Emerson Process Management NGA2000 NDIR manuel d'utilisation
- Voir en ligne ou télécharger le manuel d’utilisation
- 96 pages
- 0.89 mb
Aller à la page of
Les manuels d’utilisation similaires
-
Carbon Monoxide Alarm
Emerson Process Management MAN-0114
24 pages 1.16 mb -
Network Card
Emerson Process Management Helicoid Gauges
1 pages 0.23 mb -
Welding Consumables
Emerson Process Management X-STREAM X2 Series
102 pages 8.23 mb -
Network Card
Emerson Process Management ControlWave LP
11 pages 0.72 mb -
Welding Consumables
Emerson Process Management MMI-20014120
78 pages 3.49 mb -
Automobile Electronics
Emerson Process Management rev. 4
32 pages 0.88 mb -
Projection Television
Emerson Process Management MMI-20020964
242 pages 2.93 mb -
Marine Sanitation System
Emerson Process Management LF-Series
104 pages 3.25 mb
Un bon manuel d’utilisation
Les règles imposent au revendeur l'obligation de fournir à l'acheteur, avec des marchandises, le manuel d’utilisation Emerson Process Management NGA2000 NDIR. Le manque du manuel d’utilisation ou les informations incorrectes fournies au consommateur sont à la base d'une plainte pour non-conformité du dispositif avec le contrat. Conformément à la loi, l’inclusion du manuel d’utilisation sous une forme autre que le papier est autorisée, ce qui est souvent utilisé récemment, en incluant la forme graphique ou électronique du manuel Emerson Process Management NGA2000 NDIR ou les vidéos d'instruction pour les utilisateurs. La condition est son caractère lisible et compréhensible.
Qu'est ce que le manuel d’utilisation?
Le mot vient du latin "Instructio", à savoir organiser. Ainsi, le manuel d’utilisation Emerson Process Management NGA2000 NDIR décrit les étapes de la procédure. Le but du manuel d’utilisation est d’instruire, de faciliter le démarrage, l'utilisation de l'équipement ou l'exécution des actions spécifiques. Le manuel d’utilisation est une collection d'informations sur l'objet/service, une indice.
Malheureusement, peu d'utilisateurs prennent le temps de lire le manuel d’utilisation, et un bon manuel permet non seulement d’apprendre à connaître un certain nombre de fonctionnalités supplémentaires du dispositif acheté, mais aussi éviter la majorité des défaillances.
Donc, ce qui devrait contenir le manuel parfait?
Tout d'abord, le manuel d’utilisation Emerson Process Management NGA2000 NDIR devrait contenir:
- informations sur les caractéristiques techniques du dispositif Emerson Process Management NGA2000 NDIR
- nom du fabricant et année de fabrication Emerson Process Management NGA2000 NDIR
- instructions d'utilisation, de réglage et d’entretien de l'équipement Emerson Process Management NGA2000 NDIR
- signes de sécurité et attestations confirmant la conformité avec les normes pertinentes
Pourquoi nous ne lisons pas les manuels d’utilisation?
Habituellement, cela est dû au manque de temps et de certitude quant à la fonctionnalité spécifique de l'équipement acheté. Malheureusement, la connexion et le démarrage Emerson Process Management NGA2000 NDIR ne suffisent pas. Le manuel d’utilisation contient un certain nombre de lignes directrices concernant les fonctionnalités spécifiques, la sécurité, les méthodes d'entretien (même les moyens qui doivent être utilisés), les défauts possibles Emerson Process Management NGA2000 NDIR et les moyens de résoudre des problèmes communs lors de l'utilisation. Enfin, le manuel contient les coordonnées du service Emerson Process Management en l'absence de l'efficacité des solutions proposées. Actuellement, les manuels d’utilisation sous la forme d'animations intéressantes et de vidéos pédagogiques qui sont meilleurs que la brochure, sont très populaires. Ce type de manuel permet à l'utilisateur de voir toute la vidéo d'instruction sans sauter les spécifications et les descriptions techniques compliquées Emerson Process Management NGA2000 NDIR, comme c’est le cas pour la version papier.
Pourquoi lire le manuel d’utilisation?
