Genesis Advanced Technologies 2006 manuel d'utilisation
- Voir en ligne ou télécharger le manuel d’utilisation
- 22 pages
- 0.23 mb
Aller à la page of
Les manuels d’utilisation similaires
-
Cordless Saw
Genesis Advanced Technologies GCS100
11 pages 0.67 mb -
Speaker
Genesis Advanced Technologies G 928
1 pages 0.22 mb -
Portable Speaker
Genesis Advanced Technologies 350
2 pages 3.02 mb -
Speaker
Genesis Advanced Technologies Gen 928
2 pages 2.11 mb -
Cordless Sander
Genesis Advanced Technologies GMT15A
12 pages 0.79 mb -
Portable Speaker
Genesis Advanced Technologies 350SE
2 pages 1.64 mb -
Bathroom Aids
Genesis Advanced Technologies 2002
22 pages 0.23 mb -
Speaker System
Genesis Advanced Technologies Monitor 2
8 pages 7.65 mb
Un bon manuel d’utilisation
Les règles imposent au revendeur l'obligation de fournir à l'acheteur, avec des marchandises, le manuel d’utilisation Genesis Advanced Technologies 2006. Le manque du manuel d’utilisation ou les informations incorrectes fournies au consommateur sont à la base d'une plainte pour non-conformité du dispositif avec le contrat. Conformément à la loi, l’inclusion du manuel d’utilisation sous une forme autre que le papier est autorisée, ce qui est souvent utilisé récemment, en incluant la forme graphique ou électronique du manuel Genesis Advanced Technologies 2006 ou les vidéos d'instruction pour les utilisateurs. La condition est son caractère lisible et compréhensible.
Qu'est ce que le manuel d’utilisation?
Le mot vient du latin "Instructio", à savoir organiser. Ainsi, le manuel d’utilisation Genesis Advanced Technologies 2006 décrit les étapes de la procédure. Le but du manuel d’utilisation est d’instruire, de faciliter le démarrage, l'utilisation de l'équipement ou l'exécution des actions spécifiques. Le manuel d’utilisation est une collection d'informations sur l'objet/service, une indice.
Malheureusement, peu d'utilisateurs prennent le temps de lire le manuel d’utilisation, et un bon manuel permet non seulement d’apprendre à connaître un certain nombre de fonctionnalités supplémentaires du dispositif acheté, mais aussi éviter la majorité des défaillances.
Donc, ce qui devrait contenir le manuel parfait?
Tout d'abord, le manuel d’utilisation Genesis Advanced Technologies 2006 devrait contenir:
- informations sur les caractéristiques techniques du dispositif Genesis Advanced Technologies 2006
- nom du fabricant et année de fabrication Genesis Advanced Technologies 2006
- instructions d'utilisation, de réglage et d’entretien de l'équipement Genesis Advanced Technologies 2006
- signes de sécurité et attestations confirmant la conformité avec les normes pertinentes
Pourquoi nous ne lisons pas les manuels d’utilisation?
Habituellement, cela est dû au manque de temps et de certitude quant à la fonctionnalité spécifique de l'équipement acheté. Malheureusement, la connexion et le démarrage Genesis Advanced Technologies 2006 ne suffisent pas. Le manuel d’utilisation contient un certain nombre de lignes directrices concernant les fonctionnalités spécifiques, la sécurité, les méthodes d'entretien (même les moyens qui doivent être utilisés), les défauts possibles Genesis Advanced Technologies 2006 et les moyens de résoudre des problèmes communs lors de l'utilisation. Enfin, le manuel contient les coordonnées du service Genesis Advanced Technologies en l'absence de l'efficacité des solutions proposées. Actuellement, les manuels d’utilisation sous la forme d'animations intéressantes et de vidéos pédagogiques qui sont meilleurs que la brochure, sont très populaires. Ce type de manuel permet à l'utilisateur de voir toute la vidéo d'instruction sans sauter les spécifications et les descriptions techniques compliquées Genesis Advanced Technologies 2006, comme c’est le cas pour la version papier.
Pourquoi lire le manuel d’utilisation?
Tout d'abord, il contient la réponse sur la structure, les possibilités du dispositif Genesis Advanced Technologies 2006, l'utilisation de divers accessoires et une gamme d'informations pour profiter pleinement de toutes les fonctionnalités et commodités.
Après un achat réussi de l’équipement/dispositif, prenez un moment pour vous familiariser avec toutes les parties du manuel d'utilisation Genesis Advanced Technologies 2006. À l'heure actuelle, ils sont soigneusement préparés et traduits pour qu'ils soient non seulement compréhensibles pour les utilisateurs, mais pour qu’ils remplissent leur fonction de base de l'information et d’aide.
