Emerson 951C manual

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64

Go to page of

A good user manual

The rules should oblige the seller to give the purchaser an operating instrucion of Emerson 951C, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.

What is an instruction?

The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Emerson 951C one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.

Unfortunately, only a few customers devote their time to read an instruction of Emerson 951C. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.

What should a perfect user manual contain?

First and foremost, an user manual of Emerson 951C should contain:
- informations concerning technical data of Emerson 951C
- name of the manufacturer and a year of construction of the Emerson 951C item
- rules of operation, control and maintenance of the Emerson 951C item
- safety signs and mark certificates which confirm compatibility with appropriate standards

Why don't we read the manuals?

Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Emerson 951C alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Emerson 951C, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Emerson service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Emerson 951C.

Why one should read the manuals?

It is mostly in the manuals where we will find the details concerning construction and possibility of the Emerson 951C item, and its use of respective accessory, as well as information concerning all the functions and facilities.

After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.

Table of contents for the manual

  • Page 1

    Instruction Manua l 748214-V June 2009 http://www.raihome.com Model 951C NO X Analyzer[...]

  • Page 2

    WORLD HEA DQUARTERS Emerson Process Manage ment Rosemount A nalytical Inc. 6565P Davis Industrial Parkw ay Solon, OH 44139 T 440 914 1261 T 800 433 6076 F 440 914 1262 E gas.csc@emerson.co m ESSENTI A L INSTRUCT IONS READ THIS P A GE BEFO RE PROCEEDI NG! Rosem ount Anal yti cal designs, m anufac tures and tests its products t o meet m any national [...]

  • Page 3

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Contents i Model 951C HIGHLIG HTS OF CH A NGES EFFECTIVE JUNE 10, 2009 PAGE SUMM A RY Cover Updated photo, re vision num ber, and rele ase date P-6 Rem oved Com pliance secti on 1-1 Add ed Mid Range 1-2 Add ed Mid Range 3-1 Added Mid R ange, B[...]

  • Page 4

    Instruction Manua l 748214-V June 2009 ii Contents Rosemount Analytical Inc. A D ivision of Emerson Process Man agement Model 951C TABLE O F CON TENTS PREFACE .................................................................................................................................. ......... P-1 Definitions ..................................[...]

  • Page 5

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Contents iii Model 951C 4-1 Nitric Oxide D eterm ination b y Chemilum inescence Me thod ............................................. .... 4-1 4-2 Analyzer F low S y stem ........................................................................[...]

  • Page 6

    Instruction Manua l 748214-V June 2009 iv Contents Rosemount Analytical Inc. A D ivision of Emerson Process Man agement Model 951C LIST OF I LLUSTR ATIONS Figure 2-1. Power Supp ly Board Voltage Sel ect Switch es .................................................... ...... 2-2 Figure 2-2. T em perature Control Board .............................. ..[...]

  • Page 7

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Contents v Model 951C[...]

  • Page 8

    [...]

  • Page 9

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Preface P-1 Model 951C PREF A CE The pu rpose of this m anual is to p rovide inf ormatio n concerning the com ponen ts, function s, installation and main tenance o f the 951 C NO X Analyzer. Some section s ma y describe equipm ent no t used in[...]

  • Page 10

    Instruction Manua l 748214-V June 2009 P-2 Preface Rosemount Analyti cal Inc. A Division of Emerson Process Manage ment Model 951C S A FETY SUM MARY If this equipm ent is used in a m anner not spec ified in thes e instructions , protective s ystems m ay be impaired. AUTHORI ZED PE RSONNE L To avoid explosion, loss of life, personal injury and d ama[...]

  • Page 11

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Preface P-3 Model 951C WARNING TOXIC CH EMIC AL HAZARD This inst rument gener ates oz one w hich is toxic b y inhalation and i s a stron g irritant to th roat and lungs. Oz one is also a strong oxidizing agen t. Its presence i s detected b y a[...]

  • Page 12

    Instruction Manua l 748214-V June 2009 P-4 Preface Rosemount Analyti cal Inc. A Division of Emerson Process Manage ment Model 951C WARNING . TOXIC AND OXIDIZING GAS H AZARD The oz one generator lam p cont ains mercury. L am p breakage could result in mercu ry exposur e. Mercury is hig hly toxic if absorbed t hrough s kin or ing ested, or if v apors[...]