Tout d'abord, il contient la réponse sur la structure, les possibilités du dispositif Emerson Process Management NGA2000 NDIR, l'utilisation de divers accessoires et une gamme d'informations pour profiter pleinement de toutes les fonctionnalités et commodités.
Après un achat réussi de l’équipement/dispositif, prenez un moment pour vous familiariser avec toutes les parties du manuel d'utilisation Emerson Process Management NGA2000 NDIR. À l'heure actuelle, ils sont soigneusement préparés et traduits pour qu'ils soient non seulement compréhensibles pour les utilisateurs, mais pour qu’ils remplissent leur fonction de base de l'information et d’aide.
Table des matières du manuel d’utilisation
-
Page 1
Instruction Manual 748332-F April 2003 http://www.processanalytic.com Model NGA2000 NDIR Non-Dispersive Infrared Analyzer[...]
-
Page 2
Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com http://www.processanalytic.com ESSENTIAL INSTRUCTIONS READ THIS PAGE BE FORE PROCEEDING! Rosemount Analytical desig ns, ma nufactures and tests its produc ts to [...]
-
Page 3
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Contents i Model NGA2000 NDIR TABLE OF CONTENTS PREFACE........................................................................................................................ ................... P-1 Definitions ................................[...]
-
Page 4
Instruction Manual 748332-F April 2003 ii Contents Rosemount Analytical Inc. A Di vision of Emerson Process Management Model NGA2000 NDIR 4-0 MAINTENANCE AND TROUBLESHOOTING ......................................................................................... 4-1 4-1 Overview .....................................................................[...]
-
Page 5
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii Model NGA2000 NDIR LIST OF ILLUSTRATIONS Figure 1-1. NDIR Techno logy .................................................................................................. 1-2 Figure 1-2. NGA2000 NDIR Analyzer Module (Typical - Actual[...]
-
Page 6
Instruction Manual 748332-F April 2003 iv Contents Rosemount Anal ytical Inc. A Di vision of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 7
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1 Model NGA2000 NDIR PREFACE The purpose of this manual is to provide information concerning th e components, functions, installation and maintenance of the NGA2000 NDIR and the Sy stem Accessories of the NGA2000 System. Some sections[...]
-
Page 8
Instruction Manual 748332-F April 2003 P-2 Preface Rosemount Anal ytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR SAFETY SUMMARY If this equipment is used in a manner not specified in these in structions, protectiv e systems may be impaired. AUTHORIZED PERSONNEL To avoid explosion, loss of life, pe rsonal injury and damage [...]
-
Page 9
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3 Model NGA2000 NDIR CAUTION HAND INJURY H AZARD Dropping the front panel of the Platfo rm while hand or fingers are inside either case handle can cause serious injury. CAUTION PARTS INTEGRITY Tampering with or unauthorized substituti[...]
-
Page 10
Instruction Manual 748332-F April 2003 P-4 Preface Rosemount Anal ytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compre ssed Gas Ass ociation's "Handboo k of Compressed Gases" published in 1981 Com[...]
-
Page 11
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5 Model NGA2000 NDIR NAMUR DOCUMENTATION The following NGA2000 NDIR in stru ction materials are av ailable. Contact Customer Service Center or the local representative to order. 748332 Instruction Manual (this do cument) COMPLIANCES T[...]
-
Page 12
Instruction Manual 748332-F April 2003 P-6 Preface Rosemount Anal ytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR GLOSSARY OF TERMS Analyzer Module The module that contains all sen so r/det ector compon ents for development of a Primary Vari able signal; includes all signal conditioning an d temperature control circuitry. B[...]
-
Page 13
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-7 Model NGA2000 NDIR Softkey s The five function softkeys located belo w the front pan el display; they assume the function displayed dire ctly above each on the display, a function dictated by soft ware. System Any collection of Anal[...]
-
Page 14
Instruction Manual 748332-F April 2003 P-8 Preface Rosemount Anal ytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 15
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Proc ess Management Descrip tion and Specifications 1-1 Model NGA2000 NDIR SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 OVERVIEW This manual describe s the Non-Di spersive Infrared (NDIR) Analyzer Module of Ro semount Analytical's NGA2000 Series of gas analy[...]
