Go to page of
Similar user manuals
-
Clock Radio
Emerson CKD2330
36 pages 0.47 mb -
Fan
Emerson CF755TG05
12 pages 0.18 mb -
Fan
Emerson CF980MBX 00
16 pages 0.32 mb -
Power Supply
Emerson 1500
32 pages 1.58 mb -
Fan
Emerson CF905BS00
16 pages 0.39 mb -
DVD Recorder
Emerson EWR10D5
2 pages 0.26 mb -
Fan
Emerson CF1AB01
12 pages 0.33 mb -
Power Supply
Emerson 582145500
76 pages 15.87 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Emerson 54ea, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Emerson 54ea one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Emerson 54ea. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Emerson 54ea should contain:
- informations concerning technical data of Emerson 54ea
- name of the manufacturer and a year of construction of the Emerson 54ea item
- rules of operation, control and maintenance of the Emerson 54ea item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Emerson 54ea alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Emerson 54ea, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Emerson service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Emerson 54ea.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Emerson 54ea item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
Model 54eA Amperometric HAR T ® Analyzer/Controller Instruction Manual 51-54eA/rev .I April 2005[...]
-
Page 2
ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards. Because these instru- ments are sophisticated technical product s, you must properly install, use, and maintain them to ensure they continue to operate within their normal sp[...]
-
Page 3
[...]
-
Page 4
MODEL 54eA T ABLE OF CONTENTS MODEL 54eA ANAL YZER/CONTROLLER T ABLE OF CONTENTS Section Title Page 1.0 SPECIFICA TIONS ................................................................................................... 1 1.1 Features and Applications ....................................................................................... 1 1.2 S pe[...]
-
Page 5
9.0 CALIBRA TION - TOT AL CHLORINE ..................................................................... 60 9.1 Introduction .............................................................................................................. 60 9.2 Zeroing the Sensor ........................................................................................[...]
-
Page 6
iii MODEL 54eA T ABLE OF CONTENTS LIST OF FIGURES Section Title Page 2-1 W all Mounting .......................................................................................................... 5 2-2 Pipe Mounting .......................................................................................................... 6 2-3 Pipe and W all Mountin[...]
-
Page 7
1 MODEL 54eA SECTION 1.0 SPECIFICA TIONS SECTION 1.0 SPECIFICA TIONS 1.1 FEA TURES AND APPLICA TIONS The Model 54eA Analyzer/Controller with the appro- priate sensor monitors and controls dissolved oxy- gen (ppm and ppb level), free chlorine, total chlorine, and ozone in a variety of process liquids. The analyzer is compatible with Rosemount Analyt[...]
-
Page 8
MODEL 54eA SECTION 1.0 SPECIFICA TIONS 2 1.2 SPECIFICA TIONS - GENERAL Enclosure: Epoxy-painted (light gray) cast aluminum, NEMA4X (IP65). 144 x 144 x 132 mm (5.7 x 5.7 x 5.2 in.), DIN size. Front Panel: Membrane keypad with t actile feedback. Three green LEDs indicate alarm status. Red LED indicates fault condition. Display: Three-line, back-lit, [...]
-
Page 9
MODEL 54eA SECTION 1.0 SPECIFICA TIONS SPECIFICA TIONS — OXYGEN Measurement Range: 0-99 ppm (mg/L), 0-200% sat- uration Resolution: 0.01 ppm, 0.1 ppb for 499A T rDO sen- sor T emperature correction for membrane permeabili- ty: automatic between 0 and 50°C (can be dis- abled) Calibration: automatic air calibration or calibration against a standar[...]
-
Page 10
MODEL 54eA SECTION 1.0 SPECIFICA TIONS 4 1.3 ORDERING INFORMA TION The Model 54eA Microprocessor Analyzer measures dissolved oxygen, free chlorine, total chlorine, ozone, and pH. pH is available for free chlorine only . The analyzer has an on-board pressure sensor for automatic air calibration of oxygen sensors. Amperometric measurements are fully [...]
-
Page 11
5 MODEL 54eA SECTION 2.0 INST ALLA TION SECTION 2.0 INST ALLA TION 2.1 UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present. If items are missing, not[...]
-
Page 12
6 FIGURE 2-2. Pipe Mounting MODEL 54eA SECTION 2.0 INST ALLA TION 2.2.3 Pipe mounting 9.52 .375 SIDE VIEW FRONT VIEW 4 MOUNTING HOLES WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH PIPE MOUNTING PN 2002577 W ALL MOUNTING FIGURE 2-3. Pipe and W all Mounting Dimensions DWG . NO. REV . 40005402 E[...]
-
Page 13
MODEL 54eA SECTION 2.0 INST ALLA TION FIGURE 2-4. Panel Mounting 2.2.4 Panel mounting WHEN INCH AND METRIC DIMS ARE GIVEN MILLIMETER INCH FIGURE 2-5. Panel Mounting Dimensions 144 5.7 35 1.38 50.8 2 1.52 .06 33 1.3 66 2.6 28.72 1.17 139.7 5.5 139.7 5.5 155.7 6.13 137.9 5.43 68.96 2.715 137.9 5.43 144 5.7 FRONT VIEW BOTTOM VIEW SIDE VIEW P ANEL SUPP[...]
