Go to page of
Similar user manuals
-
Speaker
Fluke Rev. 4
118 pages 1.93 mb -
Switch
Fluke 2205A
154 pages 7.29 mb -
DVD Recorder
Fluke 1750
60 pages 3.42 mb -
Thermometer
Fluke 233
16 pages 0.65 mb -
CD Player
Fluke 1652C
2 pages 0.02 mb -
Computer Monitor
Fluke RPM4
150 pages 2.38 mb -
Thermometer
Fluke 5623A
17 pages 0.09 mb -
Network Card
Fluke 2645A
353 pages 11.42 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Fluke 9150, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Fluke 9150 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Fluke 9150. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Fluke 9150 should contain:
- informations concerning technical data of Fluke 9150
- name of the manufacturer and a year of construction of the Fluke 9150 item
- rules of operation, control and maintenance of the Fluke 9150 item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Fluke 9150 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Fluke 9150, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Fluke service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Fluke 9150.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Fluke 9150 item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
9150 Portable Furnace User’ s Guide Rev. 5B1601 Hart Scientific[...]
-
Page 2
Rev. 5B1601 Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. • Copyright © 2005 • Printed in USA[...]
-
Page 3
T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W ARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 CA UTIONS [...]
-
Page 4
8.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 8.2.2 Set-point V alue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 8.2.3 T emperature Scale Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 8.3 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8[...]
-
Page 5
10.2.1 Stabilization and Accurac y . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 11 Calibration Procedur e . . . . . . . . . . . . . . . . . . . . . . 45 11.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . 45 11.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 45 12 Maintenance . . . . . . .[...]
-
Page 6
1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the International Electri cal Symbols. Some or all of these symbols may be used on the instrument or in this manu al. Symbol Description AC ( Alternating Current) AC- DC Batter y CE Complies with European Uni on Directiv es DC Double Insulated Electric Shock Fuse PE Ground Hot Surf ace (Bur n Haz[...]
-
Page 7
Symbol Description Canadian Standards Associati on O VERV OL T AGE (Installat ion) CA TEGORY II, P ollution Deg ree 2 per IEC1010-1 r e - f ers to the le vel of Impulse Wit hstand V oltage protect ion provided. Equipment of O VERV OL T AGE CA TEGOR Y II is energy-consuming equipment to be suppli ed from the fix ed installat ion. Examples include ho[...]
-
Page 8
DO NO T operate this unit without a properly grounded, properly polarized power c ord. DO NO T connect this unit to a non-grounded, non-polarized outlet. HIGH V OL T A GE is used in the operati on of this equipment. SEVERE IN - JUR Y OR DEA TH may result if personnel fail to observe safety precautions. Before working inside t he equipment, turn the[...]
-
Page 9
Allo w for test probe expansi on inside the well as the furnace heats. DO NO T u se fluids to clean out the well. Ne ver introduce foreign mat erial into the probe hole of the insert. Fluids, etc. can leak into the calibrator causing d amage. DO NO T change the values of the calibration constant s from the factory set values. The correct setting of[...]
-
Page 10
Chao Y ang District Beijing 100004, PRC CHIN A Phone: +86-10-6-512-3436 T elefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Of fice Service Center 60 Alexandra T errace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 T elefax: +65 6799-5588 E-mail: antng@singa.fluke.com [...]
-
Page 11
2 Intr oduction The Hart Scientif ic 9150 thermocouple furnace can be used for calibrating ther - mocouple and R TD temperature probes. Cal ibrations may be done ov er a range of 150°C to 1200°C (302°F to 2192°F). T emperature displa y of the 9150 is 0.1 degrees belo w 1000°and 1 degrees abov e 1000°. The furnace features: • Rapid heating a[...]
-
Page 12
3 Specifications and En vir onmental Conditions 3.1 Specifications Temperature Range 150–1200°C (302–2192°F) Display Resolution 0.1° to 999.9°, 1° above 1000° Stability ±0.5°C Display Accuracy ±5.0°C Well Diameter 1.25" (32 mm) Well Depth 4" (102 mm) Heating Time 35 minutes to 1200°C Cooling Time 140 minutes with block Well[...]