-
Page 16
Instruction Manual 748332-F April 2003 1-2 Description and Specifications Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR During analysis, a portion of the infrared radiation is absorbed by the compone nt of interest in the sample. The quantity of infrared radiation that is absorbed is proportional to the comp[...]
-
Page 17
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Proc ess Management Descrip tion and Specifications 1-3 Model NGA2000 NDIR 1-4 SAMPLE REQUIREMENTS Maximum allowable sample pressure is 690 hPa-gauge (10 psig) for a standard configuration NDIR that ha s a flow restri ctor which sets the flow at between 0.5 L/min[...]
-
Page 18
Instruction Manual 748332-F April 2003 1-4 Description and Specifications Rosemount An alytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Figure 1-2. NGA2000 NDIR Analy zer Module (Typical - Actual Configurati on May Vary) CASE TEMPERATURE SENSOR REFERENCE OUT REFERENCE IN PURGE GAS IN CHOPPER MOTOR SOURCE CELLS DETECTOR THER[...]
-
Page 19
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Proc ess Management Descrip tion and Specifications 1-5 Model NGA2000 NDIR 1-7 SPECIFICATIONS a. General Measurement Species: .................. Heteroatomic gases such as ammonia (NH 3 ), carbon dioxide (CO 2 ), carbon monoxide (CO), ca rbon monoxide + carbon di[...]
-
Page 20
Instruction Manual 748332-F April 2003 1-6 Description and Specifications Rosemount An alytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 21
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1 Model NGA2000 NDIR ANALY ZER M ODUL E GUI DES PIN SEATS DISENGAGED FRON T PANEL SECTION 2 INSTALLATION WARNING . Before starting to install this equipment, read the “Essential Instructions” on the inside cover and the Safet[...]
-
Page 22
Instruction Manual 748332-F April 2003 2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 2-3 LOCATION Install the NDIR Analyzer Module in a clean, non-hazardous, weather protecte d, vibration free location free from extreme temperature variations. For best results, install the instrument near the[...]
-
Page 23
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3 Model NGA2000 NDIR Table 2-1. Cell Purging Times at Atmospheric Sample Pressure Figure 2-2. NDIR Back Panel CELL LENGTH CELL VOLUME CC TOTAL VOLUME CC TIME FOR 2 VOLUMES @ 2 SCFH (1L/MIN) mm inch without inlet tube cell with in[...]
-
Page 24
Instruction Manual 748332-F April 2003 2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 2-6 ELECTRICAL CONNE CTIONS NOTE Electrical connections must be in compli- ance with National Electrical Code (ANSI/NFPA 70) and/or any applicable na- tional or electrical codes. Two electrical connectio ns a[...]
-
Page 25
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5 Model NGA2000 NDIR Figure 2-4. NDIR Wiring Diagram SA M P LE J2 -1 SO UR CE J2-2 RE F E R E NC E J 2 - 3 SO UR CE J2-4 POWER SU PPLY BOARD PRESSURE COMPENS ATI ON BOARD J7 J2 J4 J9 J3 J1 1 J8 J1 J15 J1 4 J6 J10 J5 1 2 3 4 1 2 F[...]
-
Page 26
Instruction Manual 748332-F April 2003 2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR .6 [15] 1.6 [40] 20.0 [508.0 ] STA ND AR D 24.8 [628. 7] EXTENDED 17. 41 [142.2 ] STA N DA RD 22.41 [569.2 ] EX TE NDED 1.0 [25] 1.1 [27] 8.4 [ 213 ] 1.1 [ 28 ] 5.6 [143] 2.8 [71] 4.3 [109] .5 [13] 1.1 [28] 6[...]
-
Page 27
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Startu p and Operation 3-1 Model NGA2000 NDIR SECTION 3 STARTUP AND OPERATION 3-1 OVERVIEW Prior to initial startup, t he user should leak test the module as outlined in Section 2. For the remainder of this section, Analyzer Module interconnect[...]
-
Page 28
Instruction Manual 748332-F April 2003 3-2 Startup and Operation Rosemount Analytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR Controller Board and the network "self-install" (bind together) during initial startup. If the system contains more than one Analyzer Module, the startup pr ocedure will interrogate the ne[...]
