Go to page of
Similar user manuals
-
Oxygen Equipment
Teledyne 311PC
31 pages 1.47 mb -
Oxygen Equipment
Teledyne 310
26 pages 1.47 mb -
Oxygen Equipment
Teledyne P/NM70946
72 pages 0.95 mb -
Oxygen Equipment
Teledyne 9060Z
67 pages 5.13 mb -
Oxygen Equipment
Teledyne 3000PA
69 pages 0.88 mb -
Oxygen Equipment
Teledyne LXT-220
61 pages 1.93 mb -
Oxygen Equipment
Teledyne UltraTrace 3000 Oxygen Analyzer
72 pages 0.95 mb -
Oxygen Equipment
Teledyne 9060L
51 pages 0.99 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Teledyne BDS-3960, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Teledyne BDS-3960 one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Teledyne BDS-3960. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Teledyne BDS-3960 should contain:
- informations concerning technical data of Teledyne BDS-3960
- name of the manufacturer and a year of construction of the Teledyne BDS-3960 item
- rules of operation, control and maintenance of the Teledyne BDS-3960 item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Teledyne BDS-3960 alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Teledyne BDS-3960, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Teledyne service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Teledyne BDS-3960.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Teledyne BDS-3960 item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
Oxygen Analyzer Teledyne Analytical Instruments i OPERA TING INSTRUCTIONS FOR M M M M O O O O D D D D E E E E L L L L B B B B D D D D S S S S - - - - 3 3 3 3 9 9 9 9 6 6 6 6 0 0 0 0 Oxygen Analyzer DANGER Tox ic g a se s a nd o r fla mm a bl e liq ui ds ma y be pr es e nt i n this mon it or ing s y ste m. Personal protective equipment may be requir[...]
-
Page 2
BDS 3960 Teledyne Analytical Instruments ii Copyright © 2002 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, me[...]
-
Page 3
Oxygen Analyzer Teledyne Analytical Instruments iii Specific Model Information The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument. Commonly available options are listed below, with check boxes. Any that are incorporated in the instrument for which this manual is supplied are[...]
-
Page 4
BDS 3960 Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety aler[...]
-
Page 5
Oxygen Analyzer Teledyne Analytical Instruments v IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES. This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please[...]
-
Page 6
BDS 3960 Teledyne Analytical Instruments vi Table of Contents Safety Messages .......................................................................... iv Table of Contents ......................................................................... vi List of Figures ............................................................................... ix [...]
-
Page 7
Oxygen Analyzer Teledyne Analytical Instruments vii 3.2 Mounting the Analyzer 30 3.3 Rear Panel Connections 32 3.3.1 Gas Connections 32 3.3.2 Electrical Connections 33 3.3.2.1 Primary Input Power 34 3.3.2.2 50-Pin Equipment Interface Connector 34 3.4 Electrolyte Refill of BDS Sensor 39 3.5 Testing the System 40 3.6 Powering Up the System 41 Operati[...]
-
Page 8
BDS 3960 Teledyne Analytical Instruments viii 4.9 Switching The Program Back To The Front 56 Maintenance ................................................................................. 58 5.1 Routine Maintenance 58 5.2 Adding Water to the BDS Sensor 58 5.3 Fuse Replacement 59 5.4 Battery Backup Replacement 60 5.5 Reinstalling Application software[...]
-
Page 9
Oxygen Analyzer Teledyne Analytical Instruments ix List of Figures Figure 1-1: BDS-3960 Front Panel ................................................ 15 Figure 1-2: Model BDS 3960 Rear Panel ...................................... 17 Figure 2.1: Cross Section of the BDS Oxygen Sensor ................. 20 Figure 2.2: BDS sensor output at different gas[...]
-
Page 10
BDS 3960 Teledyne Analytical Instruments x Figure 5-3: Vacuum Degassing for the BDS Oxygen Sensor ....... 64 Figure A-1: Single 19" Rack Mount (dimensions in mm) ............... 69[...]
-
Page 11
Oxygen Analyzer Teledyne Analytical Instruments xi List of Tables Table 3-1: Analog Output Connections ......................................... 35 Table 3-2: Alarm Relay Contact Pins ............................................ 36 Table 3-3: Remote Calibration Connections .................................. 37 Table 3-4: Range ID Relay Connections [...]