  • Page 13

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Preface P-5 Model 951C GENER AL PREC AUTIONS FOR H ANDLING AND STOR ING HIG H PRESSU RE G AS CYLINDE RS Edited from selected p aragraph s of the Comp resse d Gas Association's "Han dboo k of Compres sed Gases" pub lished in 1 98[...]

  • Page 14

    Instruction Manua l 748214-V June 2009 P-6 Preface Rosemount Analyti cal Inc. A Division of Emerson Process Manage ment Model 951C DOCUMEN T A TION The f ollowing 951C instruc tion m aterials are available. Con tact Cust omer Service Center or the local rep- resentative to order. 748214 Ins truction M anual (this doc ument)[...]

  • Page 15

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Preface P-7 Model 951C CONDENS ED ST ARTUP AND CALIBR A TION P ROCEDU RE The f ollowing summ arized ins tructions on startup and c alibration are intended f or opera- tors alread y familiar wit h the anal yzer. For initial start up, refer to d[...]

  • Page 16

    Instruction Manua l 748214-V June 2009 P-8 Preface Rosemount Analyti cal Inc. A Division of Emerson Process Manage ment Model 951C b. Suppl y upscale st andard gas of accu- rately kno wn NO X conte nt to rear p anel SAMPLE in let. c. Adjust f ront panel R ANGE1 Contro l so that reading on meter or recor der is equal to th e know parts- per-m illion[...]

  • Page 17

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Installation 1-1 Model 951C SECTION 1 DESCRIPTI ON AND S PECI FICATIONS 1-1 OVERVI EW The Model 9 51C NO X Analyzer is d esigned to meas ure NO X using one of three sets of ran ges designated as Hi, Mid, or Lo. The H i Range set consists of sp[...]

  • Page 18

    Instruction Manua l 748214-V June 2009 Description and Specifi cations Rosemount Analytical Inc. A Division of Emerson Process Mana gement 1-2 Model 951C 1-3 SPECIFIC ATIONS Ranges L O ......................................... 0 to 10, 0 to 25, 0 to 100, 0 to 250 ppm NO X M ID ....................................... 0 to 20, 0 to 50, 0 to 200, 0 to[...]

  • Page 19

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Installation 2-3 Model 951C SECTION 2 INST A LL ATION 2-1 UNP ACKING Carefull y examine the sh ipping car ton and c on- tents for s igns of dam age. Imm ediately notif y the shipping carrier if th e carton or its contents are dam aged. Retain [...]

  • Page 20

    Instruction Manua l 748214-V June 2009 2-2 Installation Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C Set switch wind ow for voltage required. The f ollowing power cord and/or sup port feet (f or bench top use) ar e availab le: • Power Cord 634061 • North Am erican power c ord set (10 foot) • Enclosure S uppor[...]

  • Page 21

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Installation 2-3 Model 951C EXHAUST AIR IN 20 PSI (1 38 kPa) NOMINAL RECORDER OUTPUT POWER + - G + - CUR OUTPUT VOLT OUTPUT L1/HOT L2/NEUT GND SAMPLE IN 10 PSI - 1 7 PSI (70 kPa - 120 kPa) FUSE RANGE 1 RANGE 2 RANGE 3 RANGE 4 RTN RTN +24 VDC I[...]

  • Page 22

    Instruction Manua l 748214-V June 2009 2-4 Installation Rosemount Analytical In c. A Division of Emerson Process Mana gement Model 951C Figure 2-5. R ear Pane l Wiring Con nections RECO RDER OUTPUT POWER + - G + - CUR OUTPUT VOLT OUTPUT L1/HOT L2/NEUT GND FUSE RANGE 1 RANGE 2 RANGE 3 RANGE 4 RTN RTN +24 VDC IN +24 VDC IN CHS GND CHS GND TB1 TB2 J2 [...]

