Genesis Advanced Technologies 2006 manual
- Consulta online o descarga el manual de instrucciones
- 22 páginas
- 0.23 mb
Ir a la página of
manuales de instrucciones parecidos
-
Portable Speaker
Genesis Advanced Technologies 201
2 páginas 1.38 mb -
Speaker System
Genesis Advanced Technologies G6.1c
17 páginas 0.53 mb -
Speaker
Genesis Advanced Technologies S4/8
18 páginas 1.31 mb -
Stereo Amplifier
Genesis Advanced Technologies B-200
4 páginas 9.19 mb -
Cordless Drill
Genesis Advanced Technologies GHD1260
11 páginas 0.79 mb -
Stereo System
Genesis Advanced Technologies Servo/Bass
5 páginas 0.64 mb -
Speaker
Genesis Advanced Technologies 8300F
2 páginas 4.4 mb -
Speaker System
Genesis Advanced Technologies G6.1sr
9 páginas 0.19 mb
Buen manual de instrucciones
Las leyes obligan al vendedor a entregarle al comprador, junto con el producto, el manual de instrucciones Genesis Advanced Technologies 2006. La falta del manual o facilitar información incorrecta al consumidor constituyen una base de reclamación por no estar de acuerdo el producto con el contrato. Según la ley, está permitido adjuntar un manual de otra forma que no sea en papel, lo cual últimamente es bastante común y los fabricantes nos facilitan un manual gráfico, su versión electrónica Genesis Advanced Technologies 2006 o vídeos de instrucciones para usuarios. La condición es que tenga una forma legible y entendible.
¿Qué es un manual de instrucciones?
El nombre proviene de la palabra latina “instructio”, es decir, ordenar. Por lo tanto, en un manual Genesis Advanced Technologies 2006 se puede encontrar la descripción de las etapas de actuación. El propósito de un manual es enseñar, facilitar el encendido o el uso de un dispositivo o la realización de acciones concretas. Un manual de instrucciones también es una fuente de información acerca de un objeto o un servicio, es una pista.
Desafortunadamente pocos usuarios destinan su tiempo a leer manuales Genesis Advanced Technologies 2006, sin embargo, un buen manual nos permite, no solo conocer una cantidad de funcionalidades adicionales del dispositivo comprado, sino también evitar la mayoría de fallos.
Entonces, ¿qué debe contener el manual de instrucciones perfecto?
Sobre todo, un manual de instrucciones Genesis Advanced Technologies 2006 debe contener:
- información acerca de las especificaciones técnicas del dispositivo Genesis Advanced Technologies 2006
- nombre de fabricante y año de fabricación del dispositivo Genesis Advanced Technologies 2006
- condiciones de uso, configuración y mantenimiento del dispositivo Genesis Advanced Technologies 2006
- marcas de seguridad y certificados que confirmen su concordancia con determinadas normativas
¿Por qué no leemos los manuales de instrucciones?
Normalmente es por la falta de tiempo y seguridad acerca de las funcionalidades determinadas de los dispositivos comprados. Desafortunadamente la conexión y el encendido de Genesis Advanced Technologies 2006 no es suficiente. El manual de instrucciones siempre contiene una serie de indicaciones acerca de determinadas funcionalidades, normas de seguridad, consejos de mantenimiento (incluso qué productos usar), fallos eventuales de Genesis Advanced Technologies 2006 y maneras de solucionar los problemas que puedan ocurrir durante su uso. Al final, en un manual se pueden encontrar los detalles de servicio técnico Genesis Advanced Technologies en caso de que las soluciones propuestas no hayan funcionado. Actualmente gozan de éxito manuales de instrucciones en forma de animaciones interesantes o vídeo manuales que llegan al usuario mucho mejor que en forma de un folleto. Este tipo de manual ayuda a que el usuario vea el vídeo entero sin saltarse las especificaciones y las descripciones técnicas complicadas de Genesis Advanced Technologies 2006, como se suele hacer teniendo una versión en papel.
¿Por qué vale la pena leer los manuales de instrucciones?
Sobre todo es en ellos donde encontraremos las respuestas acerca de la construcción, las posibilidades del dispositivo Genesis Advanced Technologies 2006, el uso de determinados accesorios y una serie de informaciones que permiten aprovechar completamente sus funciones y comodidades.
Tras una compra exitosa de un equipo o un dispositivo, vale la pena dedicar un momento para familiarizarse con cada parte del manual Genesis Advanced Technologies 2006. Actualmente se preparan y traducen con dedicación, para que no solo sean comprensibles para los usuarios, sino que también cumplan su función básica de información y ayuda.