-
Page 29
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Startu p and Operation 3-3 Model NGA2000 NDIR Figure 3-3. Basic Controls Menu Figure 3-4. Expert Controls and Setup Menu Figure 3-5. Technical Level Configuration Menu 23.2 % CO Analyzer XXXXXXXX Basic Controls Range upper limit: 25% Range and [...]
-
Page 30
Instruction Manual 748332-F April 2003 3-4 Startup and Operation Rosemount Analytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR Figure 3-6. Typical Help Screen 3-5 CALIBRATI ON Calibration can be executed from the Basic Controls menu. Calibration gas data can be entered only through the Expert Contro ls and Setup menu. To ca[...]
-
Page 31
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Startu p and Operation 3-5 Model NGA2000 NDIR DISPLAY MESSAGE DESCRIPTION TYPE BAROMETER System Barometer WARNING CASE TEMP Case Te mperature WARNING CHOP SPEED Choppe r Speed WARNING CRUDE NOISE Calculated Noise WARNING DET SIG Detector Signal[...]
-
Page 32
Instruction Manual 748332-F April 2003 3-6 Startup and Operation Rosemount Analytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR 3-6 LINEARIZATION The NDIR Analyzer Module can be ope rated in linear and non-linea r mo de. Linearization can be toggled ON/OFF in the Expert Controls menu. In the OFF position, linearization is di[...]
-
Page 33
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Startu p and Operation 3-7 Model NGA2000 NDIR CONCENTRATI ON NORMALIZ ED 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 READINGS , NORMAL IZED (Axis Reversed) 0 5 00 100 0 15 00 200 0 2 500 300 0 3 500 400 0 45 [...]
-
Page 34
Instruction Manual 748332-F April 2003 3-8 Startup and Operation Rosemount Analytical Inc. A Divis ion of Emerson Process Management Model NGA2000 NDIR 3-7 ROUTINE OPERATION Set the NDIR Analyzer Module for desired operating range. Zero and span the module, and then supply sample gas to the SAMPLE INLET at the rear of the module. The NDIR Analyzer [...]
-
Page 35
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 3-9 Model NGA2000 NDIR 4 Figure 3-9. Display Screens (1 of 5) Analy zer Span Are you sure? You must hav e zero gas flowi ng through the anal yzer. This control do es NOT control any auto- cal. Module boun d to this analy[...]
-
Page 36
Instruction Manual 748332-F April 2003 3-10 Startup and Operation Rosemount Analytical Inc. A Divi sion of Emerson Process Management Model NGA2000 NDIR Figure 3-10. Display Screens (2 of 5) Zero/Span diagnostic data Date of last zero: 24-08-94 Error message for last zero: 24- 08-94 Error percentag e for last zero: 5% Raw signal at last zero: 5000 [...]
-
Page 37
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 3-11 Model NGA2000 NDIR Figure 3-11. Display Screens (3 of 5) #6 Analyzer Parameter List Primar y Variable Para m eters Control mode: LOCAL Output dela y time: 0 Sec s Range 1 upper lim it: 10 ppm Range 2 upper lim it: 1[...]
-
Page 38
Instruction Manual 748332-F April 2003 3-12 Startup and Operation Rosemount Analytical Inc. A Divi sion of Emerson Process Management Model NGA2000 NDIR Figure 3-12. Display Screens (4 of 5) Service History Control module data... Analyzer m odule data... Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Module Tag Service[...]
-
Page 39
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 3-13 Model NGA2000 NDIR Figure 3-13. Display Screens (5 of 5) Analyzer Diagnostics Power supply voltages ... Primary variable parameters... Physical meas urements... Temperature cont rol param eters ... Miscellaneous co [...]
-
Page 40
Instruction Manual 748332-F April 2003 3-14 Startup and Operation Rosemount Analytical Inc. A Divi sion of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 41
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-1 Model NGA2000 NDIR SECTION 4 MAINTENANCE AND TROUBLESHOOTING DANGER ELECTRICAL SHOCK HAZARD Servicing requires access to liv e parts which can cause death or serious inju ry. Refer servicing to qualified per[...]