-
Page 12
BDS 3960 Teledyne Analytical Instruments xii DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for usage in a nonhazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist. The customer should ensure tha[...]
-
Page 13
Oxygen Analyzer Introduction Teledyne Analytical Instruments 13 Introduction 1.1 Overview The Teledyne Analytical Instruments Model BDS 3960 Oxygen Analyzer is a versatile instrument for detecting oxygen at the parts-per- billion (ppb) level in a variety of gases. This manual covers the Model BDS 3960 General Purpose flush-panel and/or rack-mount u[...]
-
Page 14
Introduction BDS 3960 Teledyne Analytical Instruments 14 • Three user definable output ranges (from 0-100 ppb through 0-100 ppm) allow best match to users process and equipment, plus a fixed 100 ppm over range. • Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user[...]
-
Page 15
Oxygen Analyzer Introduction Teledyne Analytical Instruments 15 Figure 1-1: BDS-3960 Front Panel Digital Meter Display: The meter display is a Light Emitting Diode (LED) device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-999.9 ppb and then switches to a continuous ppm read[...]
-
Page 16
Introduction BDS 3960 Teledyne Analytical Instruments 16 Ac ces s D raw er: Fo r a cce ss t o t he B DS S enso r And p r ess ur e re gul at or. T h e f r ont p anel sli de s out w he n the fo ur t humb scr ews ar e l oos ened . Open i ng t he i nt e r ior gives a cces s t o m ost ci rcu i t boards t oo. 1.5 Rear Panel (Equipment Interface) The rear[...]
-
Page 17
Oxygen Analyzer Introduction Teledyne Analytical Instruments 17 Figure 1-2: Model BDS 3960 Rear Panel[...]
-
Page 18
Operational Theory BDS 3960 Teledyne Analytical Instruments 18 Operational Theory 2.1 Introduction The analyzer is composed of four subsystems: • BDS Sensor • Sample System • Pocket PC • Electronic I/O Signal Processing and Display The sample system is designed to accept the sample gas and transport it through the analyzer without contamina[...]
-
Page 19
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 19 controlled with respect to the reference electrode. The counter electrode is used to carry the current that flow through the sensor. A potentiostat is typically constructed with several operational amplifiers. The three electrodes in an electrochemical cell and the operational am[...]
-
Page 20
Operational Theory BDS 3960 Teledyne Analytical Instruments 20 The main advantage of the BDS technology lies in the unique second potentiostat. It is designed to remove dissolved oxygen and other impurities in the electrolyte. It eliminates the internal background current which previously limited the detection process. The second potentiostat is lo[...]
-
Page 21
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 21 While the 2nd working electrode allows OH - to move through, it prevents the dissolved oxygen from the top portion of the sensor to reach the sensing electrode. The reference electrode provides a potential reference for both the sensing electrode and the 2nd working electrode. NO[...]
-
Page 22
Operational Theory BDS 3960 Teledyne Analytical Instruments 22 2.2.3 Gas Pressure The analyzer is equipped with a pressure regulator as shown in Figure 2-3. To access the pressure regulator, the front panel must be opened. The inlet pressure should be regulated anywhere between 4 to 50 psig. The sensor is not affected by pressure changes in the inl[...]
-
Page 23
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 23 Figure 2.4 Typical Purge-down Curve After Air Saturation S h or t e n i ng t h e ai r exp o s u r e wi l l al l o w a f a s t e r se ns o r r ec o v e r y. A ty p i c a l BD S se n s o r wi l l re c o v e r t o 1 pp m i n ap p r ox i m a t el y 25 m i nu t e s , t o 10 0 pp b af [...]
-
Page 24
Operational Theory BDS 3960 Teledyne Analytical Instruments 24 hydrogen, it reads 100% lower. For example, when the float ball in the flowmeter is at 0.5 SLPM, the actual flow rate of hydrogen is about 1 SLPM. The BDS oxygen sensor can tolerate exposure to acidic gases. Up to 0.2% CO 2 has no effect to ppb level oxygen measurement. 2.2.7 Stability [...]
-
Page 25
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 25 The re i s a Max line and Min Lin e clearly ma r ked on the BDS s ens or body. It is a good pr ac tice to chec k the el ectrolyte le vel every mont h and ad d de-ioni zed wate r into t h e s ensor wh enever i t is con v eni ent. Whe n runnin g dry gas throug h the se nsor, the ga[...]
