubleshooting Peak symptoms 209 Peak symptoms No peaks This is usually due to operator error; possibilities include injection on the wrong column, incorrect signal assignment, attenuation too high (peaks are present but not visible), a bent syringe needle in an automatic sampler , etc. Check system parameters for the analysis. Inv[...]

  • Seite 210

    Chromatographic Troubleshooting Peak symptoms 210 stationary phase with trace levels of O ,H O, and/or other materials present in the carrier gas. A contaminated inlet may also produce ghost peaks. Residues in the inlet are volatilized or pyrolyzed and swept onto the head of the column. T ry reducing inlet temperature; if this eliminates or reduces[...]

  • Seite 211

    Chromatographic Troubleshooting Peak symptoms 211 Deformed peaks The ideal peak, rarely occurring in chromatography , is a pure Gaussian shape. In practice, some asymmetry is always present, particularly near the baseline. 1. The peak rises normally , then drops sharply to baseline: Overloaded Peak Figure 9-1. The most likely cause is column overlo[...]

  • Seite 212

    Chromatographic Troubleshooting Peak symptoms 212 Interaction with column material is a frequent cause. Silanized support may help. An all•glasssystem may be required if metal column tubing is the source. Column overload with a gas sample often shows this effect; try injecting less. This may be a merged peak situation: Running at lower (30 C) ove[...]

  • Seite 213

    Chromatographic Troubleshooting Peak symptoms 213 4. T op (apex) of the peak is split: FID/NPD Flameout, or TCD with H (in He Carrier) Figure 9-4. V erify that this is not a merged peak situation: Reduce oven temperature 30 C and repeat the run. If the split peak becomes better resolved, it is probably a merged pair . Gross overload of an FID may c[...]

  • Seite 214

    Chromatographic Troubleshooting Troubleshooting valve systems 214 Troubleshooting valve systems Chromatographic symptoms T roubleshooting valves and their related plumbing is primarily a matter of systematic checking and verification of unimpaired mechanical operation of any moving part. This requires an understanding of how the valve functions int[...]

  • Seite 215

    Chromatographic Troubleshooting Troubleshooting valve systems 215 Loss of peaks in specific areas of the chromatogram Entire sections of chromatographic data can be lost due to a valve that does not rotate or one that rotates improperly . Other than obvious component failures (i.e., solenoid, actuator , etc.), generally improper adjustments and mis[...]

  • Seite 216

    Chromatographic Troubleshooting Locating leaks 216 Extraneous peaks Air peaks are sometimes seen in a chromatogram when leakage occurs because the valve rotor does not seal properly . These leaks may not be detectable by using the soap•bubble method. The leak test procedure is described in the Site Prep and Installation Manual . If a leak is susp[...]

  • Seite 217

    Chromatographic Troubleshooting Pressure check 217 Pressure check The pressure•check method will indicate, but sometimes not isolate, a leak in the flow path. Since this method does not necessarily isolate the leak, one of the leak•check methods may be needed to locate the leak specifically . Note that each valve in a system has two flow paths,[...]

  • Seite 218

    Chromatographic Troubleshooting Electronic pressure control 218 Electronic pressure control The electronic pressure control option provides very accurate and precise control of column head pressure, resulting in retention time reproducibility of better than 0.02% RSD when there are no column effects. The inlet pressure can be set constant, programm[...]

  • Seite 219

    Chromatographic Troubleshooting Electronic pressure control 219 Safety shutdown Systems equipped with electronic pressure programming have a safety shutdown feature to prevent gas leaks from creating a safety hazard. If the system cannot reach a pressure setpoint it beeps. After about 45 seconds the beep will stop and the message: EPPB: SAFETY SHUT[...]

  • Seite 220

    Chromatographic Troubleshooting Electronic pressure control 220 Proper configuration If the inlet is not working at all, there may be a configuration problem. 1. T urn GC power off , and remove the side panel of the GC. 2. Check if the red switches on the inlet controller board are set for your configuration. 3. T urn the GC on .[...]

  • Seite 221

    Chromatographic Troubleshooting Electronic pressure control 221 Switch setting examples EPC A or EPC B MODE A or MODE B IN A0 or IN B0 IN A1 or IN B1 LEFT, Electronic Pressure Control present / RIGHT, Electronic Pressure Control not present LEFT, (FPR) Programmable Cool On-Column Inlet & Purged Packed Inlet / RIGHT, (BPR) Split/Splitless Capill[...]

