A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Graham Field JB112-070, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Graham Field JB112-070 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Graham Field JB112-070. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Graham Field JB112-070 should contain:
- informations concerning technical data of Graham Field JB112-070
- name of the manufacturer and a year of construction of the Graham Field JB112-070 item
- rules of operation, control and maintenance of the Graham Field JB112-070 item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Graham Field JB112-070 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Graham Field JB112-070, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Graham Field service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Graham Field JB112-070.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Graham Field JB112-070 item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
R ead this manual before operating the nebulizer . Sav e this manual f or future r eference. Info: The most cur rent version of this manual can be found online a t www .gr ahameld.com . F ederal Law (USA) r estricts this de vice to sale by or on the or der of a physician. T his device should not be used unless the operator has been instr ucted b[...]
-
Page 2
JB01 12-070-INS-LAB-RevB14 2 CONTENTS INTRODUCTION ................................................................................................... 3 INTENDED USE ............................................................................................ 3 SYMBOLS KEY .............................................................................[...]
-
Page 3
JB01 12-070-INS-LAB-RevB14 3 INTRODUCTION This user manual contains important information and safety pr ecautions for the John Bunn Neb-u-T yke Neb-a-Doodle Nebulizer Compr essor . Befor e using your Nebulizer Compr essor , please r ead and understand this entir e user manual. T ake special note of all safety pr ecautions marked DANGER and W ARNING[...]
-
Page 4
JB01 12-070-INS-LAB-RevB14 4 Symbols K ey O Off I On ~ Alternating current IPX0 Enclosed equipment without protection against ingress of water Attention (read the instructions) II Class II (not applicable for UL1431) BF: This product is in accordance with the degree of protection against electric shock for type BF equipment. (not applicable for UL1[...]
-
Page 5
JB01 12-070-INS-LAB-RevB14 5 IMPORT ANT SAFETY PRECA UTIONS When using electrical pr oducts, especially when childr en ar e pr esent, always follow basic safety pr ecautions, including the following: DANGER: Indicates a potential hazard situation or unsafe practice that, if not avoided, will result in death or serious personal injury . W ARNING: In[...]
-
Page 6
JB01 12-070-INS-LAB-RevB14 6 W ARNING:T oreducetheriskofburns,electrocution,re or personal injury: 1. Never leave this device unattended when plugged in. 2. This product contains small parts that may present a choking hazard to small children, and the tubing may present a strangulation hazard. Always use close supe[...]
-
Page 7
JB01 12-070-INS-LAB-RevB14 7 13. This device has a polarized plug (one blade is wider thantheother). Asasafetyfeature,thisplugwilltinto apolarizedoutletonlyoneway .Iftheplugdoesnott fullyintotheoutlet,reversetheplug.Ifitstilldoesnott[...]
-
Page 8
JB01 12-070-INS-LAB-RevB14 8 PRODUCT DESCRIPTION F ea tures and Components Please note nebulizer featur es and components, labeled in the illustration below . reservoirtube ON/OFFswitch shapes stylus slidetab drawing board nebulizer housing pushbutton airoutletconnector fortubing filtercap adultmask nebulizercup mouthp[...]
-
Page 9
JB01 12-070-INS-LAB-RevB14 9 SETUP Info: Before initial operation, clean the nebulizer cup assembly following the instr uctions given in MAINTENANCE/Cleaning the Nebulizer Cup . W ARNING: Ensure that the nebulizer compressor is unplugged during setup. 1. Befor e each use, inspect the compressor and nebulizer cup assembly for damage or wear; r eplac[...]
-
Page 10
JB01 12-070-INS-LAB-RevB14 10 Info: The reservoir tube, shown in the illustration earlier in this manual, acts as a holding tank for the mist, reducing the amount lost during treatment. Its use is optional. If you wish to use the reservoir tube, install it now on the small end of the T adapter . 8. Install T adapter onto top of nebulizer medication[...]