-
Page 13
3.3 W arranty Fluke Corporati on, Hart Scientif ic Di vision (Hart) warrants t his product to be free from defects in materia l and workmanship under normal use and service for a period as stated in our current product catal og from the date of shipment. This warrant y extends onl y to the original purchaser and shall not apply to an y product whic[...]
-
Page 14
4 Safety Guidelines • Operate the instrument in room temperatures between 5-50°C (41-122°F). Allow suf ficient air circulation by leaving at least 6 inches of space be - tween the instrument and nearby objects. Overhead clearance needs to al - low for safe and easy insertion and removal of probes for calibration. • The furnace is a precision [...]
-
Page 15
5 Quic k Start 5.1 Unpac king Unpack the furnace carefully and insp ect it for any damage that may hav e oc - curred during shipment. If there i s shipping damage, notify the carrier immediately . V erify that the follo wing components are present: • 9150 Furnace • 3150, Insert • Insert Insulator • Power Cord • Manual • RS-232 Cable 5.2[...]
-
Page 16
5.4 Setting the T emperature Section 8 explains in detail how to set the temperature set-point on the calibra - tor using the front panel ke ys. The procedure is summarized here. (1) Press “SET” twice to access the set-point value. (2) Press “SET” to move the cursor to the units that need changing. (3) Press “UP” or “DOWN” to change[...]
-
Page 17
6 P ar ts and Contr ols The user should become familiar with the f urnace back panel, front panel, and constant temperature block assembly . 6.1 Bac k P anel The back panel (Figure 1) features the Power Entry Module (PEM) that con - tains the power cord socket, the p o w er switch, and the heater voltage switch, the serial port, and the f an. Pow e[...]
-
Page 18
Fa n - The f an inside the calibrator runs conti nuously when the unit is being op - erated to provi de cooling for the inst rument. Slots at t he top and sides of the calibrator are provi ded for airflow . The area around the calibrator must be kept clear to allow adequate ventilation. The airflo w is directed upward and can be extremely hot. 6.2 [...]
-
Page 19
EXIT – Used to exi t from a menu. When EXIT is pressed an y changes made to the displayed v alue are ignored. 6.3 Constant T emperature Block Assembly The constant temperature block assemb ly is sho wn in Figure 3 and consists of remov able inserts. 6.3.1 Constant T emperature Block The “Block” is made of aluminum-oxide and pro vides a relati[...]
-
Page 20
• Insert A Model 3150-2 (variety block): 1/2”, 1/4”, 3/8”, 3/16”, 1/8”, and 1/16” holes • Insert B Model 3150-3 (comparison block): tw o each 3/8”, 1/4”, and 3/16” holes or • Insert C Model 3150-4 (1/4” comparison block): six 1/4” holes 9150 Portable Furnace User’s Guide 18[...]
-
Page 21
7 General Operation 7.1 Changing Displa y Units The 9150 can display temperature i n Celsius or F ahrenheit. The temperature units are shipped from t he factory set to Celsi us. T o change to Fahrenheit or back to Celsius there are two ways: 1 - Press the “SET” and “DO WN” simultaneously . or 1 - Press the “SET” key three times from the[...]
-
Page 22
8 Contr oller Operation This section discusses in d etail ho w to operate the furn ace temperature control - ler using the front control panel. By using the front panel ke y-switches and LED display the user may monitor the well temperature, adjust t he set-point temperature in de grees C or F , monitor the heater output power , adjust the con - tr[...]
-
Page 23
9150 Portable Furnace User’s Guide 22 Figure 4 Controller Operation Flowchar t[...]
-
Page 24
200.0 C W ell temperatur e in de gr ees Celsius S Access set-point memory 1. 200. Set-point memory 1, 200.0°C currently used T o change the set-point memory press “UP” or “DO WN”. 5. 900. New set-point memory 5, 900.0°C Press “SET” to accept the ne w selection and access the set-point value. S Accept selected set-point memory 8.2.2 Se[...]
