Go to page of
Similar user manuals
-
Oxygen Equipment
Invacare Perfecto2
86 pages 3.95 mb -
Oxygen Equipment
Invacare 10L
86 pages 3.95 mb -
Oxygen Equipment
Invacare HF2POST6
24 pages 0.63 mb -
Oxygen Equipment
Invacare Platinum 5LX
86 pages 3.95 mb -
Oxygen Equipment
Invacare 636959
32 pages 0.96 mb -
Oxygen Equipment
Invacare Non Platinum 5LX
86 pages 3.95 mb -
Oxygen Equipment
Invacare XPO100B
48 pages 2.39 mb -
Oxygen Equipment
Invacare IRC5LXO2AW
36 pages 0.92 mb
A good user manual
The rules should oblige the seller to give the purchaser an operating instrucion of Invacare IRC5PAW, along with an item. The lack of an instruction or false information given to customer shall constitute grounds to apply for a complaint because of nonconformity of goods with the contract. In accordance with the law, a customer can receive an instruction in non-paper form; lately graphic and electronic forms of the manuals, as well as instructional videos have been majorly used. A necessary precondition for this is the unmistakable, legible character of an instruction.
What is an instruction?
The term originates from the Latin word „instructio”, which means organizing. Therefore, in an instruction of Invacare IRC5PAW one could find a process description. An instruction's purpose is to teach, to ease the start-up and an item's use or performance of certain activities. An instruction is a compilation of information about an item/a service, it is a clue.
Unfortunately, only a few customers devote their time to read an instruction of Invacare IRC5PAW. A good user manual introduces us to a number of additional functionalities of the purchased item, and also helps us to avoid the formation of most of the defects.
What should a perfect user manual contain?
First and foremost, an user manual of Invacare IRC5PAW should contain:
- informations concerning technical data of Invacare IRC5PAW
- name of the manufacturer and a year of construction of the Invacare IRC5PAW item
- rules of operation, control and maintenance of the Invacare IRC5PAW item
- safety signs and mark certificates which confirm compatibility with appropriate standards
Why don't we read the manuals?
Usually it results from the lack of time and certainty about functionalities of purchased items. Unfortunately, networking and start-up of Invacare IRC5PAW alone are not enough. An instruction contains a number of clues concerning respective functionalities, safety rules, maintenance methods (what means should be used), eventual defects of Invacare IRC5PAW, and methods of problem resolution. Eventually, when one still can't find the answer to his problems, he will be directed to the Invacare service. Lately animated manuals and instructional videos are quite popular among customers. These kinds of user manuals are effective; they assure that a customer will familiarize himself with the whole material, and won't skip complicated, technical information of Invacare IRC5PAW.
Why one should read the manuals?
It is mostly in the manuals where we will find the details concerning construction and possibility of the Invacare IRC5PAW item, and its use of respective accessory, as well as information concerning all the functions and facilities.
After a successful purchase of an item one should find a moment and get to know with every part of an instruction. Currently the manuals are carefully prearranged and translated, so they could be fully understood by its users. The manuals will serve as an informational aid.
Table of contents for the manual
-
Page 1
Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Perfecto 2 ™ Series Oxygen Concentrators without SensO 2 ® , Oxygen Concentrators with SensO 2[...]
-
Page 2
WARNING Perfecto 2 ™Series 2 Part No 1150713 WARNING DO NOT use this product or any av ailable optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner ’ [...]
-
Page 3
TABLE OF CONTENTS Part No 1150713 3 Perfecto 2 ™Series ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ............................................ 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Radio Frequency Interference .. ...............[...]
-
Page 4
SPECIAL NOTES Perfecto 2 ™Series 4 Part No 1150713 SPECIAL NOTES Signal w ords are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for ?[...]
-
Page 5
SPECIAL NOTES Part No 1150713 5 Perfecto 2 ™Series CAUTION “Caution: Statutory law can restrict this device to sal e by or on order of a physician, or any other practitioner licens ed by the law of the governmental agency in which[...]
-
Page 6
LABEL LOCATION Perfecto 2 ™Series 6 Part No 1150713 LABEL LOCATION Serial Number Lab el is located on the resonator intake assembly Specification Label is located on the back near the bottom Perfecto 2 ™ FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room[...]
-
Page 7
SECTION 1—GENERAL GUIDELINES Part No 1150713 7 Perfecto 2 ™Series SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Pe r f e c t o 2 concentrator these instructions MUST be followed. WARNING This section conta[...]
-
Page 8
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™Series 8 Part No 1150713 Operating Information For optimum performance, Inv acare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may[...]
-
Page 9
SECTION 1—GENERAL GUIDELINES Part No 1150713 9 Perfecto 2 ™Series Radio Frequency Interference This equipmen t has been tested and found to comply with EMC limits specified by IEC/EN 60601 ‐ 1 ‐ 2. These limits are designed to provide a reasonable ?[...]
-
Page 10
SECTION 2—FEATURES Perfecto 2 ™Series 10 Part No 1150713 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter REAR VIEW *HF Outlet Fitting Power Cord *NOTE: This outlet fitting is to be used on[...]
-
Page 11
SECTION 3—HANDLING Part No 1150713 11 Perfecto 2 ™Serie s SECTION 3—HANDLING The concentrator should AL W A YS be kept in the upright position to prevent cabinet damage while being transported. If the concentrator is to be reshipped by common c[...]
-
Page 12
SECTION 4—TECHNICAL DESCRIPTION Perfecto 2 ™Series 12 Part No 1150713 SECTION 4—TECHNICAL DESCRIPTION The Invacare Per f e c t o 2 concentr ator is used by patients with respiratory disorders who require supplemental oxygen. The device is not intended to susta[...]