-
Page 26
Operational Theory BDS 3960 Teledyne Analytical Instruments 26 Note that instrument air at a pressure of 70 to 80 psig is needed to activate the proper valves. Figure 2-6: Flow Diagram 2.4 Electronics and Signal Processing T h e Mo del B DS 3960 Oxyg en Anal yzer uses an 8 051 m icr ocont r ol l er w i th 32 kB of R AM a nd 8 k B o f R OM t o co nt[...]
-
Page 27
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 27 as the signal is passed to a second channel of the analog to digital converter. The raw counts of the analog to digital converter for both the oxygen sensor amplifier and temperature amplifier are sent to the PPC via RS232 interface once every second along with a status byte. The[...]
-
Page 28
Operational Theory BDS 3960 Teledyne Analytical Instruments 28 Figure 2-7: BDS 3960 Electronics Block Diagram Temperature Sensor ADC DAC 8051 processor UART Pocket PC BDS Sensor 0-1 VDC and 4-20 ma support electronics Alarm 1 Alarm 2 System Alarm LED Display Other Outputs Am p lifiers[...]
-
Page 29
Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 29[...]
-
Page 30
Installation BDS 3960 Teledyne Analytical Instruments 30 Installation Installation of the Model BDS 3960 Analyzer includes: • Unpacking • Mounting • Gas connections • Electrical connections • Filling the Sensor with Electrolyte. • Testing the system. 3.1 Unpacking the Analyzer Although the analyzer is shipped complete, certain parts, su[...]
-
Page 31
Oxygen Analyzer Installation Teledyne Analytical Instruments 31 Figure 3-1: Model BDS 3960 Front Panel Access to the electronics, the sensor, and sampling system is accomplished by loosening the thumbscrews on the front of the analyzer. The whole assembly will slide out toward the front. Allow clearance for the assembly to slide out when maintenanc[...]
-
Page 32
Installation BDS 3960 Teledyne Analytical Instruments 32 3.3 Rear Panel Connections Figure 3-3 shows the Model BDS 3960 rear panel. There are ports for gas inlet and outlet, power, communication, and analog concentration output. Figure 3-3: Rear Panel of the Model BDS 3960 3.3.1 Gas Connections T h e un i t is ma nu f a ct u r e d wi t h 1/ 4 in ch[...]
-
Page 33
Oxygen Analyzer Installation Teledyne Analytical Instruments 33 If greater sample flow is required for improved response time, install a bypass in the sampling system upstream of the analyzer input. SPAN IN: Span gas is connected to this port and it is automatically routed by solenoid valves when the analyzer goes into the span mode. Its pressure s[...]
-
Page 34
Installation BDS 3960 Teledyne Analytical Instruments 34 3.3.2.1 P RIMARY I NPUT P OWER The power cord receptacle and fuse block are located in the same assembly. Insert the power cord into the power cord receptacle. CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. The power su[...]
-
Page 35
Oxygen Analyzer Installation Teledyne Analytical Instruments 35 4–20 mA DC % Range: Current increases linearly with increasing oxygen, from 4 mA at 0 ppm to 20 mA at full scale ppm. (Full scale = 100% of programmable range.) 4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16 mA = High Range, 20 mA = 100ppm. Table 3-1: Analog Output[...]
-
Page 36
Installation BDS 3960 Teledyne Analytical Instruments 36 Threshold Alarm 2: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below threshold). • Can be configured as failsafe or nonfailsafe. • Can be configured as latching or nonlatching. • Can be configured out (defeated).[...]
-
Page 37
Oxygen Analyzer Installation Teledyne Analytical Instruments 37 Zero: • Floating input—5 to 24 V input across the + and – pins puts the analyzer into the Zero mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows Zero mode to terminate when done. A synchronous signal must open and close the ext[...]
-
Page 38
Installation BDS 3960 Teledyne Analytical Instruments 38 W h e n t h e c o n t a c t i s OP E N , t h e an a l y z e r i s an a l y z i n g , t h e R e m o t e Ca l I n p u t s ar e be i n g po l l e d , an d a ze r o or sp a n co m m a n d c a n be se n t . When the contact is CLOSED, the analyzer is already calibrating. It will ignore your reques[...]