-
Page 25
Scale units curr ently selected Press “UP” or “DO WN” to change the units. Un= F New units selected Press “EXIT” to display the well temperature or press “SET” to access the scan control. 8.3 Scan The scan rate can be set and enabl ed so that when the set-point is changed the furnace heats or cools at a specified rate (degrees per m[...]
-
Page 26
Press “SET” to accept the ne w scan rate and continue. S Accept scan rate 8.4 Ramp and Soak Pr ogram The ramp and soak program feature for the 9150 al lows the user to program a number of set-points, cycle t he furnace au tomatically between th e t emperatures at a scan rate set by the user , and hold the furnace at each temperature for a pe - [...]
-
Page 27
8.4.3 Pr ogram Set-P oints The controller all ows t he user to adjust up to eight program points. These are accessed by pressing “SET” after setting the number of program points as de - scribed in Section 8.4.2. Each program point has three associated paramet ers: the program set-point, the program scan rate, and the program hold (or soak) time[...]
-
Page 28
it can be adjusted. 00200 Pr ogram point 4 soak time set for 200 minutes Press “SET” to sav e the new soak-time v alue or “EXIT” to discard changes S Accept the pr ogram point soak time The next v alue to edit is the program scan rate. This value is ignored if scan is not enabled for the unit (See Section 8.3.1 ). Sr 4 Pr ogram point 4 scan[...]
-
Page 29
Pr ogram mode Press “SET” to adjust the program mode and the “UP” or “DO W N” b uttons to change the mode. Pf=4 New mode Press “SET” to continue or “EXIT” to continue without sa ving the ne w va lue. S Save new setting 8.4.5 Pr ogram Control The final paramete r in the program menu is the control parameter . Y ou may choose betw[...]
-
Page 30
”EXIT” simultaneously and release. The heater power is displayed as a percent - age of full power . 962.4 W ell temperatur e S+E Access heater power in secondary menu 100.0 P Heater power in percent T o exit out of t he secondary menu press “EXIT”. T o continue on to the set-point voltage setting function press “SET”. 8.7 Set-point V ol[...]
-
Page 31
troller cannot respond very well to cha nging conditions or noise in the system. If the proportional band is t oo narrow t he temperature may swing back and forth because the controller overreacts to t emperature v ariations. For best con - trol stability the proportional band must be set for the optimum width. The proportional band width is set at[...]
-
Page 32
bration parameters. The groups are selected using the “UP” and “DO WN” ke ys a n dt h e np r e s s i n g“ S E T ” . 8.10 Operating P arameters The operating parameters menu is indicated by: Par Operating parameter s menu Press “SET” to enter the menu. The operating parameters menu contains the HL (High Limit) parameter , the Soft Cu[...]
-
Page 33
1225 Flashes the curr ent value and then displays the value for adjustment 1225.0 Curr ent Soft Cutout setting Adjust this parameter by using “UP”, “DO WN”, and “SET” as each digit flashes. 1200.0 New Soft Cutout setting Press “SET” to accept the ne w temperature limit. If the temperature of the unit is e ver greater than the Soft C[...]
-
Page 34
8.11.1 B A UD Rate The B A UD rate is the first parameter in the menu . The B A UD rate setting de - termines the serial communications tran smission rate. The B A UD rate parameter is indicated by , bAUd Serial BA UD rate parameter Press “SET” to choose to set the B A UD rate. The current B A UD rate va lue is then be displayed. 2400 b Curr en[...]
-
Page 35
via the serial interface are immediately echoed or trans mitted back to the device of origin. W ith half duplex the commands are e xecut ed but not echoed. The du - plex mode parameter is i ndicated by , dUPL Serial duplex mode parameter Press “SET” to access the mode setting d=FULL Curr ent duplex mode setting The mode may be changed using “[...]
-
Page 36
CAL Calibration par ameters menu Press “SET” fi ve times to enter the me nu. The calibration parameters menu contains the parameters Hard Cutout , CT1, CE1, CT2, CE2, CT3, and CE3. 8.12.1 Har d Cutout This parameter is the temperat ure abov e which the unit shuts do wn automati - cally . The parameter is set at the f actory to approximately 126[...]