-
Page 14
8 MODEL 54eA SECTION 3.0 WIRING SECTION 3.0 WIRING 3.1 GENERAL W ARNING Electrical installation must conform to the National Electrical Code, all st ate and local codes, and all plant codes and standards for electrical equipment. Electrical inst allation and wiring must be done by qualified personnel. The five holes in the bottom of the instrument [...]
-
Page 15
9 MODEL 54eA SECTION 3.0 WIRING FIGURE 3-1. Power Input and Relay Output Wiring for Model 54eA DWG . NO. REV . 454EPH02 D[...]
-
Page 16
10 MODEL 54eA SECTION 3.0 WIRING FIGURE 3-3. Amperometric sensors with st andard cable. FIGURE 3-4. Amperometric sensors with optimum EMI/RFI cable or V ariopol cable. FIGURE 3-2. Wiring Label 3.3 SENSOR WIRING 3.3.1 General The wiring label, which is shown in Figure 3-2, is a general purpose label. It has wiring infor- mation concerning other sens[...]
-
Page 17
11 MODEL 54eA SECTION 3.0 WIRING 3.3.3 Wiring 499ACL-01 (free chlorine) sensors and pH sensors for automatic pH correction. If free chlorine is being measured and the pH of the liquid varies more than 0.2 pH unit, a continuous correction for pH must be applied to the chlorine reading. Therefore, a pH sensor must also be wired to the 54eA controller[...]
-
Page 18
12 MODEL 54eA SECTION 3.0 WIRING FIGURE 3-6. Free chlorine sensor with st andard cable and 399-14 pH sensor having internal preamplifier . FIGURE 3-7. Free chlorine sensor with st andard cable and 399-09-62 pH without internal preamplifier . Although the blue wire is connect- ed to a terminal labeled solution ground , the blue wire does not connect[...]
-
Page 19
MODEL 54eA SECTION 3.0 WIRING FIGURE 3-9. Free chlorine sensor with optimum EMI/RFI cable or V ariopol cable and 399-14 pH sensor having internal preamplifier . FIGURE 3-8. Free chlorine sensor with optimum EMI/RFI cable or V ariopol cable and 399VP-09 pH sensor without internal preamplifier . 13 Although the blue wire is connect- ed to a terminal [...]
-
Page 20
14 MODEL 54eA SECTION 3.0 WIRING FIGURE 3-10. Free chlorine sensor with optimum EMI/RFI cable or V ariopol cable and 399-09-62 pH sensor without internal preamplifier . DWG . NO. REV . 4054eA04 A[...]
-
Page 21
3.3.4 Wiring Hx438 and Gx448 sensors Hx438 and Gx448 steam-sterilizable dissolved oxygen sensors use a 22k NTC thermistor . The thermistor is wired to terminals 1 and 3 on TB5. FIGURE 3-1 1. Hx438 and Gx448 sensors. MODEL 54eA SECTION 3.0 WIRING 15 DWG . NO. REV . 4054eA07 A[...]
-
Page 22
16 MODEL 54eA SECTION 4.0 DISPLA Y AND OPERA TION SECTION 4.0 DISPLA Y AND OPERA TION 4.1 GENERAL DESCRIPTION The 54eA analyzer/controller is a normally a single input, dual output instrument. It does, however , accept a second input for pH when the main measurement is free chlorine. Figure 4-1 shows how the instrument inputs and output s can be co[...]
-
Page 23
17 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION SECTION 5.0 SOFTW ARE CONFIGURA TION The instrument is configured at the factory to measure oxygen. Figure 5-1 is an outline of the menu structure. T able 5-1 list s the default settings and the range of choices available for each setting. T o reduce the chance of error when configuring the control[...]
-
Page 24
18 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION Continued on the following page ITEM CHOICES F ACTOR Y SETTINGS CONFIGURE A. Display options (Section 5.5) 1. Measurement Oxygen, ozone, free chlorine, total chlorine, monochloramine Oxygen 2. Sensor (Oxygen only) Rosemount standard, Rosemount biopharm, or Rosemount standard other steam sterilizabl[...]
-
Page 25
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION ITEM CHOICES F ACTOR Y SETTINGS CONFIGURE (continued) C. Alarms (Section 5.7) (continued) 3. Alarm 1 setup (TPC) a. Setpoint if oxygen (ppm) -20 to 20 ppm 1 ppm if oxygen (ppb) -999 to 999 ppb 100 ppb if oxygen (% saturation) 0 to 200% 100% if chlorine, monochloramine, or ozone -9999 to 9999 ppm 1.0 p[...]
