Tektronix TDS 520A manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290

Ir para a página of

Bom manual de uso

As regras impõem ao revendedor a obrigação de fornecer ao comprador o manual com o produto Tektronix TDS 520A. A falta de manual ou informações incorretas fornecidas ao consumidor são a base de uma queixa por não conformidade do produto com o contrato. De acordo com a lei, pode anexar o manual em uma outra forma de que em papel, o que é frequentemente utilizado, anexando uma forma gráfica ou manual electrónicoTektronix TDS 520A vídeos instrutivos para os usuários. A condição é uma forma legível e compreensível.

O que é a instrução?

A palavra vem do latim "Instructio" ou instruir. Portanto, no manual Tektronix TDS 520A você pode encontrar uma descrição das fases do processo. O objetivo do manual é instruir, facilitar o arranque, a utilização do equipamento ou a execução de determinadas tarefas. O manual é uma coleção de informações sobre o objeto / serviço, um guia.

Infelizmente, pequenos usuários tomam o tempo para ler o manual Tektronix TDS 520A, e um bom manual não só permite conhecer uma série de funcionalidades adicionais do dispositivo, mas evita a formação da maioria das falhas.

Então, o que deve conter o manual perfeito?

Primeiro, o manual Tektronix TDS 520A deve conte:
- dados técnicos do dispositivo Tektronix TDS 520A
- nome do fabricante e ano de fabricação do dispositivo Tektronix TDS 520A
- instruções de utilização, regulação e manutenção do dispositivo Tektronix TDS 520A
- sinais de segurança e certificados que comprovam a conformidade com as normas pertinentes

Por que você não ler manuais?

Normalmente, isso é devido à falta de tempo e à certeza quanto à funcionalidade específica do dispositivo adquirido. Infelizmente, a mesma ligação e o arranque Tektronix TDS 520A não são suficientes. O manual contém uma série de orientações sobre funcionalidades específicas, a segurança, os métodos de manutenção (mesmo sobre produtos que devem ser usados), possíveis defeitos Tektronix TDS 520A e formas de resolver problemas comuns durante o uso. No final, no manual podemos encontrar as coordenadas do serviço Tektronix na ausência da eficácia das soluções propostas. Atualmente, muito apreciados são manuais na forma de animações interessantes e vídeos de instrução que de uma forma melhor do que o o folheto falam ao usuário. Este tipo de manual é a chance que o usuário percorrer todo o vídeo instrutivo, sem ignorar especificações e descrições técnicas complicadas Tektronix TDS 520A, como para a versão papel.

Por que ler manuais?

Primeiro de tudo, contem a resposta sobre a construção, as possibilidades do dispositivo Tektronix TDS 520A, uso dos acessórios individuais e uma gama de informações para desfrutar plenamente todos os recursos e facilidades.

Após a compra bem sucedida de um equipamento / dispositivo, é bom ter um momento para se familiarizar com cada parte do manual Tektronix TDS 520A. Atualmente, são cuidadosamente preparados e traduzidos para sejam não só compreensíveis para os usuários, mas para cumprir a sua função básica de informação

Índice do manual

  • Página 1

    User Manual TDS 520A, 524A, 540A, & 544A Digitizing Oscilloscopes 070-8710-01 Please check for change information at the rear of this manual. First Printing: July 1993 Revised: November 1993[...]

  • Página 2

    Instrument Serial Numbers Each instrument manufactured by T ektronix has a serial number on a panel insert or tag, or stamped on the chassis. The first letter in the serial number designates the country of manufacture. The last five digits of the serial number are assigned sequentially and are unique to each instrument. Those manufactured in the Un[...]

  • Página 3

    W ARRANTY T ektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, T ektronix, at its option, either will repair the defective product without charge for parts and labor , or will provide a[...]

  • Página 4

    German Postal Information Certificate of the Manufacturer/Importer We hereby certify that the TDS 520A, TDS 524A, TDS 540A, and TDS 544A Oscilloscopes and all factory-installed options complies with the RF Interference Suppression requirements of Postal Regulation Vfg. 243/1991, Amended per Vfg. 46/1992 The German Postal Service was notified that t[...]

  • Página 5

    EC Declaration of Conformity We T ektronix Holland N.V . Marktweg 73A 8444 AB Heerenveen The Netherlands declare under sole responsibility that the TDS 520A, 524A, 540A, & 544A Digitizing Oscilloscopes meet the intent of Directive 89/336/EEC for Electromagnetic Compatibility . Compliance was demonstrated to the following specifications as liste[...]

  • Página 6

    [...]

  • Página 7

    TDS 620A, 640A, & 644A User Manual W elcome This is the User Manual for the TDS Family Digitizing Oscilloscopes . The Getting Started section familiarizes you with the operation of the digitizing oscilloscope. Operating Basics covers basic principles of the operation of the oscilloscope. These articles help you understand why your instrument wo[...]

  • Página 8

    Welcome Welcome In the Getting Started and Reference sections, you will find various proce- dures which contain steps of instructions for you to perform. T o keep those instructions clear and consistent, this manual uses the following conventions: In procedures, names of front panel controls and menu labels appear in boldface print. Names also appe[...]

  • Página 9

    TDS 520A, 524A, 540A, & 544A User Manual T able of Contents Safety v . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Getting Started Product Description 1-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Start Up 1-3 . . . . . . . . . . . . . . . . . [...]

  • Página 10

    T able of Contents Contents File System (Optional on TDS 520A & TDS 540A) 3-55 . . . . . . . . . . . . . . Hardcopy 3-59 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Help 3-67 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H[...]

  • Página 11

    TDS 620A, 640A, & 644A User Manual Safety Please take a moment to review these safety precautions. They are provided for your protection and to prevent damage to the digitizing oscilloscope. This safety information applies to all operators and service personnel. These two terms appear in manuals: statements identify conditions or practices that[...]

  • Página 12

    Safety Safety Observe all of these precautions to ensure your personal safety and to pre- vent damage to either the digitizing oscilloscope or equipment connected to it. Power Source The digitizing oscilloscope is intended to operate from a power source that will not apply more than 250 V RMS between the supply conductors or between either supply c[...]

  • Página 13

    Getting Started[...]

  • Página 14

    TDS 520A, 524A, 540A, & 544A User Manual Product Description Y our T ektronix digitizing oscilloscope is a superb tool for acquiring, displaying, and measuring waveforms. Its performance addresses the needs of both benchtop lab and portable applications with the following features: 500 MHz maximum analog bandwidth. 1 Gigasample/second maximum d[...]

  • Página 15

    Product Description Getting Started[...]

  • Página 16

    TDS 620A, 640A, & 644A User Manual Start Up Before you use the digitizing oscilloscope, ensure that it is properly installed and powered on. T o ensure maximum accuracy for your most critical measurements, you should know about signal path compensation. Signal Path Compensation Be sure you compensate your oscilloscope for the surrounding temper[...]

  • Página 17

    Start Up Getting Started 4. Check that you have the proper electrical connections. The digitizing oscilloscope requires 90 to 250 V AC RMS , continuous range, 47 Hz to 63 Hz, and may require up to 300 W . 5. Connect the proper power cord from the rear-panel power connector (see Figure 1-1) to the power system. Power Connector Principal Power Switch[...]

  • Página 18

    Start Up TDS 620A, 640A, & 644A User Manual Once the digitizing oscilloscope is installed, it is typical to leave the princi- pal power switch on and use the ON/STBY button as the power switch. ON/STBY Button Figure 1-2: ON/STBY Button Self T est Check the self test results. The digitizing oscilloscope automatically performs power-up tests each[...]

  • Página 19

    Getting Started Setting Up for the Examples All the examples use the same setup. Once you perform this setup, you do not have to change the signal connections for any of the other examples. Remove all probes and signal inputs from the input BNC connectors along the lower right of the front panel. Then, using one of the probes supplied with the digi[...]

  • Página 20

    TDS 620A, 640A, & 644A User Manual Example 1: Displaying a W aveform In this first example you learn about resetting the digitizing oscilloscope, displaying and adjusting a waveform, and using the autoset function. All examples in the tutorial begin by resetting the digitizing oscilloscope to a known factory default state. Reset the oscilloscop[...]

  • Página 21

    Example 1: Displaying a W aveform Getting Started 2. Press the button directly below the Recall Factory Setup menu item. The display shows side menus along the right side of the screen. The buttons to select these side menu items are to the right of the side menu. Because an accidental instrument reset could destroy a setup that took a long time to[...]

  • Página 22

    Example 1: Displaying a W aveform TDS 620A, 640A, & 644A User Manual Figure 1-7 shows the display that results from the instrument reset. There are several important points to observe: The trigger level bar shows that the waveform is triggered at a level near 50% of its amplitude (from step 4). The trigger position indicator shows that the trig[...]

  • Página 23

    Example 1: Displaying a W aveform Getting Started The display shows the probe compensation signal. It is a 1 kHz square wave of approximately 0.5 V amplitude. Y ou can adjust the size and placement of the waveform using the front-panel knobs. Figure 1-8 shows the main VERTICAL and HORIZONT AL sections of the front panel. Each has SCALE and POSITION[...]

  • Página 24

    Example 1: Displaying a W aveform TDS 620A, 640A, & 644A User Manual 1. T o create an unstable display , slowly turn the trigger MAIN LEVEL knob (see Figure 1-9) first one direction, then the other . Observe what happens when you move the trigger level above the highest part of the displayed waveform. Leave the trigger level in that untriggered[...]

  • Página 25

    Example 1: Displaying a W aveform Getting Started Figure 1-1 1: The Display After Pressing Autoset NOTE If the corners on your displayed signal look rounded or pointed (see Figure 1-12), then you may need to compensate your probe. The Probe Compensation section on page 3-1 10 explains how to com- pensate your probe. Figure 1-12: Display Signals Req[...]

  • Página 26

    TDS 620A, 640A, & 644A User Manual Example 2: Multiple W aveforms In this example you learn how to display and control more than one waveform at a time. The VERTICAL section of the front panel contains the channel selection buttons. These are CH 1 , CH 2 , CH 3 , CH 4 , and MORE (Figure 1-13); on the TDS 620A & 524A, they are CH 1 , CH 2 , [...]

  • Página 27

    Example 2: Multiple W aveforms Getting Started 3. Press AUT OSET . 4. Press CH 2 . The display shows a second waveform, which represents the signal on channel 2. Since there is nothing connected to the CH 2 input connector , this waveform is a flat line. There are several other important things to observe: The channel readout on the display now sho[...]

  • Página 28

    Example 2: Multiple W aveforms TDS 620A, 640A, & 644A User Manual Side Menu Title Ch2 Reference Indicator Figure 1-14: The V ertical Main Menu and Coupling Side Menu Pressing a channel ( CH ) button sets the vertical controls to that channel. It also adds the channel to the display if that waveform is not already displayed. 1. Press CH 1 . Obse[...]

  • Página 29

    Example 2: Multiple W aveforms Getting Started Side Menu Title Figure 1-15: The Menus After Changing Channels Pressing the W A VEFORM OFF button removes the waveform for the current- ly selected channel. If the waveform you want to remove is not already se- lected, select that channel using the channel ( CH ) button. 1. Press W A VEFORM OFF (under [...]

  • Página 30

    TDS 620A, 640A, & 644A User Manual Example 3: Automated Measurements In this example you learn how to use the automated measurement system to get numeric readouts of important waveform characteristics. T o use the automated measurement system, you must have a stable display of your signal. Also, the waveform must have all the segments necessary[...]

  • Página 31

    Example 3: Automated Measurements Getting Started 5. If it is not already selected, press Select Measrmnt (main). The readout for that menu item indicates which channel the measurement will be taken from. All automated measurements are made on the selected channel. The Select Measurement side menu lists some of the measurements that can be taken on[...]

  • Página 32

    Example 3: Automated Measurements TDS 620A, 640A, & 644A User Manual Press here to remove menus from screen. Figure 1-17: Four Simultaneous Measurement Readouts The Measure menu lets you remove measurements you no longer want displayed. Y ou can remove any one measurement, or you can remove them all with a single menu item. Press MEASURE ➞ Re[...]

  • Página 33

    Example 3: Automated Measurements Getting Started General Purpose Knob Setting and Readout Highlighted Menu Item with Boxed Readout V alue General Purpose Knob Icon Figure 1-18: General Purpose Knob Indicators There are several important things to observe on the screen: The knob icon appears at the top of the screen. The knob icon indicates that th[...]