-
Page 13
SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 13 Perfecto 2 ™Serie s SECTION 5—TYPICAL PRODUCT PARAMETERS Alternating Current Type BF equipment Unit runnin g Unit not runnin g Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulated Not Protected Electrical Requirements: 230 VAC ± 10% (253 VAC/20 7 VAC), 50 [...]
-
Page 14
SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 14 Part No 1150713 Flow Range: 0.5 to 5 L/min (maximum ). For flowrates less than 1 L/min, we recommend th e use of the Invacare Pediatric Flowmeter Accessory (IRCPF16AW). Low Flow Alarm: 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Average Power Consumption[...]
-
Page 15
SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 15 Perfecto 2 ™Serie s No AP/APG Not suitable for use in the presence of a flammable anaesthetic mixture. Standards and Regulatory Listing: IRC5PO2AW/IRC5PAW IEC/EN61000-3-2 IEC/EN61000-3-3, IEC/EN 60601-1 , A1, A2 IEC/EN 60601-1 -2 IRC5PO2AW Models Only ISO8359 MDD 93/42/EEC, Annex I and IX [...]
-
Page 16
SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 16 Part No 1150713 IRC5PO2AW INDICATORS IRC5PAW INDI CATORS LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied[...]
-
Page 17
SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 17 Perfecto 2 ™Serie s SECTION 6— OPERATING INSTRUCTIONS Introduction Yo u r oxygen concentrator is intended for individual use. It is an electronically operated device that separates oxygen from room air . It p[...]
-
Page 18
SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 18 Part No 1150713 Yo u r oxygen concentrator will perform best when operated under the conditions outlined in the Ty p i c a l Product P arameters on page 13. Usage in environments other than those[...]
-
Page 19
SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 19 Perfecto 2 ™Serie s 3. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off (FIGURE 6.2). 4. Pull up [...]
-
Page 20
SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 20 Part No 1150713 10. After assembly , ensure that oxygen is flowing through the cannula. FIGURE 6.3 Humidifier Compartmen t Power Switch NOTE: For this pr ocedure, refe r to FIGURE 6.4. 1. Press power switch to O[...]
-
Page 21
SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 21 Perfecto 2 ™Serie s Flowrate NOTE: For this procedure, refe r to FIGURE 6.5. NOTE: Ta k e care not to set the flow above RED ring. An oxygen flow gr eater than 5 L/min will decr ease the[...]
-
Page 22
SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 22 Part No 1150713 SensO 2 Oxygen Purity Indicator - Model IRC5PO2AW This feature monitors the purity of oxygen generated by the oxygen concentrator . If purity falls below factory preset standards, indicator light[...]
-
Page 23
SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 23 Perfecto 2 ™Serie s FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights - Model IRC5PO2AW Explanation of Indicator Lights - Model IRC5PAW NOTE: For this procedure, refe r to FIGURE 6.7. IRC5PAW Model Indicator Light Explanation RED light () ‐ T[...]
-
Page 24
SECTION 7—MAINTENANCE Perfecto 2 ™Series 24 Part No 1150713 SECTION 7—MAINTENANCE WARNING The Invacare concentrators are specifically designed to minimize routine preventiv e mainte nance. Only qualified personnel should perform preventiv e maintenance on the [...]
-
Page 25
SECTION 7—MAINTENANCE Part No 1150713 25 Perfecto 2 ™Serie s 2. Clean the cabinet filter with a va c u u m cleaner or wa s h in w arm soapy wa t e r and rinse thoroughly . 3. Dry the filter thoroughly before reinstallation. FIGURE 7.1 Cleaning the Cabinet Filter[...]
-
Page 26
SECTION 7—MAINTENANCE Perfecto 2 ™Series 26 Part No 1150713 PREVENTIVE MAINTENANCE RECORD Model No. IRC ______________ Serial No.________________ On Each Inspection *NOTE: Refer to Preventive Ma intenance section of Service Manual. **NOTE: Model not available within [...]
-
Page 27
SECTION 8—TROUBLESHOOTING GUIDE Part No 1150713 27 Perfecto 2 ™Serie s SECTION 8—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord not plugged in. 1. Insert plug in to outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit break[...]
-
Page 28
SECTION 8—TROUBLESHOOTING GUIDE Perfecto 2 ™Series 28 Part No 1150713 YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2AW models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace ite[...]
-
Page 29
SECTION 9—OPTIONAL ACCESSORIES Part No 1150713 29 Perfecto 2 ™Serie s SECTION 9—OPTIONAL ACCESSORIES The following optional accessories are also av ailable: • Standard Adult Nasal Cannula, 2.1 m (7 feet) ‐ M3120 • Standard Humidifier ‐ M5120 • PreciseRX™ P ed[...]
-
Page 30
RECYCLING INFORMATION Perfecto 2 ™Series 30 Part No 1150713 RECYCLING INFORMATION This product has been supplied from an environmentally aw are manufacturer that complies with the Wa s t e Electrical and Electronic Equipment (WEEE) Directive 2002/96/CE. This p[...]
-
Page 31
LIMITED WARRANTY Part No 1150713 31 Perfecto 2 ™Serie s LIMITED WARRANTY United Kingdom: Invacare Ltd, South Road, Bridgend Industrial Estate, UK ‐ Bridge nd CF31 3PY Te l : (44) (0)1656 664 321, Fax: (44) (0)1656 667 532 UK@inv acare.com Ireland: Inv acare [...]
-
Page 32
0120 Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Services 440-329-6593 800-832-4707 EU Representative Invacare International Sarl Route de Cité Ouest 2 1196 Gland Switzerland Tel: +41 22 354 60 10 Fax: +41 22 354 60 11 Part No 1150713 Invacare Corporation 2101 E. Lake M[...]