-
Page 39
Oxygen Analyzer Installation Teledyne Analytical Instruments 39 Table 3-4: Range ID Relay Connections Pin Function 21 Range 1 ID Contact 38 Range 1 ID Contact 22 Range 2 ID Contact 39 Range 2 ID Contact 19 Range 3 ID Contact 18 Range 3 ID Contact 34 Range 4 ID Contact (Over range) 35 Range 4 ID Contact (Over range) 3.4 Electrolyte Refill of BDS Sen[...]
-
Page 40
Installation BDS 3960 Teledyne Analytical Instruments 40 7. Pour the rest of the electrolyte into the sensor. Gas bubbles in the top portion of the sensor do not affect the sensor performance. One bottle of electrolyte is sufficient to rise the electrolyte level to the MAX line. For the rest of sensor life, no further electrolyte addition is needed[...]
-
Page 41
Oxygen Analyzer Installation Teledyne Analytical Instruments 41 3.6 Powering Up the System Before powering up the system, set the pressure regulator to minimum to prevent damage due to incorrect setting. Power up the system by turning on the switch on the rear. Then turn on the PPC by pressing the button on its left side. Make sure that the Analyze[...]
-
Page 42
Maintenance BDS 3960 Teledyne Analytical Instruments 42 Operation 4.1 Introduction Once the analyzer has been installed, it can be configured for your application. To do this you will: • Set system parameters: • Calibrate the instrument. • Define the three user-selectable analysis ranges, then choose autoranging or select a fixed range of ana[...]
-
Page 43
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 43 4.2 The Analyzer application The Pocket Personal Computer (PPC) is the brains of the analyzer. The PPC runs the Analyzer program installed by TAI. For the instrument to operate, the program must be launched as instructed in section 3.6. When the program is launched the main screen will [...]
-
Page 44
Maintenance BDS 3960 Teledyne Analytical Instruments 44 • System . Pressing this button will open a second screen that consists of several variables that regulate the internal operations of the analyzer. • Training Video . This buttons launches a video player. At this point you could view a short training video regarding BDS technology. The pre[...]
-
Page 45
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 45 amplifi er . The range of this count is the s ame as the B DS s enso r, 0 000 to 4095. The l ast digit is a s tatus di git—zero ind icates t h e I/O P CB is in nor mal mod e. If th e I/ O PC B re ceiv es a remote c omma nd to z ero or s pan, it will let the PPC program know th rough t[...]
-
Page 46
Maintenance BDS 3960 Teledyne Analytical Instruments 46 Figure 4.2: Parameter Selection Box 4.3.2.1 T HE D IGIT AL F ILTER The Analyzer software performs a digital operation to reduce the noise from the BDS sensor and amplifier. This filtering is reduced to around 10 ppb. If the signal exceeds 10 ppb, the filter disengages in order to track the tra[...]
-
Page 47
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 47 * Output normalized by calibrating with a span gas in N 2 . The values in Table 4-1 are theoretical but should give a first approximation as to what the output of the sensor will be in the new background gas. The BDS 3960 will make further refinements once the proper gas factor is input[...]
-
Page 48
Maintenance BDS 3960 Teledyne Analytical Instruments 48 3. After the sensor has been purged for at least two weeks and the baseline is stable, monitor the oxygen reading and ambient temperature over a minimum period of 24 hours. Take the maximum and minimum oxygen readings, and the maximum and minimum temperature readings. 4. Calculate the coeffici[...]
-
Page 49
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 49 Filter 5 minutes Gas Factor 1.00 Temperature Coef. 0 ppb/degree centigrade Alarm 1 setpoint 1000 ppb Alarm 2 setpoint 100 ppb Pressing the Set Defaults button brings up a confirmation dialog box. Press Yes to reset to the defaults or No to keep the current values. 4.4 Calibration of the[...]
-
Page 50
Maintenance BDS 3960 Teledyne Analytical Instruments 50 A zero calibration is not recommended during the first 10 days of the operation of the cell. The ZERO button on the Main screen is used to enter the zero calibration function. Zero calibration can be performed either automatically or manually. In the automatic mode, an internal timer will purg[...]
-
Page 51
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 51 4.4.2.1 S PAN F AILURE The analyzer checks the output of the cell at the end of the span. If the raw output of the cell is less than 1.5 nA/ppb or more than 13.5 nA/ppb O 2 , the span will not be accepted. The analyzer will return to the previous calibration values, and display at the t[...]