-
Page 11
JB01 12-070-INS-LAB-RevB14 1 1 OPERA TION Info: The drawing board must be removed from the nebulizer housing before use. T o remove the drawing board, push the button and slide the drawing board in the opposite direction from the cover ar row . Nebulizer Operation s NOTICE: The Neb-u-T yke Neb-a-Doodle Nebulizer Compressor is designed for intermitt[...]
-
Page 12
JB01 12-070-INS-LAB-RevB14 12 Drawing Boar d Oper ation pushbutton arrow 2 1 T o remove the drawing boar d, push the button and slide the drawing boar d in the opposite dir ection fr om the cover arr ow . T o install the drawing board, insert the drawing board on the side with the housing arr ow that matches the cover arr ow , then push it all t[...]
-
Page 13
JB01 12-070-INS-LAB-RevB14 13 Safety Ov er load s NOTICE: Do not exceed 30 minutes of continuous operation. 1. The motor of this device has a built-in thermal overload pr otector . Should the motor over heat, the pr otector will automatically shut off the motor . Should this occur , move the ON/OFF switch to the OFF position and allow the motor to [...]
-
Page 14
JB01 12-070-INS-LAB-RevB14 14 MAINTENANCE Cleaning the Compressor’s Outer Case W ARNING: Electrical shock hazard. Do not remove the outer case of the device. All disassembly and maintenance of the device must be performed by a qualiedservicetechnician.Referservicetoqualied service personnel. W ARNING: This device does n[...]
-
Page 15
JB01 12-070-INS-LAB-RevB14 15 R eplacing the Filter s NOTICE:Changethelteronceevery30daysorwhenthe lterchangescolor . Additionallterscanbepurchased atwww .grahameld.comorfromyourGraham-Field authorized distributor . 1. Ensur e that ON/OFF switch is in the OFF[...]
-
Page 16
JB01 12-070-INS-LAB-RevB14 16 TROUBLESHOO TING If your Neb-u-T yke Neb-a-Doodle Nebulizer Compressor fails to function, consult the following T r oubleshooting guide. If the pr oblem persists, consult your Graham-Field authorized distributor . Problem Cause and Solution Device doesn’t operate Check if plug is properly t into a grounded electri[...]
-
Page 17
JB01 12-070-INS-LAB-RevB14 17 SPECIFICA TIONS Specication Description Max pressure (compressor) 32.5 PSI Max ow rate Approximately 12 LPM Nebulization rate .33 ml/min Particle dimension .5 - 5 microns Capacity 5 ml Filter Sponge Dimensions (LxWxH) 10" × 6.5" × 4.2" (255 mm x 166 mm x 108 mm) Weight 4 lb (1860 g) Electrical re[...]
-
Page 18
JB01 12-070-INS-LAB-RevB14 18 LIMITED WARRANTY SCOPE OF WARRANTY GF Health Products, Inc. (“GF”) warrants to the original pur chaser only that it will replace or r epair components, at GF’s sole discr etion, that ar e defective in material or workmanship under normal use and ser vice. All warranties are conditioned upon the pr oper use of the[...]
-
Page 19
JB01 12-070-INS-LAB-RevB14 19 ENTIRE WARRANTY , EX CLUSIVE REMED Y AND CONSEQUENTIAL D AMA GES DISCLAIMER THIS W ARRANTY IS GF’S ONL Y W ARRANTY AND IS IN LIEU OF ALL OTHER W ARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED W ARRANTIES OF ANY KIND INCLUDING ANY IMPLIED W ARRANTIES OF MERCHANT ABILITY OR FITNESS FOR A P ARTICULAR PURPOSE. IF ANY[...]
-
Page 20
Manufactur ed for: GF Health Pr oducts, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 USA T elephone: 770-368-4700 Fax: 770-368-2386 www .grahamfield.com Made in China[...]
-
Page 21
Lea antes de operar el nebulizador . Guarde este manual para r eferencia en el futur o. Info: La más reciente v er sión de este manual podrá ser encontrado en www .gr ahameld.com . La Ley F ederal (USA) r estringe este apar ato para v entas por la or den de su médico. Este aparato no debe ser utilizado si el operador no ha sido adiestr ado p[...]