-
Page 26
20 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION T ABLE 5-1. Program Settings List (continued) ITEM CHOICES F ACTOR Y SETTINGS CONFIGURE (continued) D. pH measurement (Section 5.8) 1. pH sensor Enable or disable Enable 2. pH compensation (free chlorine only) Auto or manual Auto 3. pH value 0.00 - 14.00 7.00 4. pH diagnostic On or off Off a. Glass[...]
-
Page 27
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION FIGURE 5-1. Menu T ree for the 54eA Controller Calibrate Program (see page 23) Diagnostic V ariables Main Menu Calibrate main sensor Zero main sensor Adjust temperature Calibrate pH Slope Buffer calibration S tandardize Main measurement Main sensor current Sensitivity (μA/ppm) Zero current pH reading[...]
-
Page 28
22 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION Program Calibrate (see page 21) Diagnostic V ariable (see page 21) Main Menu Alarms 1, 2, and 3 setpoints 4 mA or 0 mA 20 mA Present output current Alarm Setpoints Output setpoints T est output 1 or 2 T est alarm 1, 2, 3, or 4 Simulated tests Configure Display Main Sensor Oxygen Free Chlorine T ota[...]
-
Page 29
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION FIGURE 5-1. Menu T ree for the 54eA Controller (continued) Display (see page 22) Outputs (see page 22) Alarms (see page 22) Program Calibrate (see page 21) Diagnostic V ariable (see page 21) Main Menu Alarm Setpoints (see p age 22) Output setpoints (see page 22) Simulated tests (see page 22) Configure[...]
-
Page 30
24 5.1 CHANGING ALARM SETPOINTS 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Press Enter (F4). 4. Move the cursor to the desired alarm and press Enter (F4). 5. The screen appearing at this point depends on how the alarm was configured. 6. If the alarm is a normal (i.e., not TPC) alarm, a s[...]
-
Page 31
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.2 RANGING THE OUTPUTS 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Move the cursor to "Output setpoint s" and press Enter (F4). 4. Move the cursor to the desired output and press Enter (F4). 5. This screen confirms that changes to[...]
-
Page 32
26 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.3 CHANGING OUTPUT SETPOINTS (PID ONL Y) 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Move the cursor to "Output setpoint s" and press Enter (F4). 4. Move the cursor to the desired output and press Enter (F4). 5. This screen con[...]
-
Page 33
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.4 TESTING OUTPUTS AND ALARMS 2. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 3. Move the cursor to "Simulated test s" and press Enter (F4). 4. Move the cursor to the desired output or alarm. Both output s and all four alarms can be test[...]
-
Page 34
28 Output setpoint s Simulated test s Configure Exit Enter MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.5 CHOOSING DISPLA Y OPTIONS 3. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. With the cursor on "Display", pr[...]
-
Page 35
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.5 CHOOSING DISPLA Y OPTIONS (CONTINUED) 8. Set the remainder of the display parameters. Use the and keys to choose the desired parameter . Then press Edit (F4). Use the key to move the cursor to the desired selection. Press Save (F4) to store. SECURITY CAUTION The controller uses the tim[...]
-
Page 36
30 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.6 CHANGING OUTPUT P ARAMETERS 3. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. Move the cursor to "Outputs" and press Enter (F4). 6. Five menu headers relate to outp[...]
-
Page 37
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.6 CHANGING OUTPUT P ARAMETERS (continued) e. The display returns to the “Output: Process” screen. Press Exit (F1). The display returns to the “Output Measurement” screen. Move the cursor to "Control mode" and press Enter (F4). f. Press Edit (F4). Use the key to toggle between &[...]
-
Page 38
32 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION Integral: Integral is the number of seconds over which devia- tions from the setpoint are integrated to remove continuing off- sets. Smaller values provide higher response. Derivative: Derivative is a form of control that resists all changes in readings. Higher readings increase the derivative func[...]
-
Page 39
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.7 CHANGING ALARM P ARAMETERS 3. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 4. Move the cursor to "Configure" and press Enter (F4). 5. Move the cursor to "Alarms" and press Enter (F4). 6. Nine menu headers relate to alarms. A[...]
-
Page 40
34 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.7 CHANGING ALARM P ARAMETERS (continued) c. T o change the activation method, press Edit (F4). Use the key to scroll through the choices: "Process", "T emperature", and “pH” (if pH was enabled). "Process" means the measurement made by the main sensor (oxygen,[...]
-
Page 41
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.7 CHANGING ALARM P ARAMETERS (continued) 9. Alarm setup for TPC alarms: a. Move the cursor to the desired alarm setup and press Enter (F4). b. Use the and keys to move the cursor to the desired param- eter . Press Edit (F4). Use the arrow keys to change the setting to the desired value and p[...]
-
Page 42
36 Feed limit : Disable T imeout: 3600 sec Exit Edit 1.000 ppm 26.2°C 12.0mA Feed limit alarm 1 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.7 CHANGING ALARM P ARAMETERS (continued) Example: The setpoint is 3.00 ppm. The UR V is +5.00 ppm and the LRV is 0.00 ppm. The time period is 30 seconds. When the concentration is 4.00 ppm, the relay wil[...]
