Go to page of
Similar user manuals
-
Thermometer
Braun IRT4520USSM
42 pages 1.71 mb -
Thermometer
Braun Thermoscan 6023
86 pages -
Thermometer
Braun ThermoScan IRT 6500
20 pages 3 mb -
Thermometer
Braun PRT 1000 High speed
1 pages 1.92 mb -
Thermometer
Braun 950787
27 pages 0.19 mb -
Thermometer
Braun Thermoscan 6022
86 pages -
Thermometer
Braun PRT 2000 - Age Precision
49 pages 24.03 mb -
Thermometer
Braun 6024
19 pages 1.76 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Braun 950787, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Braun 950787 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Braun 950787. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Braun 950787 should contain:
- informations concerning technical data of Braun 950787
- name of the manufacturer and a year of construction of the Braun 950787 item
- rules of operation, control and maintenance of the Braun 950787 item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Braun 950787 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Braun 950787, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Braun service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Braun 950787.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Braun 950787 item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
Vista ™ basic 950787 Rev H 02/03 I n s t r u c t i o n s f o r U s e Rx Only[...]
-
Page 2
Vista ™ basic Instructions for Use 1/25 950787 Rev H 02/03 The CE mark confirms compliance with the “Council Directive on Medical Products 93/42/EEC” dated the 14 th of June, 1993. B. Braun Melsungen AG RESPONSIBILITY OF MANUFACT URER Manufacturer, assembly, and installation personnel or instructo rs can only be held responsi ble for any effe[...]
-
Page 3
Vista ™ basic Contents Vista ™ basic Instructions for Use 2/25 950787 Rev H 02/03 CONTENTS Warnings & C autions .................................................................. 3 Introducti on ................................................................................ 5 Preparing the In fusion ........................................[...]
-
Page 4
WARNINGS & CAUTIONS Vista ™ basic Instructions for Use 3/25 950787 Rev H 02/03 The following should be observe d to avoid patient/user injury. This Instruction s for Use manual contains detailed in structions and warning s on the use of the Vista ™ basic Infusion Pump. Please read it c o mpletely prior to using this device. This manual is i[...]
-
Page 5
WARNINGS & CAUTIONS Vista ™ basic Instructions for Use 4/25 950787 Rev H 02/03 ♦ After starting an infusion, make sure drops are falling in the drip chamber. If no drops are falling, make sure the roller clamp is open. If the roller clamp is open and still no drops are falling, replace and dispose of the set. ♦ The air detector cannot rec[...]
-
Page 6
INTRODUCTION Vista ™ basic Instructions for Use 5/25 950787 Rev H 02/03 Handle for carrying Display All essential information is displayed in plain language. Green background lighting illuminates only when connected to AC power, or, in battery operation, only when a key is pressed. Keyboard Allows data entry Clear Key Resets a value to zero Alarm[...]
-
Page 7
INTRODUCTION Vista ™ basic Instructions for Use 6/25 950787 Rev H 02/03 Door Open Button Aluminum Case Easy upkeep, drip proof, and resistant to disinfectants Function Keys Allow choices of YES, NO, OK, INFO, OPTS, etc. ON/OFF ( Power ) Ke y Alarm Silence Key Silences alarm tone for 2 minutes START/STOP Ke y[...]
-
Page 8
PREPARING THE INFUSION Vista ™ basic Instructions for Use 7/25 950787 Rev H 02/03 Prepare the Pump 1. Check for cleanliness, com pleteness, and damage. If unit appears damaged, do not use it. 2. Attach the pump to an IV pole by turning the knob on the pole clam p clockwise, or place the pump on a flat, stable surfa ce. 3. Plug the pump’s power [...]
-
Page 9
Continuous Mode Vista ™ basic Instructions for Use 8/25 950787 Rev H 02/03 When the pump is turned on, it may display: Tubing Type VISTA Pump Set Confirm you are using a Vista™ Pump Set, then press YES . The pump may then display: Is This a New Therapy? YES Begins in Continuous Mode with all data cleared. NO Returns to the beginning of the last[...]
-
Page 10
Continuous Mode Vista ™ basic Instructions for Use 9/25 950787 Rev H 02/03 chamber. VTBD and TIME 1. Press VTBD. 2. Enter bag volume in ml. 3. Press OK. (To leave screen without saving ch ang es, press EXIT.) 4. Press TIME. 5. Enter infusion time in hours and minutes. N OTE : For a 2 hour infusion, enter 2, 0, 0. The pump will sho w: Time: 02h00m[...]
-
Page 11
Continuous Mode Vista ™ basic Instructions for Use 10/25 950787 Rev H 02/03 KEEP VEIN OPEN RATE After an infusion is complete, the pump will alarm and begin pum ping at 1.0 ml/hr or the current infusion rate, whichever is less. 1. Stop the infusion. 2. Turn off the pump or reset the program. INFORMATION KEY This key is only available while the pu[...]
-
Page 12
OPTIONS Piggyback Vista ™ basic Instructions for Use 11/25 950787 Rev H 02/03 The Vista ™ basic contains many option s to allow cu stom program ming. To access the various options, press OPTS while the pu mp is sto pped. Pressing OPTS each time will move you to the next option. Eventually, the OPTS key will bring you back to the mode from which[...]
-
Page 13
OPTIONS Piggyback, Ramp/Taper Mode Vista ™ basic Instructions for Use 12/25 950787 Rev H 02/03 To Exit Piggyback 1. Press to stop the infusio n. 2. Press END. 3. Press OPTS. 4. Press OFF to turn Piggyback off 5. Press END to return to Continuous Mode. RAMP/TAPER MODE The Ramp/Taper Mode is desig ned to de liver infusions, such as Total Parenteral[...]