Índice de manuales de instrucciones
-
Página 1
USER’S MANUAL GENES I S 2002 TM GENES I S 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com 0609[...]
-
Página 2
GENESIS 2006 SHOWN 2[...]
-
Página 3
To assure proper record keeping, please retain this information for future use. DATE PURC HASED ____________________________ SERIAL # _____________________________________ DISTRIBUT OR NAME __________________________ PHONE NUMBER ______________________________ CITY, STATE __________________________________ The GENESIS THAWING SYSTEMS incorporate wa[...]
-
Página 4
CONTENTS page 1. UNPACKING ........................................................ 5 2. TO THAW BLOOD PRODUCTS ........................ 5 3. TO REMOVE A THAWED BAG ........................ 9 4. TO MAINTAIN THE DEVICE IN THE READY TO THAW STATUS .............................. 9 5. TO CLEAN THE BATH OR TO ADD WATER .. 10 6. CONTROL PANEL AND INDICAT[...]
-
Página 5
1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006 unit. C o n t r o l U n i t W a t e r B a t h T h a w i n g B a g D i s p e n s e r Plasma Overwrap Thawing Bags (1000/Box) P l a s m a B a g H o l d e r L A B a t h C L E A N TM R e f e r e n c e T h e r m o m e t e r T h a w i n g B a g W e i g h t s U s e r ' s M a n[...]
-
Página 6
indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEAN TM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prev ent the formation of turbidity and odor. 2.3 Plug the device into a 115 volt AC outlet. For GENESIS 2002 , a dedica[...]
-
Página 7
c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 ° C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other. The display should now show the new SET TEMP. When th[...]
-
Página 8
thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time. f. To change the default TIMER : Press the START/CHANGE key for the selected bag location, [...]
-
Página 9
3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder. Discard plasma thawing overwrap bag after each use, as they are designed for single use only. b. If there is another bag being thawed, press the ROCK key to resume the rocking motion for the rem ai[...]
-
Página 10
5. TO CLEAN THE BATH OR TO ADD WATER S A F E T Y : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL) IS TURNED ON. RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET. 5.1 TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF. BEFORE REMOVING THE[...]
-
Página 11
when filled with water. Do not drop the bath on a hard surface. GENESIS 2006 : Remove all, or part of the water. The bath weighs approximately thirty (30) pounds when filled with water. Carefully remove the water bath from the device. 5.6 DO NOT USE BLEACH OR HARSH CLEANERS. Depending on mineral content of water used, the bath may discolor. Pro[...]
-
Página 12
6. CONTROL PANEL AND INDICATORS GENESIS 2002 TM CONTROL PANEL 12[...]
-
Página 13
GENESIS 2006 TM CONTROL PANEL 13[...]
-
Página 14
CONTROL KEYS POWER .......... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER SWITCH on the left panel is a master on/off switch to remove electrical power completely from all circuits in the unit. SET TEMP ..... With this key switch you can adjust water temperature within the range of ambient to +[...]
-
Página 15
INDICATORS HEATING ....... This red indicator is lit when the heating element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off. FAULT ............ If the water temperature exceeds 58 ° C, an alarm will sound and the red fault light will be illuminated. This m[...]
-
Página 16
2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch. 3) The FAULT light is on. • Check the water temperature. If the w[...]
-
Página 17
7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrum ent. 8) HEATING light is flashing. • This is normal operation. At approximately one degree below the set temperature, the microprocessor begins[...]
-
Página 18
8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time setting for each bag location • Forced water circulation with gentle rocking motion • Compact size - saves valuable bench space • Quick heating • Simple to operate - easy to read LED display • Li[...]
-
Página 19
10. SPECIFICATIONS GENESIS 2002 TM GENESIS 2006 TM External Dim ension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” High Interior Bath Dimension 13.0” L x 7.0” Wide x 7.5” Deep 18.0” L x 8.0” Wide x 7.5” Deep Bath Volume 2.5 US gallons 3.7 US gallons Controlled Tem p Range Ambient to 56 o C Ambient to 56 o C [...]
-
Página 20
11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com RETURN INFORMATION : Same as above. 1. Obtain a Return Goods Authorization number. 2. Ensure all water is [...]
-
Página 21
Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006 was purchased S e r i a l # RGA # (Return Goods Authorization #) 6. Mark outside of box clearly, with your return address, and send to GenesisBPS, LLC . (as above) 21[...]
-
Página 22
GENESIS 2002 GENESIS 2006 PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006 , Plasma Thawing Water Bath System is warranted against defects in materials and workmanship for a period of one year from the date of purchase. The warranty applies only to products that have not been tampered with, used for purposes other than th[...]