-
Page 43
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.7 CHANGING ALARM P ARAMETERS (continued) 1 1. Interval timer setup: Alarm 1, 2, or 3 can be used as an interval timer . The selected relay will open and close at time intervals programmed by the user . The interval timer is useful for automatic cleaning of sensors. NOTE The alarm relay used for the [...]
-
Page 44
38 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.8 CONFIGURING THE pH MEASUREMENT 1. Press any key to enter the main menu. Move the cursor to "Program" and press Enter (F4). 2. Move the cursor to "Configure" and press Enter (F4). 3. Move the cursor to “pH” and press Enter (F4). 4. The default settings are “pH sensor: E[...]
-
Page 45
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.8 CONFIGURING THE pH MEASUREMENT (continued) b. Move the cursor to "Diagnostics". Press Edit (F4) and use the key to change "Off" to "On". Press Save (F4). NOTE Choosing "On" means the controller will display pH diagnos- tic warning messages and fault alar[...]
-
Page 46
40 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.8 CONFIGURING THE pH MEASUREMENT (continued) T emp coef f : 0.000 Operate iso: 7.00 pH Sensor iso: 7.00 pH Exit Enter St abilize pH and S tabilize time: For the controller to accept cal- ibration data, the pH must remain within a specified range for a specified period of time. The default values [...]
-
Page 47
T emp comp : Auto T emp unit s: °C Exit Edit T emp comp: Manual T emp unit s: °C T emperature : 25.0°C Exit Edit MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.9 TEMPERA TURE COMPENSA TION AND TEMPERA TURE UNITS 1. Refer to Section 6.1 for a discussion of the ways in which temperature affect s amperometric and pH meas- urements. pH T emperatu[...]
-
Page 48
42 Noise rejection : 60 Hz Exit Edit MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.10 NOISE REDUCTION 1. For maximum noise reduction the frequency of the ac power must be entered into the analyzer . T emperature Noise rejection Main sensor cal Exit Enter 2. Press any key to enter the main menu. Move the cursor to "Program" and press E[...]
-
Page 49
S tabilize : 0.050 ppm S tabilize time: 10 sec Exit Edit S tabilize: 0.050 ppm S tabilize time: 10 sec Dual range cal : Disable Exit Edit S tabilize: 0.050 ppm S tabilize time: 10 sec Salinity : 0.0 o /oo Exit Edit MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.1 1 MAIN SENSOR CALIBRA TION P ARAMETERS 1. Main sensor refers to the amperometric se[...]
-
Page 50
44 Bar meas : Auto Bar unit s: mm Hg Exit Edit MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.12 BAROMETRIC PRESSURE 1. The barometric pressure menu header appears only if an oxygen sensor (ppm or ppb level) is being used. Barometric pressure is used during air calibration. Noise rejection Barometric pressure Main sensor cal Exit Enter 2. Press [...]
-
Page 51
MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.13 SECURITY 1. The controller can be programmed to require a password for access to menus. There are three levels: Level 1: A level 1 user can 1. Zero and calibrate the main (amperometric) sensor 2. Calibrate the barometric pressure sensor 3. Calibrate the pH sensor 4. Enter a temperature slope for [...]
-
Page 52
46 MODEL 54eA SECTION 5.0 SOFTW ARE CONFIGURA TION 5.14 CONTROLLER MODE PRIORITY The Model 54eA controller can function in different modes depending on both how it is configured, what process conditions exist, and actions an operator may have made. T o reconcile these possible modes, there is a set priority that determines exactly what will happen [...]
-
Page 53
47 MODEL 54eA SECTION 6.0 CALIBRA TION - TEMPERA TURE SECTION 6.0 CALIBRA TION - TEMPERA TURE 6.1 INTRODUCTION All four amperometric sensors (oxygen, ozone, free chlorine, total chlorine, and monochloramine) are membrane- covered sensors. As the sensor operates, the analyte (the substance to be determined) diffuses through the mem- brane and is con[...]
-
Page 54
48 6.2 TEMPERA TURE CALIBRA TION MODEL 54eA SECTION 6.0 CALIBRA TION - TEMPERA TURE 2. From the main display , press any key . With the cursor on “Calibrate,” press Enter (F4). NOTE If Hold was enabled in Section 5.6, the hold screen will appear . T o activate hold, refer to Section 5.6, step 1 1. 3. Press the key twice to move the cursor t[...]
-
Page 55
MODEL 54eA SECTION 7.0 CALIBRA TION - DISSOL VED OXYGEN SECTION 7.0 CALIBRA TION - DISSOL VED OXYGEN 7.1 INTRODUCTION As Figure 7-1 shows, oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero st an- dard) a[...]