  • Página 34

    Example 3: Automated Measurements TDS 620A, 640A, & 644A User Manual 1. Press Low Ref (side). 2. On the numeric keypad, press the 2 , the 0 , and the ENTER ( ) but- tons, which sets the low measurement reference to 20%. Observe that the rise-time value has changed. 3. Press Remove Measrmnt (main) ➞ All Measurements (side). That returns the di[...]

  • Página 35

    Example 3: Automated Measurements Getting Started The snapshot measurements do not continuously update. Snapshot executes a one-time capture of all measurements and does not update those measurements unless it is performed again. 2. Press Again (side) to do another snapshot and update the snapshot measurements. 3. Press Remove Measrmnt (main) to re[...]

  • Página 36

    TDS 620A, 640A, & 644A User Manual Example 4: Saving Setups This example shows you how to save all the settings of the digitizing oscillo- scope and how to recall the setup later to quickly re-establish the previously saved state. The oscilloscope provides several storage locations where you can save the setups. With the file system (optional o[...]

  • Página 37

    Example 4: Saving Setups Getting Started Figure 1-20: Save/Recall Setup Menu 7. Press one of the T o Setup side menu buttons to store the current instru- ment settings into that setup location. Remember which setup location you selected for use later . There are more setup locations than can be listed at one time in the side menu. The –more– si[...]

  • Página 38

    Operating Basics[...]

  • Página 39

    TDS 620A, 640A, & 644A User Manual Overview This section describes the basic concepts of operating the digitizing oscillo- scope. Understanding the basic concepts of your digitizing oscilloscope will help you use it much more effectively . The first part, At a Glance , quickly shows you how the oscilloscope is orga- nized and gives some very ge[...]

  • Página 40

    Overview Operating Basics[...]

  • Página 41

    TDS 620A, 640A, & 644A User Manual At a Glance The At a Glance section contains illustrations of the display , the front and rear panels, and the menu system. These will help you understand and operate the digitizing oscilloscope. This section also contains a visual guide to using the menu system. ON/STBY Switch, page 1-3 CLEAR MENU Removes Men[...]

  • Página 42

    At a Glance Operating Basics File System, page 3-55 (Optional on TDS 620A & TDS 640A) Hardcopy , page 3-59 Autoset, page 3-10 Help, page 3-67 Status, page 3-140 Selecting Channels, page 3-136 Horizontal Control, page 3-68 Acquisition Modes, page 3-3 Cursor Measurements, page 3-17 Triggering, page 3-142 Color , page 3-12 (TDS 644A & TDS 524A[...]

  • Página 43

    At a Glance TDS 620A, 640A, & 644A User Manual Security Bracket GPIB Connector page 3-126 Centronics Connector (Optional on TDS 620A & TDS 640A) Principal Power Switch, page 1-3 Serial Number RS-232 Connector (Optional on TDS 620A & TDS 640A) Fuse, page 1-3 Power Connector , page 1-3 VGA Output (Color with TDS 524A & TDS 644A, Monoc[...]

  • Página 44

    At a Glance Operating Basics The value entered with the general purpose knob Channel level and waveform source The acquisition status, page 3-3 The main menu with choices of major actions The side menu with choices of specific actions V ertical scale, page 3-147 Horizontal scale and time base type, page 3-68 Trigger parameters, page 3-144 Cursor me[...]

  • Página 45

    At a Glance TDS 620A, 640A, & 644A User Manual 1. Press front-panel menu button. (Press SHIFT first if button label is blue.) 2. Press one of these buttons to select from main menu. 3. Press one of these buttons to select from side menu (if displayed). 4. If side menu item has an ad- justable value (shown in re- verse video), adjust it with the[...]

  • Página 46

    At a Glance At a Glance Press to display pop-ups. Press it again to make selection. Alternatively , press SHIFT first to make selection in the opposite direction. Press here to remove menus from screen. A pop-up selection changes the other main menu titles. T o Operate a Pop-Up Menu[...]

  • Página 47

    At a Glance TDS 520A, 524A, 540A, & 544A User Manual Acquire Menu (see page 3-3) Press these but- tons: Cursor Menu (see page 3-17) Delayed Trigger Menu (see page 3-22) Display Menu – Color (TDS 524A & TDS 544A) (see page 3-12 ) T o bring up these menus: Application Menu (see the Programmer manual for more details) Display Menu – Displa[...]

  • Página 48

    At a Glance Operating Basics Measure Menu (see page 3-86) Save/Recall Setup Menu (see page 3-130) Save/Recall W aveform Menu (see page 3-133) More Menu (see page 3-159) Main Trigger Menu –Pulse (see page 3-1 19) Main Trigger Menu – Logic (see page 3-78) Status Menu (see page 3-140) Press these buttons: T o bring up these menus: Main Trigger Men[...]

  • Página 49

    At a Glance TDS 620A, 640A, & 644A User Manual Press these buttons: T o bring up these menus: Utility Menu – I/O – GPIB (see page 3-126) V ertical Channel Menu ( see page 3-147) Zoom Menu ( see page 3-162) Utility Menu – Calibration ( see page ) Utility Menu – Diagnostics ( see the Service manual) Utility Menu – Config (see pages ) Ut[...]

  • Página 50

    At a Glance Operating Basics[...]

  • Página 51

    TDS 620A, 640A, & 644A User Manual T riggering This section describes the edge trigger of the main trigger system and ex- plores, in a general sense, the topic of triggering. This oscilloscope also has logic and pulse triggers in the main trigger system and a delayed trigger system. They are described in Section 3. T riggers determine when the [...]

  • Página 52

    T riggering Operating Basics AC Line V oltage — this trigger source is useful when you are looking at signals related to the power line frequency . Examples include devices such as lighting equipment and power supplies. Because the digitizing oscilloscope generates the trigger , you do not have to input a signal to create the trigger . Auxiliary [...]

  • Página 53

    T riggering TDS 620A, 640A, & 644A User Manual Automatic — this trigger mode (auto mode) lets the oscilloscope acquire a waveform even if a trigger does not occur . Auto mode uses a timer that starts after a trigger event occurs. If another trigger event is not detected before the timer times out, the oscilloscope forces a trigger anyway . Th[...]

  • Página 54

    T riggering Operating Basics Holdoff Holdof f Trigger Points Trigger Level Holdoff Acquisition Interval Acquisition Interval Triggers are Not Recognized During Holdof f Time Figure 2-2: T rigger Holdoff Time Ensures V alid T riggering Holdoff is settable from 0% (minimum holdof f available) to 100% (maximum available). T o see how to set holdoff, s[...]

  • Página 55

    T riggering TDS 620A, 640A, & 644A User Manual The adjustable trigger position defines where on the waveform record the trigger occurs. It lets you properly align and measure data within records. The part of the record that occurs before the trigger is the pretrigger portion. The part that occurs after the trigger is the posttrigger portion. T [...]

  • Página 56

    T riggering Operating Basics As mentioned earlier in this section there is also a delayed trigger system that provides an edge trigger (no pulse or logic triggers). When using the delayed time base, you can also delay the acquisition of a waveform for a user-speci- fied time or a user-specified number of delayed trigger events (or both) after a mai[...]

  • Página 57

    TDS 620A, 640A, & 644A User Manual Acquisition Acquisition is the process of sampling the analog input signal, converting it into digital data, and assembling it into a waveform record. The oscilloscope creates a digital representation of the input signal by sampling the voltage level of the signal at regular time intervals (Figure 2-4). Input [...]

  • Página 58

    Acquisition Operating Basics The digitizer can use the extra samples to perform additional processing, such as averaging or looking for minimum and maximum values. The digitizing oscilloscope creates a waveform record containing a user-spe- cified number of data points. Each record point represents a certain voltage level that occurs a determined a[...]

  • Página 59

    Acquisition TDS 620A, 640A, & 644A User Manual If you focus on only one channel at the maximum possible real-time rate, the TDS 524A and TDS 620A oscilloscopes can acquire at 500 Megasamples/ second using both its digitizers, while the TDS 644A and TDS 640A oscillo- scopes can combine all four digitizers and acquire at 1 Gigasample/second. Depe[...]

  • Página 60

    Acquisition Operating Basics 1st Acquisition Cycle 3rd Acquisition Cycle nth Acquisition Cycle 2nd Acquisition Cycle Record Points Figure 2-7: Equivalent-Time Sampling The oscilloscope takes a few samples with each trigger event and eventually constructs a waveform record using the samples from multiple acquisitions. That feature lets you accuratel[...]

  • Página 61

    Acquisition TDS 620A, 640A, & 644A User Manual T able 2-1: Sampling Mode Selection — 100 ns/Div to 50 ns/Div (When Fit to Screen is Off) Instrument and Number of Channels 100 ns/Div 50 ns/Div TDS 544A & 540A, any 1 channel Real-time Real-time TDS 544A & 540A, any 2 channels Real-time Equivalent-time or interpolated real-time TDS 544A [...]

  • Página 62

    Acquisition Operating Basics Ground (GND) coupling disconnects the input signal from the acquisition. Input impedance lets you select either 1 M or 50 impedance. NOTE If you select 50 impedance with AC coupling, the digitizing oscillo- scope will not accurately display frequencies under 200 kHz. See Scaling and Positioning Waveforms , on page 2-25.[...]

  • Página 63

    TDS 620A, 640A, & 644A User Manual Scaling and Positioning W aveforms Scaling and positioning waveforms means increasing or decreasing their displayed size and moving them up, down, right, and left on the display . T wo display icons, the channel reference indicator and the record view , help you quickly see the position of the waveform in the [...]

  • Página 64

    Scaling and Positioning W aveforms Operating Basics Y ou can adjust the vertical position of the selected waveform by moving it up or down on the display . For example, when trying to compare multiple wave- forms, you can put one above another and compare them, or you can overlay the two waveforms on top of each other . T o move the selected wavefo[...]

  • Página 65

    Scaling and Positioning W aveforms TDS 620A, 640A, & 644A User Manual Aliasing When aliasing happens, you see a waveform with a frequency lower than the actual waveform being input or a waveform is not stable even though the light next to TRIG’D is lit. Aliasing occurs because the oscilloscope cannot sample the signal fast enough to construct[...]

  • Página 66

    Scaling and Positioning W aveforms Operating Basics FastFrame TM Y ou can define and enable FastFrame TM (also called “segmented memory”) on the TDS 600A. This feature lets you capture multiple acquisitions in the acquisition memory of a single channel. Figure 2-10 shows how FastFrame combines the desired captured records into one larger record[...]

  • Página 67

    Scaling and Positioning W aveforms TDS 620A, 640A, & 644A User Manual See Autoset , on page 3-10. See Delayed T riggering , on page 3-22. See Horizontal Control , on page 3-68. See V ertical Control , on page 3-147. See Zoom , on page 3-162. For More Information[...]

  • Página 68

    Operating Basics Measurements The digitizing oscilloscope not only displays graphs of voltage versus time, it also can help you measure the displayed information (see Figure 2-1 1). Ch 1 Period 10 ns Automated Measurements Graticule Ch 1 Frequency 100 MHz Cursors Cursor Readouts Figure 2-1 1: Graticule, Cursor and Automated Measurements The oscillo[...]

  • Página 69

    Measurements TDS 620A, 640A, & 644A User Manual The snapshot selection in the Measurement menu lets you display almost all of the measurements at once. Y ou can read about snapshot under Snapshot of Measurements, on page 3-95. Automated measurements use readouts to show measurement status. These readouts are updated as the oscilloscope acquires[...]

  • Página 70

    Measurements Operating Basics T racking mode cursors operate in tandem: you move both cursors at the same time using the general purpose knob. T o adjust the solid cursor relative to the dashed cursor , you push the SELECT button to suspend cursor tracking and use the general purpose knob to make the adjust- ment. A second push toggles the cursors [...]

  • Página 71

    Reference[...]

  • Página 72

    TDS 620A, 640A, & 644A User Manual Overview This section describes the details of operating the digitizing oscilloscope. It contains an alphabetical list of tasks you can perform with the digitizing oscilloscope. Use this section to answer specific questions about instrument operation. These tasks include: Probe Cal Probe Compensation Probe Sel[...]

  • Página 73

    Overview Reference[...]

