National Instruments NI 9239 manual
- Consulta online o descarga el manual de instrucciones
- 34 páginas
- 0.41 mb
Ir a la página of
manuales de instrucciones parecidos
-
Network Card
National Instruments MXI Series
40 páginas 0.98 mb -
Network Card
National Instruments VXI-1394
74 páginas 0.65 mb -
Network Card
National Instruments Gigabit Ethernet (GigE) Interface Device NI PCIe-8231
11 páginas 0.15 mb -
Network Card
National Instruments SCXI-1163
103 páginas 0.78 mb -
Network Card
National Instruments NI USB-6009
32 páginas 0.83 mb -
Network Card
National Instruments NI PCI-6110
118 páginas 1.13 mb -
Network Card
National Instruments 370551A-01
139 páginas 1.34 mb -
Network Card
National Instruments NI 6013
109 páginas 0.81 mb
Buen manual de instrucciones
Las leyes obligan al vendedor a entregarle al comprador, junto con el producto, el manual de instrucciones National Instruments NI 9239. La falta del manual o facilitar información incorrecta al consumidor constituyen una base de reclamación por no estar de acuerdo el producto con el contrato. Según la ley, está permitido adjuntar un manual de otra forma que no sea en papel, lo cual últimamente es bastante común y los fabricantes nos facilitan un manual gráfico, su versión electrónica National Instruments NI 9239 o vídeos de instrucciones para usuarios. La condición es que tenga una forma legible y entendible.
¿Qué es un manual de instrucciones?
El nombre proviene de la palabra latina “instructio”, es decir, ordenar. Por lo tanto, en un manual National Instruments NI 9239 se puede encontrar la descripción de las etapas de actuación. El propósito de un manual es enseñar, facilitar el encendido o el uso de un dispositivo o la realización de acciones concretas. Un manual de instrucciones también es una fuente de información acerca de un objeto o un servicio, es una pista.
Desafortunadamente pocos usuarios destinan su tiempo a leer manuales National Instruments NI 9239, sin embargo, un buen manual nos permite, no solo conocer una cantidad de funcionalidades adicionales del dispositivo comprado, sino también evitar la mayoría de fallos.
Entonces, ¿qué debe contener el manual de instrucciones perfecto?
Sobre todo, un manual de instrucciones National Instruments NI 9239 debe contener:
- información acerca de las especificaciones técnicas del dispositivo National Instruments NI 9239
- nombre de fabricante y año de fabricación del dispositivo National Instruments NI 9239
- condiciones de uso, configuración y mantenimiento del dispositivo National Instruments NI 9239
- marcas de seguridad y certificados que confirmen su concordancia con determinadas normativas
¿Por qué no leemos los manuales de instrucciones?
Normalmente es por la falta de tiempo y seguridad acerca de las funcionalidades determinadas de los dispositivos comprados. Desafortunadamente la conexión y el encendido de National Instruments NI 9239 no es suficiente. El manual de instrucciones siempre contiene una serie de indicaciones acerca de determinadas funcionalidades, normas de seguridad, consejos de mantenimiento (incluso qué productos usar), fallos eventuales de National Instruments NI 9239 y maneras de solucionar los problemas que puedan ocurrir durante su uso. Al final, en un manual se pueden encontrar los detalles de servicio técnico National Instruments en caso de que las soluciones propuestas no hayan funcionado. Actualmente gozan de éxito manuales de instrucciones en forma de animaciones interesantes o vídeo manuales que llegan al usuario mucho mejor que en forma de un folleto. Este tipo de manual ayuda a que el usuario vea el vídeo entero sin saltarse las especificaciones y las descripciones técnicas complicadas de National Instruments NI 9239, como se suele hacer teniendo una versión en papel.
¿Por qué vale la pena leer los manuales de instrucciones?
Sobre todo es en ellos donde encontraremos las respuestas acerca de la construcción, las posibilidades del dispositivo National Instruments NI 9239, el uso de determinados accesorios y una serie de informaciones que permiten aprovechar completamente sus funciones y comodidades.