-
Page 14
OPTIONS Ramp/Taper Mode Vista ™ basic Instructions for Use 13/25 950787 Rev H 02/03 To Clear Ramp/Taper Mode 1. Press OPTS until display shows RAMP TAPER MODE. 2. Press CLR. 3. Press YES to clear all the data in Ramp/Taper Mode. 4. Press ON and continue to program a ne w Ramp/Taper program. To Exit Ramp/Taper Mode 1. When infusion is complete, pr[...]
-
Page 15
OPTIONS Program Mode Vista ™ basic Instructions for Use 14/25 950787 Rev H 02/03 PROGRAM MODE Program Mode is for infusions requiring up to 9 sep arate rates and volumes to be delivered one after another. This is desi gned to deliver infusions for step ped oncology drugs, circadian medication deliv ery, and ot her cu stom programs. 1. Press OPTS [...]
-
Page 16
OPTIONS Delay A Mode Vista ™ basic Instructions for Use 15/25 950787 Rev H 02/03 DELAY A MODE The start of Continuous, Piggyback, Ramp/Taper, an d Program Modes may all be delayed. The Vista ™ basic also provides an optio n to extend the Infusion Complete alarm at the end of the infusion. 1. Program the infusion. 2. Press OPTS until disp lay sh[...]
-
Page 17
OPTIONS Dosage Calculation Vista ™ basic Instructions for Use 16/25 950787 Rev H 02/03 DOSAGE CALCULATION This feature allows a rate to be calculated given the con centration and dose. 1. Press OPTS until display shows DOSAGE CALC. 2. Press ON. 3. If concentration unit desired is other than the one displayed, pre ss button under the mcg? until un[...]
-
Page 18
OPTIONS Bolus Mode, Interval Bo lus (Intermittent Mode) Vista ™ basic Instructions for Use 17/25 950787 Rev H 02/03 BOLUS MODE This feature allows a bolus to be given during a Continuous Mode infusion. 1. Press OPTS until display shows BOLUS MODE. 2. Press ON. 3. Press RATE. 4. Enter the rate at which the bolus should i nfuse, then p ress OK. 5. [...]
-
Page 19
OPTIONS Total Infused, Standby Function Vista ™ basic Instructions for Use 18/25 950787 Rev H 02/03 TOTAL INFUSED The Total Infused information screen display s the total amount of fluid in ml that has been pumped since the last time the Total Infused wa s cl eared. To Clear the Total Infused 1. Press OPTS until display shows Tot Infused=. 2. Pre[...]
-
Page 20
OPTIONS Occlusion Pressure, Drop Sensor, Battery Vista ™ basic Instructions for Use 19/25 950787 Rev H 02/03 OCCLUSION PRESSURE The pressure at which a downstream occlusion alar m will be triggered may be varied. The higher the pressure limit, the less sen sitive the pump is to changes in fluid resistan ce. 1. Press OPTS until displa y shows OCCL[...]
-
Page 21
OPTIONS Data Lock, Contrast Selection Vista ™ basic Instructions for Use 20/25 950787 Rev H 02/03 DATA LOCK This features prevents unauthorized u se by lockin g out all keys except , ,and . 1. Press OPTS until disp lay show s DATA L OCK. 2. Press ON. 3. Press END to return to Continuous Mode. To Turn Off Data Lock 1. Press OPTS until disp lay sho[...]
-
Page 22
OPTIONS Adjust Clock, Loudness Vista ™ basic Instructions for Use 21/25 950787 Rev H 02/03 ADJUST CLOCK This feature allows the time and date to be changed. 1. Press OPTS until displa y show s ADJUST CLO CK. 2. Press DAT to enter the date. (Press C to clear the current data and re- enter new data.) 3. Enter the date using two digits for month , d[...]
-
Page 23
OPTIONS Alarms Vista ™ basic Instructions for Use 22/25 950787 Rev H 02/03 In alarm situations, press to silence the alarm. Correct the proble m. Resume the infusion by pressing . Alarm Possible Cause Action Air-In-Line! Air bubble in air-in-line sensor Remove air from system Check Drop Sensor! Free Flow! Not Enough Drops! Too Many Drops! Solutio[...]
-
Page 24
AC, 12 V, BATTERY, & CC INTERFACE Vista ™ basic Instructions for Use 23/25 950787 Rev H 02/03 AC and 12 V ♦ Check AC voltage as per rating label. ♦ Insert/screw on power lea d (12 V lead in ambulance car) on back. Battery Operation ♦ Switching to internal battery is automatic in the event of power failure or if 12 V or AC is not connect[...]
-
Page 25
MAINTENANCE, STORAGE, & SERVICE Vista ™ basic Instructions for Use 24/25 950787 Rev H 02/03 Maintenance Clean the pump with a soft, lint-free cloth or swab dampened with soap and water, a general non staining chemical disinfectant, isopropyl alcohol (15%), derivatives of phenylphenol (5%) or ammonium chloride (10%). As a general recommendatio[...]
-
Page 26
SPECIFICATIONS Vista ™ basic Instructions for Use 25/25 950787 Rev H 02/03 Mechanism Linear peristaltic volumetric infusion pump Dimensions WxHxD 5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm) Weight 6.8 lbs. (3.1 kg) Power Requirements 100-120 VAC, 50-60 Hz, 13 Watts Leakage Current Max 30µA Battery 7.2V, 1.2 Ah NiCd (rechargeable) Battery Opera[...]
-
Page 27
Vista ™ basic Instructions for Use 950787 Rev H 02/03 ©2001 B. Braun Medical Inc. B. Braun Medical Inc. 824 Twelfth Ave. Bethlehem, PA 18018 Tel 800-227-2862 www.bbraunusa.com[...]