-
Page 52
Maintenance BDS 3960 Teledyne Analytical Instruments 52 2. Are either or both of the alarms to be configured as failsafe? In failsafe mode, the alarm relay de-energizes in an alarm condition. For non-failsafe operation, the relay is energized in an alarm condition. You can set either or both of the concentration alarms to operate in failsafe or non[...]
-
Page 53
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 53 4.6 The Range Function The Range function allows the operator to program up to four concentration ranges to correlate with the DC analog outputs. If no ranges are defined by the user, the instrument defaults to: Range 1 = 0–100 ppb Range 2 = 0–1000 ppb Range 3 = 0–10000 ppb Range [...]
-
Page 54
Maintenance BDS 3960 Teledyne Analytical Instruments 54 Figure 4.3: Range Options List Box To Set the ranges to values different than the default, click on ‘Set Range’ and the Set Range screen will pop up. Figure 4.4: Range Setup Screen The first screen allows you to set range 1. If you click on the Range button, it will take you to the setup o[...]
-
Page 55
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 55 as 0–1000 ppb, range 3 cannot be set as 0– 500 ppb since it is lower than range 2. Note: Refer to Section 4.3.2.2 to find maximum working range. 4.6.2 Fixed Range Analysis The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single p[...]
-
Page 56
Maintenance BDS 3960 Teledyne Analytical Instruments 56 60 0.6 13.6 70 0.7 15.2 80 0.8 16.8 90 0.9 18.4 100 1.0 20.0 The analog output signal has a voltage which depends on the oxygen concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument i[...]
-
Page 57
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 57 displayed. This is because windows CE® does not allow two instances of the same program. If this occurs, tap on the blue the corner of the blue bar (blue bar appears either at the top of the screen or at the bottom, depending on which application is at the front), next to the windows l[...]
-
Page 58
Maintenance BDS 3960 Teledyne Analytical Instruments 58 Maintenance 5.1 Routine Maintenance Aside from normal cleaning and checking for leaks at the gas connections, routine maintenance is limited to refilling sensor with deionized water, replace burned fuses, replace backup batteries for PPC, and recalibration. For recalibration, see Section 4.4 C[...]
-
Page 59
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 59 Figure 5.1 Adding Water into the BDS sensor WARNING: THE SENSOR USED IN THE MODEL BDS 3960 OXYGEN ANALYZER USES ELECTROLYTE WHICH CONTAINS POTASSIUM HYDROXIDE, THAT CAN BE HARMFUL IF TOUCHED, SWALLOWED, OR INHALED. AVOID CONTACT WITH ANY FLUID OR POWDER IN OR AROUND THE UNIT. WHAT MAY A[...]
-
Page 60
Maintenance BDS 3960 Teledyne Analytical Instruments 60 Figure 5-2: Removing Fuse Block from Housing 5.4 Battery Backup Replacement The software application running on the PPC is mantained when power is removed from the analyzer by a single 3 volt lithium battery. If the unit is powered down and the battery power is low, you run the risk of losing [...]
-
Page 61
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 61 5.5 Reinstalling Application software to PPC. If the applications software on the PPC becomes corrupt or is lost, it can be restored by Downloading the Analyzer program and components. Before installing the software, you must have a software version of ActiveSync® on a separate PC comp[...]
-
Page 62
Maintenance BDS 3960 Teledyne Analytical Instruments 62 10. Disconnect the serial cable and run the Analyzer program from the Start pull down menu to check for successful installation. 5.5 Major Internal Components The Se ns or is a ccess e d by l ooseni ng the thumbscrew on the f ront panel and slidi ng it o ut, as described e arlier. Oth er int e[...]
-
Page 63
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 63 Read higher than expected (1), (2), (3) Read lower than expected (2), (3) Read negative (3), (4) Noise signal (3), (5) Slow response (5) Causes and solution keys: (1) Gas leak: Make sure to use new VCR gaskets, high quality valves and gas regulator for the sampling system. Tighten each [...]
-
Page 64
Maintenance BDS 3960 Teledyne Analytical Instruments 64 Figure 5-3: Vacuum Degassing for the BDS Oxygen Sensor[...]
-
Page 65
Oxygen Analyzer Maintenance Teledyne Analytical Instruments 65[...]