  • Página 74

    TDS 620A, 640A, & 644A User Manual Acquisition Modes The acquisition system has several options for converting analog data into digital form. The Acquisition menu lets you determine the acquisition mode, whether or not to permit equivalent time sampling, and how to start and stop acquisitions. The digitizing oscilloscope supports five acquisiti[...]

  • Página 75

    Acquisition Modes Reference Waveform Drawn on CRT Samples Acquired in Four Acquisition Intervals Displayed Record Points Acquisition Mode Use for fastest acquisition rate. This is the default mode. Use to reveal aliasing and for glitch detection. Provides the benefits of enveloping with the speed of a single acquisition. Use to reduce apparent nois[...]

  • Página 76

    Acquisition Modes TDS 620A, 640A, & 644A User Manual Hi Res Mode In Hi Res mode, the digitizing oscilloscope averages all samples taken during an acquisition interval to create a record point. That average results in a higher-resolution, lower-bandwidth waveform. This mode only works with real-time, non-interpolated sampling. If you set the tim[...]

  • Página 77

    Acquisition Modes Reference After each trigger event, the oscilloscope acquires data and then compares the min/max values from the current acquisition with those stored from pre- vious acquisitions. The final display shows the most extreme values for all the acquisitions for each point in the waveform record. A verage Mode Average mode lets you acq[...]

  • Página 78

    Acquisition Modes TDS 620A, 640A, & 644A User Manual T o bring up the acquisition menu (Figure 3-2) press SHIFT ACQUIRE MENU . Acquisition Mode T o choose how the digitizing oscilloscope will create points in the waveform record: Press SHIFT ACQUIRE MENU ➞ Mode (main) ➞ Sample , Peak Detect , Hi Res , Envelope, or A verage (side). When you [...]

  • Página 79

    Acquisition Modes Reference Figure 3-3: Acquire Menu — Stop After RUN/ST OP button only (side) lets you start or stop acquisitions by toggling the RUN/STOP button. Pressing the RUN/STOP button once will stop the acquisitions. The upper left hand corner in the display will say Stopped and show the number of acquisitions. If you press the button ag[...]

  • Página 80

    Acquisition Modes TDS 620A, 640A, & 644A User Manual Limit T est Condition Met (side) lets you acquire waveforms until wave- form data exceeds the limits specified in the limit test. Then acquisition stops. At that point, you can also specify other actions for the oscillo- scope to take, using the selections available in the Limit T est Setup m[...]

  • Página 81

    Reference Autoset The autoset function lets you quickly obtain and display a stable waveform of usable size. Autoset automatically sets up the front panel controls based on the characteristics of the input signal. It is much faster and easier than a manual control-by-control setup. Autoset makes adjustments in these areas: Acquisition Display Horiz[...]

  • Página 82

    Autoset TDS 620A, 640A, & 644A User Manual T able 3-2: Autoset Defaults Control Changed by Autoset to Selected channel Numerically lowest of the displayed channels Acquire Mode Sample Acquire Repetitive Signal On Acquire Stop After RUN/ST OP button only Display Style V ectors Display Intensity — Overall (TDS 640A & TDS 620A) If less than [...]

  • Página 83

    Reference Color (TDS 524A & TDS 644A) The TDS 524A & TDS 644A can display information in different colors. The Color menu lets you choose palettes of colors and decide what colors to assign to what pieces of information. T o bring up the Color menu: 1. Press DISPLA Y to show the Display menu. 2. Press Settings in the main menu until you sel[...]

  • Página 84

    Color TDS 620A, 640A, & 644A User Manual Choose Palette Y ou can choose a palette of 13 colors from a menu of pre-set palettes. 1. Choose the starting palette by selecting Palette from the main menu. 2. Select one of the available palettes in the side menu. Choose from Nor- mal , Bold , Hardcopy Preview or Monochrome . 3. If you are using a per[...]

  • Página 85

    Color Reference ScrTxt Figure 3-5: Display Menu — Palette Colors 2. Select one of the 13 colors by pressing (repeatedly) Color in the side menu. 3. If you want to use the factory default for this color , press the side menu Reset to Factory Color . 4. Choose Hue from the side menu and use the general purpose knob or keypad to select the desired h[...]

  • Página 86

    Color TDS 620A, 640A, & 644A User Manual 3. If you want to assign the selected math waveform to a specific color , press Color and cycle through the choices. 4. If you want the selected math waveform to be the same color as the waveform it is based on, select Color Matches Contents . If the math waveform is based on dual waveforms, the math wav[...]

  • Página 87

    Color Reference Select Options T o define what color to show where a waveform crosses another waveform: 1. Press the Options main menu item. 2. Select that you wish to use a special color to mark collision zones by toggling Collision Contrast in the side menu to ON . Restore Colors T o restore colors to their factory default settings: 1. Press the [...]

  • Página 88

    TDS 620A, 640A, & 644A User Manual Cursor Measurements Use the cursors to measure the difference (either in time or voltage) between two locations in a waveform record. Cursors are made up of two markers that you position with the general pur- pose knob. Y ou move one cursor independently or both cursors in tandem, depending on the cursor mode.[...]

  • Página 89

    Cursor Measurements Reference NOTE When cursors measure certain math waveforms, the measurement may not be of time, frequency , or voltage. Cursor measurement of those math waveforms that are not of time, frequency , or voltage is described in Waveform Math, which begins on page 3-159. There are two cursor modes: independent and tracking (see Figur[...]

  • Página 90

    Cursor Measurements TDS 620A, 640A, & 644A User Manual In FastFrame mode, the @ shows the time position of the selected cursor relative to the trigger point of the frame that the selected cursor is in. The shows the time difference between the two cursors only if both cursors are in the same frame. Paired: the value after one shows the voltage [...]

  • Página 91

    Cursor Measurements Reference Non-selected Cursor (Dashed V ertical Bar) Cursor Readout (Paired) Selected Cursor (Solid V ertical Bar) Position of V ertical Bar Cursors (Useful for Locating Cursors Outside the Display) Figure 3-1 1: Paired Cursor Menu and Readouts Mode Select the cursor mode you want using the Mode menu item. 1. Press CURSOR ➞ Mo[...]

  • Página 92

    Cursor Measurements TDS 620A, 640A, & 644A User Manual Time Units Y ou can choose to display vertical bar cursor results in units of time or fre- quency . If you have Option 5 V ideo, you can also display the results in terms of video line number . Press CURSOR ➞ T ime Units (main) ➞ seconds or 1/seconds (Hz) or , with Option 5, video line [...]

  • Página 93

    Reference Delayed T riggering The TDS 600A Series oscilloscopes provide a main time base and a delayed time base. The delayed time base, like the main time base, requires a trigger signal and an input source dedicated to that signal. Y ou can only use delay with respect to the main edge trigger and certain classes of main pulse trig- gers. There ar[...]

  • Página 94

    Delayed T riggering TDS 620A, 640A, & 644A User Manual After Time waits the user-specified time, then waits for the next delayed trigger event, and then acquires. After Events waits for the specified number of delayed trigger events and then acquires. After Events/Time waits for the specified number of delayed trigger events, then waits the use[...]

  • Página 95

    Delayed T riggering Reference Time Delay (From Delay Trig Menu) Time Delay (From Delay Trig Menu) Start Posttrigger Acquisition (T rigger on nth Delayed Trigger Event) Time Delay (From Horiz Menu) Time Delay (From Delay Trig Menu) Waiting for nth Event (Where n=5) Main Trigger Point Start Posttrigger Acquisition Main Trigger Source Delayed Triggera[...]

  • Página 96

    Delayed T riggering TDS 620A, 640A, & 644A User Manual Delayed T riggerable Y ou must make sure that the Main Trigger menu settings are compatible with Delayed T riggerable. 1. Press TRIGGER MENU . 2. If T ype is set to Logic , press T ype (main) to change it to either Edge or Pulse as fits your application. Logic type is incompatible with Dela[...]

  • Página 97

    Delayed T riggering Reference The Source menu lets you select which input will be the delayed trigger source. 7. Press Source (main) ➞ Ch1, Ch2, Ch3 ( Ax1 on the TDS 524A & TDS 620A), Ch4 ( Ax2 on the TDS 524A & TDS 620A), or Auxiliary (not available on the TDS 524A & TDS 620A) (side). Figure 3-15: Delayed T rigger Menu 8. Press Coupl[...]

  • Página 98

    Delayed T riggering TDS 620A, 640A, & 644A User Manual NOTE When you set the V ertical SCALE smaller than 200 mV , the oscillo- scope reduces the Set to TTL or Set to ECL trigger levels below standard TTL and ECL levels. That happens because the trigger level range is fixed at center . At 100 mV (the next smaller setting after 200 mV) the trigg[...]

  • Página 99

    Reference Display Modes The digitizing oscilloscope can display waveform records in different ways. The Display menu lets you adjust the oscilloscope display style, intensity level, graticule, and format. T o bring up the Display menu: 1. Press DISPLA Y to show the Display menu. 2. On the TDS 644A & TDS 524A, press Setting in the main menu unti[...]

  • Página 100

    Display Modes TDS 620A, 640A, & 644A User Manual V ectors has the display draw vectors (lines) between the record points. Dots display waveform record points as dots. Intensified Samples also displays waveform record points as dots. However , the points actually sampled are displayed in the Zone color (TDS 644A & TDS 524A) or intensified re[...]

  • Página 101

    Display Modes Reference Display Readout Readout options control whether the trigger indicator , trigger level bar , and current date and time appear on the display . The options also control what style trigger level bar , long or short, is displayed. 1. Press DISPLA Y ➞ Readout (main). 2. T oggle Display ‘T’ @ T rigger Point (side) to select [...]

  • Página 102

    Display Modes TDS 620A, 640A, & 644A User Manual Filter T ype The display filter types are sin(x)/x interpolation and linear interpolation. For more information see the Concepts section, page 2-21. Press DISPLA Y ➞ Filter (main) ➞ Sin(x)/x Interpolation or Linear Interpo- lation (side). NOTE When the horizontal scale is set to rates faster [...]

  • Página 103

    Display Modes Reference XY format compares the voltage levels of two waveform records point by point. That is, the digitizing oscilloscope displays a graph of the voltage of one waveform record against the voltage of another waveform record. This mode is particularly useful for studying phase relationships. T o set the display axis format: Press DI[...]

  • Página 104

    Display Modes TDS 620A, 640A, & 644A User Manual See Acquisition, on page 2-19. See Color , on page 3-12. See Measurements, on page 2-30. For More Information[...]

  • Página 105

    Reference Edge T riggering An edge trigger event occurs when the trigger source passes through a specified voltage level in a specified direction (the trigger slope). Y ou will likely use edge triggering for most of your measurements. Y ou can select the edge source, coupling, slope, level, and mode (auto or normal). The T rigger readout shows some[...]

  • Página 106

    Edge T riggering TDS 620A, 640A, & 644A User Manual Figure 3-19: Main T rigger Menu — Edge T ype Coupling T o select the coupling you want: Press TRIGGER MENU ➞ T ype (main) ➞ Edge (pop-up) ➞ Cou- pling (main) ➞ DC , AC , HF Rej , LF Rej , or Noise Rej (side). DC passes all of the input signal. In other words, it passes both AC and DC[...]

  • Página 107

    Edge T riggering Reference Slope T o select the slope that the edge trigger will occur on: 1. Press the TRIGGER MENU ➞ T ype (main) ➞ Edge (pop-up) ➞ Slope (main). 2. Alternatives for slope are the rising and falling edges. Level Press the TRIGGER MENU ➞ T ype (main) ➞ Edge (pop-up) ➞ Level (main) ➞ Level , Set to TTL , Set to ECL , o[...]

  • Página 108

    Edge T riggering TDS 620A, 640A, & 644A User Manual Mode & Holdoff Y ou can change the holdoff time and select the trigger mode using this menu item. See T riggering on page 2-13 for more details. 1. Press the TRIGGER MENU ➞ Mode & Holdoff (main) ➞ Auto or Nor- mal (side). In Auto mode the oscilloscope acquires a waveform after a sp[...]

  • Página 109

    Reference Fast Fourier T ransforms Advanced DSP Math (optional on TDS 620A & TDS 640A), provides the Fast Fourier T ransform (FFT). The FFT allows you to transform a waveform from a display of its amplitude against time to one that plots the amplitudes of the various discrete frequencies the waveform contains. Further , you can also display the[...]

