Go to page of
Similar user manuals
-
Thermometer
Omega Engineering OS534
78 pages 1.5 mb -
Thermometer
Omega Engineering HH802W
2 pages 0.56 mb -
Thermometer
Omega Engineering OMEGAMETER HHM57
20 pages 0.42 mb -
Thermometer
Omega Engineering DialtempTM A
2 pages 0.09 mb -
Thermometer
Omega Engineering OS562
2 pages 0.41 mb -
Thermometer
Omega Engineering HH507R
21 pages 0.25 mb -
Thermometer
Omega Engineering OS561
2 pages 0.28 mb -
Thermometer
Omega Engineering OMEGASCOPE OS523
78 pages 1.57 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Omega Engineering HH504, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Omega Engineering HH504 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Omega Engineering HH504. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Omega Engineering HH504 should contain:
- informations concerning technical data of Omega Engineering HH504
- name of the manufacturer and a year of construction of the Omega Engineering HH504 item
- rules of operation, control and maintenance of the Omega Engineering HH504 item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Omega Engineering HH504 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Omega Engineering HH504, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Omega Engineering service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Omega Engineering HH504.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Omega Engineering HH504 item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
OMEGA HH504 Digital Thermometer[...]
-
Page 2
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.com e-mail: info@omega.com omeg[...]
-
Page 3
1 WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. The symbol on the instrument indicates that the operator must refer to an explanation in this manual. INTRODUCTION This instrume[...]
-
Page 4
2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Pt385(100 W ) -200°C to 800°C, (-328°F to 1472°F) Pt3926(100 W ) -200°C to 630°C, (-328°F to 1166°F) Resolution: 0.1°C or 0.2°F Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F[...]
-
Page 5
3 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto[...]
-
Page 6
2 3 1 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale (Main display) Reading the main displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set[...]
-
Page 7
5 4 6 5 T1/ T2/T1-T2 Main display Input Selection The input selection indicates which input is selected for main display; T1 probe, T2 probe or the difference between the two probes (T1-T2), when the thermometer is turned on, it is set to T1. When main display input selected T1, then T1 input can select alternate of probe by pressing Pt385/Pt3926 k[...]
-
Page 8
6 7 The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated value is retained as the average reading, but the actual minimum and maximum reading will continue to be captured. In the this mode, press HOLD key to stop the recording of[...]
-
Page 9
8 9 °C/°F Selecting the Temperature Scale (Second display) Readings the second displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C/°F key. SET mode (Relati[...]
-
Page 10
8 10 11 12 4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits set value that was in use when thermometer was last turned off set values. T1/T2 /T1-T2 second display Input Selection The input selection indicates which input is selected for second display; T1 probe, T2 probe or the difference between the two probes (T1-T2),[...]
-
Page 11
9 RTD PROBE CONNECTION P/N : 17650-0505 RW1=RW2=RW3 RW= WIRE RESISTANCE G + - RTD 100 W Pt385 (Pt3926) WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the rmometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD [...]
-
Page 12
10 TEMPERATURE VS RESISTANCE TABLE(ITS90) °C Pt385 Pt3926 -200°C 18.521 W 16.996 W -100°C 60.256 W 59.479 W 0°C 100.000 W 100.000 W 100°C 138.505 W 139.272 W 200°C 175.856 W 177.362 W 300°C 212.052 W 214.275 W 400°C 247.092 W 250.018 W 500°C 280.977 W 284.591 W 600°C 313.708 W 317.994 W 700°C 345.280 W - 800°C 375.700 W -[...]
-
Page 13
1 HH 504 / HH 505 CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter. 1. T urn of f the meter then set the adjusted jumper to J1. 2. Insert the 383.00 Ω c[...]
-
Page 14
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC . warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers [...]
-
Page 15
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermocouple, RTD & Thermistor þ Calibrators & Ice Point References þ Recorders, Controllers & Process Monitors þ Infrared Pyrometers PRESSURE[...]