Agilent Technologies Agilent E3617A Serials KR83502522 инструкция обслуживания
- Просмотреть online или скачать инструкцию
- 43 страниц
- 1.45 mb
Идти на страницу of
Похожие руководства по эксплуатации
-
Video Gaming Accessories
Agilent Technologies 6681A
156 страниц 3.22 mb -
Video Gaming Accessories
Agilent Technologies 6683A
156 страниц 3.22 mb -
Video Gaming Accessories
Agilent Technologies Serials 3004A-05881 to 05885 3023A-06181 and above Agilent 6033A
129 страниц 0.69 mb -
Video Gaming Accessories
Agilent Technologies Model 6682A: 3339A-00101
133 страниц 1.05 mb -
Video Gaming Accessories
Agilent Technologies 6812B
91 страниц 0.4 mb -
Video Gaming Accessories
Agilent Technologies Serials 2934A-01825 to 01829 3023A-01925
129 страниц 0.69 mb -
Video Gaming Accessories
Agilent Technologies Serials 3038A-00101 and above Agilent 6038A
129 страниц 0.69 mb -
Video Gaming Accessories
Agilent Technologies 90B
125 страниц 1.52 mb
Хорошее руководство по эксплуатации
Законодательство обязывает продавца передать покупателю, вместе с товаром, руководство по эксплуатации Agilent Technologies Agilent E3617A Serials KR83502522. Отсутствие инструкции либо неправильная информация, переданная потребителю, составляют основание для рекламации в связи с несоответствием устройства с договором. В законодательстве допускается предоставлении руководства в другой, чем бумажная форме, что, в последнее время, часто используется, предоставляя графическую или электронную форму инструкции Agilent Technologies Agilent E3617A Serials KR83502522 или обучающее видео для пользователей. Условием остается четкая и понятная форма.
Что такое руководство?
Слово происходит от латинского "instructio", тоесть привести в порядок. Следовательно в инструкции Agilent Technologies Agilent E3617A Serials KR83502522 можно найти описание этапов поведения. Цель инструкции заключается в облегчении запуска, использования оборудования либо выполнения определенной деятельности. Инструкция является набором информации о предмете/услуге, подсказкой.
К сожалению немного пользователей находит время для чтения инструкций Agilent Technologies Agilent E3617A Serials KR83502522, и хорошая инструкция позволяет не только узнать ряд дополнительных функций приобретенного устройства, но и позволяет избежать возникновения большинства поломок.
Из чего должно состоять идеальное руководство по эксплуатации?
Прежде всего в инструкции Agilent Technologies Agilent E3617A Serials KR83502522 должна находится:
- информация относительно технических данных устройства Agilent Technologies Agilent E3617A Serials KR83502522
- название производителя и год производства оборудования Agilent Technologies Agilent E3617A Serials KR83502522
- правила обслуживания, настройки и ухода за оборудованием Agilent Technologies Agilent E3617A Serials KR83502522
- знаки безопасности и сертификаты, подтверждающие соответствие стандартам
Почему мы не читаем инструкций?
Как правило из-за нехватки времени и уверенности в отдельных функциональностях приобретенных устройств. К сожалению само подсоединение и запуск Agilent Technologies Agilent E3617A Serials KR83502522 это слишком мало. Инструкция заключает ряд отдельных указаний, касающихся функциональности, принципов безопасности, способов ухода (даже то, какие средства стоит использовать), возможных поломок Agilent Technologies Agilent E3617A Serials KR83502522 и способов решения проблем, возникающих во время использования. И наконец то, в инструкции можно найти адресные данные сайта Agilent Technologies, в случае отсутствия эффективности предлагаемых решений. Сейчас очень большой популярностью пользуются инструкции в форме интересных анимаций или видео материалов, которое лучше, чем брошюра воспринимаются пользователем. Такой вид инструкции позволяет пользователю просмотреть весь фильм, не пропуская спецификацию и сложные технические описания Agilent Technologies Agilent E3617A Serials KR83502522, как это часто бывает в случае бумажной версии.
Почему стоит читать инструкции?
Прежде всего здесь мы найдем ответы касательно конструкции, возможностей устройства Agilent Technologies Agilent E3617A Serials KR83502522, использования отдельных аксессуаров и ряд информации, позволяющей вполне использовать все функции и упрощения.