-
Page 66
Appendix BDS 3960 Teledyne Analytical Instruments 66 Appendix A-1 Specifications Packaging: General Purpose • Flush panel mount (Standard). Sensor: Teledyne BDS Sensor, patent pending. Sample System: All wetted parts of 316 stainless steel with built-in in pressure regulator, oxygen scrubber and calibration valves. 90 % Response Time: Less than 9[...]
-
Page 67
Oxygen Analyzer Appendix Teledyne Analytical Instruments 67 Operating Temperature: 5-40 °C Accuracy: ±2% of full scale for all ranges at constant temperature. All accuracy specifications are contingent upon the completion of zero and span calibration. All accuracy is established at constant pressure and equilibrium has been established. Analog ou[...]
-
Page 68
Appendix BDS 3960 Teledyne Analytical Instruments 68 A-2 Recommended 2- Year Spare Parts List Qty. Part Number Description 1 C65507B Back Panel Board 1 C72914A Front Panel Board 1 C72000A Preamplifier Board (Instruction) 1 B71997A Main Computer Board 2 F1296 Fuse, 2A, 250V 5x20mm (European) Slow Blow 1 CP1798 50 pin D-sub interface connector 50 CP1[...]
-
Page 69
Oxygen Analyzer Appendix Teledyne Analytical Instruments 69 A-3 Drawing List D-73299 Outline Diagram A-4 19-inch Relay Rack Panel Mount Figure A-1: Single 19" Rack Mount (dimensions in mm)[...]
-
Page 70
Appendix BDS 3960 Teledyne Analytical Instruments 70 A-5 Application notes Pressure and flow recommendations: 3960 series analyzers require reasonably regulated sample pressures. While the 3960 analyzers are not sensitive to variations of incoming pressure (provided they are properly vented to atmospheric pressure), the pressure must be maintained [...]
-
Page 71
Oxygen Analyzer Appendix Teledyne Analytical Instruments 71 Material Safety Data Sheet Section I - Product Identification Product Name: Micro-fuel Cells Mini-Micro-fuel Cells Super Cell, all classes except T-5F Electrochemical Oxygen Sensors, all classes Manufacturer: Teledyne Electronic Technologies Analytical Instruments Address: 16380 Chestnut S[...]
-
Page 72
Appendix BDS 3960 Teledyne Analytical Instruments 72 Section III -Physical Hazards Potential for fire and explosion: The electrolyte in the Micro-fuel Cells is not flammable. There are no fire or explosion hazards associated with Micro-fuel Cells. Potential for reactivity: The sensors are stable under normal conditions of use. Avoid contact between[...]
-
Page 73
Oxygen Analyzer Appendix Teledyne Analytical Instruments 73 Section V - Emergency and First Aid Procedures Eye Contact: Flush eyes with water for at least 15 minutes and get immediate medical attention. Skin Contact: Wash affected area with plenty of water and remove contaminated clothing. If burning persists, seek medical attention. Ingestion: Giv[...]
-
Page 74
Appendix BDS 3960 Teledyne Analytical Instruments 74 Index ac ces s d oo r, 16 accuracy, 67 ActiveSync, 61 adding electrolyte, 39 adding water, 59 address, 68, See company address Ag/Ag 2 O electrode, 19 air saturation, 22 alarm, 51 , 66 concentration, 14, 16 failure, 14, 51 high, 36 low, 36 relay, 35, 51 relay contact pins, 36 resetting, 52 system[...]
-
Page 75
Oxygen Analyzer Index Teledyne Analytical Instruments 75 digital to analog converter, 26 dimensions, 67 , 69 d is pl ay , 26, 66 dissolved oxygen, 20 door. See access door door clearance, 31 drawer, 14 drawings, 69 dri ft, 25 electrochemical cell, 19 electrochemical device, 18 electrochemical reaction, 19 electrode, 18 electrolyte, 19, 20, 25, 30, [...]
-
Page 76
Appendix BDS 3960 Teledyne Analytical Instruments 76 reference electrode, 18, 21 refilling the sensor, 39 regulator, 32 remote calibration connection, 37 remote calibration protocol, 37 remote probe, 16 response time, 33, 66 reticulated vitreous carbon, 20 RFI, 14 ROM, 13, 26 safety information, iv sample flow. See flowrate sample system, 18, 22, 2[...]