-
Page 42
Instruction Manual 748332-F April 2003 4-2 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR TOP VIEW SIDE VIEW POWER SUPPLY BOARD MICRO BOARD PRESSURE COMPENSATION BOARD SIGNAL BOARD Figure 4-1. Printed Circuit Board Fold-Out Panel Views 4-3 POWER FUSE REPLACE MENT Remove power t[...]
-
Page 43
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-3 Model NGA2000 NDIR AIR FLOW 4-4 MODULE FAN REPLACEMENT The Analyzer Module fan assembly is disassembled as sh own in Figure 4-3 b elow. Before doing so, the entire optical ben ch must be removed. See Figure [...]
-
Page 44
Instruction Manual 748332-F April 2003 4-4 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 4-5 CHOPPER MOTOR REPLACEMENT See Figure 4-4 below for a exploded view of the Motor/Source Assembly. The entire optical bench must be removed to perform this replacement. 4-6 SOURCE REPLAC[...]
-
Page 45
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-5 Model NGA2000 NDIR Power Transistor/Heatsink Motor/Source Assembly Detector Assembly Oscillator Tune Adjust Flow Sensor 4-7 DETECTOR REMOVAL See Figure 4-5 below and Figure 4-8 on page 4-9 for breakdowns of [...]
-
Page 46
Instruction Manual 748332-F April 2003 4-6 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 4-8 FLOW SENSOR REPLACEMENT See Figure 4-5 on page 4-5 for Flow Sensor location. To replace Fl ow Se nsor, remove all connecting hardwa re and undo connections to the sample line. 4-9 CASE[...]
-
Page 47
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-7 Model NGA2000 NDIR Resistive Temperature Detector Thermal Fuse Oscillator Board Detector Detector Heater Figure 4-6. Oscillator Tune, S ource Balance Shutter Adjustments Figure 4-7. Detector Block (Exploded [...]
-
Page 48
Instruction Manual 748332-F April 2003 4-8 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 4-12 CLEANING CELLS When zero gas is introduced in the SAMPLE IN inlet and the Detector signal is gre ater 1 .2V, the cells and windows re qui re cleaning. To do this, follow this procedur[...]
-
Page 49
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-9 Model NGA2000 NDIR Spacer Motor/Source Assembly Sample Cell Shutter Block Assembly Detector Assembly Desiccant Holder SHORT PATH CELLS 4mm TO 2.5 INCHES Motor/Source Assembly Sample Cell Shutter Block Assemb[...]
-
Page 50
Instruction Manual 748332-F April 2003 4-10 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR GAS DESICCANT PART NUMBER CO 2 Cardoxide 096218 CO Mg (C10 4 )2 096217 H 2 0 Mg (C10 4 )2 096217 SOS Mg (C10 4 )2 096217 CH 4 Mg (C10 4 )2 096217 Hexane Mg (C10 4 )2 096217 CO + CO 2 Card[...]
-
Page 51
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Pr ocess Management Maintenance and Troubleshooting 4-11 Model NGA2000 NDIR 4-14 MODULATION CHECK The modulation check is a measure of the energy absorption capacity of a gas mixture, where total absorbency is 100% mod ula tion and zero absorbency is 0% modulatio[...]
-
Page 52
Instruction Manual 748332-F April 2003 4-12 Maintenance and Troubleshooting Rosemount Analyt ical Inc. A Division of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 53
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 5-1 Model NGA2000 NDIR SECTION 5 REPLACEMENT PARTS WARNING PARTS INTEGRITY Tampering with or unauthorized substitution of componen ts may adversely affect safe ty of this product. Use only factory -approved components for repa[...]
-
Page 54
Instruction Manual 748332-F April 2003 5-2 Replacement Parts Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR 5-2 REPLACEMENT PARTS While the following sect ions list parts which are common to all NDIR Analyzer Modul es, the configuration number is required when ordering parts whi ch a re specifi c to an indivi[...]
-
Page 55
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emerson Proc ess Management Return of Material 6-1 Model NGA2000 NDIR SECTION 6 RETURN OF MATERIAL 6-1 RETURN OF MATERIAL If factory repair of defective equipment is required, proceed as follo ws: 1. Secure a return authori zati on from a Rosemount Analytical Inc. Sales [...]