-
Page 56
50 1. Place the sensor in a fresh solution of 5% sodium sulfite (Na 2 SO 3 ) in water . Be sure air bubbles are not trapped against the membrane. The current will drop rapidly at first and then gradually reach a st able zero value. T o monitor the sensor current, go to the main display . Press any key to obtain the main menu. Press the key once[...]
-
Page 57
1. Remove the sensor from the process liquid. Use a soft tissue and a stream of water from a wash bottle to clean the membrane. Blot dry . The membrane must be dry during air calibration. 2. Pour some water in a beaker and suspend the sensor with the membrane about 0.5 inch (1 cm) above the water surface. T o avoid drift caused by temperature chang[...]
-
Page 58
52 MODEL 54eA SECTION 7.0 CALIBRA TION - DISSOL VED OXYGEN 9. During calibration, the analyzer stores the measured current and cal- culates the sensitivity . Sensitivity is the sensor current in nA divided by the saturation concentration of oxygen in ppm. The t able gives typical sensitivity for Rosemount Analytical dissolved oxygen sensors. T o vi[...]
-
Page 59
The analyzer and sensor can be calibrated against a standard instrument. For oxygen sensors inst alled in aera- tion basins in waste treatment plants, calibration against a second instrument is of ten preferred. For an accurate calibration be sure that… 1. The standard instrument has been zeroed and calibrated against water-saturated air followin[...]
-
Page 60
54 1. If the barometric pressure measured by the controller does not agree with the local barometric pressure, cal- ibrate the pressure sensor . A pressure error of 3 mm Hg introduces an error of about 0.5% in the final meas- urement. When calibrating the pressure reading, be sure to use the actual barometric pressure. Weather fore- casters and air[...]
-
Page 61
MODEL 54eA SECTION 8.0 CALIBRA TION - FREE CHLORINE SECTION 8.0 CALIBRA TION - FREE CHLORINE 8.1 INTRODUCTION As Figure 8-1 shows, a free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero st an- dard) a[...]
-
Page 62
56 1. Place the sensor in the zero standard (see Section 8.1). Be sure no air bubbles are trapped against the mem- brane. The sensor current will drop rapidly at first and then gradually reach a st able zero value. T o monitor the sensor current, go to the main display . Press any key to obtain the main menu. Press the key once to high- light ?[...]
-
Page 63
1. Place the sensor in the process liquid. If automatic pH correction is being used, calibrate the pH sensor (see Section 1 1.0) and place it in the process liquid. If manual pH correction is being used, measure the pH of the process liquid and enter the value. See Section 5.8. Adjust the sample flow until it is within the range recommended for the[...]
-
Page 64
58 MODEL 54eA SECTION 8.0 CALIBRA TION - FREE CHLORINE Figure 8-2 shows the principle of dual slope cali- bration. Between zero and concentration C1, the sensor response is linear . When the concentration of chlorine becomes greater than C1, the response is non-linear . In spite of the non-linearity , the response can be approximated by a straight [...]
-
Page 65
MODEL 54eA SECTION 8.0 CALIBRA TION - FREE CHLORINE 7. Once the reading is stable, the screen at lef t appears. Sample the process liquid. Make a note of the reading before taking the sample. Immediately determine free chlorine. Note the controller reading again. If the present reading (X) differs from the reading when the sample was taken (Y), cal[...]
-
Page 66
60 MODEL 54eA SECTION 9.0 CALIBRA TION - TOT AL CHLORINE SECTION 9.0 CALIBRA TION - TOT AL CHLORINE 9.1 INTRODUCTION T ot al chlorine is the sum of free and combined chlorine. The continuous determination of total chlorine requires two steps. See Figure 9-1. First, the sample flows into a conditioning system (SCS 921) where a pump continuously adds[...]
-
Page 67
1. Complete the startup sequence described in the SCS921 instruction manual. Adjust the sample flow to between 80 and 100 mL/min, and set the sample pressure to between 3 and 5 psig. 2. Remove the reagent uptake tube from the reagent bottle and let it dangle in air . The perist altic pump will sim- ply pump air into the sample. 3. Let the system ru[...]
-
Page 68
62 1. If the sensor was just zeroed, place the reagent uptake tube back in the bottle. Once the flow of reagent start s, it takes about one minute for the sensor current to begin to increase. It may take an hour or longer for the reading to st abilize. Be sure the sample flow stays between 80 and 100 mL/min and the pressure is between 3 and 5 p sig[...]
-
Page 69
MODEL 54eA SECTION 9.0 CALIBRA TION - TOT AL CHLORINE 9.4 DUAL SLOPE CALIBRA TION Figure 9-3 shows the principle of dual slope calibration. Between zero and con- centration C1, the sensor response is linear . When the concentration of chlo- rine becomes greater than C1, the response is non-linear . In spite of the non-linearity , the response can b[...]
-
Page 70
64 MODEL 54eA SECTION 9.0 CALIBRA TION - TOT AL CHLORINE 7. Once the reading is stable, the screen at lef t appears. Sample the process liquid. Make a note of the reading before taking the sample. Immediately determine total chlorine. Note the controller reading again. If the present reading (X) differs from the reading when the sample was taken (Y[...]