После удачной покупки оборудования/устройства стоит посвятить несколько минут для ознакомления с каждой частью инструкции Agilent Technologies Agilent E3617A Serials KR83502522. Сейчас их старательно готовят или переводят, чтобы они были не только понятными для пользователя, но и чтобы выполняли свою основную информационно-поддерживающую функцию.
Содержание руководства
-
Страница 1
A gilen t E361x A 60W B ENCH SERIES DC POWER SUPPLIES OPERA TING A N D SER VICE M A NUA L FOR M ODELS: A gilent E361 4A , Serials KR835 03035 and above A gilent E361 5A , Serials KR835 06197 and above A gilent E361 6A , Serials KR835 02651 and above A gilent E361 7A , Serials KR835 02522 and above For instruments with higher Serial Numbers than abo[...]
-
Страница 2
1-2 SA FETY SUMMARY The following gener al safety precaut i ons must be observed during all phas es of operation, service, and repair of t his instrumen t. Failure to c omply with these precautions or with s pecific warnings elsewhere in this manual violates sa fety standards of design , manufactur e, and intended use of the instrument . Agilent T [...]
-
Страница 3
1-3 T able of Contents SAFETY SUMMARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 GENERAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -4 INTRODUCTION . . . . . . . . . . . . . . . . [...]
-
Страница 4
1-4 GENERA L INFORMA TION INTRODUC TION This manual describes all models in the Agilent E 361xA 6 0W Bench Power Supply family and unless stated other wise, the information in this manual applies to all models. SA FET Y REQU IREM ENTS This product is a Safety Class I instrument, which means that it is provided with a protective earth ground termina[...]
-
Страница 5
1-5 b e g rounded or the power suppl y can be operated float- in g at up to a maximum of 240 V olts o f f g round. T otal out- put voltage to g round must not exceed 240 Vdc. LINE FUSE Line V oltage Fuse Agilent Part No. 100/ 1 15 V ac 2.0 A T 2 1 10-0702 230 V ac 1.0 A T 2 1 10-0457 SPECIFIC A TIONS Detailed specifications for the power suppl y ar[...]
-
Страница 6
1-6 T able 1. Specifications and Operati n g Characteristics (Cont’d) *REMOTE PROGRAMMING SPEED Maximum time required for output volta g e to chan g e from initial value to within a tolerance band (0.1%) of the new l y pro g rammed value followin g the onset of a step chan g e in the pro g rammin g input volta g e. Full load No load Up: E3614A: 3[...]
-
Страница 7
1-7 e . Replace the cover and mark the suppl y clearl y with a ta g or label indicatin g the correct line volta g e and fuse that is in use. Fig ure 2. Line V olta g e Selector (set for 1 15 V ac) Power Cord T o protect operatin g personnel, the instrument should be g rounded. This instrument is equipped with a three conductor power cord. The third[...]
-
Страница 8
1-8 that the current limit value can be set from zero to maximum rated value. OPERA TING MODES The settin g of the rear panel switch determines the operatin g modes of the power suppl y . The local operatin g mode is set so the power suppl y senses the output volta g e directl y at the output terminals (local sensin g ) for operation usin g the fro[...]
-
Страница 9
1-9 REMOTE OPERA TING MODES Remote operatin g modes discussed below are remote volta g e sensin g and remote volta g e pro g rammin g . Y ou can set up the unit for remote operatin g modes b y chan g in g the settin g s of the rear panel switch and connectin g the leads from the rear panel termi- nals to the load or the external volta g e. Solid co[...]
-
Страница 10
1-10 pl y above 100% rated output. Limit y our pro g rammin g volta g e to 10 Vdc. Remote Pro g rammin g Connections. Remote pro g rammin g requires chan g in g settin g s of the switch and connectin g external volta g es to + and - terminals of "CV" or "CC" on the rear panel. An y noise picked up on the pro g rammin g leads wil[...]
-
Страница 11
1-11 Settin g V olta g e and Current. T urn the slave unit's CURRENT control full y clockwise. Adjust the master unit's controls to set the desired output volta g e and current. The master suppl y operates in a completel y normal fashion and ma y be set up for either con- stant volta g e or constant current operation as required. V erif y[...]
-
Страница 12
1-12 AUTO-SERIES OPERA TION Auto-series operation permits equal or proportional volta g e sharin g , and allows control of output volta g e from one master unit. The volta g e of the slaves is determined b y the settin g of the front panel VOL T AGE control on the master and volta g e divider resistor . The master unit must be the most positive sup[...]
