Genesis Advanced Technologies 2006 инструкция обслуживания
- Просмотреть online или скачать инструкцию
- 22 страниц
- 0.23 mb
Идти на страницу of
Похожие руководства по эксплуатации
-
Bathroom Aids
Genesis Advanced Technologies 2006
22 страниц 0.23 mb -
Stereo Amplifier
Genesis Advanced Technologies Reference Amplifier
2 страниц 0.98 mb -
Speaker
Genesis Advanced Technologies G7.1p
1 страниц 0.22 mb -
Speaker
Genesis Advanced Technologies 8200F
2 страниц 4.4 mb -
Portable Speaker
Genesis Advanced Technologies 350
2 страниц 3.02 mb -
Speaker
Genesis Advanced Technologies G6.1e
1 страниц 0.42 mb -
Stereo Amplifier
Genesis Advanced Technologies B-200
4 страниц 9.19 mb -
Portable Speaker
Genesis Advanced Technologies 500
9 страниц 8.94 mb
Хорошее руководство по эксплуатации
Законодательство обязывает продавца передать покупателю, вместе с товаром, руководство по эксплуатации Genesis Advanced Technologies 2006. Отсутствие инструкции либо неправильная информация, переданная потребителю, составляют основание для рекламации в связи с несоответствием устройства с договором. В законодательстве допускается предоставлении руководства в другой, чем бумажная форме, что, в последнее время, часто используется, предоставляя графическую или электронную форму инструкции Genesis Advanced Technologies 2006 или обучающее видео для пользователей. Условием остается четкая и понятная форма.
Что такое руководство?
Слово происходит от латинского "instructio", тоесть привести в порядок. Следовательно в инструкции Genesis Advanced Technologies 2006 можно найти описание этапов поведения. Цель инструкции заключается в облегчении запуска, использования оборудования либо выполнения определенной деятельности. Инструкция является набором информации о предмете/услуге, подсказкой.
К сожалению немного пользователей находит время для чтения инструкций Genesis Advanced Technologies 2006, и хорошая инструкция позволяет не только узнать ряд дополнительных функций приобретенного устройства, но и позволяет избежать возникновения большинства поломок.
Из чего должно состоять идеальное руководство по эксплуатации?
Прежде всего в инструкции Genesis Advanced Technologies 2006 должна находится:
- информация относительно технических данных устройства Genesis Advanced Technologies 2006
- название производителя и год производства оборудования Genesis Advanced Technologies 2006
- правила обслуживания, настройки и ухода за оборудованием Genesis Advanced Technologies 2006
- знаки безопасности и сертификаты, подтверждающие соответствие стандартам
Почему мы не читаем инструкций?
Как правило из-за нехватки времени и уверенности в отдельных функциональностях приобретенных устройств. К сожалению само подсоединение и запуск Genesis Advanced Technologies 2006 это слишком мало. Инструкция заключает ряд отдельных указаний, касающихся функциональности, принципов безопасности, способов ухода (даже то, какие средства стоит использовать), возможных поломок Genesis Advanced Technologies 2006 и способов решения проблем, возникающих во время использования. И наконец то, в инструкции можно найти адресные данные сайта Genesis Advanced Technologies, в случае отсутствия эффективности предлагаемых решений. Сейчас очень большой популярностью пользуются инструкции в форме интересных анимаций или видео материалов, которое лучше, чем брошюра воспринимаются пользователем. Такой вид инструкции позволяет пользователю просмотреть весь фильм, не пропуская спецификацию и сложные технические описания Genesis Advanced Technologies 2006, как это часто бывает в случае бумажной версии.
Почему стоит читать инструкции?
Прежде всего здесь мы найдем ответы касательно конструкции, возможностей устройства Genesis Advanced Technologies 2006, использования отдельных аксессуаров и ряд информации, позволяющей вполне использовать все функции и упрощения.
После удачной покупки оборудования/устройства стоит посвятить несколько минут для ознакомления с каждой частью инструкции Genesis Advanced Technologies 2006. Сейчас их старательно готовят или переводят, чтобы они были не только понятными для пользователя, но и чтобы выполняли свою основную информационно-поддерживающую функцию.
Содержание руководства
-
Страница 1
USER’S MANUAL GENES I S 2002 TM GENES I S 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com 0609[...]
-
Страница 2
GENESIS 2006 SHOWN 2[...]
-
Страница 3
To assure proper record keeping, please retain this information for future use. DATE PURC HASED ____________________________ SERIAL # _____________________________________ DISTRIBUT OR NAME __________________________ PHONE NUMBER ______________________________ CITY, STATE __________________________________ The GENESIS THAWING SYSTEMS incorporate wa[...]