-
Page 71
MODEL 54eA SECTION 10.0 CALIBRA TION - MONOCHLORAMINE SECTION 10.0 CALIBRA TION - MONOCHLORAMINE 65 10.1 INTRODUCTION As Figure 10-1 shows, a monochloramine sensor generates a current directly proportional to the concentration of monochloramine in the sample. Calibrating the sensor requires exposing it to a solution containing no monochlo- ramine ([...]
-
Page 72
66 MODEL 54eA SECTION 10.0 CALIBRA TION - MONOCHLORAMINE 1. Place the sensor in the zero standard. Be sure no air bubbles are trapped against the membrane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. T o monitor the sensor current, go to the main display . Press any key to obtain the main menu. Press [...]
-
Page 73
MODEL 54eA SECTION 10.0 CALIBRA TION - MONOCHLORAMINE 1. Place the sensor in the process liquid. Adjust the sample flow until it is within the range recommended for the sensor . Refer to the sensor instruction sheet. 2. Adjust the monochloramine concentration until it is near the upper end of the control range. W ait until the con- troller reading [...]
-
Page 74
68 MODEL 54eA SECTION 1 1.0 CALIBRA TION - OZONE SECTION 1 1.0 CALIBRA TION - OZONE 1 1.1 INTRODUCTION As Figure 1 1-1 shows, an ozone sensor generates a current directly proportional to the concentration of ozone in the sample. Calibrating the sensor requires exposing it to a solution containing no ozone (zero st andard) and to a solution containi[...]
-
Page 75
1. Place the sensor in the zero standard. Be sure no air bubbles are trapped against the membrane. The sensor current will drop rapidly at first and then gradually reach a stable zero value. T o monitor the sensor current, go to the main display . Press any key to obtain the main menu. Press the key once to highlight “Diagnostic vari- ables.?[...]
-
Page 76
70 1. Place the sensor in the process liquid. Adjust the sample flow until it is within the range recommended for the sensor . Refer to the sensor instruction sheet. 2. Adjust the ozone concentration until it is near the upper end of the control range. W ait until the controller read- ing is stable before st arting the calibration. 1 1.3 FULL SCALE[...]
-
Page 77
MODEL 54eA SECTION 12.0 CALIBRA TION - pH SECTION 12.0 CALIBRA TION - pH 12.1 INTRODUCTION A new pH sensor must be calibrated before use. Regular recalibration is also necessary . A pH measurement cell (pH sensor and the solution to be measured) can be pictured as a battery with an extreme- ly high internal resistance. The voltage of the battery de[...]
-
Page 78
72 4.02 pH Buf1done: 4.00 @25°C Abort Cont 4.02 pH Autocal buf fer 1 Abort Cont Adjust temperature Calibrate pH Output trim Exit Enter Buf fer calibration S tandardize pH pH slope Exit Enter 1. Be sure the pH feature has been enabled. See Section 5.8. 2. Obtain two buffer solutions. Ideally the buf fer pH values should bracket the range of pH valu[...]
-
Page 79
MODEL 54eA SECTION 12.0 CALIBRA TION - pH 9. Remove the sensor from buffer 1, rinse it with water , and place it in buffer 2. Swirl the sensor . Press Cont (F3). "W ait" flashes until the reading is stable. 10. The screen at left appears once the reading is stable. Use the or key to change the reading to the nominal pH of the buff[...]
-
Page 80
74 1. Be sure the pH feature has been enabled. See Section 5.8. 2. The controller comes from the factory set for automatic pH calibration. T o do a manual calibration, the factory default setting must be changed. Refer to Section 5.8. 3. Obtain two buffer solutions. Ideally the buf fer pH values should bracket the range of pH values to be meas- ure[...]
-
Page 81
MODEL 54eA SECTION 12.0 CALIBRA TION - pH 10. The screen at left appears if the pH reading in step 9 was changed. Press Pt2 (F3). Go to step 1 1. 1 1. Rinse the sensor and thermometer with water and place them in buffer 2. Be sure the glass bulb and junction are completely sub- merged. Swirl the sensor . Press Cont (F3). "Wait" flashes un[...]
-
Page 82
76 1. The pH measured by the controller can be changed to match the reading from a second or referee instrument. The process of making the two reading agree is called standardization, or one-point calibration. 2. During standardization, the dif ference between the two pH values is converted to the equivalent voltage. The voltage, called the referen[...]
-
Page 83
1. If the electrode slope is known from other measurements, it can be entered directly into the controller . The slope must be entered as the slope at 25°C. T o calculate the slope at 25°C from the slope at temperature t°C, use the equation: slope at 25°C = (slope at t°C) Changing the slope overrides the slope determined from the previous buff[...]