  • Page 23

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Installation 2-5 Model 951C 2-5 GAS REQ UIREMENT S The instrum ent requires t wo gases norm ally supplied f rom c ylinders. The y are: a. Air (U.S.P. Br eathing G rade) This is us ed as both (a) an ox ygen source for generat ion of the ozo ne [...]

  • Page 24

    Instruction Manua l 748214-V June 2009 2-6 Installation Rosemount Analytical In c. A Division of Emerson Process Mana gement Model 951C 1. Rem ove plugs a nd caps fr om all inlet and outlet fittings. ( See Fig ure 2-4, pag e 2-3.) 2. Conn ect EXHAUST outlet t o externa l vent via tubing with O.D. of 1/4-inc h (6.3 mm ) or larger. Use on ly stainl e[...]

  • Page 25

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Operation 3- 1 Model 951C Converter Temp Check (S4) Range Select Switch (SW 1) SIGNAL BOAR D (See Figure 3-2) Ozone Indicator Lamp (Signal Board DS 2) Range1 Contro l (Signal B oard R101) Zero Control (Signal Board R100) Display (Signal Board [...]

  • Page 26

    Instruction Manua l 748214-V June 2009 3-2 Operation Ro semount Analytical Inc. A Division of Emerson Proces s Management Model 951C e. Zero and Span Potent iometers See Figure 3- 1 (page 3- 1) and F igure 3-2 below. Scr ewdriver ac cess holes through the front pane l allow a djustm ents of the ZERO, RANG E1 and R ANGE2 p otenti- ometer s (R100, R1[...]

  • Page 27

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Operation 3- 3 Model 951C 3-2 ST ARTUP PROCEDU RE The f ollowing are deta iled instruc tions on startup and c alibration. 1. Suppl y electrical p ower to t he analyzer. The anal yzer will require approx imately two hours f or tem perature equi[...]

  • Page 28

    Instruction Manua l 748214-V June 2009 3-4 Operation Ro semount Analytical Inc. A Division of Emerson Proces s Management Model 951C 3. Establish c orrect pressure f or air b y the fol- lowing: a. Adjust O ZONE Pres sure Regu lator so that OZONE Pressure Ga uge ind i- cates 20 to 2 5 psig (13 8 to 172 k Pa). b. To e st a b l i sh co rr ec t p re s [...]

  • Page 29

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Operation 3- 5 Model 951C Figure 3-9. Po wer Sup ply Board[...]

  • Page 30

    Instruction Manua l 748214-V June 2009 3-6 Operation Ro semount Analytical Inc. A Division of Emerson Proces s Management Model 951C 3-3 CALIBR ATION a. Zero Calibratio n 1. On the Signal Boar d (Figur e 3-2, page 3-2), set PPM RANG E Switch (Figure 3-6, p age 3-3) f or the sam e range that w ill be used d uring sam ple analysis. Set front panel RA[...]

  • Page 31

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Operation 3- 7 Model 951C Initiall y, the temper ature of the peak of the converter ef ficienc y starts at a relati vely high value because signific ant heat m ust be sup- plied to m ake the con verter active e nough to reduce the i nput nitro[...]

  • Page 32

    Instruction Manua l 748214-V June 2009 3-8 Operation Ro semount Analytical Inc. A Division of Emerson Proces s Management Model 951C NOT E Converter t emperatur e is not a direct measure of co nverte r efficienc y. Temperat ure measur ement is fo r reference purposes onl y.[...]

  • Page 33

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Operation 3- 9 Model 951C 3-6 M EASUREMENT OF CO NVERTER EFFICIENCY It is the respo nsibilit y of the user to m easure efficienc y of the NO 2 -to-NO converter dur ing initial start up, and there after at interv als ap- propriate to th e appli[...]

  • Page 34

    [...]

  • Page 35

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Theory 4-1 Model 951C SECTION 4 THEORY 4-1 NITRIC OXIDE DET ERMINATION BY CHEM ILUMINESCENC E M ETHOD The chem iluminescenc e method f or detection of nitric oxide (NO) is base d on its react ion with ozone (O 3 ) to prod uce nitro gen dioxide[...]