Table des matières du manuel d’utilisation
-
Page 1
USER’S MANUAL GENES I S 2002 TM GENES I S 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com 0609[...]
-
Page 2
GENESIS 2006 SHOWN 2[...]
-
Page 3
To assure proper record keeping, please retain this information for future use. DATE PURC HASED ____________________________ SERIAL # _____________________________________ DISTRIBUT OR NAME __________________________ PHONE NUMBER ______________________________ CITY, STATE __________________________________ The GENESIS THAWING SYSTEMS incorporate wa[...]
-
Page 4
CONTENTS page 1. UNPACKING ........................................................ 5 2. TO THAW BLOOD PRODUCTS ........................ 5 3. TO REMOVE A THAWED BAG ........................ 9 4. TO MAINTAIN THE DEVICE IN THE READY TO THAW STATUS .............................. 9 5. TO CLEAN THE BATH OR TO ADD WATER .. 10 6. CONTROL PANEL AND INDICAT[...]
-
Page 5
1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006 unit. C o n t r o l U n i t W a t e r B a t h T h a w i n g B a g D i s p e n s e r Plasma Overwrap Thawing Bags (1000/Box) P l a s m a B a g H o l d e r L A B a t h C L E A N TM R e f e r e n c e T h e r m o m e t e r T h a w i n g B a g W e i g h t s U s e r ' s M a n[...]
-
Page 6
indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEAN TM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prev ent the formation of turbidity and odor. 2.3 Plug the device into a 115 volt AC outlet. For GENESIS 2002 , a dedica[...]
-
Page 7
c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 ° C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other. The display should now show the new SET TEMP. When th[...]
-
Page 8
thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time. f. To change the default TIMER : Press the START/CHANGE key for the selected bag location, [...]
-
Page 9
3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder. Discard plasma thawing overwrap bag after each use, as they are designed for single use only. b. If there is another bag being thawed, press the ROCK key to resume the rocking motion for the rem ai[...]
-
Page 10
5. TO CLEAN THE BATH OR TO ADD WATER S A F E T Y : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL) IS TURNED ON. RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET. 5.1 TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF. BEFORE REMOVING THE[...]
-
Page 11
when filled with water. Do not drop the bath on a hard surface. GENESIS 2006 : Remove all, or part of the water. The bath weighs approximately thirty (30) pounds when filled with water. Carefully remove the water bath from the device. 5.6 DO NOT USE BLEACH OR HARSH CLEANERS. Depending on mineral content of water used, the bath may discolor. Pro[...]
-
Page 12
6. CONTROL PANEL AND INDICATORS GENESIS 2002 TM CONTROL PANEL 12[...]
-
Page 13
GENESIS 2006 TM CONTROL PANEL 13[...]
-
Page 14
CONTROL KEYS POWER .......... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER SWITCH on the left panel is a master on/off switch to remove electrical power completely from all circuits in the unit. SET TEMP ..... With this key switch you can adjust water temperature within the range of ambient to +[...]
-
Page 15
INDICATORS HEATING ....... This red indicator is lit when the heating element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off. FAULT ............ If the water temperature exceeds 58 ° C, an alarm will sound and the red fault light will be illuminated. This m[...]
-
Page 16
2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch. 3) The FAULT light is on. • Check the water temperature. If the w[...]
-
Page 17
7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrum ent. 8) HEATING light is flashing. • This is normal operation. At approximately one degree below the set temperature, the microprocessor begins[...]
-
Page 18
8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time setting for each bag location • Forced water circulation with gentle rocking motion • Compact size - saves valuable bench space • Quick heating • Simple to operate - easy to read LED display • Li[...]
-
Page 19
10. SPECIFICATIONS GENESIS 2002 TM GENESIS 2006 TM External Dim ension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” High Interior Bath Dimension 13.0” L x 7.0” Wide x 7.5” Deep 18.0” L x 8.0” Wide x 7.5” Deep Bath Volume 2.5 US gallons 3.7 US gallons Controlled Tem p Range Ambient to 56 o C Ambient to 56 o C [...]
-
Page 20
11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com RETURN INFORMATION : Same as above. 1. Obtain a Return Goods Authorization number. 2. Ensure all water is [...]
-
Page 21
Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006 was purchased S e r i a l # RGA # (Return Goods Authorization #) 6. Mark outside of box clearly, with your return address, and send to GenesisBPS, LLC . (as above) 21[...]
-
Page 22
GENESIS 2002 GENESIS 2006 PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006 , Plasma Thawing Water Bath System is warranted against defects in materials and workmanship for a period of one year from the date of purchase. The warranty applies only to products that have not been tampered with, used for purposes other than th[...]