-
Page 84
78 MODEL 54eA SECTION 13.0 CALIBRA TION - CURRENT OUTPUTS SECTION 13.0 CALIBRA TION - CURRENT OUTPUTS 13.1 INTRODUCTION Although the controller outputs are calibrated at the factory , they can be trimmed in the field to match the reading from a standard current meter . Both the low output (0 or 4 mA) and the high output (20 mA) can be trimmed. 13.2[...]
-
Page 85
79 MODEL 54eA SECTION 14.0 PID AND TPC CONTROL SECTION 14.0 PID AND TPC CONTROL 14.1 PID CONTROL (CODE -20) PID Control The Model 54eA current outputs can be programmed for PID control. PID control is used with a control device that is capable of varying it s output from 0 to 100 percent in response to a changing signal in mil- liamps. Automated co[...]
-
Page 86
80 MODEL 54eA SECTION 14.0 CALIBRA TION - CONTROL Proportional (Gain) Plus Integral (Reset) For the automatic elimination of deviation, I (Integral mode), also referred to as Reset, is used. The propor- tional function is modified by the addition of automatic reset. With the reset mode, the controller continues to change its output until the deviat[...]
-
Page 87
MODEL 54eA SECTION 14.0 CALIBRA TION - CONTROL 6. When sufficient dat a have been collected, return the output signal to its original value using the sim- ulate test function. Maintain the controller in this manual mode until you are ready to initiate auto- matic PID control, after you have calculated the tuning constant s. Once these steps are com[...]
-
Page 88
82 MODEL 54eA SECTION 14.0 CALIBRA TION - CONTROL concentration, temperature) compared to the measurement range, the dif ference between the 20 mA (Hi) and 4 (or 0) mA (Lo) setpoint s, which you determined when configuring the analog output. In the example shown in Figure 14-1: The percent change in pH was: x 100% = = 33.3% The change in the output[...]
-
Page 89
MODEL 54eA SECTION 14.0 CALIBRA TION - CONTROL 14.2 TIME PROPORTIONAL CONTROL (TPC) MODE (Code -20) In the TPC mode, you must est ablish the following parameters which will determine how the Model 54eA controller responds to your system (see Section 5.7): • Setpoint • T ime period • UR V point (or 100% on) • LR V point (or 0% on) • Propor[...]
-
Page 90
84 MODEL 54eA SECTION 15.0 TROUBLESHOOTING SECTION 15.0 TROUBLESHOOTING 15.1 OVERVIEW The 54eA controller continuously monitors it self and the sensor for faults. When the controller detects a fault in the amperometric or pH sensor or in the instrument itself it displays a fault message . If alarm 4 was enabled, the red F AIL LED will also light an[...]
-
Page 91
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.2.1 High input current Excessive sensor current implies that the amperometric sensor is miswired or the sensor has failed. V erify that wiring is correct, including connections through a junction box. See Section 3.3. If wiring is correct, try replacing the sensor . 15.2.2 Check sensor zero The sensor curr[...]
-
Page 92
86 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.2.8 Sense line open Most Rosemount Analytical sensors use a Pt100 or a Pt1000 in a three-wire configuration (see Figure 15-4). The in and return leads connect the RTD to the measuring circuit in the analyzer . A third wire, called the sense line, is connected to the return lead. The sense line allows th[...]
-
Page 93
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.4.1 Zero current is too high A. Is the sensor properly wired to the analyzer? See Section 3.3. B. Is the membrane completely covered with zero solution and are air bubbles not trapped against the mem- brane? Swirl and tap the sensor to release air bubbles. C. Is the zero solution fresh and properly made? Z[...]
-
Page 94
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.4.5 Barometric pressure reading is too high or too low . A. Is the pressure inside the enclosure equal to ambient pressure? The pressure sensor is inside the controller enclo- sure. When cable glands are in place and the front panel is tightly closed, the enclosure is moderately airtight. Therefore, as the[...]
-
Page 95
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.5 TROUBLESHOOTING WHEN NO F AUL T MESSAGE IS SHOWING - FREE CHLORINE 15.5.1 Zero current is too high A. Is the sensor properly wired to the controller . See Section 3.3. B. Is the zero solution chlorine-free? T ake a sample of the solution and test it for free chlorine level. The con- centration should be [...]
-
Page 96
90 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.5.3 Sensor can be calibrated, but the current is too low A. Is the temperature low or is the pH high? Sensor current is a strong function of pH and temperature. The sen- sor current decreases about 3% for every °C drop in temperature. Sensor current also decreases as pH increases. Above pH 7, a 0.1 uni[...]
-
Page 97
MODEL 54eA SECTION 14.0 TROUBLESHOOTING 15.5.6 Sensor does not respond to changes in chlorine level. A. Is the grab sample test accurate? Is the grab sample representative of the sample flowing to the sensor? B. Is the pH compensation correct? If the controller is using manual pH correction, verify that the pH value in the controller equals the act[...]
-
Page 98
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.7 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING — MONOCHLORAMINE 15.7.1 Zero current is too high A. Is the sensor properly wired to the analyzer? See Section 3.3. B. Is the zero solution monochloramine-free? T ake a sample of the solution and test it for monochloramine level. The concentration should [...]