Tras una compra exitosa de un equipo o un dispositivo, vale la pena dedicar un momento para familiarizarse con cada parte del manual National Instruments NI 9239. Actualmente se preparan y traducen con dedicación, para que no solo sean comprensibles para los usuarios, sino que también cumplan su función básica de información y ayuda.
Índice de manuales de instrucciones
-
Página 1
OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9229/9239 4-Channel, ±60/±10 V, 2 4-Bit Simultaneous, Channel-to-Chan nel Isolated Analog In put Modules ni.com/manuals Deutsch Français[...]
-
Página 2
NI 9229/9239 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9229 and National Instr u ments 9239 and incl u des specif ications and terminal assignments. In t his doc u ment, the NI 9229 and NI 9239 are referred to incl u si v ely as the NI 9229/9239. V isit ni.com/info and enter rd s oftwarever s ion to determine which [...]
-
Página 3
© National Instruments Corp. 3 NI 9229/9239 Safety Guidelines Operate the NI 92 29/ 9239 only as described in these operating instr u ctions. Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines for Hazardous V oltages If hazardo u s voltages are connected to th[...]
-
Página 4
NI 9229/9239 4 ni.com Caution When mod u le terminals are hazardo u s voltage LIVE (>42.4V pk /60 VDC), yo u m u st ens u re that devices and circ u its connected to the mod u le are properly ins u lated from h u man contact . Y o u m u st u se the NI 9971 connector backshell kit to ens u re that the terminals are not accessible. Fig u re 1 sho [...]
-
Página 5
© National Instruments Corp. 5 NI 9229/9239 Safety Guidelines fo r Hazardous Loca tions The NI 922 9/9239 is s u itable for u se in Class I, Division 2, Gro u ps A, B, C, D, T4 hazardo u s locations; C lass I, Zone 2, AEx nA IIC T4, and Ex nA IIC T4 hazardo u s locations; and nonhazardo u s locations on ly . Foll o w these g u idelines if yo u are[...]
-
Página 6
NI 9229/9239 6 ni.com Caution For Zone 2 applications, install a protection device between the inp u t signal and the NI 9229/9239 inp u t terminal. The device m u st pre vent the inp u t channel-to-COM voltage from exceeding 85 V if there is a transient overv oltage condition. Caution For Zone 2 applications, connected signals m u st be within the[...]
-
Página 7
© National Instruments Corp. 7 NI 9229/9239 Special Conditions fo r Marine Applications Some mod u les are Lloyd’ s Register (LR) T ype Approv ed for marine applications. T o verify Lloyd’ s Register certif ication, go to ni.com/certification and search for the LR certificate, or look for the Lloyd’ s Re gister mark on the mod u le. Caution [...]
-
Página 8
NI 9229/9239 8 ni.com Connecting the NI 9229/9239 The NI 9229/ 9239 provides connections for fo u r sim u ltaneo u sly sampled, isolated analog inp u t channels. Figure 2. NI 9229/923 9 T erminal Assignmen ts AI0+ AI0– AI1+ AI1– AI2+ AI2– AI3+ AI3– 0 1 0 1 0 1 0 1[...]
-
Página 9
© National Instruments Corp. 9 NI 9229/9239 The NI 922 9/9239 has fo u r 2-terminal de tachable scre w-terminal connectors. Note Yo u m u st u se 2-wire ferr u les to create a sec u re connection when connecting mo re than one wire to a sin gle ter min al o n th e NI 9229/9239. Yo u can connect gro u nd-referenced or floating signal so u rces to t[...]
-
Página 10
NI 9229/9239 10 ni.com Figure 3. Connecting a Grounded Signal So urce to the NI 9229/9239 Figure 4. Connecting a Floating Signa l Source to the NI 9229/9239 NI 9229/9239 Signal Source AI– + – AI+ NI 9229/9239 Signal Source AI– + – AI+[...]
-
Página 11
© National Instruments Corp. 11 NI 9229/9239 The NI 9229/9239 analog inp u t channels are floating with respect to earth gro u nd and each other . The incoming analog signal on each channel is conditioned, b u ffered, and th en sampled by a 24-bit Delta-Sigma ADC. Each channel pro vides an independen t signal path and ADC, enabling yo u to sample [...]