-
Page 22
JB01 12-070-INS-LAB-RevB14 2 GF Health Products, Inc. no es r esponsable de errores tipográcos. Para obtener la información más actualizada y vigente relativa a los envases, garantías, productos y especicaciones, incluyendo la versión más actualizada de estas instrucciones, por favor visite nuestro sitio web en www .grahameld.com. Gr[...]
-
Page 23
JB01 12-070-INS-LAB-RevB14 3 INTRODUCCIÓN Este manual del usuario contiene información importante y pr ecauciones de seguridad para el John Bunn Neb-u-T yke Neb-a-Doodle Compr esor de Nebulizador . Antes de usar su Compr esor de Nebulizador , favor de leer y entender el manual de usuario. T ome nota especial en todas las pr ecauciones mar cadas P[...]
-
Page 24
JB01 12-070-INS-LAB-RevB14 4 Símbolos O Apagado I Encendido ~ Corriente alternando IPX0 El equipo no tiene protección contra sumergiendo en agua Atención (lea las instrucciones) II Clase II (no aplicable para UL1431) BF: Este producto está en concordancia con el grado de protección sobre el riesgo de descarga eléctrica por equipo de tipo BF ([...]
-
Page 25
JB01 12-070-INS-LAB-RevB14 5 DIRECTIV AS IMPORT ANTES DE SEGURID AD PELIGRO: Indica una situación de peligro o una practica insegura que, si no es evitada, resultará en muerte o heridas serias a la persona. ADVERTENCIA: Indica una situación de peligro o una practica insegura que, si no es evitada, puede resultar en muerte o heridas serias a la p[...]
-
Page 26
JB01 12-070-INS-LAB-RevB14 6 ADVERTENCIA: Para reducir el riesgo de quemadura, electrocución, fuego o heridas a la persona. 1. Este producto nunca debería ser desatendido mientras esté conectado. 2. Este producto contiene piezas pequeñas que presenta riesgosdeasxiaparaniños,ylamangapuedepresentar riesgo de estr[...]
-
Page 27
JB01 12-070-INS-LAB-RevB14 7 12. Conecte este producto solamente en un enchufe de pared correcto. 13 . Este aparato tiene un enchufe polarizado (un aspa es más ancha que la otra). Para reducir el riesgo de electrocución, este enchufe es diseñado para poder ser posicionado de una sola manera en una toma eléctrica polarizada. Si el enchufe no ent[...]
-
Page 28
JB01 12-070-INS-LAB-RevB14 8 DESCRIPCIÓN DEL PRODUCT O Características y componentes Por favor de tomar nota de las características y los componentes del nebulizador descritos en la ilustración abajo: mangadereserva interruptordeON/OFF tablade dibujar compartimento denebulizador conectordel enchufedeaire tapadel[...]
-
Page 29
JB01 12-070-INS-LAB-RevB14 9 ENSAMBLAJE Aviso: Antes de operación inicial, limpie la taza del nebulizador siguiendo las instr ucciones escritos en la sección Mantenimiento/Limpieza de la T aza del Nebulizador . ADVERTENCIA: Antes de conectar el cordón de electricidad, asegure que el interruptor de ON/OFF está en la posición OFF . El enchufe ta[...]
-
Page 30
JB01 12-070-INS-LAB-RevB14 10 Info: La manga de reserva, en la ilustración superior del manual, es como un tanque de retención para reducir la cantidad de medicamento perdido durante tratamiento. Es opcional. Si desea usar la manga de reserva, posicione en la parte inferior del adaptador T . 8. Instale el adaptador T sobr e la tapa de la taza del[...]
-
Page 31
JB01 12-070-INS-LAB-RevB14 1 1 OPERACIÓN Info: La tabla de dibujar tiene que ser apartado del nebulizador antes de uso. Para quitar la tabla de dibujar , empuje el botón y deslice la tabla de dibujar en la dirección opuesto de la flecha en la tapa. Operación del Compr esor de Nebulizador s A VISO: El Compresor de Nebulizador está diseñado pa[...]