  • Página 110

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual zontal scale for FFT math waveforms is always expressed in frequency per division with the beginning (left-most point) of the waveform representing zero frequency (DC). The FFT waveform is based on digital signal processing (DSP) of data, which allows more versatility in measuring the [...]

  • Página 111

    Fast Fourier T ransforms Reference Figure 3-21: Define FFT W aveform Menu 5. If the selected math waveform is not FFT , press Change Math Definition (side) ➞ FFT (main). See Figure 3-21. 6. Press Set FFT Source to (side) repeatedly until the channel source selected in step 1 appears in the menu label. 7. Press Set FFT V ert Scale to (side) repeat[...]

  • Página 112

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Hamming — V ery good window for resolving frequencies that are very close to the same value with somewhat improved amplitude accuracy over the rectangular window . Hanning — V ery good window for measuring amplitude accuracy but degraded for resolving frequencies. Blackman-Harris ?[...]

  • Página 113

    Fast Fourier T ransforms Reference Cursor Measurements of an FFT Once you have displayed an FFT math waveform, use cursors to measure its frequency amplitude or phase angle. 1. Be sure MORE is selected in the channel selection buttons and that the FFT math waveform is selected in the More main menu. 2. Press CURSOR ➞ Mode (main) ➞ Independent ([...]

  • Página 114

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Figure 3-23: Cursor Measurement of an FFT W aveform 10. Press Function (main) ➞ Paired (side). 1 1. Use the technique just outlined to place the vertical bar of each paired cursor to the points along the horizontal axis you are interested in. 12. Read the amplitude between the X of t[...]

  • Página 115

    Fast Fourier T ransforms Reference The FFT Frequency Domain Record The following topics discuss the relation of the source waveform to the record length, frequency resolution, and frequency range of the FFT frequency domain record. (The FFT frequency domain waveform is the FFT math waveform that you display .) FFT s May Not Use All of the W aveform[...]

  • Página 116

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Zero Phase Reference W aveform Record > 10 K W aveform Record ≤ 10 K FFT Time Domain Record = Waveform Record Zero Phase Reference FFT Time Domain Record = 10k Figure 3-24: W aveform Record vs. FFT Time Domain Record FFT s T ransform Time Records to Frequency Records — The FFT t[...]

  • Página 117

    Fast Fourier T ransforms Reference Where: F is the frequency resolution. Sample Rate is the sample rate of the source waveform. FFT Length is the length of the FFT Time Domain wave- form record. The sample rate also determines the range these frequencies span; they span from 0 to the sample rate is often referred to as the Nyquist frequency or poin[...]

  • Página 118

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Record Length Most often, you will want to use a short record length because more of the FFT waveform can be seen on screen and long record lengths can slow oscilloscope response. However , long record lengths lower the noise relative to the signal and increase the frequency resolution[...]

  • Página 119

    Fast Fourier T ransforms Reference Zoom always uses either sin(x)/x or linear interpolation when expanding displayed waveforms. T o select the interpolation method: press DISPLA Y ➞ Setting (main) ➞ Display (pop-up) ➞ Filter (main) ➞ Sin(x)/x or Linear (side), or if your oscilloscope does not have color , press DISPLA Y ➞ Fil- ter (main) [...]

  • Página 120

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Nyquist Frequency Point Frequency Aliased Frequencies Actual Frequencies Figure 3-26: How Aliased Frequencies Appear in an FFT Considerations for Phase Displays When you set up an FFT math waveform to display the phase angle of the frequencies contained in a waveform, you should take i[...]

  • Página 121

    Fast Fourier T ransforms Reference For records with a 15 K length, set the trigger position to 33%. Use the horizontal position knob to move the trigger T on screen to the center horizontal graticule line. For records with 30 K, 50 K, or 60 K lengths (not all lengths are available for all TDS models — consult your User manual), set the trigger po[...]

  • Página 122

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Windowing Process — The oscilloscope multiplies the FFT time domain record by one of four FFT windows before it inputs the record to the FFT function. Figure 3-27 shows how the time domain record is processed. The FFT windowing acts like a bandpass filter between the FFT time domain [...]

  • Página 123

    Fast Fourier T ransforms Reference FFT Time Domain Record FFT Time Domain Record After Windowing FFT Window FFT Frequency Domain Record FFT Xs Figure 3-27: Windowing the FFT T ime Domain Record Y ou can often determine the best window empirically by first using the window with the most frequency resolution (rectangular), then proceeding toward that[...]

  • Página 124

    Fast Fourier T ransforms TDS 620A, 640A, & 644A User Manual Window Characteristics — When evaluating a window for use, you may want to examine how it modifies the FFT time domain data. Figure 3-28 shows each window , its bandpass characteristic, bandwidth, and highest side lobe. Consider the following characteristics: The narrower the central[...]

  • Página 125

    Fast Fourier T ransforms Reference Rectangular Window Hanning Window Hamming Window Blackman-Harris Window FFT Window T ype Bandpass Filter 0.89 -3 dB Bandwidth Highest Side Lobe 1.28 1.28 1.28 -13 dB –32 dB –43 dB –94 dB Figure 3-28: FFT Windows and Bandpass Characteristics Care should be taken when using bell shaped widows to be sure that t[...]

  • Página 126

    TDS 620A, 640A, & 644A User Manual File System (Optional on TDS 620A & TDS 640A) The File Utilities menu, which comes with the Hardcopy , Save Setup, and Save W aveforms menus, gives you a variety of features for managing the floppy disk. The File Utilities menu lets you delete, rename, copy , print files, create a new directory , operate t[...]

  • Página 127

    File System Reference NOTE The amount of free space on the disk is shown in the upper right corner of the display . The digitizing oscilloscope shows the amount in K bytes. T o convert the amount to bytes, you simply multiply the K bytes amount times 1024. Thus, the 71 1 kB shown in Figure 3-29 = 71 1 Kbytes * 1024 bytes/K = 728,064 bytes. Delete T[...]

  • Página 128

    File System TDS 620A, 640A, & 644A User Manual Figure 3-30: File System — Labelling Menu Copy T o copy a file or directory , turn the general purpose knob until it scrolls the cursor over the name of the file to copy . Then, press the side menu Copy button. The file menu will reappear with the names of directories to copy to. Select a directo[...]

  • Página 129

    File System Reference The labelling menu should appear . T urn the general purpose knob or use the main-menu arrow keys to select each letter . Press Enter Char from the main menu to enter each letter . When you have entered the name, press the side menu OK Accept item. (See Figure 3-30) Confirm Delete T o turn on or off the confirm delete message,[...]

  • Página 130

    TDS 520A, 524A, 540A, & 544A User Manual Hardcopy Y ou can get a copy of the digitizing oscilloscope display by using the hardco- py feature. Depending on the output format you select, you create either an image or a plot. Images are direct bit map representations of the digitizing oscilloscope display . Plots are vector (plotted) representatio[...]

  • Página 131

    Hardcopy Reference Before you make a hardcopy , you need to set up communications and hard- copy parameters. This discussion assumes that the hardcopy device is already connected to the GPIB port on the rear panel. If that is not the case see Connection Strategies on page 3-63. Setting Communication Parameters T o set up the communication parameter[...]

  • Página 132

    Hardcopy TDS 520A, 524A, 540A, & 544A User Manual Setting Hardcopy Parameters T o specify the hardcopy format, layout, and type of port using the hardcopy menu: 1. Press SHIFT HARDCOPY MENU to bring up the Hardcopy menu. 2. Press Format (main) ➞ Thinkjet , Deskjet , Laserjet , Epson , DPU-41 1, DPU-412 , PCX , PCX Color (TDS 544A), TIFF , BMP[...]

  • Página 133

    Hardcopy Reference T o stop and discard the hardcopy being sent, press HARDCOPY again while the hardcopy in process message is still on screen. T o add additional hardcopies to the printer spool, press HARDCOPY again after the hardcopy in process message is removed from the screen. Y ou can add hardcopies to the spool until it is full. When the spo[...]

  • Página 134

    Hardcopy TDS 520A, 524A, 540A, & 544A User Manual Date and Time Display Figure 3-33: Date and Time Display 5. Use the general purpose knob or the keypad to set the parameter you have chosen to the value desired. (The format when using the keypad is day .month. For example, use 23.6 for the 23 rd of June.) 6. Repeat steps 4 and 5 to set other pa[...]

  • Página 135

    Hardcopy Reference Strategies for actually printing a copy include: Send output straight to a printer/plotter . Send the data to a computer to print from there and/or to import into your favorite desktop publishing or other application package. Send your data to a floppy disk file (optional on the TDS 520A & TDS 540A) for later printing from a [...]

  • Página 136

    Hardcopy TDS 520A, 524A, 540A, & 544A User Manual GPIB Cable Digitizing Oscilloscope PC Compatible Centronics or RS-232 Cable Hardcopy Device Figure 3-35: Connecting the Digitizing Oscilloscope and Hardcopy Device Via a PC If your controller is PC-compatible and it uses the T ektronix GURU or S3FG210 (National Instruments GPIB-PCII/IIA) GPIB pa[...]

  • Página 137

    Hardcopy Reference 5. T ype: IBRDF <Filename> where <Filename> is a valid DOS file name you want to call your hardcopy information. It should be 8 characters long with up to a 3 character extension. For example, you could type “ibrdf screen1” . 6. Exit the IBIC program by typing: EXIT 7. T ype: COPY <Filename> <Output port&[...]

  • Página 138

    TDS 620A, 640A, & 644A User Manual Help The on-line help system provides brief information about each of the digitizing oscilloscope controls. T o use the on-line help system: Press HELP to provide on-screen information on any front panel button, knob or menu item (see Figure 3-36). When you press that button, the instrument changes mode to sup[...]

  • Página 139

    Reference Horizontal Control Y ou can control the horizontal part of the display (the time base) using the horizontal menu and knobs. By changing the horizontal scale, you can focus on a particular portion of a waveform. By adjusting the horizontal position, you can move the waveform right or left to see different portions of the waveform. That is [...]

  • Página 140

    Horizontal Control TDS 620A, 640A, & 644A User Manual At the top of the display , the Record View shows the size and location of the waveform record and the location of the trigger relative to the display (see Figure 3-38). The Time Base readout at the lower right of the display shows the time/division settings and the time base (main or delaye[...]

  • Página 141

    Horizontal Control Reference Y ou also can select Delayed Runs After Main or Delayed T riggerable . For more information on how to use these two menu items, see Delayed T rigger- ing on page 3-22. T rigger Position T o define how much of the record will be pretrigger and how much posttrigger information using the T rigger Position menu item: Press [...]

  • Página 142

    Horizontal Control TDS 620A, 640A, & 644A User Manual Y ou can also control whether changing the horizontal position setting affects all displayed waveforms, just the live waveforms, or only the selected wave- form. The Horizontal Lock setting in the Zoom menu determines which wave- forms the horizontal position knob adjusts whether zoom is on [...]

  • Página 143

    Horizontal Control Reference Figure 3-39: Horizontal Menu — FastFrame Setup FastFrame Interactions — Envelope, Average, and HiRes form the envelope or average following the last frame of the concatenated record. For example, if average or HiRes acquisition modes are selected and the frame count is 10, segments 1 through 10 will show sample or H[...]

  • Página 144

    TDS 620A, 640A, & 644A User Manual Limit T esting Limit testing provides a way to automatically compare each incoming or math waveform against a template waveform. Y ou set an envelope of limits around a waveform and let the digitizing oscilloscope find waveforms that fall outside those limits (see Figure 3-40). When it finds such a waveform, t[...]

  • Página 145

    Limit T esting Reference NOTE The template will be smoother if you acquire the template waveform using A verage acquisition mode. If you are unsure how to do this, see Acquisition Modes on page 3-7. Once you have selected a source, select a destination for the template. 2. Press T emplate Destination (side) ➞ Ref1 , Ref2 , Ref3 , or Ref4 . Figure[...]

  • Página 146

    Limit T esting TDS 620A, 640A, & 644A User Manual If you wish to create another limit test template, store it in another des- tination to avoid overwriting the template you have just created. If you wish to view the template you have created, press the MORE button. Then press the button corresponding to the destination reference memory you have[...]

  • Página 147

    Limit T esting Reference If you want to send a hardcopy command when waveform data exceeds the limits set, toggle Hardcopy if Condition Met (side) to ON . Y ou can set the hardcopy system to send the hardcopy to the file system (optional on the TDS 620A & TDS 640A). (Do not forget to set up the hardcopy system. See Hardcopy on page 3-59 for det[...]