  • Page 36

    Instruction Manua l 748214-V June 2009 4-2 Theory Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C 4-3 SIGN AL CONDIT IONING AND DI SPL AY A block diagram of the signal-proces sing elec- tronics is sho wn in Figur e 4-1, pag e 4-3. The signa l conditioning and disp lay board pro- vides the f ollowing func tions: • Si[...]

  • Page 37

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Theory 4-3 Model 951C recorder out put adjustm ent to m atch dis- play. U5 is the recor der output am plifier. T he out- put signal m ay be trim med by R25 to pre- cisely matc h the recorder calibrat ion. Note that this adj ustm ent does not a[...]

  • Page 38

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Theory 4-5 Model 951C Figure 4-1. Analyzer B lock Diag ram Display Range Sw itch Span Adjustm ents R101, R102 Voltage Output Control Voltage-to-Curr ent Converter High Voltage Supply Zero Control PMT Electrometer Gain Amp li fier Span Am plifi[...]

  • Page 39

    Instruction Manua l 748214-V June 2009 4-6 Theory Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C Figure 4-2. Analyzer T hermal S ystem Fa n Hea ter CASE HEATER Cool er Fa n Cooli ng Fi ns PMT SOLID-STATE COOLER EXHAUST FAN Top Vie w of An aly z er FRONT PANEL INLET VENT HOLES[...]

  • Page 40

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Theory 4-7 Model 951C[...]

  • Page 41

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Troubleshooting 5-1 Model 951C SECTION 5 ROUTINE S ERVIC ING WARNING ELECTRIC AL SHOCK H AZARD Servicing req uires acc ess to liv e parts which can cause death or seriou s injury. Refer servicin g to qual ified perso nnel. WARNING INTERN AL UL[...]

  • Page 42

    Instruction Manua l 748214-V June 2009 5-2 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C c. Ozone O utput WARNING TOXIC G AS HAZ ARD Use extrem e caution in trou bleshooting the ozon e generator. O zone is toxi c. To check for adequate output f rom the ozone lam p, a conven ient techni que is to ca[...]

  • Page 43

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Routine Servicing 5-3 Model 951C 1. T urn off instrum ent power a nd shut of f all gases. 2. Ref er to Figure 6-1 (page 6 -4) and Figure 6-2 (pa ge 6-5). Co ver and shade the f ittings on the r eaction cham ber with a dark cloth or other light[...]

  • Page 44

    Instruction Manua l 748214-V June 2009 5-4 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C a. Removal To rem ove the photom ultiplier tube/reactio n cham ber assem bly from the analyzer, do t he follo w: 1. Disc onnect pow er from the analyzer. 2. Releas e pressure f rom SAMPLE and AIR supplies. 3. U[...]

  • Page 45

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Routine Servicing 5-5 Model 951C Com pany under its cata log num ber S-195650- A. Using Alco nox and clean, soft f acial tissue (NOT industrial wipes) , carefull y clean the quartz wind ow. Vigorous ly flush react ion cham ber and quart z wind[...]

  • Page 46

    Instruction Manua l 748214-V June 2009 5-6 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C 5. Or ient the m etal shield and black plastic insu lator as noted in step 2. 6. Caref ully rotate and insert t he tube, shield and co ver into the housing. Orient as note d in ste p 1. 5-4 OZONE G ENER A TION [...]

  • Page 47

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Routine Servicing 5-7 Model 951C c. Power Supp ly Remov al To rem ove the Power S upply, do the f ol- lowing: Refer to Fig ure 6-1. 1. Rem ove the lam p and hous ing as in Section 5-4b. 2. Disc onnect the po wer lead from the Power Supp ly Boa[...]

  • Page 48

    Instruction Manua l 748214-V June 2009 5-8 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C[...]

  • Page 49

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Replacement Parts 6-1 Model 951C SECTION 6 REPL A CEM EN T P A RTS WARNING PARTS INT EGRIT Y Tamperin g with or unauthorized su bstitutio n of co mponents m ay adverse ly affect saf ety of t his produ ct. Use only f actory-app roved comp onent[...]

  • Page 50

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Replacement Parts 6-3 Model 951C 6-2 CIRCUIT BOARD R EPLACEM ENT POL ICY In mos t situations invol ving a m alfunction of a circuit board, it is m ore practical t o replace t he board than to attem pt isolation and replace- ment of the individ[...]