-
Page 99
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.7.3 Sensor can be calibrated, but the current is too low A. Is the temperature low? The sensor current decreases about 5% for every °C drop in temperature. B. Sensor current depends on the rate of sample flow past the sensor tip. If the flow is too low , monochloramine readings will be low . Refer to the [...]
-
Page 100
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.7.7 Readings are too low . A. W as the sample tested as soon as it was t aken? Monochloramine solutions are moderately unstable. T est the sample immediately after collecting it. Avoid exposing the sample to sunlight. B. When was the sensor fill solution last replaced? The monochloramine sensor loses sensi[...]
-
Page 101
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.8 TROUBLESHOOTING WHEN NO F AUL T MESSAGE IS SHOWING - OZONE 15.8.1 Zero current is too high A. Is the sensor properly wired to the controller . See Section 3.3. B. Is the zero solution ozone free? T est the zero solution for ozone level. The concentration should be less than 0.02 ppm. C. Has adequate time[...]
-
Page 102
96 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.8.4 Process readings are erratic A. Readings are often erratic when a new sensor or a rebuilt sensor is first placed in service. The current usual- ly stabilizes af ter a few hours. B. Is the sample flow within the recommended range? High sample flow may cause erratic readings. Refer to the sensor instr[...]
-
Page 103
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.9 TROUBLESHOOTING WHEN NO F AUL T MESSAGE IS SHOWING - pH 15.9.1 W arning or error message during two-point calibration. Once the two-point (manual or automatic) calibration is complete, the controller automatically calculates the sen- sor slope (at 25°C). If the slope is less than 45 mV/pH, the controlle[...]
-
Page 104
98 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.9.2 W arning or error message during two-point calibration. During standardization, the millivolt signal from the pH cell is increased or decreased until it agrees with the pH reading from a reference instrument. A unit change in pH requires an of fset of about 59 mV . The controller limits the offset t[...]
-
Page 105
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.9.6 Calibration was successful, but process pH is grossly wrong and/or noisy . Grossly wrong or noisy readings suggest a ground loop (measurement system connected to earth ground at more than one point), a floating system (no earth ground), or noise being brought into the analyzer by the sensor cable. The [...]
-
Page 106
100 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.10 TROUBLESHOOTING NOT RELA TED TO MEASUREMENT PROBLEMS Problem Action Display segments missing Replace display board Alarm relays are chattering 1. Check alarm setpoints. 2. Increase hysteresis time delay settings (see Section 5.7) Incorrect current output 1. V erify that output load is less than 600 [...]
-
Page 107
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.12 SIMULA TING INPUTS - OTHER AMPEROMETRIC MEASUREMENTS T o check the performance of the controller , use a decade box and a battery to simulate the current from the sen- sor . The battery , which opposes the polarizing voltage, is necessary to ensure that the sensor current has the cor- rect sign. A. Disc[...]
-
Page 108
102 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.13 SIMULA TING INPUTS - pH 15.13.1 General This section describes how to simulate a pH input into the controller . T o simulate a pH measurement, connect a standard millivolt source to the controller . If the controller is working properly , it will accurately measure the input voltage and convert it t[...]
-
Page 109
MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.14 SIMULA TING TEMPERA TURE 15.14.1 General. The 54eA controller accepts either a Pt100 R TD (for pH, 499ADO, 499A T rDO, 499ACL-01, 499ACL-02, 499ACL-03, and 499AOZ sensors) or a 22k NTC thermistor (for Hx438 and Gx448 DO sensors and most steam-sterilizable DO sensors from other manufacturers). The Pt100 [...]
-
Page 110
104 MODEL 54eA SECTION 15.0 TROUBLESHOOTING 15.15 MEASURING REFERENCE VOL T AGE Some processes contain subst ances that poison or shift the potential of the reference electrode. Sulfide is a good example. Prolonged exposure to sulfide converts the ref- erence electrode from a silver/silver chloride electrode to a silver/silver sulfide electrode. Th[...]
-
Page 111
MODEL 54eA SECTION 16.0 MAINTENANCE SECTION 16.0 MAINTENANCE REPLACEMENT P ARTS P ART NUMBER DESCRIPTION 23540-05 Enclosure, Front with Keyboard 23848-00 Power Supply Circuit Board Shield 23849-00 Half Shield, Power Supply 23969-02 PCB, CPU and power supply , calibrated, 1 15/230 V ac 23969-06 PCB, CPU and power supply , calibrated, 24 Vdc 33281-00[...]
-
Page 112
MODEL 54eA SECTION 17.0 RETURN OF MA TERIAL SECTION 17.0 RETURN OF MA TERIAL 17.1 GENERAL. T o expedite the rep air and return of instruments, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a R e t urn Materials Authorization (RMA) number . 17.2 W ARRANTY REP [...]
-
Page 113
W ARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defect s in workman- ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller . Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be [...]
-
Page 114
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r[...]