-
Page 37
9 Digital Comm unication Interface The furnace is capable of communicatin gw i t ha n db e i n gc o n t r o l l e db yo t h e r equipment through the digital serial interface. W ith a digital interface the instrument may be connected to a comput er or other equipment. Th is allo ws the user to set the set-point temperatur e, monitor the temperature[...]
-
Page 38
ary menu. Press “SET” repeat edly until the display reads “ CAL ”. Press “UP” until the serial interface menu is indicated with “ SErIAL ”. Finally press “SET” to enter the serial parameter menu. In th e serial interface parameters menu are the B A UD rate, the sample rate, the duplex mode, and the linefeed parameter . 9.1.2.1 B[...]
-
Page 39
temperature set-point and vie w or program the vari ous parameters. The inter - face commands are discussed in Sectio n 9.2. All commands are ASCII charac - ter strings terminated with a carriag e-return character (CR, ASCII 13). 9.2 Interface Commands The various commands for accessing the ca librator functions via the digital in - terfaces are li[...]
-
Page 40
9150 Portable Furnace User’s Guide 40 Command Description Command Format Command Example Returned Returned Example Acceptable Values Display Temperatur e Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.0 C Set current set-point t o n s[etpoint]=n s=450 Instrument Range Set temperature units: u[nits]=c/f Co rF Set temperature un[...]
-
Page 41
41 9 Digital Communication Interface Interface Commands Command Description Command Format Command Example Returned Returned Example Acceptable Values Stop program pc=s[top] pc=s Continue program pc=c[ ont] pc=c Read program function pf pf pf: 9 pf: 3 Set program function to n pf= n p f = 2 1t o4 Configuration Menu Operating Paramet ers Menu Read h[...]
-
Page 42
9150 Portable Furnace User’s Guide 42 Command Description Command Format Command Example Returned Returned Example Acceptable Values These commands are only used for f actory testing. Miscellaneous (not on menus) Read firmware ver sion number *ver[sion] * ver ver.9999,9.99 ver.9150,2.20 Read structure of all commands h[elp] h list of commands Leg[...]
-
Page 43
10 T est Probe Calibration For optimum accurac y and stability , allow the calibrator to warm up for 10 minutes after power -up and then allo w adequate stabilization time after reach - ing the set-point temperature. Af ter completing operation of the calibrator, al - low t he well to cool by setting the temperature t o 150°C or less for one-half [...]
-
Page 44
1200°C takes 10 minutes to be within 0.5°C of its settled point and tak es 15 minutes to achieve maximum stability . Speeding up the calibration process can be accomplished by kno wing how soon to make the measurement. It is recommended that typical measurements be made at the desired temperatures with th e desired test probes to establish th ese[...]
-
Page 45
11 Calibration Pr ocedure At times the user may want to calibrate th e unit to improv e th e temperature set-point accuracy . Calibration is done by adjusting the controller probe cali - bration constants CE1, CE2, and C E3 so that the te mperature of the unit as measured with a standard thermocouple agrees more closel y with the set-point. The the[...]
-
Page 46
ture, CEn is the old value for calibration error , and CEm is the new v alue for calibration error 6. Enter the ne w CEm value in the calibration parameter menu using either the ke ypad or through the serial port. 9150 Portable Furnace User’s Guide 46[...]
-
Page 47
12 Maintenance • The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. A void operating the instrument in an oily , wet, dirty , or dusty environment.[...]
-
Page 48
13 T r oubleshooting This section contains information on troubl eshooting, CE Comments, and a wiring diagram. 13.1 T roub leshooting Pr oblems, P ossible Causes, and Solutions In the ev ent that the instrument appears to function abnormally , this section may help to fi nd and solve the problem. Se veral possible problem conditions are described a[...]
-
Page 49
Problem Possible Causes and Solutions The display shows any of the following: err 1 , err 2 , err 3 , err 4 , Err 5 , Err 6 , Err 7 ,o r Err 8 Controller pr oblem. The error messages si gnify the followi ng problems with the controller. Err 1 - a RAM error Err 2 - a NVRAM error Err 3 - a RAM error Err 4 - an ADC setup e rror Err 5 - an ADC ready er[...]