  • Página 148

    Limit T esting TDS 620A, 640A, & 644A User Manual Multiple W aveform Comparisons When comparing one or more waveforms, each against a common template or against its own template, consider the following operating characteristics: Y ou should set Horizontal Lock to None in the Zoom side menu (push ZOOM and press (repeatedly) Horizontal Lock to No[...]

  • Página 149

    Reference Logic T riggering There are two classes of logic triggering: pattern and state. A pattern trigger occurs when the logic inputs to the logic function you select cause the function to become TRUE (or at your option F ALSE). When you use a pattern trigger , you define: The precondition for each logic input — logic high, low , or do not car[...]

  • Página 150

    Logic T riggering TDS 620A, 640A, & 644A User Manual At the bottom of the display , the T rigger readout shows some of the key parameters of the logic trigger (see Figure 3-42). Trigger Class = State Ch 1, 2, 3 Inputs = High, Don’t Care, High Ch 4 Input = Rising Edge Logic = OR Figure 3-42: Logic T rigger Readouts NOTE When Logic is the selec[...]

  • Página 151

    Logic T riggering Reference T able 3-4: Logic T riggers Pattern State Definition 1,2 AND Clocked AND If all the preconditions selected for the logic inputs 3 are true, then the oscilloscope triggers. NAND Clocked NAND If not all of the preconditions se- lected for the logic inputs 3 are true, then the oscilloscope trig- gers. OR Clocked OR If any o[...]

  • Página 152

    Logic T riggering TDS 620A, 640A, & 644A User Manual Figure 3-43: Logic T rigger Menu T rigger When This menu item lets you determine if the oscilloscope will trigger when the logic condition is met ( Goes TRUE ) or when the logic condition is not met ( Goes F ALSE ). (The T rue when less than and T rue when greater than menu items are only use[...]

  • Página 153

    Logic T riggering Reference Mode & Holdoff Y ou can change the holdoff time and select the trigger mode using this menu item. 1. Press TRIGGER MENU ➞ T ype (main) ➞ Logic (pop-up) ➞ Class (main) ➞ Pattern or State (pop-up) ➞ Mode & Holdoff (main) ➞ Auto or Normal (side). In Auto mode the oscilloscope acquires a waveform after a [...]

  • Página 154

    Logic T riggering TDS 620A, 640A, & 644A User Manual Define a T ime Qualified Pattern T rigger Y ou can also time qualify a pattern logic trigger . That is, you specify a time that the boolean logic function (AND, NAND, OR, or NOR) must be TRUE (logic high). Y ou also choose the type of time qualification (greater or less than the time limit sp[...]

  • Página 155

    Logic T riggering Reference Logic Function (AND) Becomes TRUE Logic Function Becomes F ALSE and Triggers Acquisition Time Logic Function is TRUE Time Logic Function Must be TRUE Figure 3-44: Logic T rigger Menu — Time Qualified TRUE When you select State logic triggering, the oscilloscope uses channel 4 (Aux 2 on the TDS 620A & TDS 524A) as a[...]

  • Página 156

    Logic T riggering TDS 620A, 640A, & 644A User Manual Define Logic T o choose the type of logic function you want applied to the input channels: Press TRIGGER MENU ➞ T ype (main) ➞ Logic (pop-up) ➞ Class (main) ➞ State (pop-up) ➞ Define Logic (main) ➞ AND , OR , NAND , or NOR (side). See T riggering , on page 2-13. See T riggering , [...]

  • Página 157

    Reference Measurement System There are various ways to measure properties of waveforms. Y ou can use graticule, cursor , or automatic measurements. This section describes auto- matic measurements ; cursors and graticules are described elsewhere. (See Cursor Measurements on page 3-17 and Measurements on page 2-30.) Automatic measurements are general[...]

  • Página 158

    Measurement System TDS 620A, 640A, & 644A User Manual T able 3-5: Measurement Definitions (Cont.) Name Definition Delay T iming measurement. The time between the MidRef crossings of two different traces or the gated region of the traces. Fall Time T iming measurement. T ime taken for the falling edge of the first pulse in the waveform or gated [...]

  • Página 159

    Measurement System Reference T able 3-5: Measurement Definitions (Cont.) Name Definition Period T iming measurement. T ime it takes for the first complete signal cycle to happen in the waveform or gated region. The reciprocal of frequency . Measured in seconds. Positive Duty Cycle T iming measurement of the first cycle in the waveform or gated regi[...]

  • Página 160

    Measurement System TDS 620A, 640A, & 644A User Manual The readout area for measurements is on the right side of the waveform window . Y ou can display and continuously update as many as four measure- ments at any one time. When menus are displayed, the readouts appear in the graticule area. If the menu area is empty , then the readouts are disp[...]

  • Página 161

    Measurement System Reference T o vary the source for measurements, simply select the other channel and then choose the measurements you want. Be careful when taking automatic measurements on noisy signals. Y ou might measure the frequency of the noise and not the desired waveform. Y our digitizing oscilloscope helps identify such situations by disp[...]

  • Página 162

    Measurement System TDS 620A, 640A, & 644A User Manual Gated Measurements The gating feature lets you limit measurements to a specified portion of the waveform. When gating is Off , the oscilloscope makes measurements over the entire waveform record. When gating is activated, vertical cursors are displayed. Use these cursors to define the sectio[...]

  • Página 163

    Measurement System Reference NOTE Cursors are displayed relative to the selected waveform. If you are making a measurement using two waveforms, this can be a source of confusion. If you turn off horizontal locking and adjust the hori- zontal position of one waveform independent of the other , the cursors appear at the requested position with respec[...]

  • Página 164

    Measurement System TDS 620A, 640A, & 644A User Manual 2. Press High Ref , Mid Ref , Low Ref , or Mid2 Ref (side). High Ref — Sets the high reference level. The default is 90%. Mid Ref — Sets the middle reference level. The default is 50%. Low Ref — Sets the low reference level. The default is 10%. Mid2 Ref — Sets the middle reference le[...]

  • Página 165

    Measurement System Reference 2. Press Measure Delay to (side) repeatedly to choose the delay to wave- form. The choices are Ch1 , Ch2 , Ch3, Ch4 (on the TDS 644A & TDS 640A); Ch1, Ch2, Ax1 , Ax2 (on the TDS 524A & TDS 620A); and Math1 , Math2 , Math3 , Ref1 , Ref2 , Ref3 , and Ref4. Figure 3-49: Measure Delay Menu — Delay T o Delay Edges [...]

  • Página 166

    Measurement System TDS 620A, 640A, & 644A User Manual Creating the Delay Measurement — Once you have specified the waveforms you are measuring between and which edges to use, you need to notify the digitizing oscilloscope to proceed with the measurement. Press Delay T o (main) ➞ OK Create Measurement (side). T o exit the Measure Delay menu [...]

  • Página 167

    Measurement System Reference T o use snapshot, obtain a stable display of the waveform to be measured. Pressing AUTOSET may help. 1. Press MEASURE ➞ SNAPSHOT (main). 2. Press either SNAPSHOT (main) or AGAIN (side) to take another snap- shot. NOTE The snapshot display tells you the channel that the snapshot is being made on. 3. Push Remove Measrmn[...]

  • Página 168

    TDS 520A, 524A, 540A, & 544A User Manual Probe Accessories The probe you use and how you connect it to a signal source affect the oscilloscope acquisition of the waveform record. T wo important factors are ground lead inductance (introduced by the probe) and the physical layout of your circuit and component devices. For an amplitude measurement[...]

  • Página 169

    Probe Accessories Reference 4 Inch Ground Lead Low-Inductance Ground Lead Figure 3-52: Signal V ariation Introduced by Probe Ground Lead (1 ns/division) The following descriptions explain how to use many of the accessories that came with your probe. Figure 3-53 shows both standard and optional probe accessories and how they attach to your probe. Th[...]

  • Página 170

    Probe Accessories TDS 520A, 524A, 540A, & 544A User Manual Marker Ring (Standard) Marker Ring (Standard) IC Protector Tip (Optional) Dual Lead Adapter (Optional) Probe Tip-to-Circuit Board Adapter (Standard) Probe Tip-to-Chassis Adapter (Optional) Retractable Hook Tip (Standard) KlipChip TM (Standard) Slip-on Ground Lead (Standard) Alligator Cl[...]

  • Página 171

    Probe Accessories Reference Marker Rings The marker rings help you keep track of individual probes and signal sources when you have a complicated test setup. Use the marker rings whenever you want to identify a particular probe. Long Ground Leads Use long ground leads when a long reach is important and high-frequency information is not. Long ground[...]

  • Página 172

    Probe Accessories TDS 520A, 524A, 540A, & 544A User Manual Probe-Tip-to-Circuit Board Adapters The probe-tip-to-circuit board adapters let you design minimum inductance test points into your next circuit board. That adapter provides maximum performance for the probe, because it virtually eliminates ground inductance effects. Instructions for in[...]

  • Página 173

    Probe Accessories Reference Probe-Tip-to-Chassis Adapter The probe-tip-to-chassis adapter makes your test point accessible without removing instrument covers or panels. It provides an easy-access, low-induc- tance test point anywhere on your circuit. The probe-tip-to-chassis adapter has the same low inductance properties as the probe-tip-to-circuit[...]

  • Página 174

    Probe Accessories TDS 520A, 524A, 540A, & 544A User Manual Dual-Lead Adapter The dual-lead adapter makes an easy connection to 0.025 diameter connec- tor pins (see Figure 3-54). One lead connects to a ground reference pin, and the other to the signal pin. The adapter prevents burring and pin damage that can result when a retractable hook tip is[...]

  • Página 175

    Reference Probe Cal This oscilloscope lets you compensate the entire signal path, from probe tip to digitized signal, to improve the gain and offset accuracy of the probe. By executing Probe Cal on a channel with its probe installed, you can optimize the oscilloscope capability to make accurate measurements using that chan- nel and probe. Run a Pro[...]

  • Página 176

    Probe Cal TDS 620A, 640A, & 644A User Manual 1. Install the probe on the input channel on which it is to be used. 2. Power on the digitizing oscilloscope and allow a 20 minute warm-up before doing this procedure. 3. Press SHIFT UTILITY ➞ System (main) ➞ Cal (pop-up). 4. Look at the status label under Signal Path in the main menu. If the sta[...]

  • Página 177

    Probe Cal Reference Figure 3-55: Probe Cal Menu and Gain Compensation Display 1 1. If the Probe Offset Compensation message is displayed, continue with step 15; otherwise, continue with step 12. 12. If the Compensation Error message is displayed, continue with step 13; otherwise continue with step 18. 13. Press SHIFT UTILITY ➞ System (main) ➞ D[...]

  • Página 178

    Probe Cal TDS 620A, 640A, & 644A User Manual 19. If desired, repeat this procedure beginning at step 1 to compensate for other probe/channel combinations. But before you do so, be sure you take note of the following requirements: Remember to first low frequency compensate any passive probe you connect (see Prerequisites at the beginning of this[...]

  • Página 179

    Probe Cal Reference Figure 3-56: Re-use Probe Calibration Data Menu If the Re-use Probe Calibration data? menu is displayed, you can choose one of the following options: Press OK Use Existing Data (side) to use the Probe Cal data last stored to compensate the probe. Press OK Erase Probe Cal Data (side) to erase the Probe Cal data last stored and us[...]

  • Página 180

    Probe Cal TDS 620A, 640A, & 644A User Manual T able 3-6: Probe Cal Status Probe User T ype Probe Connected 2 Cal’d? 1 Action Simple Interface 3 Complex Interface 4 No Doesn’t Matter Initialized Initialized Ye s Power off Initialized (probe data is retained) Initialized (probe data is retained) Ye s Power on Can not detect different probe: D[...]

  • Página 181

    Reference Probe Compensation Passive probes require compensation to ensure maximum distortion-free input to the digitizing oscilloscope and to avoid high frequency amplitude errors (see Figure 3-57). Probe Compensated Correctly Probe Overcompensated Probe Undercompensated Figure 3-57: How Probe Compensation Affects Signals T o compensate your probe[...]