  • Page 51

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Replacement Parts 6-3 Model 951C Sample Regula tor 815187 Sample Press ure Gauge 644055 I/O Assembly 654390 Sample Flow Restri ctor 623936 Air Restrictor 655519 (on Ozone Generator inlet) Transformer/In ductor Assembly 654878 Fan Assembl y 654[...]

  • Page 52

    Instruction Manua l 748214-V June 2009 6-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C Socket Assembl y 654086 Reaction Cham ber 654381 Photomultiplier Tube and Magnetic Shield (PN 63091 6) to be Flush .3 Photomultiplier Tube Magnetic Shiel d 630916 PVC Tube Note: Silicone lu bricant to be applie[...]

  • Page 53

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Replacement Parts 6-5 Model 951C c. Converter Assembly 65 4070 Refer to Fig ure 6-3. 632784 Conn ector, Blu e Silicone 657127 Heater 632782 T emperatur e Sensor 632795 T ube Figure 6-3. C onverter Assembl y Connectors, Glass Tube 632784 Heater[...]

  • Page 54

    Instruction Manua l 748214-V June 2009 6-6 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C d. Temperat ure Control Assembl y 654068 Refer to Fig ure 6-4. 622733 Fan 622732 Heater 655335 Temperature Contro l Board 900492 Thermal Fuse Figure 6-4. C ase Heate r T emperature Cont rol Ass embly[...]

  • Page 55

    Instruction Manua l 748214-V June 2009 7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C SECTION 7 RETURN O F M ATERI A L 7-1 RETURN OF M ATERIAL If fac tory repair of defec tive equipm ent is re- quired, proce ed as f ollows: 1. Secure a return aut horizati on from a Rosem ount Anal yti cal I nc. [...]

  • Page 56

    Instruction Manua l 748214-V June 2009 7-2 Return of Material Rosemount Analytical Inc. A Divi sion of Emerson Process Mana gement Model 951C[...]

  • Page 57

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Index 8-1 Model 951C SECTION 8 INDEX 2 230 VAC con version, 2- 3 A Air (U.S.P. Bre athing G rade), 2-5 AIR inlet adequate a ir pressure, 3- 3 setpoint press ure, 3-3 analyzer installation l ocation, 2-3 mount near s ample, 2-3 operating tem pe[...]

  • Page 58

    Instruction Manua l 748214-V June 2009 8-4 Index Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C reaction cham ber, 4-1, 4- 2, 5-1, 5- 2, 5-3, 5- 4, 5-5, 5- 6 reaction cham ber, alter nate c leaning proce dure, 5-5 reaction cham ber, s tandard c leaning proce dure, 5-5 Return Author ization, 7- 2 S sample gas establis[...]

  • Page 59

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management List of Draw ings Model 951C LIST OF DR AWINGS Drawing 1-1. Panel Cu tout / Inst allation Draw ing[...]

  • Page 60

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C Drawing 1-2 Flow Diagram, Lo/M id Ran ge List of Draw ings[...]

  • Page 61

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management List of Draw ings Model 951C Drawing 1-3 Flow Diagram, Hi Rang e[...]

  • Page 62

    Instruction Manua l 748214-V June 2009 Rosemount Analytical Inc. A Division of Emerson Pro cess Management Model 951C W A RR A NT Y Goods and p art(s) ( excluding consum ables) m anuf actured b y Seller are w arrante d to be free f rom defects in workm anship and m aterial under norm al us e and s ervice for a perio d of twelve (12) months from the[...]

  • Page 63

    [...]

  • Page 64

    Instruction Manua l 748214-V June 2009 © Rosemount Analytical Inc . 2009 Model 951C WORLD HEA DQUARTERS A ND A MERICAS Emerson Process Manage ment Rosemount A nalytical Inc. 6565 P Davis Industrial Par kway Solon, OH 44139 USA T 440.914.1261 Toll Free in US and Canada 800.4 33.6076 F 440.914.1271 e-mail: gas.csc@Emerson.co m www.raiho me.c om Nati[...]