-
Страница 13
1-13 Remote Sensin g . T o remote sense with auto-series operation, set SENSE switch of the master unit and set SENSE switch of the slave unit to remote. Remote Analo g V olta g e Pro g rammin g . T o remote analo g pro- g ram with auto-series operation, connect pro g ram (external) volt- a g es to the "CV" or "CC"" termina[...]
-
Страница 14
1-14 LOAD CONSIDERA TIONS This section provides information on operatin g y our suppl y with various t y pes of loads connected to its output. PULSE LOADING The power suppl y will automaticall y cross over from constant- volta g e to constant current operation in response to an increase (over the preset limit) in the output current. Althou g h the [...]
-
Страница 15
A-1 SERVICE INFORMA TION Figure A-1. Block D iagram PRINCIPLES OF OPE R A TION (Block Diagram Overview ) Throughout t his discussion, refer t o both the block d i agram of Figure A-1 and t he schemat ic diagrams at the rear of the manual. The input ac line voltage is first applied to the prereg- ulator w hich operate s in c onj unction with t he SC[...]
-
Страница 16
A-2 series pass transistors. Whenever the output volta g e is below the slopin g V1 line, the control circuit inhibits four SCRs and the input capacitors char g e to a volta g e determined b y N1. Fi g ure A-2 indicates the windin g s that are connected as a result of the other volta g e decisions. Fi g ure A-2. Output Power Plot The series re g ul[...]
-
Страница 17
A-3 a . Perform turn-on checkout procedure g iven in pa g e 1-7. b . Perform the CV and CC Load Re g ulation perfor- mance tests g iven in the followin g para g raphs respectivel y . PERFORMANCE TESTS The followin g para g raphs provide test procedures for verif y- ing the power suppl y 's compliance with the specifications of T able 1. Please[...]
-
Страница 18
A-4 CONST ANT VOL T AGE (CV) TESTS CV Setup. For all CV tests set the output current at full rated output to assure CV operation. The onset of constant current can cause a drop in output volta g e, increased ripple, and other performance chan g es not properl y ascribed to the con- stant volta g e operation of the suppl y . Load Regulation (Load Ef[...]
-
Страница 19
A-5 Fi g ure A-5. Load T ransient Response Time W aveform P ARD(Ripple and Noise) Definition: P ARD is the Periodic and Random Deviation of the dc output volta g e from its avera g e value, over a specified bandwidth and with all other parameters maintained constant. Constant volta g e P ARD is measured in the root-mean- square(rms) or peak-to-peak[...]
-
Страница 20
A-6 a. Connect the test equipment as shown in Fi g ure A-7. b. T urn the suppl y 's power on and turn CURRENT con- trol full y clockwise. c. T urn up output volta g e to the full rated value. Check that the suppl y 's CV indicator remains li g hted. Reduce VOL T AGE control if not li g hted. d. Set the oscilloscope to AC mode and bandwidt[...]
-
Страница 21
A-7 e. Record output volta g e across Rs and convert it to current b y dividin g this volta g e b y Rs. f. Adjust autotransformer to the hi g h line volta g e (127 V ac for nominal 1 15 V ac, 1 10 V ac for nominal 100 V ac, and 253 V ac for nominal 230 V ac). g . When the readin g settles, record the volta g e across Rs a g ain and convert it curre[...]
-
Страница 22
A-8 Ammeter and CC Set Calibration T o calibrate ammeter and CC set, proceed as follows: a. Connect test setup on Fi g ure A-9. b. T urn VOL T AGE and CURRENT control full y clock- wise. c. T urn on the suppl y and to calibrate ammeter adjust R5 on the displa y board until front panel AMPS dis- pla y reads exactl y DVM value divided b y Rs. d. T o [...]
-
Страница 23
A-9 After the trouble has been isolated to one of the feedback loops, troubleshootin g can proceed as described in T ables A- 4, A-5, or A-6. Series Re g ulatin g Feedback Loop. When troubleshootin g the series re g ulatin g loop, it is useful to open the loop since measurements made an y where within a closed loop ma y appear abnormal. With a loop[...]