  • Página 182

    Probe Compensation TDS 620A, 640A, & 644A User Manual 4. If you need to change the input impedance, press Coupling (main). Then toggle the side menu selection to get the correct impedance. 5. Press SHIFT ACQUIRE MENU ➞ Mode (main) ➞ Hi Res (side). 6. Adjust the probe until you see a perfectly flat top square wave on the display . Figure 3-5[...]

  • Página 183

    Reference Probe Selection The probes included with your digitizing oscilloscope are useful for a wide variety of tasks. However , for special measurement situations you sometimes need different probes. This section helps you select the right probe for the job. Once you have decided the type of probe you need, use T able 3-7 (page 3-1 17) to determi[...]

  • Página 184

    Probe Selection TDS 620A, 640A, & 644A User Manual Low Impedance (Z O ) Probes Low impedance probes measure frequency more accurately than general purpose probes, but they make less accurate amplitude measurements. They offer a higher bandwidth to cost ratio. These probes must be terminated in a 50 scope input. Input capacitance is much lower t[...]

  • Página 185

    Probe Selection Reference Active voltage probes, sometimes called “FET” probes, use active circuit elements such as transistors. There are three classes of active probes: High speed active Differential active Fixtured active Active voltage measuring probes use active circuit elements in the probe design to process signals from the circuit under[...]

  • Página 186

    Probe Selection TDS 620A, 640A, & 644A User Manual amplifiers) to precisely connect your instrument to your device-under-test. These probes have the same electrical characteristics as high speed, active probes but use a smaller mechanical design. Current probes enable you to directly observe and measure current wave- forms, which can be very di[...]

  • Página 187

    Probe Selection Reference Optical probes let you blend the functions of an optical power meter with the high-speed analog waveform analysis capability of an oscilloscope. Y ou have the capability of acquiring, displaying, and analyzing optical and electrical signals simultaneously . Applications include measuring the transient optical properties of[...]

  • Página 188

    Probe Selection TDS 620A, 640A, & 644A User Manual T able 3-7 lists TDS 600A compatible probes classified by type. T able 3-7: TDS 600A Compatible Probes Probe T ype T ektronix Model Description Passive, high impedance voltage P6139A (std.) P6101A 10X, 500 MHz 1X, 15 MHz Passive, SMD P6563AS 20X, 500 MHz Passive, low impedance Z O P6156 10X, 3.[...]

  • Página 189

    Probe Selection Reference Another way to classify probes is by application. Different applications de- mand different probes. Use T able 3-8 to select a probe for your application. T able 3-8: Probes by Application Probe T ype T elecommuni- cations & High-Speed Logic Industrial Electronics Consumer/ Computer Electronics High Energy Pulsed Power[...]

  • Página 190

    TDS 620A, 640A, & 644A User Manual Pulse T riggering Pulse triggering can be very useful. For example, you might be testing a product with a glitch in the power supply . The glitch appears once a day . So instead of sitting by and waiting for it to appear , you can use pulse triggering to automatically capture your data. There are three classes[...]

  • Página 191

    Pulse T riggering Reference T able 3-9: Pulse T rigger Definitions Name Definition Glitch positive T riggering occurs if the oscilloscope detects positive spike widths less than the specified glitch time. Glitch negative T riggering occurs if the oscilloscope detects negative spike widths less than the specified glitch time. Glitch either T riggeri[...]

  • Página 192

    Pulse T riggering TDS 620A, 640A, & 644A User Manual The pulse trigger menus let you define the pulse source, select the mode (auto or normal), and adjust the holdoff. T o bring up the Pulse T rigger menu: Press TRIGGER MENU ➞ T ype (main) ➞ Pulse (pop-up) ➞ Class (main) ➞ Glitch , Runt , or Width (pop-up) (see Figure 3-62). Figure 3-62[...]

  • Página 193

    Pulse T riggering Reference In Normal mode the oscilloscope acquires a waveform only if there is a valid trigger . (Y ou can force a single acquisition by pressing FORCE TRIGGER .) 2. T o change the holdoff time, press Holdoff (side). Use the general pur- pose knob or the keypad to enter the value in percent. When you select the pulse class Glitch [...]

  • Página 194

    Pulse T riggering TDS 620A, 640A, & 644A User Manual If you select Set to 50% , you cause the digitizing oscilloscope to search for the point halfway between the peaks of the trigger source signal and set the trigger level to that point. When you select the pulse class Runt , the oscilloscope will trigger on a short pulse that crosses one thres[...]

  • Página 195

    Pulse T riggering Reference Selected Trigger Bar at Upper Threshold Runt Pulse Crosses First Threshold Only , Recrosses First Threshold Level, and Triggers Acquisition Unselected T rigger Bar at Lower Threshold Figure 3-63: Main T rigger Menu—Runt Class Thresholds T o set the two threshold levels used in detecting a runt pulse: 1. Press TRIGGER M[...]

  • Página 196

    Pulse T riggering TDS 620A, 640A, & 644A User Manual Regardless of the polarity setting, triggering occurs at the point the runt pulse recrosses its first threshold. When you select the pulse class Width , the oscilloscope will trigger on a pulse narrower (or wider) than some specified range of time (defined by the upper limit and lower limit).[...]

  • Página 197

    Reference Remote Communication Y ou may want to integrate your oscilloscope into a system environment and remotely control your oscilloscope or exchange measurement or waveform data with a computer . Y ou can control your oscilloscope remotely via the IEEE Std 488.2–1987 (GPIB) interface. GPIB enables data transfers between instruments that suppo[...]

  • Página 198

    Remote Communication TDS 620A, 640A, & 644A User Manual GPIB Device GPIB Device GPIB Device GPIB Device GPIB Device GPIB Device GPIB Device Figure 3-64: T ypical GPIB Network Configuration Cables — An IEEE Std 488.1–1987 GPIB cable (available from T ektronix, part number 012–0991–00) is required to connect two GPIB devices. Connector ?[...]

  • Página 199

    Remote Communication Reference T o set up remote communications, ensure that your oscilloscope is physically cabled to the controller and that the oscilloscope parameters are correctly set. Plug an IEEE Std 488.2–1987 GPIB cable into the GPIB connector on the oscilloscope rear panel and into the GPIB port on your controller (see Fig- ure 3-66). R[...]

  • Página 200

    Remote Communication TDS 620A, 640A, & 644A User Manual GPIB Configuration Menu Figure 3-67: Utility Menu See Hardcopy , on page 3-59. See the TDS Family Digitizing Oscilloscopes Programmer Manual . For More Information[...]

  • Página 201

    Reference Saving and Recalling Setups Y ou may want to save and reuse setups for many reasons. For example, after changing the setting during the course of an experiment, you may want to quickly return to your original setup. Y ou can save and recall up to ten instru- ment setups from internal oscilloscope memory . The information is retained even [...]

  • Página 202

    Saving and Recalling Settings TDS 620A, 640A, & 644A User Manual T o store a setup to disk (optional on the TDS 620A & TDS 640A), press To File . Then use the general purpose knob to select the exact file from the resulting scrollbar list. Finally , press the side-menu Save T o Selected File to complete the operation. Recalling a Setup T o [...]

  • Página 203

    Saving and Recalling Setups Reference If the checksum calculation is unsuccessful, displays a warning message; if the checksum calculation is successful, displays a confirmation mes- sage. Running File Utilities T o run file utilities (optional on the TDS 620A & TDS 640A), see the File System article on page 3-55. See T utorial Example 4: Savin[...]

  • Página 204

    TDS 620A, 640A, & 644A User Manual Saving and Recalling W aveforms Y ou can store a waveform in any of the four internal reference memories of the digitizing oscilloscope. That information is retained even when you turn the oscilloscope off or unplug it. Y ou can save any combination of different size waveform records. The digitizing oscillosco[...]

  • Página 205

    Saving and Recalling W aveforms Reference Figure 3-69: Save W aveform Menu Deleting W aveforms Y ou can choose the Delete Refs main menu item and then select the refer- ences you no longer need from the side menu ( Delete Ref1 , Delete Ref2 , Delete Ref3 , Delete Ref4 , or Delete All Refs ). Deleting All W aveforms and Setups Y ou can remove all st[...]

  • Página 206

    Saving and Recalling W aveforms TDS 620A, 640A, & 644A User Manual resulting scrollbar list. Only files with .WFM extensions are displayed. Finally , press from the side-menu T o Ref1 , T o Ref2 , T o Ref3, or T o Ref4 choices to complete the operation. Figure 3-70: More Menu Autosave (TDS 640A and 644A only) T o use autosave, press Autosave (m[...]

  • Página 207

    Reference Selecting Channels The selected channel is the channel that the digitizing oscilloscope applies all waveform-specific activities to (such as measurements or vertical scale and position). The channel readout shows the selected channel in inverse video in the lower left corner of the display . The channel reference indicator for the selecte[...]

  • Página 208

    Selecting Channels TDS 620A, 640A, & 644A User Manual T o selecting a channel: Pressing CH 1, CH 2, CH 3 ( AUX 1 on the TDS 620A & TDS 524A), or CH 4 ( AUX 2 on the TDS 620A & TDS 524A) turns the channel on if it is not al- ready on. Y ou do not use the channel selection buttons when triggering. Instead you select the trigger source in [...]

  • Página 209

    Reference Signal Path Compensation This oscilloscope lets you compensate the internal signal path used to acquire the waveforms you acquire and measure. By executing the signal path com- pensation feature (SPC), you can optimize the oscilloscope capability to make accurate measurements based on the ambient temperature. Run an SPC anytime you wish t[...]

  • Página 210

    Signal Path Compensation TDS 620A, 640A, & 644A User Manual Figure 3-73: Performing a Signal Path Compensation[...]

  • Página 211

    Reference Status The Status menu lets you see information about the oscilloscope state. T o operate the Status menu: Press SHIFT ST A TUS ➞ Status (main) ➞ System , Display , T rigger , W aveforms , or I/O (side). Note: some oscilloscopes do not have a main Status menu. On these instruments, press SHIFT ST A TUS ➞ System , Display , T rigger [...]

  • Página 212

    Status TDS 620A, 640A, & 644A User Manual T o display the banner (firmware version, options, copyright, and patents): NOTE Some TDS 644A oscilloscopes do not have a Status main menu with a banner . However , all instruments display the banner briefly at power-on. Press SHIFT ST A TUS ➞ Banner (main) (see Figure 3-75). Figure 3-75: Banner Disp[...]

  • Página 213

    Reference T riggering T riggers determine when the digitizing oscilloscope starts acquiring and displaying a waveform. The TDS 600A has four types of trigger: edge, logic, pulse, and, with option 5, video. Although these triggers are unique, they have some common characteristics that can be defined and modified using the T rigger menu, buttons, and[...]

  • Página 214

    T riggering TDS 620A, 640A, & 644A User Manual T o Set to 50% Y ou can quickly obtain an edge or pulse trigger (except for the Runt class) by pressing SET LEVEL TO 50%. The oscilloscope sets the trigger level to the halfway point between the peaks of the trigger signal. Y ou can also set the level to 50% in the Trigger menu under the main menu [...]

  • Página 215

    T riggering Reference When READY is lighted, it means the digitizing oscilloscope can accept a valid trigger event, and the digitizing oscilloscope is waiting for that event to occur . When ARM is lighted, it means the trigger circuitry is filling the pretrigger portion of the waveform record. When both TRIG’D and READY are lighted, it means the [...]

  • Página 216

    T riggering TDS 620A, 640A, & 644A User Manual The trigger point indicator shows position. It can be positioned horizontally off screen, especially with long record length settings. The trigger level bar shows only the trigger level. It remains on screen, regardless of the horizontal position, as long as the channel providing the trigger source[...]

  • Página 217

    T riggering Reference See the Option 05 Video T rigger Interface Instruction Manual , T ektronix part number 070–8748–00.[...]

  • Página 218

    TDS 620A, 640A, & 644A User Manual V ertical Control Y ou can control the vertical position and scale of the selected waveform using the vertical menu and knobs. By changing the vertical scale, you can focus on a particular portion of a waveform. By adjusting the vertical position, you can move the waveform up or down on the display . That is p[...]

  • Página 219

    V ertical Control Reference V ertical Readout Figure 3-79: V ertical Readouts and Channel Menu Coupling T o choose the type of coupling for attaching the input signal to the vertical attenuator for the selected channel and to set its input impedance: Press VERTICAL MENU ➞ Coupling (main) ➞ DC , AC , GND , or (side). DC coupling shows both the A[...]