-
Page 32
JB01 12-070-INS-LAB-RevB14 12 Operación de la T abla de Dibujar botón flecha 2 1 Para quitar la tabla de dibujar , empuje el botón y deslice la tabla de dibujar en la dir ección opuesto que la echa en la tapa. Para instalar la tabla de dibujar , inser te la tabla de dibujar en el lado con la echa en el compartimento que machea la echa [...]
-
Page 33
JB01 12-070-INS-LAB-RevB14 13 Se guridad en caso de sobrecar ga s A VISO: No debe exceder más de 30 minutos de operación continuo. 1. Este compr esor contiene un dispositivo de protección termal. Si el motor comienza a sobr ecalentarse, se apagará automáticamente. Si esto ocurr e, posicione el interruptor a la posición OFF y permita que el mo[...]
-
Page 34
JB01 12-070-INS-LAB-RevB14 14 MANTENIMIENT O Limpieza de la Carcasa del Compr esor ADVERTENCIA: Riesgo de electrocución. No remueve carcasa externa de este aparato. T odo desensamblaje y mantenimiento de este aparato debe ser realizado por un técnicodeserviciocualicado. ADVERTENCIA: Este aparato no necesita aceite. No lubrique piezas [...]
-
Page 35
JB01 12-070-INS-LAB-RevB14 15 Instr ucciones par a el cambio del ltro de air e Aviso: Cambie el filtro cada 30 días o cuando el filtro cambia color . Filtros adicionales pueden ser comprados en www .grahamfield.com o de su distribuidor autorizado de Graham-Field. 1. Remueva la tapa del ltr o de aire localizado al lado del interruptor ON/O[...]
-
Page 36
JB01 12-070-INS-LAB-RevB14 16 DETECT ANDO PR OBLEMAS Si su Neb-u-T yke Neb-a-Doodle Compresor de Nebulizador falla, consulte a la guía de Detectando Pr oblemas debajo. Si el pr oblema continua, llame a su distribuidor autorizado de Graham-Field. Problema Causa y Solución El aparato no funciona V erique si el enchufe está conectado correctamen[...]
-
Page 37
JB01 12-070-INS-LAB-RevB14 17 ESPECIFICACIONES Especicación Descripción Presión máxima del compresor 32.5 PSI Flujo máximo Aproximadamente 12 LPM T asa de nebulización 0.33 ml/min T amaño de partícula .5 - 5 microns Capacidad 5 ml Filtro Esponja Dimensión (largo x anchura x altura) 10" × 6.5" × 4.2"(255 mm x 166 mm x 10[...]
-
Page 38
JB01 12-070-INS-LAB-RevB14 18 GARANTÍA LIMIT AD A AMBIT O DE GARANTÍA GF Health Products, Inc. (“GF”) garantía al comprador original o arr endatario (cual quiera es el “Cliente”) solamente, que va reemplazar o r eparar componentes, a la decisión solo de GF , cuales son defectos en material o calidad utilizado normalmente y con ser vicio[...]
-
Page 39
JB01 12-070-INS-LAB-RevB14 19 GARANTÍA ENTERO , REMEDIO EX CL USIV O Y DESCARGO DE D AÑOS CONSIGUIENTE EST A GARANTIA ES LA UNICA GARANTIA DE GF Y ES EN VEZ DE TODAS OTRAS GARANTIAS, EXPRESIDAS O IMPLICIT ADAS. GF NO HACE GARANTIAS ENTENTIDOS DE CUAL QUIER MANERA INCLUYENDO GARANTIAS ENTENDIDOS DE COMERCIALIZACION O IDONEIDAD P ARA UN PROPOSITO P[...]
-
Page 40
Fabricado para: GF Health Pr oducts, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 USA T elephone: 770-368-4700 Fax: 770-368-2386 www .grahamfield.com Fabricado en China[...]