-
Страница 24
A-10 SYMPTOM CHECKS AND PROBABLE CAUSES Poor Load Re g ulation (Constant V olta g e) a. Refer to "Measurements T echniques" para g raph. b. Check +10 V reference volta g e. c. Ensure that the suppl y is not g oin g into current limit. Poor Load Re g ulation (Constant Current) a. Check +10 V reference volta g e. b. CR1, CR19, CR20, C2, C31[...]
-
Страница 25
A-11 STEP ACTION RESPONSE PROBABLE CAUSE 4 Check volta g e from pin 13 to pin 12 of U9. a. Measured volta g e is positive. b. Measured volta g e is ne g ative. a. Check U9A is defective. b. Check U10 and U9D is defective. Check R85 is open. 5 Check volta g e from pin 6 to pin 5 of U9. a. Measured volta g e is positive. b. Measured volta g e is ne g[...]
-
Страница 26
A-12 REPL A CEA B LE P A R TS IN TROD UCTI ON This section contains informat ion for ordering replacement parts. T able A-10 lists p arts by reference designators and pro- vides the following inf ormation: a. Reference designat ors. Refer to T able A-8. b. Agilent T echnologies Part Number . c. T otal quantity used in that assembly. d. Description.[...]
-
Страница 27
Table A -10. Replaceable P arts List Reference Designator Number Q'ty Description Model Mfr. P/N Mfr. Code E3614A 1 60W BENCH POWER SUPPLY - E3614A MODEL E3615A 1 60W BENCH POWER SUPPLY - E3615A MODEL E3616A 1 60W BENCH POWER SUPPLY - E3616A MODEL E3617A 1 60W BENCH POWER SUPPLY - E3617A MODEL E3614-60005 1 MAIN BODY ASSY 14 28480 E3615-60005 [...]
-
Страница 28
Table A -10. Replaceable P arts List (Co nt'd) Reference Designator Agilent Part Number Q'ty Description Model Mfr. P/N Mfr. Code C18,21,24,27 0160-7077 4 CAP-FXD .1UF +-10% 630V POLYE-FL 14,15 28480 C19,22 0160-4822 2 CAP-FXD 1000PF +-5% 100V CER COG 14,15 28480 C20,23 0180-3970 2 CAP-FXD 1UF +-20% 50V AL-ELECTLT 14,15 28480 C28 0160-622[...]
-
Страница 29
Table A -10. Replaceable P arts List (Co nt'd) Reference Designator Agilent Part Number Q'ty Description Model Mfr. P/N Mfr. Code R24,26,27,37,38, 64,88,117,120 0757-0442 9 RESISTOR 10K +-1% .125W TF TC=0+-100 ALL 28480 R25,30,33 0698-8824 3 RESISTOR 562K +-1% .125W TF TC=0+-100 ALL 28480 R28,111 0698-3228 2 RESISTOR 49.9K +-1% .125W TF T[...]
-
Страница 30
R62 0698-4514 1 RESISTOR 105K +-1% .125W TF TC=0+-100 15 28480 R62 0757-0481 1 RESISTOR 475K +-1% .125W TF TC=0+-100 16 28480 Table A -10. Replaceable P arts List (Co nt'd) Reference Designator Agilent Part Number Q'ty Description Model Mfr. P/N Mfr. Code R62 0698-8826 1 RESISTOR 825K +-1% .125W TF TC=0+-100 17 28480 R63 0698-8827 1 RESIS[...]
-
Страница 31
R109 0698-3153 1 RESISTOR 3.83K +-1% .125W TF TC=0+-100 ALL 28480 R115 0757-0462 1 RESISTOR 75K +-1% .125W TF TC=0+-100 14 28480 Table A -10. Replaceable P arts List (Co nt'd) Reference Designator Agilent Part Number Q'ty Description Model Mfr. P/N Mfr. Code R115 0757-0463 1 RESISTOR 82.5K +-1% .125W TF TC=0+-100 15 28480 R115 0757-0462 1[...]
-
Страница 32
Table A -10. Replaceable P arts List (Co nt'd) Reference Designator Agilent Part Number Q'ty Description Model Mfr. P/N Mfr. Code CR21,22,23,24,25, 26,27,28,29,30 1901-0033 10 DIODE-GEN PRP 180V 200MA DO-35 ALL 1N645 27014 VR1,2,3 1902-0579 3 DIODE-ZNR 5.1V 5% PD=1W IR=10UA ALL 1N4733APL 04713 RT1,2 0837-0261 2 DIODE-VARISTOR ALL V275LA20[...]