  • Seite 222

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 223

    10 Test Sample Chromatograms[...]

  • Seite 224

    224 Test Sample Chromatograms This chapter contains typical examples of test sample chromatograms. They may be used as a general guide to instrument performance. It is assumed that both the instrument and proper test column are installed, that general keyboard control is understood (temperature control, defining signal output, etc.), and that speci[...]

  • Seite 225

    Test Sample Chromatograms Test sample chromatograms 225 Test sample chromatograms Detector Type FID (or FIDw/MUG) Temp 250 DEGREES C Inlet Type PACKED (OR PURGED PACKED). Temp 200 DEGREES C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Temp Programmed (1 ramp) Init Temp 110 DEGREES C Init Time 0 min Ramp Rate 15 Fin Temp 150 [...]

  • Seite 226

    Test Sample Chromatograms Test sample chromatograms 226 IF Detector Type NPD (or NPDw/MUG) Temp 220 DEGREES C Inlet Type PACKED (OR PURGED PACKED). Temp 170 DEGREES C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Isothermal Init Temp 170 DEGREES C Init Time 3.0 min Ramp Rate 0 Fin Temp Fin Time Range 0 COLUMN: Part No. 19095S[...]

  • Seite 227

    Test Sample Chromatograms Test sample chromatograms 227 Detector Type ECD(or ECDw/MUG) Temp 300 DEGREES C Inlet Type PACKED (OR PURGED PACKED). Temp 200 DEGREES C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Temp Isothermal Init Temp 160 DEGREES C Init Time N/A min Ramp Rate 0 Fin Temp Fin Time Range 2 COLUMN: Part No. 19095[...]

  • Seite 228

    Test Sample Chromatograms Test sample chromatograms 228 Detector Type TCD(or TCDw/MUG) Temp 300 DEGREES C Inlet Type PACKED (OR PURGED PACKED). Temp 250 DEGREES C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Temp Programmed (1 ramp) Init Temp 110 DEGREES C Init Time 1 min Ramp Rate 15 Fin Temp 150 Fin Time 1 Range 0 COLUMN: [...]

  • Seite 229

    Test Sample Chromatograms Test sample chromatograms 229 Detector Type FIDw/MUG Temp 250 DEGREES C Inlet Type Ded On-Col Cap Oven Track On Temp N/A C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Temp Temp Programmed (1 ramp) Init Temp 60 DEGREES C Init Time 0.5 min Ramp Rate 20 Fin Temp 185 Fin Time 1 Flow Param (EPP) Constan[...]

  • Seite 230

    Test Sample Chromatograms Test sample chromatograms 230 Detector Type FIDw/MUG Temp 250 DEGREES C Inlet Type SPLIT ONLY OR SPLIT/SPLITLESS Temp 200 DEG C Operating Mode SPLIT(PURGE ON) Purge Time On 0 min Purge Time Off 0 min Oven Temp Programmed (1 ramp) Init Temp 100 DEGREES C Init Time 0 min Ramp Rate 20 Fin Temp 180 Fin Time 1 Range 7 COLUMN: P[...]

  • Seite 231

    Test Sample Chromatograms Test sample chromatograms 231 Detector Type NPD w/MUG Temp 220 DEGREES C Inlet Type Split only or split/splitless Temp 200 DEGREES C Operating Mode Split(Purge on) Purge Time On 0 min Purge Time Off 0 min Oven Isothermal Init Temp 180 DEGREES C Init Time 5.0 min Ramp Rate 0 Fin Temp Fin Time Range 0 COLUMN: Part No. 19095Z[...]

  • Seite 232

    Test Sample Chromatograms Test sample chromatograms 232 Detector Type ECDw/MUG Temp 300 DEGREES C Inlet Type Split only or splitless Temp 200 DEGREES C Operating Mode Split(Purge on) Purge Time On 0 min Purge Time Off 0 min Oven Temp Isothermal Init Temp 170 DEGREES C Init Time N/A min Ramp Rate 0 Fin Temp Fin Time Range 0 COLUMN: Part No. 19095S(#[...]

  • Seite 233

    Test Sample Chromatograms Test sample chromatograms 233 Detector Type TCDw/MUG Temp 300 DEGREES C Inlet Type Split only or split/splitless Temp 250 DEG C Operating Mode Split(Purge on) Purge Time On 0 min Purge Time Off 0 min Oven Temp Programmed Init Temp 100 DEGREES C Init Time 1 min Ramp Rate 10 Fin Temp 150 Fin Time 2 Range 0 COLUMN: Part No. 1[...]

