Teledyne 319R manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23

Go to page of

A good user manual

The rules should oblige the seller to give the purchaser an operating instrucion of Teledyne 319R, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.

What is an instruction?

The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Teledyne 319R one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.

Unfortunately, only a few customers devote their time to read an instruction of Teledyne 319R. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.

What should a perfect user manual contain?

First and foremost, an user manual of Teledyne 319R should contain:
- informations concerning technical data of Teledyne 319R
- name of the manufacturer and a year of construction of the Teledyne 319R item
- rules of operation, control and maintenance of the Teledyne 319R item
- safety signs and mark certificates which confirm compatibility with appropriate standards

Why don't we read the manuals?

Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Teledyne 319R alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Teledyne 319R, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Teledyne service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Teledyne 319R.

Why one should read the manuals?

It is mostly in the manuals where we will find the details concerning construction and possibility of the Teledyne 319R item, and its use of respective accessory, as well as information concerning all the functions and facilities.

After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.

Table of contents for the manual

  • Page 1

    TELEDYNE BROWN EGINEERING Analytical Instruments P/N M59608 ECO# 95-416 7/10/95 M ODEL 319R O XYGEN D ETECTOR I NSTRUCTION M ANUAL[...]

  • Page 2

    TELEDYNE BROWN EGINEERING Analytical Instruments Copyright © 1993 Teledyne Brown Engineering Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, wheth[...]

  • Page 3

    TELEDYNE BROWN EGINEERING Analytical Instruments Table of Contents 1.0 Introduction Main Features ............................................................................................... 1 Micro-Fuel Cell ............................................................................................ 2 Z-2 Cell Warranty Conditions ..............[...]

  • Page 4

    TELEDYNE BROWN EGINEERING Analytical Instruments[...]

  • Page 5

    1 319R Introduction 1.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Introduction The Teledyne Brown Engineering Analytical Instruments (TBE) Model 319R Oxygen Detector measures the percent of oxygen in a wide range of gases and gas mixtures. Modular and compact design provide outstanding performance and reliable on-line monitoring and make th[...]

  • Page 6

    1.0 Introduction 319R 2 TELEDYNE BROWN ENGINEERING Analytical Instruments 4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months. 5. Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc., at 0–10 VDC, ne[...]

  • Page 7

    3 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Installation & Operation Precautions 1. Read this chapter in its entirety before installing the unit. 2. Upon receipt of the instrument, inspect the entire unit for damage. Notify the shipper if any damage is found. Check the unit, probe, and included acce[...]

  • Page 8

    4 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments The analysis unit should be installed in the process at a maximum distance of 6 feet away from the main unit, depending upon the level of electromagnetic noise in the operating environment. Analysis Unit Connection The receptacle for the analysis unit cable is[...]

  • Page 9

    5 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Alarm Terminal Connection SPDT relay contacts are provided for each of the 3 alarms. There are 9 terminals available on the back panel of the system for alarm circuit connection. The explanation for the relay terminals abbreviations is as follows: NC = Normall[...]

  • Page 10

    6 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Setting Alarms The alarms can be adjusted over the full scale range, and they are set as a percent of full scale. Thus, an alarm set to indicate 192 on the display while on the 200 PPM O 2 range will activate at 192 PPM O 2 on the 200 PPM O 2 range and at 1920[...]

  • Page 11

    7 319R Installation & Operation 2.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Calibration To calibrate the 319R: 1. Place the power switch in the ON position. 2. Allow the LCD reading to stabilize. 3. Expose the sensor to ambient air (20.9%) or instrument grade air. If the analyzer goes overrange, the display will go blank and the LED t[...]

  • Page 12

    8 2.0 Installation & Operation 319R TELEDYNE BROWN ENGINEERING Analytical Instruments[...]

  • Page 13

    9 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Maintenance Replacing the Fuse If the unit will not turn on when plugged in and switched ON (via the power switch), or you suspect a blown fuse, replace the fuse using the following procedure: 1. Disconnect the AC power and place the power switch located on the rear panel in t[...]

  • Page 14

    3.0 Maintenance 319R 10 TELEDYNE BROWN ENGINEERING Analytical Instruments 5. Replace the fuse holder into its receptacle, pushing in firmly until it clicks. In units with DC power, the fuse is located on the rear panel above the ON/OFF switch. Open the fuse holder by unscrewing and removing the knob marked FUSE. The fuse will be located inside the [...]

  • Page 15

    11 319R Maintenance 3.0 TELEDYNE BROWN ENGINEERING Analytical Instruments Warranty Conditions The Class Z-2 cell used in the 319R is warranted for six (6) months of service from the date of shipment. Customers having warranty claims must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty mat[...]

  • Page 16

    3.0 Maintenance 319R 12 TELEDYNE BROWN ENGINEERING Analytical Instruments[...]

  • Page 17

    13 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Appendix Specifications Ranges : 0–200 PPM, 0–1999 PPM O 2 , and 0– 25% calibration range Sensitivity: 1% of full scale Signal Output : Standard: 0–10 VDC, negative ground 4-20 mA, negative ground plus contact closures for range identification. Display : Digital 7 segment, 3[...]

  • Page 18

    14 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Spare Parts List QTY. P/N DESCRIPTION 1 C57304 PC board, main 1 A33748 Thermistor assembly 1 C6689-Z2 Micro-Fuel Cell, Z-2 4 F1189 Fuse, 1A (AC) 4 F51 Fuse, ½A (DC) 1 A59264 Probe to analyzer cable, 6 feet A minimum charge of US $20.00 is applicable to spare parts orders. IMPORTANT[...]

  • Page 19

    15 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Drawing List C-57490 Outline diagram A-58154 Probe interconnection diagram D-57302 Electronic PCB schematic diagram C-57305 Power supply PCB schematic diagram C-57307 Power supply PCB assembly diagram B-58123 Analysis unit (probe) outline Note: 319R analyzers with 25% and 100% range[...]

  • Page 20

    16 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Material Safety Data Sheet Product Name: Micro-Fuel Cells and Super Cells, all classes except A-2C, A-3, and A-5. Electrochemical Oxygen Sensors, all classes except R-19. Mini-Micro-Fuel Cells, all classes. Manufacturer: Teledyne Brown Engineering Analytical Instruments Address: 168[...]

  • Page 21

    17 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Flash Point: na Flammable Limits: na LEL: na UEL: na Extinguishing Media: Use extinguishing media appropriate to surrounding fire conditions. No specific agents recommended. Wear NIOSH/OSHA approved self-contained breathing apparatus and protective clothing to prevent contact with s[...]

  • Page 22

    18 Appendix 319R TELEDYNE BROWN ENGINEERING Analytical Instruments Contact of electrolyte with skin or eyes will cause a burning sensation and/or feel soapy or slippery to touch. Other symptoms of exposure to lead include loss of sleep, loss of appetite, metallic taste and fatigue. Carcinogenicity: Lead is classified by the IARC as a class 2B carci[...]

  • Page 23

    19 319R Appendix TELEDYNE BROWN ENGINEERING Analytical Instruments Section VIII – Control Measures Eye Protection: Chemical splash goggles Hand Protection: Rubber gloves Other Protective Clothing: Apron, face shield Ventilation: na Section IX – Disposal Both lead and potassium hydroxide are considered poisonous substances and are regulated unde[...]