  • Página 220

    V ertical Control TDS 620A, 640A, & 644A User Manual Bandwidth T o eliminate the higher frequency components, change the bandwidth of the selected channel: Press VERTICAL MENU ➞ Bandwidth (main) ➞ Full , 100 MHz , or 20 MHz (side). Fine Scale Press VERTICAL MENU ➞ Fine Scale (main) to make fine adjustments to the vertical scale using the [...]

  • Página 221

    Reference W aveform Differentiation Advanced DSP Math (optional on TDS 620A & TDS 640A), provides wave- form differentiation that allows you to display a derivative math waveform that indicates the instantaneous rate of change of the waveform acquired. Such waveforms are used in the measurement of slew rate of amplifiers and in educational appl[...]

  • Página 222

    W aveform Differentation TDS 620A, 640A, & 644A User Manual Derivative Math Waveform Source Waveform Figure 3-80: Derivative Math W aveform 5. Press Set Function to (side). Repeatedly press the same button (or use the general purpose knob) until diff appears in the menu label. 6. Press OK Create Math Wfm (side) to display the derivative of the [...]

  • Página 223

    W aveform Differentation Reference Figure 3-81: Peak-Peak Amplitude Measurement of a Derivative W aveform Cursor Measurement of a Derivative W aveform Y ou can also use cursors to measure derivative waveforms. Use the same procedure as is found under Waveform Integration on page 3-155. When using that procedure, note that the amplitude measurements[...]

  • Página 224

    W aveform Differentation TDS 620A, 640A, & 644A User Manual When using the vertical scale knob to scale the source waveform, note that it also scales your derivative waveform. Because of the method the oscilloscope uses to scale the source waveform before differentiating that waveform, the derivative math waveform may be too large vertically to[...]

  • Página 225

    Reference W aveform Integration Advanced DSP Math (optional on TDS 620A & TDS 640A), provides wave- form integration that allows you to display an integral math waveform that is an integrated version of the acquired waveform. Such waveforms find use in the following applications: Measuring of power and energy , such as in switching power suppli[...]

  • Página 226

    W aveform Integration TDS 620A, 640A, & 644A User Manual 5. Press Set Function to (side). Repeatedly press the same button until intg appears in the menu label. 6. Press OK Create Math W aveform (side) to turn on the integral math waveform. Y ou should now have your integral math waveform on screen. See Figure 3-82. Use the V ertical SCALE and [...]

  • Página 227

    W aveform Integration Reference 6. Read the integrated voltage over time between the cursors in volt- seconds from the : readout. Read the integrated voltage over time between the selected cursor and the reference indicator of the math waveform from the @: readout. See Figure 3-83. Source Waveform Integral Math Waveform Figure 3-83: H Bars Cursors [...]

  • Página 228

    W aveform Integration TDS 620A, 640A, & 644A User Manual Read the integrated voltage over time between the X of the selected cursor and the reference indicator of the math waveform from the @: readout. Read the time difference between the long vertical bars of the paired cursors from the : readout. Automated Measurements of a Integral W aveform[...]

  • Página 229

    W aveform Integration Reference Zoom Once you have your waveform optimally displayed, you may magnify (or reduce) it vertically and horizontally to inspect any feature you desire. Just be sure the integrated waveform is the selected waveform. (Press MORE , then select the integrated waveform in the More main menu. Then use the V ertical and Horizon[...]

  • Página 230

    TDS 620A, 640A, & 644A User Manual W aveform Math Y ou can mathematically manipulate your waveforms. For example, you might have a waveform clouded by background noise. Y ou can obtain a cleaner waveform by subtracting the background noise from your original waveform. This section describes the invert, add, subtract, divide, and multiply wavefo[...]

  • Página 231

    W aveform Math Reference NOTE If your digitizing oscilloscope is equipped with Advanced DSP Math (optional on TDS 620A & TDS 640A), the menu item FFT will be at the same brightness as the menu items Single Wfm Math and Dual Wfm Math ; otherwise, FFT will be dimmed. See pages 3-38, 3-150, and 3-154 for information on FFT s and other advanced mat[...]

  • Página 232

    W aveform Math TDS 620A, 640A, & 644A User Manual Figure 3-85: Dual W aveform Math Main and Side Menus Dual Wfm Math 1. Select the sources with MORE ➞ Math1 , Math2 , or Math3 (main) ➞ Change Math waveform definition (side) ➞ Dual Wfm Math (main) ➞ Set 1st Source to and Set 2nd Source to (side). Enter the sources by repeatedly pressing [...]

  • Página 233

    Reference Zoom At times, you may want to expand or compress a waveform on the display without changing the acquisition parameters. Y ou can do that with the zoom feature. When you zoom in on a waveform on the display , you expand a portion of the waveform. The digitizing oscilloscope may need to show more points for that portion than it has acquire[...]

  • Página 234

    Zoom TDS 620A, 640A, & 644A User Manual NOTE Although zoom must be turned on to control which waveforms zoom affects, the setting for Horizontal Lock affects which waveforms the horizontal control positions whether zoom is on or off. The rules for the three settings are listed in step 2. Only the selected waveform (the top one) changes size. Fi[...]

  • Página 235

    Zoom Reference Reset Zoom T o reset all zoom factors to their defaults (see T able 3-10), press ZOOM ➞ Reset Zoom Factors (side). T able 3-10: Zoom Defaults Parameter Setting Zoom V ertical Position 0 Zoom V ertical Gain 1X Zoom Horizontal Position T racking Horizontal Position Zoom Horizontal Gain 1X Press ZOOM ➞ Off (side) to return to normal[...]

  • Página 236

    Appendices[...]

  • Página 237

    TDS 520A, 524A, 540A, & 544A User Manual Appendix A: Options and Accessories This section describes the various options as well as the standard and option- al accessories that are available for the TDS 500A Digitizing Oscilloscopes. The following options are available: Options A1–A5: International Power Cords Besides the standard North Americ[...]

  • Página 238

    Appendix A: Options and Accessories Appendices W arranty-Plus Service Options The following options add to the services available with the standard warranty . (The standard warranty appears following the title page in this manual.) Option M2: When Option M2 is ordered, T ektronix provides five years of warranty/remedial service. Option M3: When Opt[...]

  • Página 239

    Appendix A: Options and Accessories TDS 520A, 524A, 540A, & 544A User Manual Option 05: Video T rigger With this option, T ektronix ships the instrument with tools for investigating events that occur when a video signal generates a horizontal or vertical sync pulse. It allows you to investigate a range of NTSC, P AL, SECAM, and high definition [...]

  • Página 240

    Appendix A: Options and Accessories Appendices constant, or a ratio calibration technique. The calibration is in compliance with US MIL–STD–45662A. This option also includes a test data report for the instrument. The following standard accessories are included with the digitizing oscillo- scope: T able A-2: Standard Accessories Accessory Part N[...]

  • Página 241

    Appendix A: Options and Accessories TDS 520A, 524A, 540A, & 544A User Manual T able A-3: Probe Accessories (Cont.) Accessory Part Number Six-inch Alligator Clip Ground Lead 196–3305–00 Screwdriver: adjustment tool, metal tip 003–1433–00 SMT KlipChip TM 206–0364–00 Accessory Pouch 016–0708–00 Y ou can also order the following opt[...]

  • Página 242

    Appendix A: Options and Accessories Appendices P6204 Active, high speed digital voltage probe. FET . DC to 1 GHz. DC offset. 50 input. Use with 1 103 T ekProbe Power Supply for offset control. P6563AS Passive, SMD probe, 20X, 500 MHz P6046 Active, differential probe, 1X/10X, DC to 100 MHz, 50 input. A6501 Buffer Amplifier (active fixtured), 1 GHz, [...]

  • Página 243

    Appendix A: Options and Accessories TDS 520A, 524A, 540A, & 544A User Manual T able A-5: Probe Accessories Accessory Part Number 013–0226–00 Connector , BNC: 50 , BNC to Probe T ip Adapter 013–0227–00 Connector , Probe: Package of 100, compact 131–4244–00 Connector , Probe: Package of 25, compact 131–5031–00 Screwdriver Adjustme[...]

  • Página 244

    Appendix A: Options and Accessories Appendices Accessory Software The following optional accessories are T ektronix software products recom- mended for use with your digitizing oscilloscope: T able A-6: Accessory Software Software Part Number EZ-T est Program Generator S45F030 W avewriter: AWG and waveform creation S3FT400 T ekTMS: T est management[...]

  • Página 245

    TDS 620A, 640A, & 644A User Manual Appendix B: Algorithms TDS 600A Digitizing Oscilloscopes can take 25 automatic measurements. By knowing how they make these calculations, you may better understand how to use your TDS 600A and how to interpret your results. TDS 600A Digitizing Oscilloscopes use a variety of variables in their calcula- tions. T[...]

  • Página 246

    Appendix B: Algorithms Appendices 2. It splits the histogram into two sections at the halfway point between and (also called ). 3. The level with the most points in the upper histogram is the value, and the level with the most points in the lower histogram is the value. (Choose the levels where the histograms peak for and ) If gives the largest pea[...]

  • Página 247

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual End — is the location of the end of the measurement zone (X-value). It is ( – 1.0) samples unless you are making a gated measurement. When you use gated measurements, it is the location of the right vertical cursor . Hysteresis — The hysteresis band is 10% of the waveform amplitude[...]

  • Página 248

    Appendix B: Algorithms Appendices MidRef + (Hysteresis x Amplitude) MidRef – (Hysteresis x Amplitude) MidRef MCross1 (StartCycle) MCross2 MCross3 (EndCycle) Figure A-1: MCross Calculations W aveform[<0.0 ... RecordLength–1.0>] — holds the acquired data. TPOS — is the location of the sample just before the trigger point (the time refer[...]

  • Página 249

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual For details of the integration algorithm, see page A-19. Cycle Area Amplitude (voltage) measurement. The area over one waveform cycle. For non-cyclical data, you might prefer to use the Area measurement. If = then return the (interpolated) value at . = For details of the integration algo[...]

  • Página 250

    Appendix B: Algorithms Appendices For details of the integration algorithm, see page A-19. Delay T iming measurement. The amount of time between the and crossings of two different traces, or two dif ferent places on the same trace. Delay measurements are actually a group of measurements. T o get a specific delay measurement, you must specify the ta[...]

  • Página 251

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual High HighRef LowRe f Lo w TL F THF Fall Time Figure A-2: Fall Time Frequency T iming measurement. The reciprocal of the period. Measured in Hertz (Hz) where 1 Hz = 1 cycle per second. If = 0 or is otherwise bad, return an error . = 1/ High 100% (highest) voltage reference value. (See “[...]

  • Página 252

    Appendix B: Algorithms Appendices Maximum Amplitude (voltage) measurement. The maximum voltage. T ypically the most positive peak voltage. Examine all samples from to inclusive, and set equal to the greatest magnitude value found. Mean The arithmetic mean for one waveform. Remember that one waveform is not necessarily equal to one cycle. For cyclic[...]

  • Página 253

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual Negative Width T iming measurement. The distance (time) between (default = 50%) amplitude points of a negative pulse. If = ‘–’ then = – else = – Peak to Peak Amplitude measurement. The absolute difference between the maximum and minimum amplitude. = – Period T iming measureme[...]

  • Página 254

    Appendix B: Algorithms Appendices If the target waveform leads the reference waveform, phase is positive; if it lags, negative. Phase is not available in the Snapshot display . Positive Duty Cycle T iming measurement. The ratio of the positive pulse width to the signal peri- od, expressed as a percentage. is defined in Positive Width , following. I[...]

  • Página 255

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual 3. From , continue the search, looking for a crossing of . Update if subsequent crossings are found. If a crossing is found, it becomes the high rise time or . (Use linear interpolation if necessary .) 4. = – High HighRef LowRe f Lo w THR TLR Rise Time Figure A-3: Rise Time RMS: Amplit[...]

  • Página 256

    Appendix B: Algorithms Appendices W(t) is the sampled waveform is the continuous function obtained by linear interpolation of W(t) A and B are numbers between 0.0 and –1.0 If A and B are integers, then: where s is the sample interval. Similarly , is approximated by where: W(t) is the sampled waveform is the continuous function obtained by linear [...]