-
Страница 33
Table A-11. Component Val ue Component Model E3614A E3615A E3616A E3617A C2 1000UF 25V +-20% AL-ELECTLT 470UF 50V +-20% AL-ELECTLT 330UF 50V +-20% AL-ELECTLT 220UF 100V +-20% AL-ELECTLT C7,8 39000UF 25V +-20% AL-ELECTLT 12000UF 63V +-20% AL-ELECTLT 5600UF 100V +-20% AL-ELECTLT 2700UF 160V +-20% AL-ELECTLT C11,16,17,25,26 0.01UF 100V +-10% CER X7R 0[...]
-
Страница 34
Manual Supple ment Suppleme nt Agilent Part Numbe r : 5959-53 36, Edition 4 Supple ment Print Date : 14 A pril, 2000 This supplement update s th e following docume nt: Agilen t E 361XA 60W Serie s Lab Bench DC Power Supplie s Manual Agile nt Part Nu mber : 5959 -5310 What is a manual suppleme nt? A manual suppleme nt keeps your manual up-to-date . [...]
-
Страница 35
1-10-1 Voltage and Current Programming of the E3614A/15A/16A/ 17A with a Voltage and Current Source Remote analog voltage programming permits control of the regulated output voltage or current by means of a remotely varied voltage or current. The stability of the programming voltages directly affects the stability of the output. The voltage control[...]
-
Страница 36
1-10-2 Programming V oltage Common to the Minus Output Figure 2 Set the CV switch down on the rear panel, and all others up. V in = 1/A x V out V out = A x V in Where V out is the power supply output voltage. V in is the programming voltage. A is the gain factor and the values of each model are as below . Alternative Voltage Programming Using Resis[...]
-
Страница 37
1-10-3 Programming V oltage Common to the Minus Output Figure 4 The output will always be the same or less than the programming voltage. The M/S2 switch must be in the down position. For best results, place a 0.1 µ F capacitor in parallel with R2. V in = (R1 R2) / R2 x V out V out = R2 / R1+R2) x V in Where V out is the power supply output[...]
-
Страница 38
1-10-4 Constant Current Mode The E3614A/15A/16A/17A may be programmed for constant current with an analog voltage or current. Constant current with analog voltage programming can only be achieved with a voltage source that is common with the positive output terminal. Constant Current with V oltage Programming Figure 5 Set the CC switch down the rea[...]
-
Страница 39
1-10-5 Figure 6 Set the CC switch down, and all others up. I in = 1/A x I out I out = A x V in Where I out is the power supply output current in amps. I in is the programming current in µ amps. A is the gain. Programming currents can be increased by adding a resistor across the CC+ and CC-. A 10 volts drop across R1 represents full scale current o[...]
-
Страница 40
1-10-6 Voltage and Current Programming of the E3614A/15A/16A/17A with Resistors Remote programming with resistors permits control of the regulated output or current by means of a remotely varied resistor . The sum of the resistance of external programming resistors (R1 + R2) should be more than 40 kohm. T o have mor e pr ecise output voltage , use [...]
-
Страница 41
1-10-7 Remote Resistor Programming, Constant Current Figure 8 Set the CC switch down on the rear panel, and all others up. I out = A x [V REF x {R/(R + R2 + 100)}] Where I out is the power supply output current. A is the gain factor and the values of each model are as below . V REF is between 10.1 1 V and 11.40 V . R = (92800 x R1)/(92800 + R1) R1 [...]
-
Страница 42
CERTIFICATION Agilent T echno l ogies certifies that this product met its pub li shed specificat i ons at t i me of shipmen t fro m t he fa ctory . Agilent T ech nologies furthe r certifies that its calibrat ion meas ureme nts are trac eable to the United St ates Na tional Instit ute of Stan- dards and T echnology (fo rmerly N ational Bureau of S t[...]
-
Страница 43
DECLA RA TION OF CONFORMITY accordi ng to ISO/IEC Guide 22 an d EN 45014 Manufactu rer’s Name: Agilent Tec hnologies, Inc. Manufactu rer’s A ddress : 345-15, Ka san-dong, Kumch on-ku, Seoul 15 3-023 Korea declares, that the products Product Na me: DC Power Supply Model Numbers : E3614A, E3615 A, E3614A, E36 15A, and E3617A Product Opt ions: All[...]