  • Seite 234

    Test Sample Chromatograms Test sample chromatograms 234 Detector Type NPD w/MUG) Temp 220 DEGREES C Inlet Type Ded On-Col Cap Oven Track On Temp N/A C Operating Mode N/A Purge Time On min Purge Time Off min Oven Isothermal Init Temp 170 DEGREES C Init Time 5.0 min Ramp Rate 0 Fin Temp Fin Time Flow Param (EPP) Constant Flow Off Range 0 COLUMN: Part[...]

  • Seite 235

    Test Sample Chromatograms Test sample chromatograms 235 Detector Type TCDw/MUG Temp 300 DEGREES C Inlet Type Ded On-Col Oven Track On Temp N/A C Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Temp Programmed (1 ramp) Init Temp 60 DEGREES C Init Time 0.5 min Ramp Rate 20 Fin Temp 180 Fin Time 1 Flow Param (EPP) Constant Flow Of[...]

  • Seite 236

    Test Sample Chromatograms Test sample chromatograms 236 Detector Type ECDw/MUG Temp 300 DEGREES C Inlet Type Ded On-Column Oven Track On Temp N/A Operating Mode N/A Purge Time On N/A min Purge Time Off N/A min Oven Isothermal Init Temp 170 DEGREES C Init Time N/A min Ramp Rate 0 Fin Temp Fin Time Flow Param (EPP) Constant Flow Off Range 2 COLUMN: P[...]

  • Seite 237

    Test Sample Chromatograms Test sample chromatograms 237 SAMPLE: Type FPD Sample Inj Volume 1 l Part No. 19395•60580 COMPOSITION: 20 ng/ l (20.0 ppm W/V) each of 1-dodecanethiol and tributylphosphate in isooctane Detector Type FPD Temp 200 DEGREES C Inlet Type PACKED OR PURGED PACKED Temp 200 DEGREES C Operating Mode N/A Purge Time On N/A min Purg[...]

  • Seite 238

    Test Sample Chromatograms Test sample chromatograms 238 SAMPLE: Type FPD Sample Inj Volume 2 l Part No. 19305•60580 COMPOSITION: 20 ng/ l (20.0 ppm W/V) each of 1-dodecanethiol and tributylphosphate in isooctane Detector Type FPD Temp 200 DEGREES C Inlet Type SPLIT ONLY OR SPLIT/SPLITLESS Temp 200 DEGREES C Operating Mode SPLIT (PURGE ON) Purge T[...]

  • Seite 239

    Test Sample Chromatograms Test sample chromatograms 239 SAMPLE: Type FPD Sample Inj Volume 1 l Part No. 19395•60580 COMPOSITION: 20 ng/ L (20.0 ppm W/V) each of 1-dodecanethiol and tributylphosphate in isooctane Detector Type FPD Temp 200 DEGREES C Inlet Type DED ON-COL CAP Oven Track On Temp N/A DEGREES C Operating Mode N/A Purge Time On N/A min[...]

  • Seite 240

    T his pag e inte ntional l y le ft bl ank.[...]

  • Seite 241

    241 Index A adapters, 17 detector , 22 ECD, 23 installation, 24 TCD, 23 alphanumeric display , 33 B baseline problems noise, 204 position, 202 spiking, 206 wander and drift, 203 C calibration electronic flow sensor , 60 oven temperature, 54 capillary columns, metal, 30 clear dot function, 39 cold trapping, 1 16 collector replacing NPD, 181 type B N[...]

  • Seite 242

    Index 242 electronic flow sensor (EFS) calibration, 60 packed inlet, 102, 106 electronic pressure control troubleshooting, 218 entering setpoints, 34 F fault: messages, 52 ferrules, 14 FID, 171 ignition problems, 177 jet replacement, 30, 172, 173 on•column test chromatogram, 229 split inlet test chromatogram, 230 test chromatogram, 225 FID flameo[...]

  • Seite 243

    Index 243 L leaks ECD, 189 FPD with EFS, 198 FPD without EFS, 199 packed column inlet, 160 pressure checking, 217 split/splitless capillary inlet, 165 valves, 216 LED display , 33 lighting problems, FID, 177 liners, 17 care of, 169 detector , 22 installation, 24 metal, 171 lock keyboard, 38 M metal capillary columns, 30 N noise problems, 204 NPD, 1[...]

  • Seite 244

    Index 244 S septum, changing packed column inlet, 159 split/splitless capillary inlet, 163 septum purge, packed inlet, 105 septum purged packed column inlet, 103 setpoint protection, 38 setpoints displaying, 33 entering, 34 ranges, 46 solvent effect, 1 14 solvent purity , 1 18 spiking problems, 206 split capillary inlet inserts, 27 split ratio, 1 1[...]