-
Page 56
Instruction Manual 748332-F April 2003 6-2 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 57
Instruction Manual 748332-F April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-1 Model NGA2000 NDIR SECTION 7 INDEX A Analyzer Mo dule applying power, 3-1 calibration procedure, 3-4 function, 1-1 installation into Platform, 2-1 installation location, 2-2 more that one in system, 3-1 operating temperature, 2-2 una[...]
-
Page 58
Instruction Manual 748332-F April 2003 7-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 59
WARRANTY Goods and part(s) (excluding con su m ab les) manufac tured by Seller are warranted to be free from defects in workman ship and material under normal use and se rvice for a period of twelve (12) months from the date of shipment by Seller. Consumable s, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warra n t[...]
-
Page 60
Instruction Manual 748332-F April 2003 © Rosemount Analytical Inc. 2003 EMERSON PROCESS MAN A GEMENT Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd. 1 Pandan Crescent Singapore 128461 Re[...]
-
Page 61
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-11 Model NGA2000 NDIR Menu: 29 LINRANGE4 Menu: 30 LINRANGE0 Menu: 31 AMPWRI1[...]
-
Page 62
Instruction Manual 748417-B April 2003 7-12 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 32 FLOCHEK1I1 Menu: 33 FILTER Menu: 34 AM1VI1[...]
-
Page 63
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-13 Model NGA2000 NDIR Menu: 35 AMTEMPI1 Menu: 36 AM2VA Menu: 37 PLIMITSA[...]
-
Page 64
Instruction Manual 748417-B April 2003 7-14 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 38 TLIMITSA Menu: 39 AMMISCI1 Menu: 40 MODTEST[...]
-
Page 65
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-15 Model NGA2000 NDIR Menu: 41 ANALSIMPLE Menu: 42 FILTERI1 Menu: 43 LINRANGE0I1[...]
-
Page 66
Instruction Manual 748417-B April 2003 7-16 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 44 CALFACTORS Menu: 45 R1FACTORS Menu: 46 RN2FACTORS[...]
-
Page 67
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-17 Model NGA2000 NDIR Menu: 47 RN3FACTORS Menu: 48 RN4FACTORS Menu: 49 AMHELPINDEX[...]
-
Page 68
Instruction Manual 748417-B April 2003 7-18 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 50 LINRANGE1I1 Menu: 51 AMTRENDI1 Menu: 52 CALFACTORSI1[...]
-
Page 69
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-19 Model NGA2000 NDIR Menu: 53 APARLST2 Menu: 54 APARLST4 Menu: 55 APARLST5[...]
-
Page 70
Instruction Manual 748417-B April 2003 7-20 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 56 APARLST6 Menu: 57 LISTNOTES Menu: 58 MPARMS2[...]
-
Page 71
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-21 Model NGA2000 NDIR Menu: 59 DISPLAY Menu: 60 ANALSET2 Menu: 61 ESCAPE_HELP[...]
-
Page 72
Instruction Manual 748417-B April 2003 7-22 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 62 MPARMSI1 Menu: 63 TLIMITSIAI1 Menu: 64 BAROM_PARMS[...]
-
Page 73
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-23 Model NGA2000 NDIR Menu: 65 ANALSETI2 Menu: 66 AMHELPINDEX2 Menu: 67 MODTESTI1[...]
-
Page 74
Instruction Manual 748417-B April 2003 7-24 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 68 BAROM_PARMI1 Menu: 69 SW_DIAG Menu: 70 CALI1[...]
-
Page 75
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-25 Model NGA2000 NDIR Menu: 71 AM1VHIST Menu: 72 LINFUNCT Menu: 73 POLYSETUP[...]
-
Page 76
Instruction Manual 748417-B April 2003 7-26 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 74 MIDPOINT1 Menu: 75 MIDPOINT2 Menu: 76 MIDPOINT3[...]
-
Page 77
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-27 Model NGA2000 NDIR Menu: 77 MIDPOINT4 Menu: 78 POLYGAS1 Menu: 79 POLYGAS2[...]