  • Página 257

    Appendix B: Algorithms TDS 620A, 640A, & 644A User Manual If some samples in the waveform are missing or off-scale, the measurements will linearly interpolate between known samples to make an “appropriate” guess as to the sample value. Missing samples at the ends of the measure- ment record will be assumed to have the value of the nearest k[...]

  • Página 258

    Appendix B: Algorithms Appendices For example, if is set directly , then would not change even if samples were out of range. However , if was chosen using the % choice from the Set Levels in % Units selection of the Measure menu, then could give a “CLIPPING” warning. NOTE When measurements are displayed using Snapshot, out of range warnings are[...]

  • Página 259

    TDS 620A, 640A, & 644A User Manual Appendix C: Packaging for Shipment If you ship the digitizing oscilloscope, pack it in the original shipping carton and packing material. If the original packing material is not available, package the instrument as follows: 1. Obtain a corrugated cardboard shipping carton with inside dimensions at least 15 cm [...]

  • Página 260

    Appendix C: Packaging for Shipment Appendices[...]

  • Página 261

    TDS 620A, 640A, & 644A User Manual Appendix D: Factory Initialization Settings The factory initialization settings provide you a known state for the digitizing oscilloscope. Factory initialization sets values as shown in T able A-7. T able A-7: Factory Initialization Defaults Control Changed by Factory Init to Acquire mode Sample Acquire repeti[...]

  • Página 262

    Appendix D: Factory Initialization Settings Appendices T able A-7: Factory Initialization Defaults (Cont.) Control Changed by Factory Init to Delay trigger average # 16 Delay trigger envelope # 10 Delay time 16 ns Delay events, triggerable after main 2 Delayed, delay by ... Delay by T ime Delayed, time base mode Delayed Runs After Main Display cloc[...]

  • Página 263

    Appendix D: Factory Initialization Settings TDS 620A, 640A, & 644A User Manual T able A-7: Factory Initialization Defaults (Cont.) Control Changed by Factory Init to Display variable persistence 500 ms Edge trigger coupling DC Edge trigger level 0.0 V Edge trigger slope Rising Edge trigger source Channel 1 GPIB parameters No change Hardcopy For[...]

  • Página 264

    Appendix D: Factory Initialization Settings Appendices T able A-7: Factory Initialization Defaults (Cont.) Control Changed by Factory Init to Logic state trigger Ch4 (Ax2) input Rising edge Logic trigger input (pattern and state) Channel 1 = H (high), Channels 2 & 3 (Ax1) = X (don’t care) Logic trigger pattern time qualification Lower limit U[...]

  • Página 265

    Appendix D: Factory Initialization Settings TDS 620A, 640A, & 644A User Manual T able A-7: Factory Initialization Defaults (Cont.) Control Changed by Factory Init to Pulse runt high threshold 1.2 V Pulse runt low threshold 0.8 V Pulse runt trigger polarity Positive Pulse trigger class Glitch Pulse glitch filter state On (Accept glitch) Pulse gl[...]

  • Página 266

    Appendix D: Factory Initialization Settings Appendices T able A-7: Factory Initialization Defaults (Cont.) Control Changed by Factory Init to Zoom state Of f Zoom vertical (all channels) 1.0X Zoom vertical position (all channels) 0 divisions[...]

  • Página 267

    Glossary[...]

  • Página 268

    TDS 620A, 640A, & 644A User Manual Glossary AC coupling A type of signal transmission that blocks the DC component of a signal but uses the dynamic (AC) component. Useful for observing an AC signal that is normally riding on a DC signal. Accuracy The closeness of the indicated value to the true value. Acquisition The process of sampling signals[...]

  • Página 269

    Glossary Glossary Automatic trigger mode A trigger mode that causes the oscilloscope to automatically acquire if triggerable events are not detected within a specified time period. Autoset A function of the oscilloscope that automatically produces a stable waveform of usable size. Autoset sets up front-panel controls based on the characteristics of[...]

  • Página 270

    Glossary TDS 620A, 640A, & 644A User Manual Cycle RMS The true Root Mean Square voltage over one cycle. DC coupling A mode that passes both AC and DC signal components to the circuit. Available for both the trigger system and the vertical system. Delay measurement A measurement of the time between the middle reference crossings of two different[...]

  • Página 271

    Glossary Glossary Gated Measurements A feature that lets you limit automated measurements to a specified portion of the waveform. Y ou define the area of interest using the vertical cursors. General purpose knob The large front-panel knob with an indentation. Y ou can use it to change the value of the assigned parameter . Glitch positive trigger T [...]

  • Página 272

    Glossary TDS 620A, 640A, & 644A User Manual Holdoff, trigger A specified amount of time after a trigger signal that elapses before the trigger circuit will accept another trigger signal. T rigger holdoff helps ensure a stable display . Horizontal bar cursors The two horizontal bars that you position to measure the voltage parameters of a wavefo[...]

  • Página 273

    Glossary Glossary Low The value used as 0% in automated measurements (whenever high ref, mid ref, and low ref values are needed as in fall time and rise time measurements). May be calculated using either the min/max or the histogram method. With the min/max method (most useful for general waveforms), it is the minimum value found. With the histogra[...]

  • Página 274

    Glossary TDS 620A, 640A, & 644A User Manual NOR A logic (Boolean) function in which the output of the OR function is complemented (true becomes false, and false becomes true). On the digitizing oscilloscope, that is a trigger logic pattern and state func- tion. OR A logic (Boolean) function in which the output is true if any of the inputs are t[...]

  • Página 275

    Glossary Glossary Positive width A timing measurement of the distance (time) between two amplitude points — rising-edge MidRef (default 50%) and falling-edge MidRef (default 50%) — on a positive pulse. Posttrigger The specified portion of the waveform record that contains data acquired after the trigger event. Pretrigger The specified portion o[...]

  • Página 276

    Glossary TDS 620A, 640A, & 644A User Manual Sample acquisition mode The oscilloscope creates a record point by saving the first sample during each acquisition interval. That is the default mode of the acquisition. Sample interval The time interval between successive samples in a time base. For real-time digitizers, the sample interval is the re[...]

  • Página 277

    Glossary Glossary T rigger level The vertical level the trigger signal must cross to generate a trigger (on edge trigger mode). V ertical bar cursors The two vertical bars you position to measure the time parameter of a waveform record. The oscilloscope displays the value of the active (moveable) cursor with respect to the trigger and the time valu[...]

  • Página 278

    Index[...]

  • Página 279

    TDS 520A, 524A, 540A, & 544A User Manual Index Numbers 1/seconds (Hz) , Cursor menu, 3-21 100 MHz , V ertical menu, 3-149 20 MHz , V ertical menu, 3-149 A AC coupling, 2-16–2-17 , Glossary-1 AC line voltage, trigger input, 2-14 AC , Main T rigger menu, 3-35 Accept Glitch , Main T rigger menu, 3-122 Accessories, A-1–A-8 Optional, A-5 Probes,[...]

  • Página 280

    Index Index DISPLA Y , 3-12 , 3-28 FORCE TRIG , 3-143 HARDCOPY , 3-55 , 3-61 , 3-128 HELP , 3-67 HORIZONT AL MENU , 2-18 , 3-23 MEASURE , 3-89 MORE , 3-134 , 3-136 , 3-159 ON/STBY , 1-4 , 2-3 Save/Recall SETUP , 1-7 , 3-55 , 3-130 Save/Recall W A VEFORM , 3-55 , 3-133 SELECT , 2-31 , 3-20 , Glossary-9 SET LEVEL TO 50% , 3-143 SINGLE TRIG , 3-8 , 3-[...]

  • Página 281

    Index TDS 520A, 524A, 540A, & 544A User Manual Create Limit T est T emplate , Acquire menu, 3-73 Create Measrmnt , Measure Delay menu, 3-95 Cross Hair , Display menu, 3-31 Current probes, 3-1 15 Cursor Horizontal bar , 2-31 , 3-17 Measurements, 2-31 Mode, 2-31–2-32 Independent, 2-31–2-32 T racking, 2-32 Paired, 2-31 , 3-17 V ertical bar , 2[...]

  • Página 282

    Index Index Display , Display menu, 3-28 Display , Status menu, 3-140 Dots, 3-29 Dots style, Display menu, 3-75 Dots , Display menu, 3-29 DPU41 1–II , Hardcopy menu, 3-61 DPU412 , Hardcopy menu, 3-61 Dual Wfm Math , More menu, 3-160 Dual-lead adapter , 3-103 Duty cycle, 1-18 , Glossary-6 , Glossa- ry-7 E Edge trigger , 2-14 , 3-34 , Glossary-3 Re[...]

  • Página 283

    Index TDS 520A, 524A, 540A, & 544A User Manual G Gated Measurements, 3-91 , Glossa- ry-4 Gating , Measure menu, 3-91 General purpose (high input resis- tance) probes, 3-1 12 General purpose knob, 1-19 , 2-7 , Glossary-4 Glitch trigger , 3-1 19 , 3-120 , Glossa- ry-4 Glitch , Main T rigger menu, 3-121 , 3-122 Goes F ALSE , Main Trigger menu, 3-8[...]

  • Página 284

    Index Index I I/O , Status menu, 3-140 I/O , Utility menu, 3-60 , 3-128 IC protector tip, 3-102 Icons, 1-1 Independent Mode, Cursor , 2-31–2-32 Independent , Cursor menu, 3-20 Infinite Persistence , Display menu, 3-29 Installation, 1-3–1-4 Integral math waveform, 3-154 applications, 3-154 automated measurements of, 3-157 derivation of, 3-154 ma[...]

  • Página 285

    Index TDS 520A, 524A, 540A, & 544A User Manual T rigger When , 3-81 , 3-83 T rue for less than , 3-83 T rue for more than , 3-83 T ype , 3-34 , 3-121 , 3-123 , 3-145 Width , 3-122 MAIN TRIGGER OUTPUT , BNC, 2-5 Map Math , Color menu, 3-14 Map Reference , Color menu, 3-15 Marker rings, 3-100 , A-7 Math W aveform Differential, A-3 FFT , A-3 Integ[...]

  • Página 286

    Index Index Set FFT V ert Scale to: , 3-40 Set FFT Window to: , 3-40 Set Function to , 3-160 Set Function to: , 3-151 , 3-155 Set operator to , 3-161 Set Single Source to , 3-160 Set Single Source to: , 3-150 , 3-154 Single Wfm Math , 3-150 , 3-154 , 3-160 N NAND, Glossary-6 NAND , Main T rigger menu, 3-82 , 3-85 Negative duty cycle, 3-87 Negative [...]

  • Página 287

    Index TDS 520A, 524A, 540A, & 544A User Manual Low-inductance spring-tips, 3-101 , A-7 Marker rings, 3-100 , A-7 Probe tip-to-chassis adapter , 3-102 , A-7 Probe tip-to-circuit board adapters, 3-101 Retractable hook tip, 3-98 SMT KlipChip, 3-101 , A-7 Probe Cal, 3-104–3-109 Probe tip-to-chassis adapter , 3-102 , A-7 Probe tip-to-circuit board[...]

  • Página 288

    Index Index S Safety , v Symbols, v Sample acquisition mode, 3-3 , Glos- sary-9 Sample interval, Glossary-9 Sample , Acquire menu, 3-7 Sampling, 2-20 , Glossary-9 Sampling and acquisition mode, 2-22 Saturation , Color menu, 3-14 Save, Setups, 3-130–3-132 Save Current Setup , Save/Recall Setup menu, 3-130 Save W aveform , Save/Recall W ave- form m[...]

  • Página 289

    Index TDS 520A, 524A, 540A, & 544A User Manual T ek Secure, 3-131 , Glossary-9 T ek Secure Erase Memory , Utility menu, 3-131 T emperature compensation, 3-138–3-139 T emperature , Color menu, 3-13 T emplate Source , Acquire menu, 3-73 T ext/Grat , Display menu, 3-29 Thinkjet, 3-59 Thinkjet , Hardcopy menu, 3-61 Thresholds , Main T rigger menu[...]

  • Página 290

    Index Index W W ARNING, statement in manual, v W aveform, Glossary-10 Interval, Glossary-10 Math, 3-159–3-161 Off priority , 3-137 W aveform clipping. See Clipping W aveform differentiation, 3-150 W aveform FFT s, 3-38 W aveform integration, 3-154 W aveform memory , 3-134 W A VEFORM OFF button, 1-16 , 3-32 , 3-137 W aveform record FFT , 3-44 FFT [...]