-
Página 12
NI 9229/9239 12 ni.com Wiring for High-Vib ration Applications If an application is s u bject to high vibrat ion, National Instr u ments recommends that yo u either u se ferr u les to terminate wires to the detachable scre w-ter minal connector or u se the NI 9971 backshell kit to protect the connections. Refer to Fig u re 6 for an ill u stration o[...]
-
Página 13
© National Instruments Corp. 13 NI 9229/9239 Understanding NI 9229/9239 Filtering The NI 9229/9239 u ses a combination of analog and digital filte ring to pro vide an acc u rate representation of in-band signals while rejecting o u t-of-band signals. The filters di scriminate between signals based on the freq u ency range, or bandwidth, of the sig[...]
-
Página 14
NI 9229/9239 14 ni.com Figure 7. T ypical Passband Flatness for the NI 9229/9239 F requency/Data Rate 0.5 0.4 0.025 0.000 –0.025 –0.050 0.3 0.2 0.1 0 Gain (dB)[...]
-
Página 15
© National Instruments Corp. 15 NI 9229/9239 Stopband The filter signif icantly atten u ates all signals abo ve the stop band freq u ency . The primary goal of the fi lter is to pre vent aliasing. Therefore, the stopband freq u ency scales precisel y with the data rate. The stopband rejection is the minim u m amo u nt of att en u ation applied by [...]
-
Página 16
NI 9229/9239 16 ni.com the mod u les m u st share a single master timebase so u rce. Refer to the software help for information abo u t conf ig u ring the master timebase so u rce for the NI 9229/9239. V isit ni.com/info and enter c s erie s doc for information abo u t C Series doc u mentation. The follo w ing eq u ation pro vides the av ailable da[...]
-
Página 17
© National Instruments Corp. 17 NI 9229/9239 Sleep Mode This mod u le s u pports a low-po wer sleep mode. S u pport for sleep mode at the system le vel depe nds on the chassis that the mod u le is pl u gged into. Refer to the chassis man u al for information abo u t s u pport for sleep mode. If the chassis s u pports sleep mode, refer to the softw[...]
-
Página 18
NI 9229/9239 18 ni.com Specifications The follo wing specifications a re typical for the range – 40 to 70 °C u nless otherwise noted. All v oltages are relativ e to the AI– signal on each channel u nless otherwise noted. The specific ations are the same for the NI 9229 and the NI 9239 u nless otherwise noted. Input Characteristics N u mber of [...]
-
Página 19
© National Instruments Corp. 19 NI 9229/9239 Data rate range ( f s ) u sing external master timebase Minim u m ...................... ......... ..... 390.625 S/s Maxim u m ......... ............ ......... ..... 51.2 kS/s Data rates 1 ( f s ) . ........... ......... ......... ..... , n = 1, 2, …, 31 Inp u t vo ltage ranges (AI+ to AI–) 2 Overv [...]
-
Página 20
NI 9229/9239 20 ni.com Acc u racy , NI 9229 Measurement Conditions Percent of Reading (Gain Error) Percent of Range * (Offset Error) Calibrated max (– 40 to 70 °C) ±0.13% ±0.05% Calibrated typ (25 °C, ±5 °C) ±0.03% ±0.008% Uncalibrated max (– 40 to 70 °C) ±1.2% ±0.55% Uncalibrated typ (25 °C, ±5 ° C) ±0.3% ±0.11% * Range eq u al[...]
-
Página 21
© National Instruments Corp. 21 NI 9229/9239 Acc u racy , NI 9239 Inp u t noise NI 9229 ........... ........... ............ ..... 320 μ V rms NI 9239 ........... ........... ............ ..... 70 μ V rms Stability Gain drift ...... ........... ......... .......... ±5 ppm /°C Offset drift NI 9229 ......... .................. ..... ±150 μ V/?[...]
-
Página 22
NI 9229/9239 22 ni.com Post calibration gain matc h (ch-to-ch, 20 kHz) ...... ........... ........... . 0.22 dB max Crosstalk (1 kHz) ......... ........... .......... –130 dB Phase mismatch (ch-to-ch) NI 9229 ......... ...................... ........ 0.045°/kHz max NI 9239 ......... ...................... ........ 0.075°/kHz max Phase mismatch [...]