-
Страница 4
CONTENTS page 1. UNPACKING ........................................................ 5 2. TO THAW BLOOD PRODUCTS ........................ 5 3. TO REMOVE A THAWED BAG ........................ 9 4. TO MAINTAIN THE DEVICE IN THE READY TO THAW STATUS .............................. 9 5. TO CLEAN THE BATH OR TO ADD WATER .. 10 6. CONTROL PANEL AND INDICAT[...]
-
Страница 5
1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006 unit. C o n t r o l U n i t W a t e r B a t h T h a w i n g B a g D i s p e n s e r Plasma Overwrap Thawing Bags (1000/Box) P l a s m a B a g H o l d e r L A B a t h C L E A N TM R e f e r e n c e T h e r m o m e t e r T h a w i n g B a g W e i g h t s U s e r ' s M a n[...]
-
Страница 6
indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEAN TM each time the water is changed to control bacterial and algae growth. The bath cleaning agent also helps to prev ent the formation of turbidity and odor. 2.3 Plug the device into a 115 volt AC outlet. For GENESIS 2002 , a dedica[...]
-
Страница 7
c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 ° C, in one degree increments. To adjust, press the SET TEMP key - then press the numeric key pad to indicate the temperature desired. All keys must be pressed within 5 seconds of each other. The display should now show the new SET TEMP. When th[...]
-
Страница 8
thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel can set the desired thawing time for each unit to be thawed. They do not have to all be set for the same thaw time. f. To change the default TIMER : Press the START/CHANGE key for the selected bag location, [...]
-
Страница 9
3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder. Discard plasma thawing overwrap bag after each use, as they are designed for single use only. b. If there is another bag being thawed, press the ROCK key to resume the rocking motion for the rem ai[...]
-
Страница 10
5. TO CLEAN THE BATH OR TO ADD WATER S A F E T Y : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL) IS TURNED ON. RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET. 5.1 TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF. BEFORE REMOVING THE[...]
-
Страница 11
when filled with water. Do not drop the bath on a hard surface. GENESIS 2006 : Remove all, or part of the water. The bath weighs approximately thirty (30) pounds when filled with water. Carefully remove the water bath from the device. 5.6 DO NOT USE BLEACH OR HARSH CLEANERS. Depending on mineral content of water used, the bath may discolor. Pro[...]
-
Страница 12
6. CONTROL PANEL AND INDICATORS GENESIS 2002 TM CONTROL PANEL 12[...]
-
Страница 13
GENESIS 2006 TM CONTROL PANEL 13[...]
-
Страница 14
CONTROL KEYS POWER .......... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER SWITCH on the left panel is a master on/off switch to remove electrical power completely from all circuits in the unit. SET TEMP ..... With this key switch you can adjust water temperature within the range of ambient to +[...]
-
Страница 15
INDICATORS HEATING ....... This red indicator is lit when the heating element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off. FAULT ............ If the water temperature exceeds 58 ° C, an alarm will sound and the red fault light will be illuminated. This m[...]
-
Страница 16
2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch. 3) The FAULT light is on. • Check the water temperature. If the w[...]
-
Страница 17
7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrum ent. 8) HEATING light is flashing. • This is normal operation. At approximately one degree below the set temperature, the microprocessor begins[...]
-
Страница 18
8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time setting for each bag location • Forced water circulation with gentle rocking motion • Compact size - saves valuable bench space • Quick heating • Simple to operate - easy to read LED display • Li[...]
-
Страница 19
10. SPECIFICATIONS GENESIS 2002 TM GENESIS 2006 TM External Dim ension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” High Interior Bath Dimension 13.0” L x 7.0” Wide x 7.5” Deep 18.0” L x 8.0” Wide x 7.5” Deep Bath Volume 2.5 US gallons 3.7 US gallons Controlled Tem p Range Ambient to 56 o C Ambient to 56 o C [...]
-
Страница 20
11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 65 Commerce W ay Hackensack, NJ 07601 U.S.A. Tel: 201-708-1400 866-712-5663 Fax: 201-704-1104 www.genesisbps.com RETURN INFORMATION : Same as above. 1. Obtain a Return Goods Authorization number. 2. Ensure all water is [...]
-
Страница 21
Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006 was purchased S e r i a l # RGA # (Return Goods Authorization #) 6. Mark outside of box clearly, with your return address, and send to GenesisBPS, LLC . (as above) 21[...]
-
Страница 22
GENESIS 2002 GENESIS 2006 PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006 , Plasma Thawing Water Bath System is warranted against defects in materials and workmanship for a period of one year from the date of purchase. The warranty applies only to products that have not been tampered with, used for purposes other than th[...]