-
Page 78
Instruction Manual 748417-B April 2003 7-28 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 80 POLYGAS3 Menu: 81 POLYGAS4 Menu: 82 POLYGAS5[...]
-
Page 79
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-29 Model NGA2000 NDIR Menu: 83 POLYGAS6 Menu: 84 POLYGAS7 Menu: 85 POLYGAS8[...]
-
Page 80
Instruction Manual 748417-B April 2003 7-30 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 86 POLYGAS9 Menu: 87 POLYGAS0 Menu: 88 EXP_CAL[...]
-
Page 81
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-31 Model NGA2000 NDIR Menu: 89 ZEROI2 Menu: 90 EXP_CAL_DAT Menu: 91 EXP_CAL_DATI[...]
-
Page 82
Instruction Manual 748417-B April 2003 7-32 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 92 UNITS Menu: 93 UNITSI1 Menu: 94 RESET[...]
-
Page 83
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-33 Model NGA2000 NDIR Menu: 95 STORE Menu: 96 RFHIST1A Menu: 97 RFACTORSIA[...]
-
Page 84
Instruction Manual 748417-B April 2003 7-34 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 98 RFHIST2A Menu: 99 RFHIST3A Menu: 100 RFHIST4A[...]
-
Page 85
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-35 Model NGA2000 NDIR Menu: 101 TWEAKI1 Menu: 102 INIT Menu: 103 SW_DIAGI1[...]
-
Page 86
Instruction Manual 748417-B April 2003 7-36 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 104 STOREDPVA Menu: 105 ZERO_NOW2 Menu: 106 SPAN_NOW2[...]
-
Page 87
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-37 Model NGA2000 NDIR Menu: 107 POLYSETI1 Menu: 108 POLYSETI2 Menu: 109 POLYSETI3[...]
-
Page 88
Instruction Manual 748417-B April 2003 7-38 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 110 ANALSETI3 Menu: 111 MPARMS Menu: 112 FLOWLIMITS[...]
-
Page 89
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 7-39 Model NGA2000 NDIR Menu: 113 PLIMITSAI1 Menu: 114 ABOUT Menu: 115 ABOUT1[...]
-
Page 90
Instruction Manual 748417-B April 2003 7-40 Return of Material Rosemount A nalytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR Menu: 116 ALARM1[...]
-
Page 91
Instruction Manual 748417-B April 2003 Rosemount Analytical Inc. A Division of Emers on Process Management Index 8-1 Model NGA2000 NDIR SECTION 8 INDEX A Analyzer Mo dule applying power, 3-1 calibration procedure, 3-4 function, 1-1 installation into Platform, 2-1 installation location, 2-2 more that one in system, 3-1 operating temperature, 2-2 una[...]
-
Page 92
Instruction Manual 748417-B April 2003 8-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management Model NGA2000 NDIR[...]
-
Page 93
Addendum 748461-B April 2003 http://www.processanalytic.com Model NGA2000 PMD, NDIR VAISALA BAROMETRIC PRESSURE TRANSDUCER This addendum serves as an amen dme nt to the Model NGA 2000 PMD and NDIR instru c- tion manuals. The following information should be considere d pa rt of the manual, if your instrument has a Vaisala Barometri c Pressure Transd[...]
-
Page 94
Addendum 748461-B April 2003 Addendum – Vaisala Transducer Rosemount Anal ytical Inc. A Division of Emerson Process Management Model NGA2000 PMD, NDIR 2 of 2 5. Connect the instrument sample inlet tub- ing to the Vaisala Pressure Inlet port. 6. Enable the Barometric Pressure Com - pensation and the Use Local Readi ng op- tions via the Expert Cont[...]
-
Page 95
WARRANTY Goods and part(s) (excluding con su m ab les) manufac tured by Seller are warranted to be free from defects in workman ship and material under normal use and se rvice for a period of twelve (12) months from the date of shipment by Seller. Consumable s, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warra n t[...]
-
Page 96
Instruction Manual 748417-B April 2003 © Rosemount Analytical Inc. 2003 EMERSON PROCESS MAN A GEMENT Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd. 1 Pandan Crescent Singapore 128461 Re[...]