-
Página 23
© National Instruments Corp. 23 NI 9229/9239 Alias-free bandwidth ..... ......... .......... 0.453 · f s –3 dB prefilter band width ( f s = 50 kS/s). ......... ........... ............ ..... 24.56 kHz CMRR ( f in = 60 Hz) NI 9229 ........... ........... ............ ..... 116 dB NI 9239 ........... ........... ............ ..... 126 dB SFDR (1 [...]
-
Página 24
NI 9229/9239 24 ni.com Power Requirements Power cons u mption fr om chassis Active mode .............. ......... ........ 740 mW max Sleep mode .......... ......... ........... ... 25 μ W max Thermal dissipation (at 70 °C) Active mode .............. ......... ........ 760 mW max Sleep mode .......... ......... ......... ..... 16 mW ma x Physical [...]
-
Página 25
© National Instruments Corp. 25 NI 9229/9239 Safety Safety Voltages Connect only v oltages that are within the follo wing limits. Channel-to-earth gro u nd isolation Contin u o u s ............ ......... ............ 250 V rms , Meas u rement Category II W ithstand ................. ......... .......... 2,300 V rms , ver i fie d by a 5 s dielectri[...]
-
Página 26
NI 9229/9239 26 ni.com Meas u rement Cate gory I is for meas u rements performed on circ u its not directly connected to the electrica l distrib u tion system referred to as MAINS vo ltage. MAINS is a hazardo u s liv e electric al s u pply system that powers eq u ipment. This ca tegory is for meas u rements of voltages from spec ially protecte d se[...]
-
Página 27
© National Instruments Corp. 27 NI 9229/9239 Safety Standards This prod u ct is designed to meet the req u ire ments of the follo wing standards of safety for electrical eq u ipment for meas u rement, control, and laboratory u se: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safety certif ications, refer to the pro[...]
-
Página 28
NI 9229/9239 28 ni.com Environmental National Instr u ments C Series mod u les are intend ed for indoor u se only b u t may be u sed o u tdoors if installed in a s u itabl e enclos u re. Refer to th e man u al for the chassis yo u are u sing for more information abo u t meeting th ese specif ications. Operating temperat u re (IEC 60068-2-1, IEC 600[...]
-
Página 29
© National Instruments Corp. 29 NI 9229/9239 Shock and Vibration T o meet these specificati ons, yo u m u st panel mo u nt the system and either af fix ferr u les to the ends of the terminal wires or u se the NI 9971 backshell kit to protect the connections. Operating vibration Random (IEC 60068-2-64)......... 5 g rms , 10 to 500 Hz Sin u soidal ([...]
-
Página 30
NI 9229/9239 30 ni.com Note For EMC compliance, operate this de vice with shielded cabling. CE Compliance This prod u ct meets the essent ial req u irem ents of app licable E u ropean directi ves, as amended for CE marking, as follo ws: • 2006/95/EC; Low-V oltage Directive (safety) • 2004/108/EC; Electromagneti c Compatibility Directive (EMC) N[...]
-
Página 31
© National Instruments Corp. 31 NI 9229/9239 For additional en vironmental information, refer to the NI and the Environment W eb page at ni.com/environment . This page contains the en vironmental reg u lations and direct i ves with which NI complies, as well as othe r en vironmental information not incl u ded in this doc u ment. Waste Electrical a[...]
-
Página 32
NI 9229/9239 32 ni.com Calibration Yo u can obtain the calibration cert ificate and information abo u t calibration services for th e NI 9229/9239 at ni.com/ calibration . Calibration interval ............... ........... . 1 year Where to Go for Support The National Instr u ments W eb site is yo u r complete reso u rce for technical s u pport. At n[...]
-
Página 33
© National Instruments Corp. 33 NI 9229/9239 A u stralia 1800 300 800, A u stria 43 662 457990-0, Belgi u m 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Rep u blic 420 224 235 774, Denmark 45 45 76 2 6 00, Finland 358 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000[...]
-
Página 34
© 2007–2008 Nati onal Instruments Corp. All rights reserved. 374184E-01 Jul08 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments t rademarks. Other product and company names ment ioned herein are trad[...]